Moohyun Kim1,2, Jae Chul Hwang1,2, Sungjin Min3, Young-Geun Park1,2, Suran Kim3, Enji Kim1,2, Hunkyu Seo1,2, Won Gi Chung1,2, Jakyoung Lee1,2, Seung-Woo Cho3,2, Jang-Ung Park1,2,4. 1. Department of Materials Science and Engineering, Yonsei University, Seoul 03722, Republic of Korea. 2. Center for Nanomedicine, Institute for Basic Science (IBS), Yonsei University, Seoul 03722, Republic of Korea. 3. Department of Biotechnology, Yonsei University, Seoul 03722, Republic of Korea. 4. KIURI Institute, Yonsei University, Seoul 03722, Republic of Korea.
Abstract
Herein, we present an unconventional method for multimodal characterization of three-dimensional cardiac organoids. This method can monitor and control the mechanophysiological parameters of organoids within a single device. In this method, local pressure distributions of human-induced pluripotent stem-cell-derived cardiac organoids are visualized spatiotemporally by an active-matrix array of pressure-sensitive transistors. This array is integrated with three-dimensional electrodes formed by the high-resolution printing of liquid metal. These liquid-metal electrodes are inserted inside an organoid to form the intraorganoid interface for simultaneous electrophysiological recording and stimulation. The low mechanical modulus and low impedance of the liquid-metal electrodes are compatible with organoids' soft biological tissue, which enables stable electric pacing at low thresholds. In contrast to conventional electrophysiological methods, this measurement of a cardiac organoid's beating pressures enabled simultaneous treatment of electrical therapeutics using a single device without any interference between the pressure signals and electrical pulses from pacing electrodes, even in wet organoid conditions.
Herein, we present an unconventional method for multimodal characterization of three-dimensional cardiac organoids. This method can monitor and control the mechanophysiological parameters of organoids within a single device. In this method, local pressure distributions of human-induced pluripotent stem-cell-derived cardiac organoids are visualized spatiotemporally by an active-matrix array of pressure-sensitive transistors. This array is integrated with three-dimensional electrodes formed by the high-resolution printing of liquid metal. These liquid-metal electrodes are inserted inside an organoid to form the intraorganoid interface for simultaneous electrophysiological recording and stimulation. The low mechanical modulus and low impedance of the liquid-metal electrodes are compatible with organoids' soft biological tissue, which enables stable electric pacing at low thresholds. In contrast to conventional electrophysiological methods, this measurement of a cardiac organoid's beating pressures enabled simultaneous treatment of electrical therapeutics using a single device without any interference between the pressure signals and electrical pulses from pacing electrodes, even in wet organoid conditions.
Organoids are simplified and
miniaturized in vitro model systems of organs and
they recently have attracted a lot of attention because of their applications
in tissue development, disease modeling, and drug screening.[1] Despite rigorous advances in the culturing of
physiologically optimized organoids in the relevant field, there is
limited applicability due to inadequate tools for practical clinical
translation. The integration of engineering innovation into complex
organoid systems for more in-depth control and recording of mechanophysiological
parameters has proved to be a problematic task.[2−4] The high culturing
variability of organoids can be managed by engineering a potential
design for physical cues during the self-organization processes.[5−8] Modern organs-on-a-chip devices have established systemic characterization
and controlled synthesis of target tissue elements, which has improved
the replicability of functional readouts significantly and extended
their lifespans.[9−11] However, these devices restrict the natural self-organizing
process of growing cell aggregates. The expression of the physiological
properties is hampered when using such devices with those constraining
conditions. These cell lines fundamentally follow the mechanophysiological
properties set out by the architectural design of the respective organs-on-a-chip
devices. Hence, when paired with such devices, the physiological relevance
of these organoids ultimately is reduced.[12−14]For the
translation of the mechanophysiological parameters of cardiac
organoids, the physiological parameters associated with the beating
motions of the organoid need to be captured. Current analytic models
of cardiac organoids have been limited to the electrophysiological,
genomic, and optical characterization of their physiologic functionality.[15] It is difficult for most commonly used optical
monitoring methods to provide full, viable information about the morphological
property of the organoids.[16,17] A design that controls
the explicit locations of the organoids without restricting their
natural structure during electrophysiological characterization is
an optimal setup that resolves the limitations of translatability.Herein, we present an unconventional method for the multimodal
characterization and stimulation of human-induced pluripotent stem
cell (hiPSC)-derived cardiac organoids. Our device establishes unrestricted
development of cardiac organoids while comprehensively monitoring
the readout of their functional activity. Our device can record and
control essential cardiac functions that are indispensable for physiological
analysis without the reductionist engineering approach presented in
modern organs-on-a-chip devices. Unlike organs-on-a-chip devices,
which comprise premeditated properties of the in vivo microenvironment, our device unrestrictedly monitors the complex
cellular architecture of cardiac organoids. The variation of the pressure
of the organoid’s beating motions that were recorded using
the pressure-sensitive transistors can correspond to electrocardiogram
(ECG) traces and can generate negligible artifacts in the sensing
signal even when an electrical impulse was delivered for cardiac organoid
stimulation. Table S1 compares and differentiates
our device from previous works on pressure sensors for biomedical
applications.
