Nazia Rehman1, Shamsul Qamar1. 1. Department of Mathematics, COMSATS University, Park Road, Chak Shahzad, Islamabad 45550, Pakistan.
Abstract
In both linear and nonlinear chromatography, the lumped kinetic model is a suitable model for predicting elution bands when appropriate equilibrium functions and mass transfer coefficients are accessible. This model also works well in the case of gradient elution chromatography if variations in the equilibrium functions due to changes in the mobile phase composition are known. The rational selection of an optimum gradient is explored in this study from three different perspectives using the lumped kinetic model. Elution profiles generated by using (a) linear solvent strength, (b) quadratic solvent strength, and (c) power law are investigated. The effectiveness and reliability of the suggested numerical approach, utilizing the flux-limiting finite volume method, are demonstrated through numerical simulations. The impacts of axial dispersion, nonlinearity coefficient, Henry's constant, mass transfer coefficient, and gradient parameters are studied on single and two-component elution profiles.
In both linear and nonlinear chromatography, the lumped kinetic model is a suitable model for predicting elution bands when appropriate equilibrium functions and mass transfer coefficients are accessible. This model also works well in the case of gradient elution chromatography if variations in the equilibrium functions due to changes in the mobile phase composition are known. The rational selection of an optimum gradient is explored in this study from three different perspectives using the lumped kinetic model. Elution profiles generated by using (a) linear solvent strength, (b) quadratic solvent strength, and (c) power law are investigated. The effectiveness and reliability of the suggested numerical approach, utilizing the flux-limiting finite volume method, are demonstrated through numerical simulations. The impacts of axial dispersion, nonlinearity coefficient, Henry's constant, mass transfer coefficient, and gradient parameters are studied on single and two-component elution profiles.
In high-performance liquid
chromatography (HPLC), the mobile phase
composition mostly influences the separation of complicated mixtures,
especially when several closely related analytes are to be measured.[1] It is well-known that partitioning dominates
the retention of tiny nonpolar compounds in reversed-phase liquid
chromatography (RP-LC) mode.[2] Desirable
separation of a mixture (including wide-range retentive components)
cannot be achieved in an appropriate time frame in isocratic mode.
A gradient (in solvent strength) can be used to address such elution
issues.[3] The retention factor has the greatest
impact on the peak width in the gradient mode. A gradient separation
produces narrow and almost consistent peak widths and, therefore,
provides equivalent detection sensitivity. It is commonly employed
in a wide-range of chromatographic separations. Gradient elution chromatography
has a well-established theory,[4] and several
recent publications and reviews provide a comprehensive description
of the method from theory to application.[5,6]In recent years, the gradient elution technique has been significantly
applied for more complicated separations in analytical chromatography.
It is more commonly applied in forward-phase liquid–solid chromatography
and reversed-phase chromatography.[7−10] Optimal elution algorithms enable the separation
of multicomponent mixtures including solutes with significantly different
retention characteristics. Analysis times for complicated mixtures
with a broad range of retention factors can be significantly reduced
in gradient elution than in isocratic separation of the identical
mixtures, as mobile phase strength can be raised either gradually
(e.g., linear gradient) or abruptly (step gradient) during sample
elution.This work has two objectives. The first one is to adapt
and assess
three models of gradient elution considering mobile phase variation.
The second one is to compare their ability to predict retention behavior
in liquid chromatography. It is critical to build models (stepwise,
linear, and nonlinear solvent strength) that describe the retention
factors as functions of mobile-phase composition, having the aim to
obtain required separations efficiently. Numerous efforts have been
made to create retention-time models since the late 1970s. Various
retention-time models in RP-LC have been created, each with its own
set of expressions defining the relationship between retention and
mobile phase composition, e.g., the linear solvent strength model
(LSS),[11] the quadratic retention-time model
(QSS),[12−14] and the Neue model.[15] The
most popular model is the LSS, which Snyder and Dolan developed in
the 1990s. It characterizes isocratic retention in RP-LC as[11]where kHr, is the reference (extrapolated) value of K for ϕ0 (specifically in pure water), and α
is the solvent strength parameter that is only appropriate in a limited
range of ϕ values.[15] To accommodate
this issue and enhance retention-time modeling, multimodal retention
mechanisms have been introduced, such as the quadratic model[6] and the Neue and the Kuss model.[16]In nonlinear chromatography, a variety of kinetic
models have been
presented.[16−18] Various chromatographic models try to model band
migration in numerous ways. The general rate model takes into account
a more or less sophisticated set of kinetic equations and performs
a comprehensive study of the many steps in the chromatographic process.
