Andee M Beierle1, Colin H Quinn2, Hooper R Markert2, Adam Carr3, Raoud Marayati2, Laura V Bownes2, Sara Claire Hutchins4, Jerry E Stewart2, Benjamin Hill5, Michael Ohlmeyer6, Nigel F Reuel3, Elizabeth A Beierle2. 1. Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama 35233, United States. 2. Division of Pediatric Surgery, Department of Surgery, University of Alabama at Birmingham, Birmingham, Alabama 35233, United States. 3. Department of Chemical and Biological Engineering, Iowa State University, Ames, Iowa 50111, United States. 4. Division of Pediatric Hematology/Oncology, Department of Pediatrics, University of Alabama at Birmingham, Birmingham, Alabama 35233, United States. 5. Division of Pathology, Children's Hospital of Alabama, Birmingham, Alabama 35233, United States. 6. Atux Iskay LLC., Plainsboro, New Jersey 08536, United States.
Abstract
Cancer continues to be a significant cause of non-traumatic pediatric mortality. Diagnosis of pediatric solid tumors is paramount to prescribing the correct treatment regimen. Recent efforts have focused on non-invasive methods to obtain tumor tissues, but one of the challenges encountered is the ability to obtain an adequate amount of viable tissue. In this study, a wireless, inductor-capacitor (LC) sensor was employed to detect relative permittivity of pediatric tumor tissues. There is a comparison of resonant frequencies of tumor tissues between live versus dead tissues, the primary tumor tissue versus tissue from the organs of origin or metastasis, and treated versus untreated tumors. The results show significant shifts in resonant frequencies between the comparison groups. Dead tissues demonstrated a significant shift in resonant frequencies compared to alive tissues. There were significant differences between the resonant frequencies of normal tissues versus tumor tissues. Resonant frequencies were also significantly different between primary tumors compared to their respective metastases. These data indicate that there are potential clinical applications of LC technology in the detection and diagnosis of pediatric solid tumors.
Cancer continues to be a significant cause of non-traumatic pediatric mortality. Diagnosis of pediatric solid tumors is paramount to prescribing the correct treatment regimen. Recent efforts have focused on non-invasive methods to obtain tumor tissues, but one of the challenges encountered is the ability to obtain an adequate amount of viable tissue. In this study, a wireless, inductor-capacitor (LC) sensor was employed to detect relative permittivity of pediatric tumor tissues. There is a comparison of resonant frequencies of tumor tissues between live versus dead tissues, the primary tumor tissue versus tissue from the organs of origin or metastasis, and treated versus untreated tumors. The results show significant shifts in resonant frequencies between the comparison groups. Dead tissues demonstrated a significant shift in resonant frequencies compared to alive tissues. There were significant differences between the resonant frequencies of normal tissues versus tumor tissues. Resonant frequencies were also significantly different between primary tumors compared to their respective metastases. These data indicate that there are potential clinical applications of LC technology in the detection and diagnosis of pediatric solid tumors.
Cancer continues to be a significant cause
of non-traumatic childhood
mortality. For pediatric solid tumors, rapid and accurate diagnosis
is critical to prescribing the correct treatment regimen as treatments
vary widely not only between tumor types but also between different
biologic variables characterizing the tumors associated with staging.
Current paradigms for the diagnosis of these cancers frequently involve
tumor biopsy and investigation of the specimens with histology, immunohistochemistry,
and genetic characterization. The ability to provide the pathologist
with an adequate quantity of viable tissue is essential to the successful
completion of the studies necessary for diagnosis and treatment planning.
Historically, methods to obtain tumor tissue specimens involved an
invasive procedure under general anesthesia often requiring a postoperative
hospital admission.[1] Investigators have
been focusing efforts to shift to less invasive biopsy methods, especially
for children, to reduce exposure to general anesthetics, decrease
blood loss, and avoid hospital admission.[2] These less invasive techniques, however, require immediate analysis
of viability of the tumor biopsy specimens to prevent the need for
a repeat biopsy. Depending on the environment, these examinations
may require lengthy evaluations including transporting specimens from
the operating theater to the pathology suite followed by frozen section
processing and examination. We propose that a method capable of providing
evaluation of the quality of tumor specimens in real-time in the operating
room would be useful.An inductor-capacitor (LC) resonator is
a simple circuit that has
an inherent resonant frequency at which inductively coupled power
is most efficiently transferred.[3] It may
be rendered into a sensor by engineering circuit elements such that
alterations in temperature, pressure, humidity, or biologic analytes
will change the inductance and/or capacitance of the circuit resulting
in changes in the resonant frequency of the system.[3] Shifts in resonant frequencies are then detected contact-free
by antennas connected to a vector network analyzer (VNA). The concept
of studying resonant frequency in biologic systems is not a new one.
