Eman M Alshehri1, Nawal A Alarfaj1, Salma A Al-Tamimi1, Maha F El-Tohamy1.
Abstract
Two metal oxide nanoparticles, magnesium oxide nanoparticles (MgONPs) and aluminum oxide nanoparticles (Al2O3NPs), were synthesized from green sources, Salvia officials and Cuminum cyminum seed extract, respectively. These nanoparticles were used for construction of potentiometric enhancement sensors employed for the estimation of ranitidine hydrochloride (RNT) in authentic powder and commercial products. The electroactive substance ranitidine-phosphotungstate (RNT-PT) was formed by combining RNT with phosphotungstic acid (PTA) in the presence of plasticizing material o-nitrophenyloctyl ether (o-NPOE). The outcomes showed that the enhanced MgO and Al2O3 nanosensors behaved linearly across the concentration ranges 1.0 × 10-9-1.0 × 10-2 and 1.0 × 10-10-1.0 × 10-2 mol L-1, respectively. However, the conventional sensor (RNT-PT) displayed a linearity over 1.0 × 10-6-1.0 × 10-2 mol L-1. Least square equations were calculated as EmV = (54.1 ± 0.5) log (RNT) + 762.33, EmV = (58.6 ± 0.2) log (RNT) + 696.48, and EmV = (52.2 ± 0.7) log (RNT) + 756.76 for enriched nanometal oxides modified and conventional sensors, respectively. The correlation coefficients of regression equations were 0.9997, 0.9995, and 0.9992 for the above suggested sensors, respectively. The recorded results showed excellent sensitivity and selectivity of the modified nanometal oxide sensors for the quantification of the analyzed drug in its authentic samples and commercial products.Entities:
Keywords: coated wire sensors; metal oxide nanoparticles; modified potentiometric sensors; ranitidine hydrochloride
Year: 2022 PMID: 36236096 PMCID: PMC9571185 DOI: 10.3390/polym14194150
Source DB: PubMed Journal: Polymers (Basel) ISSN: 2073-4360 Impact factor: 4.967
Figure 1Structural formula of ranitidine hydrochloride.
Figure 2Green synthesis of (a) MgONPs using Salvia officials leaves extract and (b) Al2O3NPs using Cuminum cyminum seeds extract.
Figure 3Illustration of modified RNT-PT-MONPs and the potentiometric sensor system.
Figure 4Scanning electron microscope images of (a) MgONPs and (b) Al2O3NPs.
Figure 5FT-IR spectra of the green synthesized (a) MgONPs and (b) Al2O3NPs using Salvia officials leaves and Cuminum cyminum seed extract, respectively, at a wavenumber range from 4000 to 400 cm−1.
Figure 6EDX analysis and elemental X-ray mapping of green synthesized (a) MgONPs and (b) Al2O3NPs using Salvia officials leaves and Cuminum cyminum seed extract, respectively.
Figure 7XRD patterns of green synthesized (a) MgONPs and (b) Al2O3NPs using Salvia officials leaves and Cuminum cyminum seed extract, respectively, at Cu-kα and (k = 1.54060 Å).
Electrochemical response characteristics of conventional coated wire RNT-PT, modified RNT-PT-MgONPs, and RNT-PT-Al2O3NPs sensors.
| Parameters | Conventional RNT-PT Sensor | Modified RNT-PT-MgONPs Sensor | Modified |
|---|---|---|---|
| Slope (mV. Decade−1) | 52.2 ± 0.7 | 54.1 ± 0.5 | 58.6 ± 0.2 |
| Intercept | 756.76 | 762.33 | 696.48 |
| Correlation coefficient, r | 0.9992 | 0.9997 | 0.9995 |
| Linear range (mol L−1) | 1.0 × 10−6–1.0 × 10−2 | 1.0 × 10−9–1.0 × 10−2 | 1.0 × 10−10–1.0 × 10−2 |
| LOD | 5.0 × 10−7 | 4.9 × 10−10 | 5.0 × 10−11 |
| Response time/s | 50 | 30 | 25 |
| Working pH range | 3–9 | 3–9 | 3–9 |
| Lifetime/day | 20 | 40 | 50 |
| Temperature, °C | 25 | 25 | 25 |
| Accuracy (%) | 99.04 ± 0.73 | 99.49 ± 0.40 | 99.54 ± 0.53 |
Figure 8Calibration graphs of the fabricated (a) conventional RNT-PT, (b) modified RNT-PT-MgONPs, and (c) modified RNT-PT-Al2O3NPs sensors.
Figure 9Influence of pH on the suggested conventional RNT-PT and modified metal oxide RNT-PT-MgONPs and RNT-PT-Al2O3NPs coated wire sensors using 1.0 × 10−3 mol L−1 of RNT solution.