Integration of Pressure-Sensitive Transistors and 3D Liquid-Metal
Electrodes
Figure a shows
the schematic illustrations of the sensing mechanism of our multimodal
device, which includes the pressure-sensitive transistor array, and
the integrated 3D liquid-metal electrodes. When a cardiac organoid
is located on this device, the 3D electrodes can be inserted naturally
inside the cardiac organoid. Simultaneously, the active-matrix array
of pressure-sensitive field-effect transistors (FET) can monitor the
beating motions of this cardiac organoid mechanophysiologically by
detecting the spatiotemporal change in compressive pressures. In contrast
to previous surface-type electrodes, this 3D electrode can stimulate
and record the interior area of an organoid directly by forming the
intraorganoid interface. In addition, the 3D electrodes serve as physical
cues that support the growth of the cardiac organoids. For example, Figure b shows immunofluorescence
images of a hiPSC-derived cardiac organoid on 2D (flat) and 3D electrodes.
We observed that cardiac cells cluster around these 3D electrodes
during their early initiation and conformally surround the electrodes
during their maturation process.
Figure 1
Multimodal sensing device using directly
printed 3D electrodes
and pressure-sensitive transistor arrays. (a) Schematic illustration
of the sensing mechanism of the multimodal device (top). Illustration
of the integrated 3D liquid-metal electrodes (bottom left) and pressure-sensitive
transistor array (bottom right). (b) Immunofluorescence images of
hiPSC-derived cardiac organoid that is grown separately on 2D (flat)
and 3D electrodes. Scale bar, 100 μm. (c) Plot of the maximum
height of the 3D electrode versus the speed of vertical elevation.
Inset: Schematic illustration of the electrode printing system. (d)
Plot of the diameter of the electrode versus the diameter of the nozzle.
Inset: Schematic illustration of the 3D electrode printing. Error
bars in (d) indicate the standard deviations. (e) Photograph of the
3D electrodes after high-resolution direct printing (top). Scale bar,
600 μm. The SEM image of the 3D electrode tip (indicated by
the red circle, bottom left). Scale bar, 5 μm. The photograph
of the directly printed 3D electrodes with PDMS microwell for cardiac
organoid culturing (bottom right). Scale bar, 150 μm.
Multimodal sensing device using directly
printed 3D electrodes
and pressure-sensitive transistor arrays. (a) Schematic illustration
of the sensing mechanism of the multimodal device (top). Illustration
of the integrated 3D liquid-metal electrodes (bottom left) and pressure-sensitive
transistor array (bottom right). (b) Immunofluorescence images of
hiPSC-derived cardiac organoid that is grown separately on 2D (flat)
and 3D electrodes. Scale bar, 100 μm. (c) Plot of the maximum
height of the 3D electrode versus the speed of vertical elevation.
Inset: Schematic illustration of the electrode printing system. (d)
Plot of the diameter of the electrode versus the diameter of the nozzle.