Investigators have utilized LC sensors to detect changes in dielectric
properties in several biologic systems including monitoring wound
healing,[4] gastric pressure,[5] and intracranial pressure[6] and
detecting the presence of urine[7] or sweat.[8] Other investigators have studied the changes
in relative permittivites as applied to adult cancer types, and investigators
have reported a variation in relative permittivity between normal
and cancer specimens, including liver and breast cancer.[9,10] The notion of utilizing changes in relative permittivites in pediatric
solid tumors is unique and to our knowledge has not yet been reported.To monitor changes in tissue permittivity in the current study,
we designed a passive, Archimedean spiral resonant sensor copper patch
that was embedded into a commercial wound dressing that transduced
the system resonant frequencies to an external reader antenna connected
to a VNA. We hypothesized that this passive LC sensor system could
characterize pediatric solid tumors based upon differences in the
relative permittivity of the tissues. We evaluated common pediatric
solid tumors including neuroblastoma, hepatoblastoma, neuroendocrine,
osteosarcoma, and medulloblastoma human tumors propagated in murine
models.
Results
Scanning Murine Tissue Using the LC Sensor
To determine
the ability of the LC resonant sensor to detect changes in relative
permittivity of the murine tissues, we used a configuration that has
been previously described.[3] For the current
experiments, the resonant sensor was placed on the abdomen of the
animal and VNA was positioned 1 cm above the sensor (Figure A). We obtained a minimum of
three scans prior to euthanasia (live tissue) and 5 min post-euthanasia
(dead tissue) for each animal. A representative output by the MetroVNA
of the scans from one animal demonstrates the decrease in resonant
frequency of the abdominal tissue by a left shift on the x axis of the dead abdominal tissue peaks in comparison to the alive
abdominal tissue peaks (Figure B, arrow). Compiling data from multiple animals revealed that
the resonant frequency values of murine abdominal tissues significantly
decreased after the animals were euthanized (9.48 × 107 ± 1.85 × 106 Hz vs 7.6 × 107 ± 2.55 ×106 Hz, live vs dead, n = 16, ‡‡p ≤ 1 ×10–7) (Figure C), serving as a surrogate for a dead or non-viable tissue.
Figure 1
Scanning
murine tissue using the LC sensor. (A) To scan the mouse
abdomen, the resonant sensor was placed directly onto the abdomen
of the animal and the VNA was positioned 1 cm above the sensor. Scans
were obtained prior to (alive) and 5 min following (dead) euthanasia
(n = 16). (B) Representative graph of the output
by the MetroVNA of the scans from one animal demonstrating the decrease
in resonant frequency (measured at trough, dagger symbol) of the abdominal
tissue by a left shift (arrow) on the x axis of the
dead abdominal tissue peaks in comparison to the alive abdominal tissue
peaks. (C) Compiled data from 16 animals revealed that the resonant
frequency values of abdominal tissue significantly decreased after
the animals were euthanized. (D) Flank scans were obtained by placing
the resonant sensor onto the flank of the animal and positioning the
VNA 1 cm above the sensor. Scans were obtained prior to (alive) and
5 min following (dead) euthanasia (n = 8). (E) Representative
scans from one animal. There was a left shift on the x axis of the dead flank tissue peaks compared to the alive flank
peaks (arrow). (F) There was a statistically significant decrease
in resonant frequency values of the mouse flank following euthanasia.
A minimum of three scans were obtained for each data point. Photos
obtained by the authors. Data reported as mean ± standard error
of the mean (SEM). Student’s t-test was used
to determine statistical significance. *p ≤
0.05, ‡‡p ≤ 1 ×
10–7.
