Jun Yoshitake1, Mayuko Azami2, Haruka Sei2, Daisuke Onoshima1, Kumiko Takahashi3, Akiyoshi Hirayama1,4, Koji Uchida5,6, Yoshinobu Baba1,7,8,9, Takahiro Shibata1,2,10. 1. Institute of Nano-Life-Systems, Institutes of Innovation for Future Society, Nagoya University, Furo-cho, Chikusa-ku, Nagoya 464-8603, Japan. 2. Graduate School of Bioagricultural Sciences, Nagoya University, Furo-cho, Chikusa-ku, Nagoya 464-8601, Japan. 3. Materials Integration Laboratories, AGC Inc., 1-1, Suehiro-cho, Tsurumi-ku, Yokohama, Kanagawa 230-0045, Japan. 4. Institute for Advanced Biosciences, Keio University, 246-2 Mizukami, Kakuganji, Tsuruoka, Yamagata 997-0052, Japan. 5. Graduate School of Agricultural and Life Sciences, The University of Tokyo, Yayoi 1-1-1, Bunkyo-ku, Tokyo 113-8657, Japan. 6. Japan Agency for Medical Research and Development, CREST, 1-7-1 Otemachi, Chiyoda-ku, Tokyo 100-0004, Japan. 7. Department of Biomolecular Engineering, Graduate School of Engineering, Nagoya University, Furo-cho, Chikusa-ku, Nagoya 464-8603, Japan. 8. Institute of Quantum Life Science, National Institutes for Quantum and Radiological Science and Technology, Anagawa 4-9-1, Inage-ku, Chiba 263-8555, Japan. 9. College of Pharmacy, Kaohsiung Medical University, 100, Shin-Chuan 1st Rd., Kaohsiung 807, Taiwan. 10. Institute for Glyco-Core Research (iGCORE), Nagoya University, Furo-cho, Chikusa-ku, Nagoya 464-8601, Japan.
Abstract
Extracellular vesicles (EVs) are nanoscale lipid bilayer vesicles released by almost all cell types and can be found in biological fluids, such as blood and urine. EVs play an important role in various physiological and pathological processes via cell-cell communication, highlighting their potential applications as diagnostic markers for diseases and therapeutic drug delivery carriers. Although various methods have been developed for the isolation of EVs from biological fluids, most of them exhibit major limitations, including low purity, long processing times, and high cost. In this study, we developed a size-exclusion chromatography (SEC) column device using hydrophilic porous silica gel (PSG). Owing to the resistance to pressure of the device, a rapid system for EV isolation was developed by connecting it to a flash liquid chromatography system furnished with a UV detector and a fraction collector. This system can be used for the real-time monitoring of eluted EVs by UV absorption without further analysis and separation of high-purity EVs from urine samples with high durability, reusability, and reproducibility. In addition, there were no significant differences between the PSG column- and conventional SEC column-isolated EVs in the proteome profiles and cellular uptake activities, suggesting the good quality of the EVs isolated by the PSG column. These findings suggest that the PSG column device offers an effective and rapid method for the isolation of intact EVs from biological fluids.
Extracellular vesicles (EVs) are nanoscale lipid bilayer vesicles released by almost all cell types and can be found in biological fluids, such as blood and urine. EVs play an important role in various physiological and pathological processes via cell-cell communication, highlighting their potential applications as diagnostic markers for diseases and therapeutic drug delivery carriers. Although various methods have been developed for the isolation of EVs from biological fluids, most of them exhibit major limitations, including low purity, long processing times, and high cost. In this study, we developed a size-exclusion chromatography (SEC) column device using hydrophilic porous silica gel (PSG). Owing to the resistance to pressure of the device, a rapid system for EV isolation was developed by connecting it to a flash liquid chromatography system furnished with a UV detector and a fraction collector. This system can be used for the real-time monitoring of eluted EVs by UV absorption without further analysis and separation of high-purity EVs from urine samples with high durability, reusability, and reproducibility. In addition, there were no significant differences between the PSG column- and conventional SEC column-isolated EVs in the proteome profiles and cellular uptake activities, suggesting the good quality of the EVs isolated by the PSG column. These findings suggest that the PSG column device offers an effective and rapid method for the isolation of intact EVs from biological fluids.