The lumped rate model takes into account just one kinetic process,
known as the rate-controlling step, or a few such procedures, and
it includes the involvement of the kinetics of the other processes
in the rate constant. The EDM is the simplest and basic, combining
all kinetic effects into a single apparent dispersion value. It is
generally known that under some fundamental assumptions simpler models
may be obtained from the comprehensive general rate model.[19,20]While gradient elution studies frequently require the use
of tiny
and dilute samples, in preparative gradient elution chromatography,
the column is frequently overloaded. As a result, the adsorption isotherms
are nonlinear and competitive, and interference effects become significant.
Consequently, mass transfer resistances can be quite high, particularly
for macromolecules.Many researchers have already developed
their simulation programs
to grasp the specifics of chromatography modeling. These programs,
however, seldom attain sufficient numerical performance since they
are frequently created for particular purposes. A specific amount
of software engineering, project management, maintenance, and support
is necessary for publishing a code. Most open-source programs were
created in academic institutions or government research facilities.
For instance, CADET is one of the freely available multitask simulation
frameworks for column liquid chromatography.[21]When it comes to the simulation of chromatographic systems,
there
are certain unique challenges. Sharp fronts are likely to develop
in a variety of situations, including when the column is extremely
efficient[18] and whenever there are self-sharpening
effects (Langmuir-type isotherm). When there are no analytical solutions
to the model equations, which is the case in most of the scenarios,
to ensure precision, stability, and speed when dealing with sharp
fronts, appropriate numerical techniques must be explored. It has
been demonstrated that traditional numerical techniques, such as simple
finite difference (FD), are incapable of efficiently capturing the
real sharp fronts.[22]Thus, in this
paper, a high-resolution flux-limiting finite volume
scheme is proposed to solve the model equations.[23−25] This numerical
approach is especially useful for convection-dominated problems in
which sharp peaks or fronts are generated. It preserves the mass-conservation
property of the current model equations and is capable of capturing
sharp fronts and peaks in the solutions.[23−25] The conventional
method for resolving discontinuities and sharp fronts has been the
flux estimation technique. Because the finite volume (FV) approach
incorporates the physical concept of flux, many flux estimate techniques
may be easily applied to it.[23−25] In all techniques, the last step
of the solution is the same: depending on the kind of discretization,
they provide a set of algebraic equations or ordinary differential
equations (ODEs).The remaining portion of this article is organized
in the steps
outlined here. Section introduces the nonlinear lumped kinetic model (LKM) for three separate
gradient elution strategies. In Section , the suggested finite volume method for
solving the given model equations is formulated. Section contains discussions on numerical
case studies that demonstrate the model’s and numerical scheme’s
efficacy. Section presents conclusions based on the outcomes of the discussion.
Lumped Kinetic Model
Instead of examining
the entire interparticle concentration profile,
the lumped kinetic model uses a linear driving force in the solid
phase and only examines one extra parameter to supplement the axial
dispersion coefficient. It combines the effects of internal and external
mass transport resistances (inside one mass transfer coefficient).
Two kinetic parameters, the axial dispersion coefficient (D) and the
mass-transfer rate coefficient (KL,i),
are related to the overall mass-transfer rate in a column. The dynamical
characteristics of the gradient elution chromatographic separation
process are studied using the LKM.The following mass balance
equation is used in the mobile phase:For each component, i, c, q, and q* denote the solute concentration
in the mobile phase, nonequilibrium solute concentration in the stationary
phase, and the equilibrium solute concentration in the stationary
phase, respectively. u is interstitial velocity; t and z symbolize the time coordinate and
the axial coordinate; ϕ ≔ ϕ(t, z) is the solvent concentration; and F =
(1 – ε/ε) is the phase ratio in which ε represents
the external porosity.In this model, it is assumed that the
adsorption–desorption
and the diffusion processes in the mobile phase are very rapid. Therefore,
in order to complete the model, the accumulation in the solid phase
is evaluated using the fundamental linear driving force approach shown
below:Relationships between the liquid and solid
phase equilibrium adsorbate concentrations are termed as isotherms.