Scanning
murine tissue using the LC sensor. (A) To scan the mouse
abdomen, the resonant sensor was placed directly onto the abdomen
of the animal and the VNA was positioned 1 cm above the sensor. Scans
were obtained prior to (alive) and 5 min following (dead) euthanasia
(n = 16). (B) Representative graph of the output
by the MetroVNA of the scans from one animal demonstrating the decrease
in resonant frequency (measured at trough, dagger symbol) of the abdominal
tissue by a left shift (arrow) on the x axis of the
dead abdominal tissue peaks in comparison to the alive abdominal tissue
peaks. (C) Compiled data from 16 animals revealed that the resonant
frequency values of abdominal tissue significantly decreased after
the animals were euthanized. (D) Flank scans were obtained by placing
the resonant sensor onto the flank of the animal and positioning the
VNA 1 cm above the sensor. Scans were obtained prior to (alive) and
5 min following (dead) euthanasia (n = 8). (E) Representative
scans from one animal. There was a left shift on the x axis of the dead flank tissue peaks compared to the alive flank
peaks (arrow). (F) There was a statistically significant decrease
in resonant frequency values of the mouse flank following euthanasia.
A minimum of three scans were obtained for each data point. Photos
obtained by the authors. Data reported as mean ± standard error
of the mean (SEM). Student’s t-test was used
to determine statistical significance. *p ≤
0.05, ‡‡p ≤ 1 ×
10–7.Since we noted significant differences in resonant
frequency values
with the abdominal scans, the experiment was repeated focusing on
the mouse flank to determine if tissues in other areas of the body
would yield similar results. Scans were obtained using the same methods
as for the abdominal tissue, with the only difference being the location
of placement of the resonant sensor on the animal, e.g., on the flank
(Figure D). Similar
to the scans of the abdomen, there was a left shift on the x axis of the dead mouse flank tissue peaks compared to
the alive mouse flank peaks (Figure E, arrow) and a statistically significant decrease
in resonant frequency values (measured at trough, dagger symbol (†))
of the mouse flank following euthanasia (1.02 × 108 ± 3.41 × 106 Hz vs 9.18 × 107 ± 3.56 × 106 Hz, live vs dead, n = 8, *p ≤ 0.05) (Figure F). The resonant sensor data obtained for
both the abdominal and flank tissues of the murine models demonstrated
the ability of the LC sensor to distinguish between live and dead
tissues in proximal regions.
Scanning In Vivo Tumors Using the LC Sensor
Once the resonant sensor system could detect differences in resonant
frequency between alive and dead benign tissues, studies were advanced
to evaluate tumor tissues. The resonant frequency of established solid
tumors (osteosarcoma, neuroblastoma, medulloblastoma, and neuroendocrine)
located in the flanks of athymic nude mice was measured. At a tumor
volume of 2000 mm3, a minimum of three scans were obtained
prior to euthanasia using the scheme as previously described (Figure A). Five minutes
post-euthanasia, the resonant frequency of the tumor tissue was measured
with a minimum of three scans for each animal. Figure B depicts a representative output of the
data obtained from the MetroVNA for one of the animals in this data
set with an established osteosarcoma. As shown in the experiments
with non-tumor tissues, there was a significant difference in the
resonant frequency between alive (pre-euthanasia) and dead (post-euthanasia)
flank tumor tissues with a shift of the x axis to
the left (1.06 × 108 ± 1.18 × 106 Hz vs 1.02 × 108 ± 1.28 × 106 Hz, live vs dead, *p ≤ 0.05, n = 9) (Figure C).
Figure 2
Scanning in vivo tumors using the LC sensor. (A)
An LC sensor was used to determine the resonant frequency of osteosarcoma,
neuroblastoma, medulloblastoma, and neuroendocrine tumors located
in the flanks of athymic nude mice. At a tumor volume of 2000 mm3, the LC sensor was placed onto the flank tumor and the VNA
was positioned 1 cm above the sensor. Scans were obtained prior to
(alive) and 5 min following (dead) euthanasia. (B) Representative
scan from a single animal bearing an osteosarcoma tumor. There was
a significant decrease in the resonant frequency (measured at trough,
dagger symbol) between viable and non-viable flank osteosarcoma tumor
tissues (n = 9) seen by a left shift (arrow) on the x axis. (C) There was a statistically significant decrease
in resonant frequency values of the flank tumors following euthanasia.