Extracellular vesicles (EVs)
are nanoscale lipid bilayer-enclosed structures released by almost
all types of cells and tissues; they can be found in various biological
fluids, such as blood, urine, milk, cerebrospinal fluid, and saliva,
under both physiological and pathological conditions.[1,2] EVs carry various biological molecules, such as proteins, mRNA,
miRNA, and lipids, which reflect the conditions of the host cells
and participate in a variety of biological processes through cell–cell
communication.[2−5] Recently, EVs in body fluids have emerged as promising sources of
biomarkers for specific diseases and treatment outcomes.[1,2,6,7] In
addition, EVs can be potentially used as specific and stable drug
delivery systems.[7−10]As the clinical exploitation of EVs has increased, various
strategies
have been established to isolate EVs from biological fluids.[2,11,12] Currently, the gold standard
for the isolation of EVs is ultracentrifugation (UC),[13] which is carried out in multiple steps at different centrifugation
speeds.[14,15] However, the UC-based isolation method is
time-consuming, labor-intensive, and requires expensive equipment.
In addition, UC occasionally produces EVs in lower yield and purity
owing to contamination with co-precipitated protein aggregates.[16]In addition to UC, several other methods,
such as isolation through
membrane affinity spin columns and precipitation using polymeric compounds,
have been developed. Although these isolation methods are easy to
use, processing multiple samples using these methods is expensive
and results in low purity.[12] Recently,
using size-exclusion chromatography (SEC) for the isolation of EVs
has attracted increasing interest.[12,17,18] SEC has been used to separate EVs from various biological
fluids, such as urine, plasma, seminal plasma, saliva, and milk.[19−23] Several studies have demonstrated that SEC-based purification methods
have more advantages than conventional methods such as UC and precipitation
with chemicals. SEC exhibits a lower risk of EV aggregation and the
formation of protein complexes than UC and precipitation methods.[12] Because of this and the fact that phosphate-buffered
saline (PBS) is commonly used as a mobile phase, SEC can yield EVs
with better functionality and integrity.[24,25] However, after fractionation by SEC, western blot analysis or enzyme-linked
immunosorbent assay (ELISA) with antibodies against EV markers is
needed to confirm the EV elution position, thereby consuming more
processing time. Moreover, almost all commercially available SEC devices
for EV isolation are fabricated using acrylamide or agarose gels,
which are unsuitable for fast flow rate and repeated usage because
of their low pressure-resistance and low rigidity. Therefore, it is
difficult to adopt the conventional SEC devices to a liquid chromatography
system for repeated continuous isolation.In this study, we
prepared hydrophilic porous silica gel (PSG)
and developed a PSG-based SEC column device for EV purification. Because
PSG can withstand high pressures and flow rates, the PSG column can
be connected to a flash liquid chromatography (FLC) system with a
UV detector and a fraction collector for rapid purification of EVs
from biological fluids. The yields and purities of the EVs isolated
by the PSG column were compared with those of EVs purified using other
methods. Moreover, the effectiveness of the PSG column was evaluated
using proteomic and cellular uptake analyses.
Experimental Section
Synthesis
of the Hydrophilic PSG Particles
Industrially
produced porous silica particles were used as inorganic porous particles.