Inset: Schematic illustration of the 3D electrode printing. Error
bars in (d) indicate the standard deviations. (e) Photograph of the
3D electrodes after high-resolution direct printing (top). Scale bar,
600 μm. The SEM image of the 3D electrode tip (indicated by
the red circle, bottom left). Scale bar, 5 μm. The photograph
of the directly printed 3D electrodes with PDMS microwell for cardiac
organoid culturing (bottom right). Scale bar, 150 μm.The use of liquid metals in deformable electronic
devices has been
prominent due to their good processability at ambient conditions and
superb electrical conductivities.[18−23] Their intrinsically low modulus and superb stretchability also can
be advantageous for their use as biointerfacing electrodes, as conventional
rigid solid metals can damage cellular interfaces because of the modulus
mismatch between metallic solids and biological cells.[24] We used eutectic gallium–indium (EGaIn)
because of its low toxicity and negligible volatility. Figure S1 compares the Young’s modulus
of EGaIn with conventional solid-phase conductive materials. The modulus
of EGaIn is about 1000 times lower than the values of solid conductors
and it is comparable to the modulus of human cardiac tissues.[25] Upon exposure to air, EGaIn instantaneously
forms a thin solid layer (∼1 nm) of gallium oxide on its surfaces
under atmospheric oxygen levels. This oxide skin is thin enough to
avoid damaging the cellular interfaces substantially and solid enough
to maintain its 3D shape against gravity and surface tension. The
printing setup consists of a printing nozzle connected to an ink reservoir
that is connected to an external pneumatic pressure controller (see Supporting Information). By adjusting the stage
from a fixed position to the direction of the z-axis,
the 3D electrode can be printed (Video S1). Figure c shows
the speed of the vertical elevation of a nozzle that presets the height
of the printed 3D electrode. Also, the diameter of 3D electrode can
be determined by the inner diameter of a nozzle (Figure d). For example, 3D electrodes
with a diameter of 12 μm were printed using our nozzle of 10
μm in diameter (Figure S2), and the
electrode diameter did not significantly degrade the electrode performance
(Figure S3). Figure S4 shows the printing duration required to form a single 3D
electrode, and 10 s typically were required to print a single 3D electrode
with a height of 250 μm. Figure e shows optical micrographs of the 3D electrodes printed
on a thin flexible polyimide (PI) film. The bottom-left inset shows
a scanning electron microscope (SEM) image of the conical tip part
of an EGaIn electrode. As shown in the bottom-right inset, a polydimethylsiloxane
(PDMS) layer was perforated with a hole using a laser ablation machine
as a microwell. Here, 3D electrodes were placed within the vacant
hole of this PDMS microwell to promote the localization of the cell
aggregate around the 3D electrodes during organogenesis.
Performance Characterization of the Multimodal Sensing Device
The schematic layouts of this multimodal characterization device
for a cardiac organoid are shown in Figure a. This device is composed of two main components:
(i) the bottom part comprises an array of pressure-sensitive FETs
(as pressure sensors) for detecting the mechanophysiological compressive
pressures, and (ii) the top part comprises 3D electrodes for the recording
of electrophysiological signals and delivering of electrical stimulation.
The image of the device and its relevant components are shown in Figure S5. Figure S6 illustrates the overall fabrication process of this device. We used
an ultrathin, single-crystalline Si membrane as the FET channel by
etching the top Si layer of a silicon-on-insulator wafer (see Supporting Information for detailed fabrication
steps). For the elastomer layer, the embedding of glycerol microdroplets
inside PDMS can drastically lower the value of the modulus of the
elastomeric layer, which leads to large elastic deformation even at
low compressive pressures.[26] Then, hollow
holes were produced inside the elastomer via laser ablation. Figure S7 shows the PI film with the gate electrodes
and Pt interconnects. The 3D electrodes were printed directly onto
the open area of the Pt interconnects.
Figure 2
Characterization of the
multimodal sensing device for cardiac organoid
detection. (a) Schematic layouts of the multimodal sensor that is
composed of an active-matrix pressure-sensitive transistor array with
integrated soft liquid-metal electrodes. (b) Optical micrograph of
the two adjacent transistors. S and D denote the source electrode
and drain electrode, respectively. Scale bar, 20 μm. (c) Optical
micrograph after the gate electrode (denoted as G) assembly. Scale
bar, 50 μm. (d) Representative transfer characteristics of Si
FET (VD = 5 V) of the device at ambient
conditions. (e) Output characteristic of an air-dielectric FET (VG = 0 to 40 V, 10 V step) at ambient conditions.
(f) Real-time measurements of normalized ID changes for the applied compressive pressures during electrical
stimulation at different electrical fields (VG = 20 V, VD = 5 V). (g) Plot of
the relative changes in ID versus applied
compressive pressure to define pressure sensitivity (denoted as S).
Error bars indicate the standard deviations. (h) Plot of the pressure
sensitivity versus the different magnitudes of the electric field.