(D) Resonant frequencies were compared between flank with no tumor
and flanks bearing osteosarcoma tumors. Resonant frequency of flanks
with (n = 18) tumors was significantly increased
when compared to flanks without tumors (n = 9). (E)
Resonant frequencies of flanks with (n = 18) and
without (n = 9) tumors following euthanasia (dead)
were measured. Findings were similar to the alive data, with tumor
bearing flanks having a significantly greater resonant frequency than
those without a tumor as denoted. A minimum of three scans were obtained
for each data point. Photo taken by the authors. Data reported as
mean ± SEM. Student’s t-test was used
to determine statistical significance. *p ≤
0.05, ****p ≤ 0.0001.
Scanning in vivo tumors using the LC sensor. (A)
An LC sensor was used to determine the resonant frequency of osteosarcoma,
neuroblastoma, medulloblastoma, and neuroendocrine tumors located
in the flanks of athymic nude mice. At a tumor volume of 2000 mm3, the LC sensor was placed onto the flank tumor and the VNA
was positioned 1 cm above the sensor. Scans were obtained prior to
(alive) and 5 min following (dead) euthanasia. (B) Representative
scan from a single animal bearing an osteosarcoma tumor. There was
a significant decrease in the resonant frequency (measured at trough,
dagger symbol) between viable and non-viable flank osteosarcoma tumor
tissues (n = 9) seen by a left shift (arrow) on the x axis. (C) There was a statistically significant decrease
in resonant frequency values of the flank tumors following euthanasia.
(D) Resonant frequencies were compared between flank with no tumor
and flanks bearing osteosarcoma tumors. Resonant frequency of flanks
with (n = 18) tumors was significantly increased
when compared to flanks without tumors (n = 9). (E)
Resonant frequencies of flanks with (n = 18) and
without (n = 9) tumors following euthanasia (dead)
were measured. Findings were similar to the alive data, with tumor
bearing flanks having a significantly greater resonant frequency than
those without a tumor as denoted. A minimum of three scans were obtained
for each data point. Photo taken by the authors. Data reported as
mean ± SEM. Student’s t-test was used
to determine statistical significance. *p ≤
0.05, ****p ≤ 0.0001.Next, the resonant frequencies of the flanks without
a tumor were
compared to those with a tumor. First, a viable murine flank tissue
was compared to the viable flank tumor tissue. There was a significant
difference between the two groups, with the live flank tumor tissue
having a larger resonant frequency value than the flanks without tumors
(1.01 × 108 ± 3.03 × 106 Hz vs
1.07 × 108 ± 1.08 × 106 Hz, live
flank with a tumor (n = 18) vs live flank without
a tumor (n = 9), *p ≤ 0.05)
(Figure D). In the
second experiment, flank tumors were scanned over time and demonstrated
a significant change in resonant frequency as the tumors increased
in size (Figure S1). Next, the dead (post-euthanasia)
flank tissue was compared to the dead flank tumor tissue. Again, there
was a significant difference between the two groups, with the dead
flank tumor tissue having a larger resonant frequency value (9.18
× 107 ± 3.56 × 106 Hz vs 1.04
× 108 ± 1.51 × 106 Hz, dead flank
with a tumor (n = 18) vs dead flank without a tumor
(n = 9), ****p ≤ 0.0001)
(Figure E). These
results validate the ability of the LC sensor to consistently detect
changes in viability of tissue regardless of the tissue of origin
and detect differences between tumor and non-tumor tissues.
Detection of Tumor Type Using the LC Sensor
Different
tumor types may show a wide variation in characteristics such as size,
necrosis, stromal content, calcifications, and vasculature. As an
example, histologic sections of neuroblastoma and osteosarcoma (Figure S2) show a vast difference in morphology.
Thus, we suspected that these variations would affect the relative
permittivity of the tissue and thus the resonant frequency of the
coupled LC sensor leading to the ability to distinguish between tumor
types. To test this hypothesis, scans of viable flank tumor tissues
from neuroblastoma (n = 4), hepatoblastoma (n = 3), medulloblastoma (n= 3), osteosarcoma
(n = 3), and neuroendocrine (n =
2) tumors were obtained by scanning the tumor-bearing flank, as shown
in Figure A. Notably,
tumors of the same classification clustered around a similar resonant
frequency value. An analysis of variance (ANOVA) comparison revealed
a significant difference between the peak frequencies of each of the
tumor types (***p ≤ 0.001) (Figure A).
