| Literature DB >> 36199805 |
Gabriel A Quesada1, Rodrigo A Quesada1, Jason J Jones2, Benjamin J K Straker3, Wuchen Zhao3, Linda Tsai3, Srividhya Vilupuru3.
Abstract
Purpose: To evaluate the reproducibility of magnitude of postoperative IOL rotation following implantation of a toric intraocular lens (IOL) with modified haptics, in comparison with a Proof-of-Concept (POC) study of prototype IOLs featuring the same haptic design. Patients andEntities:
Keywords: IOL rotation; astigmatism; axis misalignment; cataract; surgery; toric IOL
Year: 2022 PMID: 36199805 PMCID: PMC9529011 DOI: 10.2147/OPTH.S373976
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1High magnification photograph of the squared and frosted haptic, a design feature of both the TECNIS Toric II and Prototype 2 IOLs.
TECNIS® Toric II Models ZCU, IOL Astigmatism Correction Range
| Cylinder Powers (D) | |||||||
|---|---|---|---|---|---|---|---|
| IOL Model | ZCU150 | ZCU225 | ZCU300 | ZCU375 | ZCU450 | ZCU525 | ZCU600 |
| 1.50 | 2.25 | 3.00 | 3.75 | 4.50 | 5.25 | 6.00 | |
| 1.03 | 1.54 | 2.06 | 2.57 | 3.08 | 3.60 | 4.11 | |
Notes: *The corresponding cylinder values at the corneal plane have been calculated based on the average pseudophakic eye.
Abbreviations: D, diopters; IOL, intraocular lens.
Participant Demographics at Baseline
| TECNIS Toric II (Model ZCU) | Prototype 2 (Model MER004) | ||
|---|---|---|---|
| N = 125 | N = 98 | ||
| Age, years | Mean (SD) | 68.4 (8.3) | 67.3 (8.7) |
| Range | 31 to 88 | 43 to 88 | |
| Age group, years n (%) | <60 | 17 (13.6) | 15 (15.2) |
| 60–69 | 47 (37.6) | 43 (43.9) | |
| 70–79 | 53 (42.4) | 35 (35.7) | |
| ≥80 | 8 (6.4) | 5 (5.1) | |
| Sex, n (%) | Male | 53 (42.4) | 30 (30.6) |
| Female | 72 (57.6) | 68 (69.4) | |
| Race, n (%) | American Indian/Alaska Native | 0 | 0 |
| Asian (including Indian) | 2 (1.6) | 43 (43.9) | |
| Black | 9 (7.2) | 3 (3.1) | |
| Native Hawaiian/Pacific Islander | 0 | 0 | |
| Caucasian | 112 (89.6) | 1 (1.0) | |
| Other | 2 (1.6) | 51 (52.0) | |
| Ethnicity, n (%) | Hispanic/Latino | 7 (5.6) | 56 (57.1) |
| Not Hispanic/Latino | 118 (94.4) | 42 (42.9) | |
| Iris color, n (%) | Blue/Gray | 46 (36.8) | 1 (1.0) |
| Brown/Black | 49 (39.2) | 97 (99.0) | |
| Green/Hazel | 30 (24) | 0 | |
| Axial length, mm | Mean | 24.49 | 23.33 |
| SD | 3.23 | 0.92 | |
| Median | 24.32 | 23.21 | |
| Min | 20.67 | 21.75 | |
| Max | 29.52 | 25.93 |
Abbreviations: Min, minimum; Max, maximum.
Mean Absolute Lens Rotation (°, Degrees) at the Postoperative 1-Day and 1-Week Visits
| Postoperative Time Point | IOL | N | Mean Degrees | SD | Median | Min | Max | ≤5° (%) | <10° (%) |
|---|---|---|---|---|---|---|---|---|---|
| 1-Day | TECNIS Toric II | 189 | 0.823 | 0.996 | 0.567 | 0.007 | 8.323 | 98.9 | 100 |
| Prototype version 2 | 89 | 0.727 | 0.753 | 0.448 | 0.001 | 3.298 | 100 | 100 | |
| 1-Week | TECNIS Toric II | 191 | 0.847 | 0.921 | 0.590 | 0.004 | 7.628 | 99.5 | 100 |
| Prototype version 2 | 91 | 0.713 | 0.690 | 0.460 | 0.006 | 2.742 | 100 | 100 |
Abbreviations: IOL, intraocular lens; Min, minimum; Max, maximum.
Figure 2Boxplots of absolute IOL rotation in degrees (°) from operative to post-operative (A) 1-day and (B) 1-week visits.
Figure 3Absolute IOL rotation by degree (°) using photographic method at postoperative (A) 1-day and (B) 1-week visits.
Figure 4Scatterplot of absolute IOL rotation by degree (°) at postoperative 1-week versus axial length (millimeters, mm) for the TECNIS Toric II IOL.
Figure 5Frequency of UCDVA and BCDVA at postoperative 1-week with the TECNIS Toric II IOL.