The porous silica particles had an average particle diameter of 35.5
μm, average pore diameter of 72.7 nm, pore volume of 1.73 mL/g,
and specific surface area of 94 m2/g. The average particle
diameter of the particles was measured by laser light scattering using
an LA-950 V2 instrument (HORIBA, Ltd., Kyoto, Japan). The average
pore diameter, pore volume, and specific surface area were measured
using the mercury intrusion technique with an Autopore IV9510 (Shimadzu
Corporation, Kyoto, Japan). To 50 g of the porous silica particles,
400 mL of toluene, 35.2 mL of diisopropylethylamine, and 45.7 mL of
coupling agent γ-glycidoxypropyltrimethoxysilane were added,
and the mixture was refluxed at 100 °C for 4.5 h. After cooling,
the mixture was filtered, washed with 1000 mL of toluene, 500 mL of
tetrahydrofuran (THF), and 660 mL of methanol and dried at 70 °C
overnight to obtain porous silica particles with an epoxy group-containing
adhesive layer. Twenty-one grams of the coupling agent-treated porous
silica particles were immersed in 84 mL of a hydrochloric acid solution
for 1 day at room temperature. The particles were then filtered, washed
with 630 mL of distilled water and 630 mL of methanol, and dried at
70 °C for 1 day to obtain porous silica particles with a diolated
adhesive layer. To increase hydrophilicity, 10 g of the obtained diolated
porous silica particles was mixed with 0.86 g of Denacol EX-521 (polyglycerol
polyglycidyl ether; Nagase ChemteX Corporation, Osaka, Japan) in 17.4
mL of methanol for 30 min at room temperature; then, the mixture was
dried at 70 °C for 1 day. This mixture was mixed with 40 mL of
decane and 28.3 mg of tris(pentafluorophenyl) borane and stirred at
100 °C for 4 h. After cooling, the mixture was filtered, washed
with 200 mL of toluene, 200 mL of methanol, 200 mL of the aqueous
hydrochloric acid solution, and 200 mL of methanol, and dried at 70
°C for 1 day to obtain porous silica particles with a hydrophilic
organic layer and an adherent layer on the surface. The obtained PSG
was dry-packed into a column (inner diameter: 16 mm, length: 60 mm).
Scanning Electron Microscopy
PSG particles were immobilized
on the sample holder by a carbon double-sided tape and coated with
platinum using plasma chemical vapor deposition to prevent charging
effects during imaging. Scanning electron microscopy (SEM) images
of PSG were obtained using an S-4800 apparatus (Hitachi High-Tech
Co., Ltd., Tokyo, Japan) operating at 2 kV.
Preparation of Rat Urine
for EV Isolation
All urine
samples were directly collected into 50 mL tubes over 24 h from Wistar
rats and pooled after the addition of a protease inhibitor mixture
(Nacalai Tesque, Kyoto, Japan). The pooled urine was centrifuged at
200 × g for 20 min at room temperature to remove
debris and then stored at −80 °C. After thawing, pooled
urine was centrifuged at 3000 × g for 20 min
at 4 °C. The supernatant was further centrifuged at 17,000 × g for 20 min at 4 °C. The supernatant was saved, and
the pellets were resuspended in PBS supplemented with 200 mg/mL dithiothreitol,
followed by incubation at 37 °C for 10 min. During incubation,
samples were mixed every 2 min and then centrifuged at 17,000 × g for 20 min at 4 °C. The two resulting supernatants
were combined and concentrated by ultrafiltration (Amicon Ultra-15
filter, 100 kDa MWCO, Merck Millipore). The concentrated pooled urine
sample was filtered through a 0.22 μm syringe filter (Merck
Millipore) to remove debris before the isolation of EVs. All animal
protocols were approved by the Animal Experiment Committee of the
Graduate School of Bioagricultural Sciences, Nagoya University.
Isolation of Urinary EVs Using the PSG Column
To isolate
EVs using the PSG column, a flash liquid chromatography (FLC) system
(SYS16078; TOKYO RIKAKIKAI CO., LTD, Tokyo, Japan) was used. The concentrated
pooled urine was separated into 40 fractions using FLC on the PSG
column. The separation was performed in the isocratic mode with PBS
as the mobile phase at a flow rate of 2 mL/min, with UV monitoring
at 280 nm using the attached UV detector.
Tunable Resistive Pulse
Sensing Analysis
The size and
concentration of particles were measured using a qNano (IZON Science
Ltd.) with nanopore NP150 (70–420 nm range). The stretch applied
to the nanopores was 47 mm. The concentration of particles was standardized
with 230 nm carboxylated polystyrene beads at a concentration of 9.2
× 1011 particles/mL.
