Samuel A Minaker1,2, Ryan H Mason1,2, David R Chow1,2.
Abstract
Purpose: To evaluate the effect of surgeon-controlled parameters on the color performance of the Ngenuity 3-dimensional (3D) visualization system. Design: A calibrated reference target was placed inside a model eye to assess the Ngenuity 3D camera under different settings. The Ngenuity 3D display was assessed with a commercial colorimeter.Entities:
Keywords: 3D visualization; 3D, 3-dimensional; AFX, air-fluid exchange; Color accuracy; HDR, high dynamic range; Medical imaging; Ngenuity; OLED, organic light-emitting diode; White balance; sRGB, standard red-green-blue
Year: 2021 PMID: 36247816 PMCID: PMC9559094 DOI: 10.1016/j.xops.2021.100054
Source DB: PubMed Journal: Ophthalmol Sci ISSN: 2666-9145
Figure 1Experimental setup for color analysis: (A) delta E calculation factoring differences in chroma, hue, lightness, and saturation in a standardized color space to quantify perceived color differences; (B) ColorGauge 6 × 5 target; (C) model eye with light pipe illumination source; (D) photograph of the color target inside the model eye; and (E) example of data output from the Imatest Master analysis software.
Summary of Delta E Values with Their Associated Level of Human Perception
| ΔE00 | Perception |
|---|---|
| < 1.0 | Not perceptible by human eyes |
| 1–2 | Perceptible through close observation |
| 3–10 | Perceptible at a glance |
| 11–49 | Colors are more similar than opposite |
| 100 | Colors are exact opposites |
Figure 2Graph showing the delta E mean and standard deviation of experiments 1 through 9, where ˆ represents the standard technique or methodology and ∗ represents those experiments that significantly differed at the 95% confidence level. AFX = air fluid exchange.
Summary of Delta E and Delta C Values for Experiments 1 through 9
| Category | Method | ΔE00 | ΔC00 | ||
|---|---|---|---|---|---|
| White balance target | Ngenuity test card | 12.81 ± 1.67 | — | 6.17 ± 0.24 | — |
| Grey card | 15.57 ± 2.04 | 0.066 | 6.21 ± 0.21 | 0.785 | |
| 4 × 4 gauze | 14.08 ± 2.31 | 0.373 | 6.14 ± 0.15 | 0.789 | |
| White paper | 15.28 ± 1.22 | 0.011 | 6.69 ± 0.10 | <0.001 | |
| Operator’s hand | 17.71 ± 2.03 | 0.009 | 11.33 ± 0.92 | 0.002 | |
| White balance technique | Standard technique | 12.81 ± 1.67 | — | 6.17 ± 0.24 | — |
| Card flat | 13.18 ± 1.70 | 0.734 | 6.65 ± 0.45 | 0.114 | |
| Low magnification | 14.77 ± 1.50 | 0.079 | 6.76 ± 0.18 | 0.001 | |
| Card in focus | 15.78 ± 1.63 | 0.026 | 7.05 ± 0.09 | <0.001 | |
| Card moving | 13.29 ± 0.61 | 0.492 | 6.39 ± 0.06 | 0.036 | |
| Room lights on | 15.53 ± 2.01 | 0.067 | 7.00 ± 0.19 | <0.001 | |
| White balance stability | Day 0 | 12.97 ± 0.49 | — | 5.34 ± 0.07 | — |
| Day 180 | 13.69 ± 0.59 | 0.176 | 9.02 ± 0.04 | <0.001 | |
| Laser filter | Laser filter on/on | 12.81 ± 1.67 | — | 6.17 ± 0.24 | — |
| Laser filter off/off | 9.28 ± 0.25 | <0.001 | 5.18 ± 0.25 | <0.001 | |
| Laser filter off/on | 16.59 ± 1.17 | 0.002 | 9.11 ± 0.56 | <0.001 | |
| Aperture | 30% | 12.81 ± 1.67 | — | 6.17 ± 0.24 | — |
| 50% | 10.46 ± 1.33 | 0.029 | 5.78 ± 0.28 | 0.064 | |
| 75% | 13.18 ± 2.35 | 0.790 | 5.87 ± 0.25 | 0.089 | |
| 100% | 15.42 ± 2.83 | 0.165 | 5.79 ± 0.25 | 0.043 | |
| Illumination | 34% light pipe | 12.81 ± 1.67 | — | 6.17 ± 0.24 | — |
| 20% light pipe | 21.67 ± 2.50 | 0.003 | 8.06 ± 0.06 | <0.001 | |
| 10% light pipe | 29.97 ± 2.40 | <0.001 | 9.18 ± 0.20 | <0.001 | |
| 50% chandelier | 14.06 ± 2.90 | 0.367 | 7.02 ± 0.27 | 0.003 | |
| 25% chandelier | 17.60 ± 3.00 | 0.041 | 7.77 ± 0.21 | <0.001 | |
| Imaging mode | Posterior | 12.81 ± 1.67 | — | 6.17 ± 0.24 | — |
| AFX | 14.91 ± 0.86 | 0.016 | 6.37 ± 0.19 | 0.167 | |
| Anterior | 11.31 ± 1.53 | 0.164 | 6.09 ± 0.24 | 0.623 | |
| Hemorrhage | 33.48 ± 3.68 | <0.001 | 17.89 ± 0.37 | <0.001 | |
| Macular | 12.32 ± 2.88 | 0.766 | 6.12 ± 0.31 | 0.804 | |
| Instruments | Ngenuity 1 | 12.81 ± 1.67 | — | 6.17 ± 0.24 | — |
| Ngenuity 2 | 10.69 ± 1.78 | 0.027 | 5.43 ± 0.48 | 0.003 | |
| Ngenuity 3 | 9.51 ± 0.54 | <0.001 | 5.12 ± 0.28 | <0.001 |
– = no P-value calculated.
Figure 3Analysis of display uniformity: (A) example of the static display overlays on the monitor; (B) examples of image retention; (C) display uniformity for Ngenuity system 1 as measured by delta E across 9 quadrants; (D) display uniformity for Ngenuity system 1 after manual pixel refresh; (E) display uniformity for Ngenuity system 2; and (F) display uniformity for Ngenuity system 3.