Figure 3
Detection of tumor type
using the LC sensor. (A) Scans of flank
tumor tissues (in vivo) from neuroblastoma (n = 4), hepatoblastoma (n = 3), medulloblastoma
(n = 3), osteosarcoma (n = 4), and
neuroendocrine tumors (n = 2) were obtained, as shown
in Figure A. An analysis
of variance (ANOVA) demonstrated a significant difference between
the peak frequencies of each of the tumor types between each other.
(B) A trephine was used to obtain a standard size core (2 mm) of tissues
from hepatoblastoma (n = 3), neuroblastoma (n = 5), medulloblastoma (n = 3), osteosarcoma
(n = 4), and neuroendocrine (n =
4) flank tumors. The tissue cores were placed on the LC sensor in
the exact same location (black mark indicated by a black arrow in
the upper panel) and readings obtained. (C) There was a significant
difference in the peak frequency between ex vivo tumor
types when measuring the tumor cores (ANOVA). A minimum of three scans
were obtained for each data point. Photo obtained by the authors.
Data presented as mean ± SEM. *p ≤ 0.05,
***p ≤ 0.001.
Detection of tumor type
using the LC sensor. (A) Scans of flank
tumor tissues (in vivo) from neuroblastoma (n = 4), hepatoblastoma (n = 3), medulloblastoma
(n = 3), osteosarcoma (n = 4), and
neuroendocrine tumors (n = 2) were obtained, as shown
in Figure A. An analysis
of variance (ANOVA) demonstrated a significant difference between
the peak frequencies of each of the tumor types between each other.
(B) A trephine was used to obtain a standard size core (2 mm) of tissues
from hepatoblastoma (n = 3), neuroblastoma (n = 5), medulloblastoma (n = 3), osteosarcoma
(n = 4), and neuroendocrine (n =
4) flank tumors. The tissue cores were placed on the LC sensor in
the exact same location (black mark indicated by a black arrow in
the upper panel) and readings obtained. (C) There was a significant
difference in the peak frequency between ex vivo tumor
types when measuring the tumor cores (ANOVA). A minimum of three scans
were obtained for each data point. Photo obtained by the authors.
Data presented as mean ± SEM. *p ≤ 0.05,
***p ≤ 0.001.There are several effects that may confound readings
of the resonant
sensor including temperature, sample pressure (mass), and the location
of the specimen on the sensor. The next set of experiments was designed
to mitigate some of these confounders. Following animal euthanasia
and tumor harvest, a standard size core of each tumor was obtained
using a 2 mm trephine, as outlined in the Methods
section. Tumor core tissues from hepatoblastoma (n = 3), neuroblastoma (n = 5), medulloblastoma (n = 3), osteosarcoma (n = 3), and neuroendocrine
(n = 4) tumors were each placed on the LC sensor
in the exact same location, as diagramed in Figure B (upper panel), and a minimum of three scans
were obtained for each tumor core. There was a significant difference
in the peak frequency between tumor types (*p ≤
0.05) (Figure C).
Collectively, these results indicate that the resonant frequencies
between the tumor types were significant enough to be detected by
an LC sensor both in vivo and ex vivo and that resonant frequencies clustered based upon tumor histology.
Tissue Differentiation Using the LC Sensor
Due to the
differences captured in the resonant frequency values of the tumor
tissue by the LC sensor, it was logical to next examine whether the
LC sensor could detect the resonant frequency of normal murine organs
and tissues. Following euthanasia, a trephine was used to harvest
a standard 2 mm core of the brain, heart, kidney, liver, lung, pancreas,
spleen, and bone. The tissue was scanned using the same schema described
in Figure B. For each
animal (n = 4), a minimum of three ex vivo scans of each organ core were obtained. Figure A displays the average resonant frequency
values obtained for each organ core, and the shifts along the x axis demonstrate the differing resonant frequency values
between the organ tissues (Figure A). A statistical analysis of the organ and bone data
demonstrated that each tissue type had significantly different average
resonant frequency values (Figure B). These results demonstrate the potential of the
LC sensor to distinguish between different tissue types. The results
obtained corroborate previous LC sensor studies performed on biological
tissues, which demonstrated the differences in dielectric properties
of tissues.[11,12]
Figure 4
Tissue differentiation using the LC sensor.