The equilibrium adsorption data may be simulated using the isotherms.
Researchers use isotherms to explore adsorption information, such
as adsorption mechanisms, maximum adsorption capacity, and the characteristics
of adsorbents. In the literature, a variety of different adsorption
isotherm models have been established. The multicomponent Langmuir
equation is used to model the adsorption equilibrium in this study.[26,27] When the mobile or liquid phase composition varies during the gradient
elution process as a result of variations in a modifier concentration,
the local equilibrium can be described asKH, denotes Henry’s coefficient, and the degree of nonlinearity
associated with the isotherm for the i-th component
of the mixture is quantified by b. Most of the time, it is necessary to conduct experiments
to determine the functional dependence of the two isotherm parameters
on ϕ.
Non-LSS QSS Model
The LSS model has
been used to separate tiny molecules as well as macromolecules such
as proteins and peptides. Furthermore, the LSS model is only viable
in a small range of ϕ values. Due to these constraints, the
following two gradient models are employed in this study to simulate
analyte retention as a function of solvent fraction and compare their
results to previously published work using the LSS model.[25] Polynomial equations can sometimes be utilized
to more correctly characterize retention behavior than linear models.
Non-LSS Power-Law Model
In this case,
the model parameters are dependent on the solvent concentration asHere, D, kLr,, kH, and b symbolize the reference
values of the axial dispersion, mass transfer, Henry’s, and
nonlinearity coefficients. Furthermore, ϕ is the volume fraction
of modifying nonretained solvent, and the specific solvent strength
parameters are denoted by α1, α2, γ1, and γ2.To determine
the distribution of the strong solvent of the mobile phase across
the column, the solvent is considered not to be retained. Consequently,
the following model is the most accurate for estimating changes in
the volume fraction (ϕ(t, z)) of the strong solvent in the mobile phase:[25]with initial and boundary conditions:For a linear gradient:Here, Φ and β are the implemented
profile and slope of the gradient, while ϕ0, ts and ϕe, te are the initial and final volume fractions and time
of the gradient, respectively.To complete the lumped kinetic
model, the initial and boundary
conditions are expressed asThese inlet boundary conditions (BCs) are:where i = 1, 2, 3, ... Nc. The i-th component-injected
concentration is denoted as c, and tinj is the injection
time.
Numerical Solution
There are several
known numerical methods for estimating chromatographic
models.[28,29] A semidiscrete high-resolution flux-limiting
finite volume technique is used in this section to solve the model
equations.[30] This approach has been recently
applied on two models of gradient elution chromatography.[24,25]
Domain Discretization
The first step
for implementing this technique is to discretize the computing domain.
The main purpose of this discretization is to produce a set of time-coordinated
coupled ODEs.Here, zj (mesh points)
are covered by the cells for 1 ≤ h ≤ N in the intervals such thatandThe averaged initial data are defined aswhere .Integrating eqs and 3 over Ω givesThe differential and dispersion terms in eq can be approximated
asThe solutions of eqs –11 giveThere are many different approaches and computational
schemes available to estimate concentrations (or fluxes) at cell interfaces.
Here, just the first and second order approximations are shown.
First-Order Approximation
At the
cell interfaces, concentration values are estimated in eq by a backward difference formula.
The first-order approximation for the concentration can be represented
as
Second-Order Approximation
The concentration
values at the cell’s interface c are approximately calculated
using the flux-limiting calculations given below.[29]andWe need η = 10–10.
The flux limiting function, i.e., Ψ, is employed to maintain
the numerical scheme’s local monotonicity in eq which is defined as[29]
Results and Discussion
Single- and
two-component samples are used as examples to demonstrate
the benefits of gradient elution over isocratic elution and the necessity
of choosing the right gradient technique. For simplicity, it is assumed
that the mass transfer coefficient KL, = kL; the axial dispersion
coefficients D = D; the nonlinearity
coefficient b = b; and solvent strength parameters α = γ (for
LSS), α1 = γ1, and α2 = γ2 (for QSS). All other parameters are listed
in Table . The values
of parameters in Table are chosen from the ranges typically used in HPLC. The plots display
ϕ (the modulator concentration) and c (the
solute concentration) at the column outlet z = L against the time t.