(i) Impedance of the 3D EGaIn electrode (height, 250 μm; diameter,
60 μm; red line) and flat electrode (black line) over the frequency
range from 10 Hz to 10 MHz. (j) Cyclic voltammetry plot of 3D EGaIn
electrode (height, 250 μm; diameter, 60 μm; red line)
and flat electrode (black line) over a range from −0.6 to 0.8
V at a sweep rate of 50 mV s–1.
Characterization of the
multimodal sensing device for cardiac organoid
detection. (a) Schematic layouts of the multimodal sensor that is
composed of an active-matrix pressure-sensitive transistor array with
integrated soft liquid-metal electrodes. (b) Optical micrograph of
the two adjacent transistors. S and D denote the source electrode
and drain electrode, respectively. Scale bar, 20 μm. (c) Optical
micrograph after the gate electrode (denoted as G) assembly. Scale
bar, 50 μm. (d) Representative transfer characteristics of Si
FET (VD = 5 V) of the device at ambient
conditions. (e) Output characteristic of an air-dielectric FET (VG = 0 to 40 V, 10 V step) at ambient conditions.
(f) Real-time measurements of normalized ID changes for the applied compressive pressures during electrical
stimulation at different electrical fields (VG = 20 V, VD = 5 V). (g) Plot of
the relative changes in ID versus applied
compressive pressure to define pressure sensitivity (denoted as S).
Error bars indicate the standard deviations. (h) Plot of the pressure
sensitivity versus the different magnitudes of the electric field.
(i) Impedance of the 3D EGaIn electrode (height, 250 μm; diameter,
60 μm; red line) and flat electrode (black line) over the frequency
range from 10 Hz to 10 MHz. (j) Cyclic voltammetry plot of 3D EGaIn
electrode (height, 250 μm; diameter, 60 μm; red line)
and flat electrode (black line) over a range from −0.6 to 0.8
V at a sweep rate of 50 mV s–1.For the spatiotemporal pressure mapping, a 10 ×
10 active-matrix
array of pressure-sensitive air-dielectric FETs was fabricated. On
the basis of the average size of the cardiac organoid, the total size
of the pressure-sensing region was optimized to 500 × 500 μm.
Also, the channel size of the individual FET was designed as 10 ×
10 μm, and the total distance between adjacent FETs (i.e., pixel
resolution) was 50 μm. This active-matrix circuitry consists
of scan and data lines in which a targeted pressure sensor (or pixel)
was designed to operate separately to give direct electrical responses,
according to the selection of the combination of rows and columns.[27−30]Figure b shows an
optical micrograph of the Si channel with the patterned source/drain
electrodes. Accordingly, the assembly of the gate electrode results
in the formation of the air-dielectric FETs (Figure c). Figure d,e plots the representative transfer and output curves
of this transistor at ambient conditions, respectively. The on/off
ratio (Ion/Ioff) and threshold voltage (Vth) were 105 and 10.5 V, respectively. As shown in Figure S8, the statistical measurement data fit the Gaussian
distribution profile by presenting outstanding homogeneity of sensitivity
(average sensitivity: 0.35 ± 0.01 kPa–1) in
all pixels. The clean interface between the Si channel and the air
dielectric can exhibit this outstanding homogeneity at ambient conditions.[31−33] In particular, the air-dielectric structure can result in negligible
hysteresis during the operation of the FET.[34] Our air-dielectric FET exhibited a response time of 48 ms and a
recovery time of 42 ms (Figure S9).Figure f shows
the real-time detection of the relative change in ID [ΔID/I0 (%)], where ID is the drain
current, I0 is the current at zero Pascal,
and ΔID= I – I0, which denotes the variation of the ID during stepwise pressure loading. The sensitivity (S) of the transistor was obtained by plotting the relative
change in ID concerning the applied pressure
(Figure g) and was
expressed as [ΔID/I0 (%)]/ΔP, where ΔP denotes the applied pressure. ΔID/I0 linearly increased within
the pressure range from 1 to 5 kPa, where the sensitivity was approximately
calculated as ∼0.35 kPa–1. In addition, the ΔID/I0 values
measured under desired pressure loading showed negligible variation
when an electrical field was applied simultaneously to the 3D electrodes
during pressure sensing (Figure f). As shown in Figure h, the application of an electric field did not change
the S value.The significance of the 3D electrodes
was validated by using electrochemical
impedance spectroscopy (EIS) and cyclic voltammetry (CV) analyses
to compare the impedance of the 3D EGaIn electrode (height: 250 μm)
with a flat surface Pt electrode using a potentiostat. The areal dimensions
of both electrodes, i.e., 3D EGaIn and flat Pt, were identical, but
the 3D electrode exhibited 250 μm in height. Both electrodes
were submerged in a phosphate-buffered saline (PBS) solution with
a reference electrode (Ag/AgCl) and a counter electrode (Pt sheet)
at ambient conditions (pH 7.4, 25 °C). As plotted in Figure i, the 3D electrode
and flat electrode exhibited impedance values (at 103 Hz)
of 10.5 and 20.7 kΩ, respectively. The charge storage characteristic
was tested via CV analysis conducted at a scan rate of 50 mV/s with
a potential limit range from −0.6 to 0.8 V (Figure j). The 3D electrode showed
an outstanding charge storage characteristic, which was notable in
the large area of the enclosed CV curve. Overall, the 3D electrodes
had lower impedance and a better charge storage characteristic, which
indicates efficient performance for electrode-to-cell signal transfer.