Selectivity coefficient (KPot RNT+) of conventional coated wire RNT-PT and modified RNT-PT-MgONPs and RNT-PT-Al2O3NPs sensors using the separate solution method using 1.0 × 10−3 mol L−1 RNT.
| Interfering Species | Conventional RNT-PT Sensor | Modified RNT-PT-MgONPs | Modified |
|---|---|---|---|
| Fe3+ | 3.6 × 10−3 | 1.4 × 10−4 | 4.9 × 10−4 |
| Ca2+ | 7.5 × 10−3 | 1.6 × 10−5 | 7.7 × 10−5 |
| Cr3+ | 4.1 × 10−3 | 5.2 × 10−5 | 5.4 × 10−4 |
| K+ | 5.5 × 10−3 | 5.8 × 10−4 | 4.6 × 10−4 |
| Na+ | 6.9 × 10−3 | 4.2 × 10−4 | 3.5 × 10−5 |
| Ag+ | 2.1 × 10−3 | 6.8 × 10−5 | 1.6 × 10−4 |
| Mg2+ | 6.2 × 10−3 | 8.4 × 10−5 | 8.1 × 10−4 |
| Lactose | 5.2 × 10−3 | 6.3 × 10−4 | 2.4 × 10−5 |
| Glycine | 8.5 × 10−3 | 4.5 × 10−5 | 1.5 × 10−4 |
| Histidine | 6.3 × 10−3 | 3.6 × 10−4 | 2.3 × 10−5 |
| Leucine | 7.9 × 10−3 | 9.8 × 10−4 | 3.9 × 10−4 |
| Niperotidine | 9.5 × 10−3 | 3.5 × 10−4 | 4.8 × 10−4 |
Statistical analysis of data obtained from the determination of RNT in bulk powder using conventional coated wire RNT-PT, RNT-PT-MgONPs, and RNT-PT-Al2O3NPs sensors.
| Conventional RNT-PT Coated Wire Sensor | Modified RNT-PT MgONPs Sensor | Modified RNT-PT-Al2O3NPs Sensor | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Taken | Found | % | Taken | Found | % | Taken | Found | % | |
| Statistical Analysis | 6.00 | 5.96 | 99.33 | 9.00 | 9.00 | 100.00 | 9.00 | 8.98 | 99.78 |
| 5.30 | 5.23 | 98.68 | 9.50 | 9.45 | 99.47 | 8.50 | 8.50 | 100.00 | |
| 5.00 | 5.00 | 100.00 | 8.00 | 7.97 | 99.63 | 8.00 | 7.95 | 99.38 | |
| 4.30 | 4.24 | 98.60 | 7.50 | 7.45 | 99.33 | 7.00 | 7.00 | 100.00 | |
| 4.00 | 3.95 | 98.75 | 7.00 | 6.98 | 99.71 | 6.50 | 6.45 | 99.23 | |
| 3.30 | 3.28 | 99.39 | 5.50 | 5.50 | 100.00 | 6.00 | 5.99 | 99.83 | |
| 3.00 | 3.00 | 100.00 | 4.00 | 3.91 | 98.75 | 4.00 | 4.00 | 100.00 | |
| 2.30 | 2.28 | 99.13 | 2.50 | 2.49 | 99.60 | 3.50 | 3.44 | 98.29 | |
| 2.00 | 1.95 | 97.50 | 2.00 | 1.98 | 99.00 | 3.00 | 2.98 | 99.33 | |
| Mean ± SD | 99.04 ± 0.73 | 99.49 ± 0.40 | 99.54 ± 0.53 | ||||||
| n | 9 | 9 | 9 | ||||||
| Variance | 0.54 | 0.16 | 0.28 | ||||||
| * %SE | 0.24 | 0.13 | 0.18 | ||||||
| %RSD | 0.74 | 0.40 | 0.53 | ||||||
* SE (%Error) = %RSD/√n.
Intra- and inter-day assay of ranitidine hydrochloride by using the modified RNT-PT-MgONPs and RNT-PT-Al2O3NPs coated wire sensors.
| Precision Test | Taken | % | % | % | |
|---|---|---|---|---|---|
| RNT-PT-MgONPs | Intra-day precision | 10.00 | 99.50 ± 0.5 | 0.5 | 0.32 |
| 5.00 | 98.83 ± 1.2 | 1.2 | 0.70 | ||
| 2.00 | 99.83 ± 0.2 | 0.2 | 0.12 | ||
| Inter-day precision | 10.00 | 98.53 ± 1.2 | 1.2 | 0.72 | |
| 5.00 | 98.83 ± 0.8 | 0.8 | 0.50 | ||
| 2.00 | 97.67 ± 0.6 | 0.6 | 0.36 | ||
| RNT-PT-Al2O3NPs | Intra-day precision | 9.00 | 99.63 ± 0.3 | 0.3 | 0.17 |
| 6.00 | 99.23 ± 0.7 | 0.7 | 0.40 | ||
| 3.00 | 98.37 ± 0.8 | 0.8 | 0.45 | ||
| Inter-day precision | 9.00 | 99.11 ± 0.4 | 0.4 | 0.24 | |
| 6.00 | 98.33 ± 1.3 | 1.3 | 0.73 | ||
| 3.00 | 98.79 ± 0.6 | 0.6 | 0.33 | ||
a Mean of three determinations, b %RSD = (SD/Mean) ×100, c % Error = %RSD/√n.