(A) A standard 2 mm
core of the brain, heart, kidney, liver, lung, pancreas, spleen, and
bone was scanned, as described in Figure B. For each mouse (n = 4),
we obtained a minimum of three ex vivo scans of each
organ core. The graph displays the average resonant frequency values
obtained for each organ core, and the shifts along the x axis demonstrate the differing resonant frequency values between
the organ tissues. (B) Compiled data (n = 4 animals).
Each tissue type had significantly different average resonant frequency
values (ANOVA). (C–E) The resonant frequencies of tumors (2
mm cores) were compared to their tissues (2 mm cores) of origin. (C)
Medulloblastoma (n = 3) had a significantly higher
resonant frequency than the brain (n = 4). (D) Hepatoblastoma
(n = 3) had a significantly greater resonant frequency
than a normal liver (n = 4). (E) Osteosarcoma (n = 4) had a significantly lower resonant frequency than
a normal bone (n = 4). A minimum of three scans were
obtained for each data point. Data reported as mean ± SEM. For
statistical comparisons, analysis of variance (ANOVA) was used for
the tissue cores and Student’s t-test for
tissue vs tumor types. *p ≤ 0.05, **p ≤ 0.01, ‡‡p ≤ 1 × 10–7.
Tissue differentiation using the LC sensor.
(A) A standard 2 mm
core of the brain, heart, kidney, liver, lung, pancreas, spleen, and
bone was scanned, as described in Figure B. For each mouse (n = 4),
we obtained a minimum of three ex vivo scans of each
organ core. The graph displays the average resonant frequency values
obtained for each organ core, and the shifts along the x axis demonstrate the differing resonant frequency values between
the organ tissues. (B) Compiled data (n = 4 animals).
Each tissue type had significantly different average resonant frequency
values (ANOVA). (C–E) The resonant frequencies of tumors (2
mm cores) were compared to their tissues (2 mm cores) of origin. (C)
Medulloblastoma (n = 3) had a significantly higher
resonant frequency than the brain (n = 4). (D) Hepatoblastoma
(n = 3) had a significantly greater resonant frequency
than a normal liver (n = 4). (E) Osteosarcoma (n = 4) had a significantly lower resonant frequency than
a normal bone (n = 4). A minimum of three scans were
obtained for each data point. Data reported as mean ± SEM. For
statistical comparisons, analysis of variance (ANOVA) was used for
the tissue cores and Student’s t-test for
tissue vs tumor types. *p ≤ 0.05, **p ≤ 0.01, ‡‡p ≤ 1 × 10–7.The resonant frequencies of excised tumors to their
tissues of
origin were then compared. Looking at the medulloblastoma tumor tissue
compared to a healthy brain tissue (n = 3) revealed
a significantly higher resonant frequency value in the tumor tissue
(9.31 × 107 ± 4.28 × 105 Hz vs
1.10 × 108 ± 1.78 × 106 Hz, medulloblastoma
vs brain, ‡‡p ≤ 1
× 10–7) (Figure C). The second set of scans compared the excised hepatoblastoma
tumor to the healthy liver tissue (n = 3), and again,
the tumor demonstrated a significantly higher resonant frequency (9.38
× 107 ± 9.22 × 105 Hz vs 1.01
× 108 ± 2.56 × 106 Hz, hepatoblastoma
vs liver, **p ≤ 0.01) (Figure D). The final set of scans compared osteosarcoma
to the bone, but unlike the previous tissues, the tumor had a significantly
lower resonant frequency value (n = 3) (1.26 ×
107 ± 9.34 × 104 Hz vs 1.10 ×
108 ± 1.69 × 106 Hz, osteosarcoma
vs bone, ‡‡p ≤ 1
× 10–7) (Figure E). These comparative scans collectively displayed
that the tumor tissue has a different resonant frequency than its
non-pathologic tissue of origin, and the LC sensor detects such differences.