Formation of Human Cardiac Organoids Using Cardiomyocytes Differentiated
from hiPSCs
The
shape and duration of the action potentials are incoherent
between animal and human heart. Hence, animal modeling is often inaccurate
in translation to human clinical setting in cardiology research.[35,36] Cardiomyocyte differentiation of hiPSCs has been employed to overcome
these limitations and to recapitulate action potentials and functional
structures of human cardiomyocytes. It is more efficient to obtain
cardiomyocytes from hiPSCs than other sources since human fetal cardiac
tissues are difficult to obtain, and the ability of adult stem cells
to proliferate and differentiate into functional cardiomyocytes is
known to be limited.[37] More importantly,
when hiPSC-derived cardiomyocytes are cultivated in a 3D condition,
their electrophysiological properties are further enhanced. Their
phenotypic characteristics become closer to that of actual human heart
tissue as the 3D structural environment increases the cellular interactions
and enhances the differentiation and maturation of the cardiomyocytes.[38]hiPSCs were seeded on a 6-well plate to
differentiate hiPSCs into
cardiomyocytes, and cardiomyocyte differentiation was induced according
to the protocol established by Hoang et al. (Figure a).[39] Four days
after hiPSC seeding, the differentiation process was initiated by
replacing the culture medium with a differentiation medium (day 0).
The differentiated cells with cardiomyocyte-like morphology appeared
over culture time up to 16 days (Figure b). The immunofluorescence staining to check
cardiomyocyte differentiation revealed that three cardiac markers,
α-actinin, cardiac troponin T (cTnT), and cardiac troponin I
(cTnI), were highly expressed in the differentiated cardiomyocytes
(Figure c). Then,
cardiac organoids were generated using hiPSC-derived cardiomyocytes
that had started beating on day 17 to day 21. The cardiac organoid
development in the specially designed PDMS microwell was monitored
over culture time (Figure d). We confirmed the high level of cardiac marker (cTnT) expression
in human cardiac organoids at day 21 (Figure e). Moreover, human cardiac organoids showed
autonomous beating motion in the microwell (Video
S2). Additionally, a live/dead staining assay conducted 3 days
after the generation of cardiac organoids with the 3D electrodes showed
good biocompatibility (Figure S10).
Figure 3
Formation of
human cardiac organoids using human-induced pluripotent
stem cell (hiPSC). (a) Timeline and culture conditions for cardiomyocyte
differentiation from hiPSC. (b) Light microscopic observation of differentiated
cardiomyocytes over culture time. Scale bar, 300 μm. (c) Immunofluorescent
images of α-actinin, cardiac troponin T (cTnT), cardiac troponin
I (cTnl), and F-actin in hiPSC-derived cardiomyocytes. DAPI was used
for nucleus staining. Scale bars, 150 μm. (d) Light microscopic
observation of 3D human cardiac organoids generated with differentiated
cardiomyocytes over culture time. Scale bar, 300 μm. (e) Immunofluorescent
image of cTnT and F-actin in human cardiac organoid (left; scale bar,
200 μm) and high-magnified image of cTnT in human cardiac organoid
(right; scale bar, 30 μm). DAPI was used for nucleus staining.