Immunoblot Analysis
The EV fractions (30 μL)
or proteins (1 μg) were subjected to reducing sodium dodecyl
sulfate-polyacrylamide gel electrophoresis (SDS-PAGE) (11% gel). Immunoblot
analysis of EVs was performed as previously reported.[26] Rabbit anti-CD9 monoclonal antibody (ab92726) and rabbit
anti-Flotillin 1 polyclonal antibody (F1180) were purchased from Abcam
(Cambridge, MA, USA) and Sigma (St Louis, MO, USA), respectively.
Mouse anti-Alix (1A12) monoclonal antibody (sc-53540) and mouse anti-THP
(B-2) monoclonal antibody (sc-271022) were purchased from Santa Cruz
Biotechnology (Santa Cruz, CA, USA).
Statistical Analysis
All quantification experiments
were repeated for at least three different preparations in one experiment.
All data are expressed as mean ± S.D. Statistical significance
was evaluated using unpaired Student’s t-test or, when appropriate,
Tukey’s test.
Results and Discussion
Fabrication and Characterization
of the PSG Device
To establish a rapid and automated strategy
for EV isolation, we
prepared a silica gel-based SEC resin. We first obtained industrially
produced porous silica gels (PSG) with different particle diameters
and pore sizes. Because EVs were not eluted from the silica gel column
(Figure S1A), these PSG were coated with
hydrophilic layers to reduce nonspecific binding to their surfaces
and packed into columns. We evaluated the isolation performance of
these columns. Among the gels we tested, the hydrophilic PSG with
a particle size of ∼36 μm and pore size of ∼73
nm exhibited the best separation performance (Figure S1B, C) and was used for further study. The morphology
of the prepared hydrophilic PSG was studied using SEM. Figure A, B shows the SEM images of
PSG and its high-magnification micrograph, respectively. The images
clearly indicate that the particles exhibit very high sphericity and
porous surfaces, retaining the morphological features of the industrially
produced porous silica particles (average sphere diameter: 35.5 ±
13.3 μm (CV = 0.38); average pore entrance diameter: 72.7 ±
46.2 nm (CV = 0.64); intruded volume: 1.73 ± 0.02 mL/g (CV =
0.01); specific surface area: 94 ± 3.5 m2/g (CV =
0.04)). Elemental analysis showed that the carbon content of the particle
surface was 8.77 ± 0.32% (CV = 0.04). The column used in this
study consisted of a hydrophilic PSG resin bed of approximately 10
mL volume within a plastic column with a diameter of 16 mm and a height
of 60 mm. The column coupled with FLC with a UV detector and a fraction
collector is shown in Figure C.
Figure 1
Fabrication of the PSG column device. (A) SEM image of PSG. (B)
SEM image at high magnification. (C) Images of the PSG column device
(left) and FLC system for EV isolation (right).
Fabrication of the PSG column device. (A) SEM image of PSG. (B)
SEM image at high magnification. (C) Images of the PSG column device
(left) and FLC system for EV isolation (right).
Fractionation of Urine Using the PSG-Based Column
Pooled
rat urine samples were used to investigate the separation efficacy
of the PSG column. The urine sample was precleared by centrifugation
and concentrated by ultrafiltration with a 100 kDa molecular weight
cutoff. After the column was prewashed with methanol and equilibrated
with PBS, the concentrated urine sample (0.5 mL) was eluted using
the PSG column. The eluate was collected into tubes in 40 fractions
(0.5 mL and 15 s each); the collection took 10 min to complete. The
concentration of proteins and the number of EVs in each fraction were
analyzed. Although almost all proteins were eluted in fractions 16–40,
a major peak in the particle number was detected in fractions 11–12
(Figure A, upper panel).
In addition, the EV marker CD9 was detected in fractions 11–12
by immunoblot analysis (Figure B). However, the Tamm-Horsfall protein (THP), which is the
most abundant non-EV protein in urine, was detected in fractions 15–21
(Figure B). These
findings suggest that the PSG column can efficiently separate EVs
from urine samples. Notably, we demonstrated that the EV-containing
fractions were consistent with the peak detected at 280 nm by UV absorption
(Figure A, lower panel).