LC Sensor Detects Differences in Resonant Frequency between
Primary and Metastatic Tumors
Tumor metastasis often demonstrates
different physical properties than the primary tumors. We hypothesized
that an LC sensor would demonstrate differences in resonant frequencies
between tissues from metastases compared to the primary tumor tissue.
We also hypothesized that an LC sensor would show that the metastatic
site may have a different resonant frequency than the same site without
malignant tissues. Using a PDX model of a metastatic human neuroendocrine
tumor, both in vivo and ex vivo tissue
scans were completed for mice that harbored intraperitoneal abdominal
metastases. The method for obtaining the in vivo resonant
sensor data followed the same procedure shown in Figure A, with at least three scans
obtained for each animal. The resonant frequency values of abdomens
with metastatic neuroendocrine tumors were compared to the values
from neuroendocrine tumors in the flank of live mice. The scans of in vivo abdominal metastasis demonstrated a decrease in
resonant frequency as compared to the scans of the in vivo flank tumors, indicated by a left shift on the x axis (Figure A,
arrow). The metastatic tumors (n = 6) had a significantly
lower resonant frequency value than that of the flank tumor (n = 5) (7.31 × 107 ± 1.05 × 106 Hz vs 1.05 × 108 ± 3.41 × 106 Hz, metastasis vs flank tumor, ‡p ≤ 1 × 10–6) (Figure B). Next, tumor cores from
abdominal metastatic and flank neuroendocrine tumors were scanned
as described (Figure B). As with the in vivo scans, the ex vivo data (tissue cores) demonstrated a significant difference in frequency
between the two specimen types (Figure C), with the metastatic tumor having a lower resonant
frequency value (1.07 × 108 ± 6.92 × 105 Hz vs 1.03 × 108 ± 8.11 × 105 Hz, metastasis vs flank tumor, ***p ≤
0.001) (Figure D).
These data demonstrate a difference between metastatic and primary
tumor tissue resonant frequency and that the resonant sensor detected
these differences both in vivo and ex vivo.
Figure 5
LC sensor detects differences in resonant frequency between primary
and metastatic tumors. (A) LC sensor was used to obtain resonant frequency
scans using a PDX model of a metastatic human neuroendocrine tumor.
Scans were compared between tumors grown in the flank (primary) and
those grown as intraperitoneal metastases. Representative scans from
a single animal demonstrate a shift to the left of the x axis (arrow) of resonant frequency (measured at trough, dagger symbol)
from the flank tumor compared to the abdominal metastatic tumor. (B)
The metastatic tumors (n = 6) had a significantly
lower resonant frequency value than that of the flank tumors (n = 5). (C) Tumor tissue cores from abdominal metastatic
and flank neuroendocrine tumors were scanned (Figure B). A representative scan demonstrates the
left shift in the x axis (arrow) of resonant frequency
(measured at trough, dagger symbol) between the flank tumor cores
and the cores from abdominal metastases. (D) The metastatic tissue
cores demonstrated significantly lower frequency than the flank tumor
cores. (E) Scans of the abdomen were compared between animals without
abdominal metastases (blank) and those with abdominal metastases.
A representative scan shows the right shift of the x axis (arrow) between resonant frequency (measured at trough, dagger
symbol) of the blank abdomen (no tumor) and the abdomen bearing tumor
metastases. (F) The abdomen scans from animals bearing metastases
displayed a higher resonant frequency than the scans of abdomens from
animals without the metastatic tumor. Data reported as mean ±
SEM. Student’s t-test was employed for statistical
comparisons. *p ≤ 0.05, ***p ≤ 0.001, ‡p ≤ 1
× 10–6.
LC sensor detects differences in resonant frequency between primary
and metastatic tumors. (A) LC sensor was used to obtain resonant frequency
scans using a PDX model of a metastatic human neuroendocrine tumor.
Scans were compared between tumors grown in the flank (primary) and
those grown as intraperitoneal metastases. Representative scans from
a single animal demonstrate a shift to the left of the x axis (arrow) of resonant frequency (measured at trough, dagger symbol)
from the flank tumor compared to the abdominal metastatic tumor. (B)
The metastatic tumors (n = 6) had a significantly
lower resonant frequency value than that of the flank tumors (n = 5). (C) Tumor tissue cores from abdominal metastatic
and flank neuroendocrine tumors were scanned (Figure B). A representative scan demonstrates the
left shift in the x axis (arrow) of resonant frequency
(measured at trough, dagger symbol) between the flank tumor cores
and the cores from abdominal metastases. (D) The metastatic tissue
cores demonstrated significantly lower frequency than the flank tumor
cores. (E) Scans of the abdomen were compared between animals without
abdominal metastases (blank) and those with abdominal metastases.