Formation of
human cardiac organoids using human-induced pluripotent
stem cell (hiPSC). (a) Timeline and culture conditions for cardiomyocyte
differentiation from hiPSC. (b) Light microscopic observation of differentiated
cardiomyocytes over culture time. Scale bar, 300 μm. (c) Immunofluorescent
images of α-actinin, cardiac troponin T (cTnT), cardiac troponin
I (cTnl), and F-actin in hiPSC-derived cardiomyocytes. DAPI was used
for nucleus staining. Scale bars, 150 μm. (d) Light microscopic
observation of 3D human cardiac organoids generated with differentiated
cardiomyocytes over culture time. Scale bar, 300 μm. (e) Immunofluorescent
image of cTnT and F-actin in human cardiac organoid (left; scale bar,
200 μm) and high-magnified image of cTnT in human cardiac organoid
(right; scale bar, 30 μm). DAPI was used for nucleus staining.
Multimodal Characterization of the Cardiac Organoids
Since the average size of the cardiac organoids was no larger than
400 μm, the organoid spanned over multiple FET, which allowed
detection of spatiotemporal changes in local pressure distribution
during its beating. As shown in Figure a, our device simultaneously monitored the beating
activity of a cardiac organoid by ECG (blue), pressure (red), and
calcium imaging (black). The ECG signal (unit: mV) was recorded using
the 3D electrodes and an electrophysiological measurement system that
was comprised of a multielectrode array and a data processor with
a real-time controller. The measurement of beating pressure was obtained
by using automated instrumentation with a custom-made jig (Figure S11) connected to separate modules using
two sourcemeters, a system switch, and a relay card. The readouts
of each FET were converted to corresponding compressive pressure values.
The organoid was stained using Fluo-4 AM for calcium imaging. The
normalized fluorescence intensity at a specific region of interest
(F/F0) of the calcium flux was plotted
against time. All modes of characterization showed a consistent beating
frequency of 12.1 beats per minute (bpm) with a consecutive beating
interval of 4.94 s. Figure b presents the relative changes in the three characterization
data during a single beat (indicated as the green dashed box in Figure a). Figure c shows pseudocolor fluorescence
images of this organoid, which represents the changes of the calcium
indicator at the time phases indicated in Figure b. These images indicated the depolarization
(top) and resting state (bottom) of the organoid. The ECG signal recorded
using the 3D electrodes did not significantly deteriorate for 7 days
(Figure S12). Figure d visualizes the spatiotemporal changes in
the beating pressure distribution as 2D color gradation contour plots. Video S3 presents the synchronization of this
spatiotemporal pressure mapping with the calcium imaging during beating
motions.
Figure 4
Multimodal characterization of cardiac organoids. (a) Real-time
characterization plot of the ECG amplitude (blue), compressive pressure
(red), and normalized fluorescence intensity (black) of the hiPSC-derived
cardiac organoid by the multimodal characterization device. (b) Magnified
plot of multimodal characterization data versus time which defines
one beating cycle of the cardiac organoid, denoted as the green dotted
box in (a). (c) Pseudocolor fluorescence images of the cardiac organoid
at the instance of depolarization (top), labeled as time frame 1 in
(b) and 3.8 s afterward (bottom) labeled as time frame 2 in (b). Scale
bar, 100 μm. (d) Time-lapsed spatiotemporal color maps of compressive
pressure distribution of the hiPSC-derived cardiac organoid to capture
the instances of movement during a single beating. Each motion frame
of the cardiac organoid was taken every 174 ms.
Multimodal characterization of cardiac organoids. (a) Real-time
characterization plot of the ECG amplitude (blue), compressive pressure
(red), and normalized fluorescence intensity (black) of the hiPSC-derived
cardiac organoid by the multimodal characterization device. (b) Magnified
plot of multimodal characterization data versus time which defines
one beating cycle of the cardiac organoid, denoted as the green dotted
box in (a). (c) Pseudocolor fluorescence images of the cardiac organoid
at the instance of depolarization (top), labeled as time frame 1 in
(b) and 3.8 s afterward (bottom) labeled as time frame 2 in (b). Scale
bar, 100 μm. (d) Time-lapsed spatiotemporal color maps of compressive
pressure distribution of the hiPSC-derived cardiac organoid to capture
the instances of movement during a single beating. Each motion frame
of the cardiac organoid was taken every 174 ms.
Electrical Stimulation of the Cardiac Organoids
Unlike
previous studies that measured electrophysiological signals
of an organoid by attaching thin, flexible electrodes to the outer
surfaces of the organoid (extraorganoid sensing),[5,6,40,41] the 3D electrodes
of our device were embedded inside a cardiac organoid for the intraorganoid
electrophysiological recording. Modern methods of confinement are
microfluidic devices or arrays of microwells, which can diminish the
quality of physiological signals or are ill-equipped to make high-quality
characterizations.[42−44] In contrast, our device was advantageous for localizing
the organoid development toward the sensing site without any elements
of constraint that potentially diminish the significance of the readouts.
