| Literature DB >> 36198731 |
Satoshi Higuchi1,2, Tatsuya Nishii3, Atsushi Hirota4, Shota Harumoto3, Hiroki Horinouchi3, Emi Tateishi3, Yasutoshi Ohta3, Keisuke Kiso3,5, Kenichi Kurosaki4, Tetsuya Fukuda3.
Abstract
Patient positioning at the isocenter of the CT gantry is important for optimizing image quality and radiation dose, but accurate positioning is challenging in pediatric patients. We evaluated whether the high-resilience pad and pre-scan measurement of chest thickness allow accurate positioning in pediatric patients with congenital heart disease. Sixty-seven patients aged 7 years or younger who underwent cardiothoracic CT were enrolled. The ideal table height, defined as the position at which the scanner's and patient's isocenters coincided, was determined by radiographers either manually (manual group) or based on the pad's and chest's thickness (calculated group). The distance between the two isocenters and image quality were evaluated. The calculated group demonstrated smaller isocenter distance and standard deviation (distance: 0.2 ± 5.8 mm vs. - 8.3 ± 11.6 mm, p < 0.01; absolute value: 4.1 [1.9-8.0] mm vs. 12.3 [5.1-16.3] mm, p < 0.01), and higher signal-to-noise ratio (SNR) and dose-normalized SNR (SNRD) in the descending aorta than the manual group (SNR: 39.8 [31.0-53.7] vs. 31.9 [28.9-36.6], p = 0.048, SNRD: 39.8 [31.0-53.7] vs. 31.9 [28.9-36.6], p = 0.04). The system allowed for more accurate positioning in pediatric cardiothoracic CT, yielding higher image quality.Entities:
Mesh:
Year: 2022 PMID: 36198731 PMCID: PMC9534888 DOI: 10.1038/s41598-022-21018-5
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.996
Figure 1(a) The high-resilience pad (white mattress) is put under the vacuum device (blue mat). The total thickness of the pad system was measured at 75 mm from the top plate, which is restricted to move up to 100 mm below the isocenter of the gantry. (b) The chest thickness in the supine position was measured at the level of the nipple by a ruler. In the calculated group, the setting table height was calculated using the following formula; setting table height = 75 + measured chest thickness (mm)/2.
Figure 2Definition of patient isocenter. (a) Axial image binarized with the CT value threshold to extract the patient isocenter, defined as the middle point (yellow dotted line) between the highest and lowest point (white dotted lines) of the extracted skin surface. All images were reconstructed so that the center of the image coincides with the isocenter of the gantry (blue cross). (b) Sagittal image showing the grand truth vertical center of patients, calculated using the averaged center values of each image along the scan length (yellow line). The patient was positioned below the isocenter of the gantry (blue line).
Patient demographics.
| Manual group (n = 34) | Calculated group (n = 33) | P value | |
|---|---|---|---|
| Age (months) | 8 [2–31] | 10 [1–29] | 0.99 |
| Female | 17 | 19 | 0.63 |
| Body weight (kg) | 6.4 [3.3–8.8] | 6.5 [3.3–10.5] | 0.77 |
| Height (cm) | 64 [51–86] | 65 [52–85] | 0.67 |
| Body thickness (mm) | 104 [96–119] | 108 [94–122] | 0.76 |
| Body width (mm) | 152 [113–173] | 154 [116–184] | 0.56 |
| Effective diameter (cm) | 12.2 [10.3–14.4] | 13.0 [10.3–15.0] | 0.58 |
| Heart rate (bpm) | 120 [109–143] | 134 [121–149] | 0.07 |
| CTDIvol(16 cm) (mGy) | 0.72 [0.54–1.06] | 0.7 [0.52–1.08] | 0.6 |
| DLP (mGy・cm) | 6.0 [4.2–10.5] | 5.9 [3.7–10.0] | 0.36 |
| SSDE (mGy) | 0.76 [0.64–1.16] | 0.90 [0.64–1.38] | 0.32 |
Values are n, median [interquartile range]. * indicates statistical significance.
CTDIvol computed tomography dose index volume, DLP dose-length product, SSDE size-specific dose estimates (14).
Figure 3Box-and-whisker plots of patient positioning performance in the calculated and manual groups. The median (horizontal line within box), interquartile range (box), and nonoutlier range (whiskers) defined as 1.5 times the interquartile range. The largest deviations from the scanner isocenter, outside the nonoutlier range, are plotted as open dots. (a) Isocenter distance. The positive and negative values above and below the isocenter, respectively. (b) Absolute distance. (c) Distance ratio, defined as the ratio of the absolute isocenter distance to the body thickness.
Image quality in the aorta.
| Calculated group (n = 21) | Manual group (n = 23) | P value | |
|---|---|---|---|
| SNR | 35.6 [26.2–56.4] | 34.4 [26.3–55.3] | 0.52 |
| SNRD | 64.1 [46.7–94.2] | 56.2 [40.8–78.7] | 0.29 |
| SNR | 39.8 [31.0–53.7] | 31.6 [28.9–36.6] | 0.048* |
| SNRD | 81.1 [49.0–106.4] | 59.1 [47.5–68.0] | 0.04* |
The patients with Glenn shunt and Fontan circulation are excluded due to their specific methods of contrast agent administration. * indicates statistical significance.
SNR signal-to-noise ratio, SNRD dose-normalized signal-to-noise ratio.