Table 1
Parameters for Single-Component Elution
parameters
values
column length
L = 10 cm
interstitial velocity
u = 1.0 cm/min
porosity
ε = 0.4
reference axial dispersion coefficient
Dzr = 0.0002 cm2/min
reference Henry’s
constant
kHr = 3.5
reference
mass transfer coefficient
kLr = 10 min–1
reference nonlinearity coefficient
bbref = 2.0
gradient start time
ts = 5 min
gradient end time
te = 90 min
initial concentration
ϕo = 0.1
final concentration
ϕe = 0.9
solvent strength parameter
α = 10.0
solvent strength
parameter (for the non-LSS model (QSS))
α1 = 8.0
solvent strength parameter (for the non-LSS model (QSS))
α2 = 10.0
order of the power-law model
n = 1
Single-Component Elution
Dual comparisons
are done in these case studies, i.e., numerical profiles plotted for
different values of parameters and a comparison of three different
gradient models mentioned above. Figure shows the comparison of numerical solutions
for different values of reference Henry’s constant kH using the LSS and non-LSS
models of gradient elution chromatography. For kH = 2 the best results can be seen for all
gradient models. By inspection it is found that the non-LSS QSS model
predicts the elution profiles better; i.e., narrow and symmetric peaks
are generated. The non-LSS power-law model gives highly asymmetric
peaks (adsorption–desorption is slow), and the retention time
decreases for small values of Henry’s constant.
Figure 1
Influence of reference
Henry’s constant on nonlinear single-component
elution profiles by taking non-LSS and LSS models of gradient elution.
Influence of reference
Henry’s constant on nonlinear single-component
elution profiles by taking non-LSS and LSS models of gradient elution.The effects of reference nonlinearity coefficient bref are given in Figure . For bref =
0, the elution
profiles are shaped like a Gaussian curve; i.e., symmetric peaks are
obtained. By increasing bref, the well-known
formation of sharp adsorption and dispersed desorption fronts having
shorter retention periods are observed. Once again, the non-LSS QSS
model better predicts the elution profiles.
Figure 2
Influence of the reference
nonlinearity coefficient on nonlinear
single-component elution profiles by taking non-LSS and LSS models
of gradient elution.
Influence of the reference
nonlinearity coefficient on nonlinear
single-component elution profiles by taking non-LSS and LSS models
of gradient elution.Figure displays
the effects of reference axial dispersion coefficient D considering LSS and non-LSS models
of gradient elution. The mean retention period remains unaffected
for LSS and non-LSS models. The concentration profile gets less broadened
as the value of D decreases
for the non-LSS power-law model, but for LSS and non-LSS QSS models
decreasing D does not
show prominent changes on elution. Overall, the non-LSS QSS model
better predicts the elution profiles among others.
Figure 3
Influence of reference
dispersion coefficient on nonlinear single-component
elution profiles by taking non-LSS and LSS models of gradient elution.
Influence of reference
dispersion coefficient on nonlinear single-component
elution profiles by taking non-LSS and LSS models of gradient elution.Effects of the gradient start time on nonlinear
single-component
elution profiles can be seen in Figure . For ts = 15 min,
a distortion in peak shape appears, i.e., the peak spilt for the LSS
model. This distortion can be minimized by implementing corrective
actions like reducing the sample size or using a diluted solution
or by reverse flow of the mobile phase. If the gradient starts late,
i.e., for ts = 30 min, ts = 60 min, and ts = 80 min, the LSS and non-LSS QSS models show better
separation (elution peaks merge for both cases), whereas for the non-LSS
power-law model an increase in gradient start time results in asymmetrical
peak shapes, increased peak heights, and longer run durations, which
make the detection more difficult.
Figure 4
Influence of gradient start time on nonlinear
single-component
elution profiles by taking non-LSS and LSS models of gradient elution.