These 3D electrodes do not force organoids into a final specific shape;
rather, they naturally guide the formation of the shape of the hiPSC-derived
cardiac organoids during their growth. The SEM image of the cardiac
organoid developed on these 3D electrodes is shown in Figure a. The cross section schematic
inset in Figure b
(left) illustrates the intraorganoid interface between the cardiac
organoid and the 3D electrodes. In Figure S13, the immunofluorescence images of the cryosectioned cardiac organoid
showed the positional traces of the 3D electrodes in the cardiac organoid,
which represent the intraorganoid interfaces with no signs of bending
or displacement. Figure b shows the immunofluorescent images of the organoid with the 3D
electrodes (middle inset) and close-up of the surface of the organoid
(right).
Figure 5
Simultaneous electrical stimulation of cardiac organoids with multimodal
sensing. (a) Colorized SEM image of the hiPSC-derived cardiac organoid
developed on the 3D liquid-metal electrodes. Scale bar, 100 μm.
(b) Cross section schematics of the intraorganoid interface between
the cardiac organoid and the 3D electrodes (left). The immunofluorescence
images of the matured cardiac organoid with the 3D electrodes (middle
inset) and close-up of the surface of the hiPSC-derived cardiac organoid
(right). Scale bar, 200 and 20 μm, respectively. (c) Real-time
characterization plot of the ECG amplitude (blue), compressive pressure
(red), and normalized fluorescence intensity (black) of the hiPSC-derived
cardiac organoid before and during electrical stimulation (gray dotted
line, 0.8 V/mm). (d) Magnified single ECG signal denoted as the black
dotted box indicated in (c). (e) Electrical circuit of cardiac organoid
with 3D electrodes for electrical stimulation and characterization.
The C, RP, and RO represent the capacitance of the organoid,
the resistance of the probe, and the resistance of the organoid, respectively.
Inset: Aerial schematics of the cardiac organoid. (f) Finite element
analysis (FEA) simulation of the electric field for different cases
of electrodes for the optimization of the organoid stimulation. Case
1 and 2 define the electric field simulation of the flat electrode
at the top and bottom position, respectively. Case 3 defines the electric
field simulation of the 3D electrodes located at the bottom base.
Simultaneous electrical stimulation of cardiac organoids with multimodal
sensing. (a) Colorized SEM image of the hiPSC-derived cardiac organoid
developed on the 3D liquid-metal electrodes. Scale bar, 100 μm.
(b) Cross section schematics of the intraorganoid interface between
the cardiac organoid and the 3D electrodes (left). The immunofluorescence
images of the matured cardiac organoid with the 3D electrodes (middle
inset) and close-up of the surface of the hiPSC-derived cardiac organoid
(right). Scale bar, 200 and 20 μm, respectively. (c) Real-time
characterization plot of the ECG amplitude (blue), compressive pressure
(red), and normalized fluorescence intensity (black) of the hiPSC-derived
cardiac organoid before and during electrical stimulation (gray dotted
line, 0.8 V/mm). (d) Magnified single ECG signal denoted as the black
dotted box indicated in (c). (e) Electrical circuit of cardiac organoid
with 3D electrodes for electrical stimulation and characterization.
The C, RP, and RO represent the capacitance of the organoid,
the resistance of the probe, and the resistance of the organoid, respectively.
Inset: Aerial schematics of the cardiac organoid. (f) Finite element
analysis (FEA) simulation of the electric field for different cases
of electrodes for the optimization of the organoid stimulation. Case
1 and 2 define the electric field simulation of the flat electrode
at the top and bottom position, respectively. Case 3 defines the electric
field simulation of the 3D electrodes located at the bottom base.For the modulation of the organoid’s beating,
it was subjected
to an electrical stimulation pulse of 0.8 V/mm (pulse width, 1 ms;
frequency, 1 Hz) using an externally connected pulse generator. Figure c shows the simultaneous
multimodal characterizations (ECG, pressure, and calcium flux) of
the organoid before and after electrical stimulation (gray-dashed
box). A single ECG peak before applying the electrical impulses (black-dashed
box in Figure c) is
shown in Figure d.