This device could be used for the real-time monitoring of eluted EVs
by UV absorption without further analysis, such as immunoblotting
and ELISA, as opposed to other SEC-based columns (Figure C). Thus, the PSG device can
be applied to the rapid isolation of EVs from biological fluids. We
also confirmed that the EVs isolated by the PSG column were round
or oval membranous vesicles, using transmission electron microscopy
(TEM) with negative staining (Figure D).
Figure 2
Characterization of the PSG-based SEC isolation of urinary
EVs.
(A) SEC isolation of urinary EVs using the PSG column device. In the
upper panel, the protein concentration and particle number of each
fraction are represented by the closed circle (left axis) and yellow
bar graph (right axis), respectively. The lower panel shows the absorption
at 280 nm. (B) Immunoblot analysis of the fractions using anti-CD9
and THP antibodies. (C) Automated isolation of EVs using UV absorption
at 280 nm. Immunoblot analysis of the isolated EV fraction using the
antibodies against CD9, Alix, Flotillin 1, and THP. An equal volume
of eluent was subjected to immunoblotting. (D) Representative TEM
image of EVs isolated by the PSG column. Scale bar in the image represents
100 nm.
Characterization of the PSG-based SEC isolation of urinary
EVs.
(A) SEC isolation of urinary EVs using the PSG column device. In the
upper panel, the protein concentration and particle number of each
fraction are represented by the closed circle (left axis) and yellow
bar graph (right axis), respectively. The lower panel shows the absorption
at 280 nm. (B) Immunoblot analysis of the fractions using anti-CD9
and THP antibodies. (C) Automated isolation of EVs using UV absorption
at 280 nm. Immunoblot analysis of the isolated EV fraction using the
antibodies against CD9, Alix, Flotillin 1, and THP. An equal volume
of eluent was subjected to immunoblotting. (D) Representative TEM
image of EVs isolated by the PSG column. Scale bar in the image represents
100 nm.
Durability and Reusability
of the PSG Column and the Reproducibility
of Fractionation
To investigate the durability and reusability
of the PSG column and the reproducibility of EV separation, concentrated
rat urine was repeatedly (up to 30 times) eluted using the same column
coupled with the FLC system. As shown in Figure A, similar UV chromatographic profiles were
obtained for all EV fractions isolated using the PSG column. In addition,
immunoblot analysis showed that CD9 was detected in the EV fraction
across all trials (Figure B). These results indicate that the PSG column exhibits high
durability, reusability, and reproducibility for isolating EVs from
urine samples.
Figure 3
Evaluation of the reproducibility of the PSG column during
EV isolation.
(A) Chromatographic peak at 280 nm. (B) Immunoblot analysis of each
EV fraction obtained using the PSG device coupled with FLC.
Evaluation of the reproducibility of the PSG column during
EV isolation.
(A) Chromatographic peak at 280 nm. (B) Immunoblot analysis of each
EV fraction obtained using the PSG device coupled with FLC.
Comparative Evaluation of Urine EVs Isolated
Using Different
Methods
Next, we evaluated the PSG column by comparing it
with other isolation methods (Figure ). EVs were isolated from pooled concentrated rat urine
using four different methods, namely, ultracentrifugation (UC), polyethylene
glycol (PEG)-based precipitation, polyacrylamide gel (PAG)-based SEC
column (EVSecond), and the PSG column. The particle sizes and purities
of the EVs isolated by each method were measured using tunable resistive
pulse sensing analysis. The size distributions of all EV samples,
regardless of the isolation method, were within the expected size
range of 90–150 nm without significant differences (Figure A, B). The ratio
of the number of particles and the protein content (particles/μg
protein) is the purity index of EVs.[27] The
purity index of the EVs isolated by the PSG and PAG columns was significantly
high; however, the EVs precipitated with PEG exhibited the lowest
purity index (Figure C). Similar results were also observed in the immunoblot analysis
of EV markers and THP (Figures D, E, and S2). These results suggest
that the PSG- and PAG-based columns can isolate high-purity EVs. A
possible reason for the low purity of the PEG method could be the
precipitation of EVs with the contamination of aggregated proteins.
