Comparison of Two Scheimpflug Systems in the Measurements of Eyes with Corneal Diameter Smaller than 11.1 mm.

Lingling Niu1, Lan Ding1, Yishan Qian2, Xingtao Zhou3.   

Abstract

INTRODUCTION: This article aimed to evaluate the measurements of ectasia parameters by two Scheimpflug-based tomography devices, Pentacam and Sirius, for eyes with different corneal diameters (CDs).
METHODS: This cross-sectional research included subjects from the Fudan University EENT Hospital Refractive Center Database that were followed once a year for at least 3 years with unremarkable slit-lamp examination and normal topography. Pentacam and Sirius examinations were performed on these subjects and the ectasia indices were compared between different CD groups.
RESULTS: The right eyes of 153 subjects were included (CD ≤ 11.1 mm, n = 50; 11.2-12 mm, n = 52; > 12.0 mm, n = 51). For the ectasia parameters from Pentacam, CD had the greatest influence on the deviation of normality of back elevation (Db, R2 = 0.371, β = - 1.119, P < 0.001), overall deviation of normality (BAD-D, R2 = 0.305, β = - 0.589, P < 0.001), and minimum pachymetric progression index (PPImin, R2 = 0.282, β = - 0.131, P < 0.001). For parameters derived from Sirius, CD had the greatest influence on Baiocchi-Calossi-Versaci index of the back surface (BCVb, R2 = 0.138, β = - 0.179, P < 0.001), keratoconus vertex of the back surface (KVb, R2 = 0.099, β = - 2.273, P < 0.001), and BCV (R2 = 0.071, β = - 0.078, P = 0.001). CD had little influence on surface asymmetry index of the front (SIf) and back surface (SIb), keratoconus vertex of the front surface (KVf), Baiocchi-Calossi-Versaci index of the front surface (BCVf), and Sirius classifier (P > 0.05).
CONCLUSIONS: For Pentacam, CD mainly influenced indices related to back elevation (BE) and pachymetry progression, whereas for Sirius, CD mainly influenced indices related to BE and corneal aberration.
© 2022. The Author(s).

Entities:  

Keywords:  Corneal diameter; Forme fruste keratoconus; Scheimpflug; Tomography

Year:  2022        PMID: 36244046     DOI: 10.1007/s40123-022-00591-y

Source DB:  PubMed          Journal:  Ophthalmol Ther


  2 in total

1.  [The influence of different corneal diameters on Belin/Ambrósio enhanced ectasia display of Pentacam corneal topography].

Authors:  K W Cao; L N Liu; Y L Sun; T Zhang; J Bai; T Liu
Journal:  Zhonghua Yan Ke Za Zhi       Date:  2020-10-11

Review 2.  Galilei Corneal Tomography for Screening of Refractive Surgery Candidates: A Review of the Literature, Part II.

Authors:  Majid Moshirfar; Mahsaw N Motlagh; Michael S Murri; Hamed Momeni-Moghaddam; Yasmyne C Ronquillo; Phillip C Hoopes
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2019