As seen, the depolarization of a cardiac organoid was indicated by
a distinct P wave in the PQRS complex of the ECG signal. However,
during the electric pacing, the electric pulses significantly disrupted
the ECG trace and essentially rendered the interpretation of the ECG
impossible. There was an undesirable interference with the electric
signals due to the leakage through the organoid’s electrical
conduction system, thereby limiting the simultaneous functions of
ECG recording and electrical pacing. After the end of electrical stimulation,
the ECG signal was returned to the normal state (Figure S14). Figure e shows a diagram of the electrical circuit of this cardiac
organoid with 3D electrodes. In contrast, our mechanophysiological
sensing enabled the recording of the organoid’s beating motions
with no signal disruption during the application of electrical impulses
in electrical therapeutics. The intraorganoid interface between the
3D electrodes and the organoid enlarged the contact area relatively
to ensure efficient charge transfer. Also, as shown in the computational
simulation of the electric potential for the shape and placement of
electrodes inside a microwell (Figure f), the 3D microstructure of the electrodes for our
device can be effective to focus and locally distribute the electric
field near the pacing region (case 3), compared with the flat geometries
of the other electrodes (case 1 and case 2).Recent research
related to organoids has performed exceedingly
well in the recapitulation of cellular diversity and microscale tissue
architecture to bring out critical functions of their in vivo counterparts. However, the issues of translational relevance and
high-throughput functional readout remain crucial challenges in practical
utilization for clinical applications.[43,45] In traditional
culture systems, there is little or no integration of biophysical
and topological parameters that can support the self-organization
of organoids.[46] Hence, the growth environment
of these organoids usually is versatile and ill-defined, which leads
to high variability in organoid phenotypes. Although optical analysis
methods have been used mainly for cellular characterization, organoids
have 3D shapes and versatile motion displacements, which makes automated
live imaging very challenging. The optical imaging method also holds
little functional insight and relies on reporter lines for the assessment
of the expression of specific markers. The multimodal sensing mechanism
can provide a capacity for deeper contextual understanding of physiological
parameters for more profound analytical studies. Furthermore, the
implementation of each of the integrated methods was conducted independently
and does not effectively affect the performance of the other methods.
The straightforward microfabrication steps and the simplicity of the
direct printing of 3D liquid-metal electrodes allow a smooth transition
into large-scale microcompartmentalization for automated screenings.
This multimodal sensory device suggests the future promise of an all-in-one
analysis platform for a higher level of biomedical modeling of cardiac
organoids in various areas, such as personalized medicine, disease
modeling, and drug cardiotoxicity.
Authors: Jiuk Jang; Sangyoon Ji; G Krishnamurthy Grandhi; Han Bin Cho; Won Bin Im; Jang-Ung Park Journal: Adv Mater Date: 2021-06-17 Impact factor: 30.849
Authors: Johanna F Dekkers; Caroline L Wiegerinck; Hugo R de Jonge; Inez Bronsveld; Hettie M Janssens; Karin M de Winter-de Groot; Arianne M Brandsma; Nienke W M de Jong; Marcel J C Bijvelds; Bob J Scholte; Edward E S Nieuwenhuis; Stieneke van den Brink; Hans Clevers; Cornelis K van der Ent; Sabine Middendorp; Jeffrey M Beekman Journal: Nat Med Date: 2013-06-02 Impact factor: 53.440
Authors: Doan C Nguyen; Tracy A Hookway; Qingling Wu; Rajneesh Jha; Marcela K Preininger; Xuemin Chen; Charles A Easley; Paul Spearman; Shriprasad R Deshpande; Kevin Maher; Mary B Wagner; Todd C McDevitt; Chunhui Xu Journal: Stem Cell Reports Date: 2014-07-04 Impact factor: 7.765
Authors: Yoonseok Park; Colin K Franz; Hanjun Ryu; Haiwen Luan; Kristen Y Cotton; Jong Uk Kim; Ted S Chung; Shiwei Zhao; Abraham Vazquez-Guardado; Da Som Yang; Kan Li; Raudel Avila; Jack K Phillips; Maria J Quezada; Hokyung Jang; Sung Soo Kwak; Sang Min Won; Kyeongha Kwon; Hyoyoung Jeong; Amay J Bandodkar; Mengdi Han; Hangbo Zhao; Gabrielle R Osher; Heling Wang; KunHyuck Lee; Yihui Zhang; Yonggang Huang; John D Finan; John A Rogers Journal: Sci Adv Date: 2021-03-17 Impact factor: 14.136