However, the high purity of the SEC-based methods could be caused
by the good separation between EVs and urinary proteins. In addition,
we found that the EV recovery rate of the PSG column was 73.5 ±
14.3%, which was similar to that of the PAG column (77.5 ± 11.5%)
(Figure S3).
Figure 4
Schematic representation
of the EV isolation methods and characterization
of the EVs isolated in this study. EVs were isolated from concentrated
pooled rat urine samples using four different methods: ultracentrifugation
(UC), precipitation with PEG, PAG-based SEC column (EVSecond), and
PSG column. EVs isolated using each method were characterized by immunoblot
analysis and tunable resistive pulse sensing analysis. EVs isolated
by PAG-based and PSG-based SECs were characterized by proteomic analysis
and cellular uptake analysis.
Figure 5
Comparative
study of the EVs isolated by ultracentrifugation, PEG
precipitation, the PAG column, and the PSG column. The EVs were isolated
from the same volume of the concentrated pooled rat urine sample by
UC, PEG precipitation, the PAG column, and the PSG column, and then
subjected to tunable resistive pulse sensing analysis. The particle
size distribution (A), modal size (B), and particle/protein ratio
(C) of the isolated EV samples. (D, E) Immunoblot analysis of CD9,
Alix, Flotillin 1, and THP in EVs isolated using each method. Densitometric
analysis of three independent experiments (panel D) is shown in panel
E. The data are presented as the mean ± SD; n = 3 in each group. Differences were analyzed using Tukey’s
test. Different letters on the bars indicate significant differences
(P < 0.05).
Schematic representation
of the EV isolation methods and characterization
of the EVs isolated in this study. EVs were isolated from concentrated
pooled rat urine samples using four different methods: ultracentrifugation
(UC), precipitation with PEG, PAG-based SEC column (EVSecond), and
PSG column. EVs isolated using each method were characterized by immunoblot
analysis and tunable resistive pulse sensing analysis. EVs isolated
by PAG-based and PSG-based SECs were characterized by proteomic analysis
and cellular uptake analysis.Comparative
study of the EVs isolated by ultracentrifugation, PEG
precipitation, the PAG column, and the PSG column. The EVs were isolated
from the same volume of the concentrated pooled rat urine sample by
UC, PEG precipitation, the PAG column, and the PSG column, and then
subjected to tunable resistive pulse sensing analysis. The particle
size distribution (A), modal size (B), and particle/protein ratio
(C) of the isolated EV samples. (D, E) Immunoblot analysis of CD9,
Alix, Flotillin 1, and THP in EVs isolated using each method. Densitometric
analysis of three independent experiments (panel D) is shown in panel
E. The data are presented as the mean ± SD; n = 3 in each group. Differences were analyzed using Tukey’s
test. Different letters on the bars indicate significant differences
(P < 0.05).
Proteomic Comparison of EVs Isolated by the SEC Columns
To evaluate the functionality and integrity of the PSG device-isolated
urinary EVs, we characterized their proteomic profiles using a MS-based
analysis. Proteomic analysis of each EV sample was triplicated. Combining
the triplicates, 1113 proteins and 1118 proteins were identified from
the EVs isolated by the PSG and PAG columns, respectively (Figure A, Table S1). Almost all proteins (993) were common to the EVs
isolated from both the PSG and PAG columns. In addition, we performed
a gene ontology (GO)-based enrichment analysis and found that the
bulk of the proteins identified in the samples isolated by the PSG
and PAG columns corresponded to extracellular vesicles (Figure B).
Figure 6
Comparative study of
the proteomic profiles of the samples isolated
by the PSG and PAG columns. (A) Comparison of the number of identified
proteins in rat urinary EVs isolated by the PSG and PAG columns. (B)
Characterization of the proteome of rat urinary EVs using LC–MS/MS.
GO term enrichment analysis of EV proteome at the cellular component
level was performed using DAVID.
Comparative study of
the proteomic profiles of the samples isolated
by the PSG and PAG columns. (A) Comparison of the number of identified
proteins in rat urinary EVs isolated by the PSG and PAG columns. (B)
Characterization of the proteome of rat urinary EVs using LC–MS/MS.
GO term enrichment analysis of EV proteome at the cellular component
level was performed using DAVID.