Influence of gradient start time on nonlinear
single-component
elution profiles by taking non-LSS and LSS models of gradient elution.The effects of gradient end time on nonlinear single-component
elution profiles for LSS and non-LSS models are presented in Figure . When the gradient
ends early, i.e., for te = 20 min,
the LSS model shows narrower peaks as compared to the non-LSS QSS
model, while the non-LSS pwer-law model gives a peak with a sharp
front and pronounced Langmuir effect. Overall, when the gradient time
is increased, the peak heights are lowered. This is characteristic
of gradient elution with micromolecule samples and corresponds to
elution of each peak at a lower volume fraction of solvent values
as the gradient duration rises.
Figure 5
Influence of gradient end time on nonlinear
single-component elution
profiles by taking non-LSS and LSS models of gradient elution.
Influence of gradient end time on nonlinear
single-component elution
profiles by taking non-LSS and LSS models of gradient elution.In Figure , comparisons
of isocratic and gradient elution are shown. These comparisons are
divided in two cases for convenience, isocratic conditions with ϕ
= 0 and (i) gradient elution with ϕ0 = 0.1, ϕe = 0.9, ts = 5 min, and te = 80 min and (ii) gradient elution
with ϕ0 = 0.1, ϕe = 0.5, ts = 5 min, and te = 120 min. Overlapping peaks are obtained for non-LSS
and LSS models for isocratic cases ϕ = 0. For the non-LSS QSS
model, gradient elution is the best choice in both cases. Overall,
better separation can be seen for the gradient case as compared to
isocratic cases. It is worth noting that the peak shapes differ in
both cases due to the change in gradient techniques.
Figure 6
Comparison of gradient
and isocratic elution for nonlinear single-component
elution profiles by taking non-LSS and LSS models of gradient elution.
Comparison of gradient
and isocratic elution for nonlinear single-component
elution profiles by taking non-LSS and LSS models of gradient elution.Effects of solvent strength parameters for LSS
and non-LSS models
and n of the non-LSS power-law model are shown in Figure . It can be seen
that profiles become narrower with short retention time as solvent
strength parameters increase (i.e., the sensitivity of solvent strength
parameters to modulate concentration) for both LSS and QSS models
and, hence, improved separation or purification. Also, for the non-LSS
power-law model, increasing n results in more prominent
peak tailings.
Figure 7
Influence of solvent strength parameters α of the
LSS model,
α1 and α2 of the QSS model, and n of the non-LSS model (power) on nonlinear single-component
elution profiles.
Influence of solvent strength parameters α of the
LSS model,
α1 and α2 of the QSS model, and n of the non-LSS model (power) on nonlinear single-component
elution profiles.
Comparison and Error Analysis of the Numerical
Schemes
This test problem quantitatively analyzes the performance
of the Koren technique compared to other available flux-limiting finite
volume schemes. Figure depicts that the Koren technique generates the lowest errors. The
reference solution was obtained over a grid of 1000 mesh cells. The
magnified plots show that the first-order scheme produces the most
diffusive elution profile, whereas the Koren scheme generates the
most resolved solution. These facts lead us to the conclusion that
the Koren scheme is suitable for solving models of gradient elution
chromatography. The same performance of the Koren has already been
verified in our previous article analytically and numerically in the
case of isocratic elution.[28]
Figure 8
Comparison
and error analysis of numerical schemes for single-component
elution profiles.
Comparison
and error analysis of numerical schemes for single-component
elution profiles.
Two-Component Elution
Real separation
problems typically involve more than two components in a feed. Moreover,
as observed above, nonlinear gradient shapes have additional potential
for enhancing the process performance compared to the linear gradient.
Due to these facts, a theoretical study of two-component elution is
also included here, i.e., Nc = 2. In this
case, the retention factors of components differ significantly, which
reflects real situations. For the connection between the component-specific
adsorption equilibrium constants, mass transfer, axial dispersion
coefficient, and modulator concentrations, the LSS and non-LSS models
are used. Here we take L = 10 cm as the column length,
porosity as ε = 0.4, and injected concentrations as c1,inj = c2,inj =
1.0 mol/L, which are injected for a duration of tinj = 2.0 min at u = 0.6 cm/min.
All other parameters are listed in Table . The modulator concentration (ϕ) and
the solute concentrations (c) are shown at the column outlet over time t. In addition, the figures show the impact of gradient parameters
on the elution profiles considering the aforementioned three gradient
models.