Statistical analysis of data obtained from the determination of RNT in Ranimax® tablets, 150 mg/tablet, using conventional coated wire RNT-PT and modified RNT-PT-MgONPs and RNT-PT-Al2O3NPs sensors.
| Conventional RNT-PT Coated Wire Sensor | Modified RNT-PT MgONPs Sensor | Modified RNT-PT-Al2O3NPs Sensor | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Taken | Found | % | Taken | Found | % | Taken | Found | % | |
| Statistical Analysis | 6.00 | 5.94 | 99.0 | 9.0 | 8.98 | 99.8 | 7.0 | 6.98 | 99.7 |
| 5.00 | 4.97 | 99.4 | 8.0 | 7.97 | 99.6 | 6.0 | 5.99 | 99.8 | |
| 4.30 | 4.26 | 99.1 | 7.0 | 6.96 | 99.3 | 5.0 | 4.95 | 99.0 | |
| 4.00 | 3.96 | 99.0 | 6.0 | 5.97 | 99.5 | 4.0 | 4.00 | 100.0 | |
| 3.00 | 2.97 | 99.0 | 3.0 | 2.99 | 98.7 | 3.0 | 2.96 | 98.7 | |
| 2.00 | 1.97 | 98.5 | 2.0 | 1.99 | 99.5 | 2.0 | 1.99 | 99.5 | |
| Mean ± SD | 99.00 ± 0.26 | 99.40 ± 0.37 | 99.45 ± 0.50 | ||||||
| n | 6 | 6 | 9 | ||||||
| Variance | 0.07 | 0.14 | 0.25 | ||||||
| %SE * | 0.11 | 0.15 | 0.20 | ||||||
| %RSD | 0.26 | 0.33 | 0.50 | ||||||
| 0.569 (2.228) ** | 1.08 (2.228) ** | 1.13 (2.228) ** | |||||||
| F-test | 2.85 (5.05) ** | 1.43 (5.05) ** | 1.25 (5.05) ** | ||||||
| Reported method [ | 99.13 ± 0.45 | ||||||||
| 6 | |||||||||
| 0.25 | |||||||||
| 0.07 | |||||||||
* SE (%Error) = %RSD/√n, ** The tabulated values of ‘‘t’’ and ‘‘F’’ at confidence level p = 0.05 [28].
A comparative study between the results obtained from the determination of RNT using the Potentiometric method by using modified RNT-PT-MgONPs and RNT-PT-Al2O3NPs sensors, and the previously reported analytical techniques.
| Analytical Method | Reagent | Linearity | LOD | Reference |
|---|---|---|---|---|
| Spectrophotometry | RNT, ninhydrin | 8.98 × 103–9.90 × 104 µg L−1 | 0.0997 µg mL−1 | [ |
| Chemiluminescence | RNT, S, N co-doped carbon quantum dots | 0.5–50 μg mL−1 | 0.12 µg mL−1 | [ |
| Fluorescence | RNT, CdS quantum dots | 0.50–15.0 μg mL−1 | 0.38 µg mL−1 | [ |
| Chromatography | RNT, RP-HPLC method, ortho-phosphoric acid 0.1% and acetonitrile pH 3.5 (25:75, %v/v) | 5–25 μg mL−1 | 1.35 µg mL−1 | [ |
| Electrochemical | RNT, poly(dopamine) modified carbon | 1.0 × 10−7–7.5 × 10−6 mol L−1 | 1.9 × 10−8 mol L−1 | [ |
| RNT, modified pencil graphite electrode (PGE) modified with | 2 × 10−4–1.7 × 10−2 mol L−1 | 1.57 × 10−4 mol L−1 | [ | |
| RNT, poly (chromotrope 2B) modified activated glassy carbon electrode (PCHAGCE) | 1.0 ×10−5–4.0 ×10−4 mol L−1 | 5.4 ×10−7 mol L−1 | [ | |
| RNT, poly(thionine)-modified anodized glassy carbon electrode (PTH/GCE) | 35–500 µmol L−1 | 1.5 µ mol L−1 | [ | |
| RNT, carbon paste electrode modified with the N,N-ethylene-bis(salicyllideneiminato)oxovanadium (IV) complex ((VO(salen))) | 9.9 × 10−5–1.0 × 10−3 mol L−1 | 6.6 ×10−5 mol L−1 | [ | |
| Modified carbon paste electrode, tetraphenylborate | 1.0 ×10−6–1.0 ×10−2 mol L−1 | 1.0 ×10−6 mol L−1 | [ | |
| Proposed method | Potentiometric measurement modified RNT-PT-MgONPs and RNT-PT-Al2O3NPs sensors | 1.0 × 10−9–1.0 × 10−2 mol L−1 | 4.9 × 10−10 mol L−1 | RNT-PT-MgONPs sensor |
| 1.0 × 10−10–1.0 × 10−2 mol L−1 | 5.0 × 10−11 mol L−1 | RNT-PT-Al2O3NPs sensor |