Functional Comparison of EVs Isolated by the SEC Columns
Finally, we investigated whether the PSG-based isolation method affects
the biological properties of the EVs. Because uptake by recipient
cells is essential for exhibiting the activity, we examined the cellular
uptake of isolated EVs using the rat renal tubular epithelial cell
line NRK-52E, which was labeled with CellTracker green CMFDA and incubated
with 5 × 108 particles/mL of PKH26-labeled EVs isolated
using the PSG or PAG columns for 18 h. Confocal fluorescence microscopy
revealed comparable uptake by the cells treated with the EV samples
isolated using the PSG and the PAG columns (Figure A). Similar results were observed in the
fluorescence intensity measurements (Figure B). These uptake experiments and the proteomic
analysis (Figure )
indicated that the EVs isolated using the PSG column are similar in
purity and quality to those isolated using a commercial PAG-based
SEC column. The PSG-based device might be more useful than the PAG
column because of its shorter processing time and the possibility
of automation. To further determine the efficacy of the PSG-based
system as a tool for EV isolation, additional studies will have to
be performed to determine whether the isolated EVs can be subjected
to miRNA analysis, which is frequently performed in EV research.
Figure 7
Cellular
uptake of PKH26-labeled EVs isolated by the PSG and PAG
columns. (A) Cellular uptake of EVs was visualized by confocal microscopy.
Scale bars, 20 μm. (B) Relative fluorescence intensity. Data
are presented as the mean ± SD; n = 3 in each
group. Differences were analyzed using Tukey’s test. Different
letters on the bars indicate significant differences (P < 0.05).
Cellular
uptake of PKH26-labeled EVs isolated by the PSG and PAG
columns. (A) Cellular uptake of EVs was visualized by confocal microscopy.
Scale bars, 20 μm. (B) Relative fluorescence intensity. Data
are presented as the mean ± SD; n = 3 in each
group. Differences were analyzed using Tukey’s test. Different
letters on the bars indicate significant differences (P < 0.05).
Conclusions
We
demonstrated the ability of a novel hydrophilic PSG-based column
device for the rapid isolation of EVs. Characterization and quantification
of EVs showed that EVs were successfully separated from urine samples
with optimal yields, functionality, and integrity. Thus, this device
can be potentially used as a powerful tool for the rapid isolation
of EVs from various biological fluids. The PSG device could be extended
to other biological fluids to facilitate future research on the discovery
of diagnostic biomarkers and to elucidate the physiological/pathological
significance of EVs.
Authors: Emma A Mol; Marie-José Goumans; Pieter A Doevendans; Joost P G Sluijter; Pieter Vader Journal: Nanomedicine Date: 2017-03-30 Impact factor: 5.307
Authors: Kseniia Boriachek; Md Nazmul Islam; Andreas Möller; Carlos Salomon; Nam-Trung Nguyen; Md Shahriar A Hossain; Yusuke Yamauchi; Muhammad J A Shiddiky Journal: Small Date: 2017-12-28 Impact factor: 13.281
Authors: Hadi Valadi; Karin Ekström; Apostolos Bossios; Margareta Sjöstrand; James J Lee; Jan O Lötvall Journal: Nat Cell Biol Date: 2007-05-07 Impact factor: 28.824
Authors: Chris Gardiner; Dolores Di Vizio; Susmita Sahoo; Clotilde Théry; Kenneth W Witwer; Marca Wauben; Andrew F Hill Journal: J Extracell Vesicles Date: 2016-10-31
Authors: S Inés Lozano-Ramos; Ioana Bancu; Laura Carreras-Planella; Marta Monguió-Tortajada; Laura Cañas; Javier Juega; Josep Bonet; M Pilar Armengol; Ricardo Lauzurica; Francesc E Borràs Journal: BMC Nephrol Date: 2018-07-31 Impact factor: 2.388
Authors: Shin La Shu; Yunchen Yang; Cheryl L Allen; Edward Hurley; Kaity H Tung; Hans Minderman; Yun Wu; Marc S Ernstoff Journal: J Extracell Vesicles Date: 2019-11-20