Table 2
Parameters for Two-Component Elution
parameters
values
reference axial dispersion coefficient
Dzr = 0.0002 cm2/min
reference mass transfer coefficient
kLr = 10 min–1
reference Henry’s
constant for component I
kHr,1 = 1.0
reference Henry’s constant for component II
kHr,2 = 3.5
reference nonlinearity
coefficient for component I
b1bref = 1
reference nonlinearity coefficient for
component II
b2bref = 2
solvent strength parameter
α = 0.90
solvent strength parameter (for the non-LSS model 1 (QSS))
α1 = 0.40
solvent strength
parameter (for non-LSS model 1 (QSS))
α2 = 0.90
initial concentration
ϕ0 = 0.10
final concentration
ϕe = 0.9
gradient start time
ts = 5 min
gradient
end time
te = 80 min
Comparisons under Gradient Conditions
A comparison of different choices of gradient models is presented
in Figure . For the
situation ϕ = ϕ0 = 0.1, the QSS model achieves
generally superior separation outcomes for both components than other
models, with narrower elution peaks and improved retention time. Moreover,
the non-LSS power-law model completely fails to produce desired results
at low concentration (ϕ = ϕ0 = 0.1) of the
mobile phase, and we did not achieve the baseline separations for
component II. For the second case ϕ = ϕe, the
results are almost similar with narrower peaks having shorter retention
times for QSS and LSS models. While in the positive (β >
0)
and negative (β < 0) gradient situations for two-component
elution, the non-LSS QSS model appears to be efficient, while the
LSS and non-LSS power-law models yield better results.
Figure 9
Comparison of the LSS
and non-LSS models under gradient conditions
on nonlinear two-component elution profiles.
Comparison of the LSS
and non-LSS models under gradient conditions
on nonlinear two-component elution profiles.
Comparison of Numerical Schemes
This
case study compares the Koren scheme’s performance
to other flux-limiting finite volume schemes. Initially cinit = 0 mol/L; i.e., the column is equilibrated with
the solvent. Afterward, pulses of c1,inj = c2,inj = 1 mol/L are injected for
a duration of tinj = 2.0 min. The
column L has a length of 10 cm, ε = 0.4, u = 1 cm/ min, and N = 300. In addition, a grid of 50 mesh cells was
utilized. Figure shows the numerical outcomes at the column outlet. All results are
obtained in MATLAB R2015a with an intel core (TM) i5-6200U CPU, RAM
8.00 GB, 2.30 GHz, and Window 10 pro graphic card. For all LSS and
non-LSS models of gradient elution, the Koren scheme gives more resolved
peaks for both components. As far as computational time is considered
the Koren scheme has minimum computational time as compared to other
flux-limiting finite volume schemes. According to the aforementioned
observations, the Koren technique is preferable for solving these
gradient elution models.
Figure 10
Comparison of different schemes for two components
with a nonlinear
isotherm.
Comparison of different schemes for two components
with a nonlinear
isotherm.
Conclusion
The separation in overload
columns utilizing three different techniques
of gradient elution chromatography was simulated using the nonlinear
lumped kinetic model. The influences of different model parameters
were considered to be dependent on the modulator’s concentration.
Comparisons of LSS and non-LSS models were demonstrated, and their
possible applications in gradient elution chromatography were explored.
Although most of our test problems were focused on single-component
elution, real separation problems typically involve more than two
components in a feed. Therefore, a theoretical study of two-component
elution was also included in the test problems. It was observed that
nonlinear gradient shapes have additional potential for enhancing
the process performance compared to the linear gradient. In the current
study, it was observed that gradient elution enhances the production
rate by decreasing the retention and cycling times and by possibly
increasing the loading factors. The results obtained verify that gradient
elution has the potential to outperform isocratic operation in both
analytical and preparative chromatography. It is particularly suitable
for the separation of mixtures containing components to have strong
retentions. In addition to productivity enhancement, the time required
for column regeneration is relatively short in the gradient elution
chromatography. In nonlinear multicomponent chromatography, the gradient
elution conditions can be influenced by the dependence of separation
factors between the target component and its neighbors on the solvent
concentration. Thus, the considered gradient models are rigorous and
allow the evaluation of the influence of gradient elution on concentration
profiles for a wide variety of operating conditions which are generally
difficult to analyze in experimental research.