A representative scan shows the right shift of the x axis (arrow) between resonant frequency (measured at trough, dagger
symbol) of the blank abdomen (no tumor) and the abdomen bearing tumor
metastases. (F) The abdomen scans from animals bearing metastases
displayed a higher resonant frequency than the scans of abdomens from
animals without the metastatic tumor. Data reported as mean ±
SEM. Student’s t-test was employed for statistical
comparisons. *p ≤ 0.05, ***p ≤ 0.001, ‡p ≤ 1
× 10–6.Since the location of the metastasis may affect
resonant frequency,
animals without peritoneal metastasis were compared to those with
metastatic disease. The abdomen scans from animals bearing metastases
displayed a higher resonant frequency than the scans of abdomens in
animals without metastases as demonstrated by a right shift on the x axis (Figure E, arrow). Statistical analysis confirmed a significantly
higher resonant frequency in abdomens of animals with metastatic tumors
(7.31 × 107 ± 1.05 × 106 Hz vs
6.63 × 107 ± 8.11 × 105 Hz, metastasis
vs no metastasis, *p ≤ 0.05) (Figure F). These data highlight the
capacity of the LC sensor to detect the presence of tumor metastasis in vivo.
Chemotherapy Alters Resonant Frequency
Last, we investigated
whether the LC sensor could detect changes in tumor tissues following
treatment. Animals bearing flank tumors of SK-N-AS human neuroblastoma
cells were treated with a protein phosphatase 2A (PP2A) activator
(ATUX-792). In brief, once tumor volumes reached 100 mm3, animals were randomized to receive either vehicle (n = 5) or ATUX-792 (n = 7, 50 mg kg–1 via gavage twice per day). At completion of therapy, the flank tumors
were scanned, and scans were compared between treated and untreated
tumors. The left shift on the x axis represents the
decrease in resonant frequency (measured at trough, dagger symbol
(†)) of the ATUX-792 treated tumors compared to the untreated
tumors (Figure A,
arrow). Statistical analysis confirmed that the treated tumors had
a significantly lower resonant frequency than the untreated tumors
(1.20 × 108 ± 9.51 × 105 Hz vs
1.12 × 108 ± 3.10 × 106 Hz, untreated
vs treated, **p ≤ 0.01) (Figure B). Finally, scans of treated
and untreated tumor cores were compared. Findings were consistent
with those of the flank scans, demonstrating a left shift on the x axis for resonant frequency (Figure C, arrow). The ATUX-792 treated tumor cores
had a significantly lower resonant frequency value than the cores
of untreated tumors (1.21 × 108 ± 3.81 ×
106 Hz vs 1.11 × 108 ± 1.86 ×
106 Hz, treated core vs untreated core, **p ≤ 0.01) (Figure D). Since these trends followed those between viable and non-viable
tissues (Figure ),
the ATUX-792 treated tumors were evaluated further. Untreated tumors
showed a significant increase in tumor necrosis (Figure S3). These data demonstrate that the resonant sensor
has the ability to detect changes in resonant frequencies between
treated and untreated tumors and viable and necrotic tumor tissues.
Figure 6
Chemotherapy
alters resonant frequency. (A) Animals bearing SK-N-AS
neuroblastoma flank tumors were treated with a small-molecule protein
phosphatase 2A (PP2A) activator (ATUX-792). Once tumor volumes reached
100 mm3, animals were randomized to receive either vehicle
(n = 5) or ATUX-792 (n = 7) via
oral gavage. Prior to animal euthanasia, the flank tumors were scanned.
Representative scan of a treated (ATUX-792) and control (vehicle)
tumor demonstrated a left shift in the x axis (arrow)
of the resonant frequency (measured at trough, dagger symbol) of the
treated tumor. (B) Statistical analysis confirmed that the treated
tumors had a significantly lower resonant frequency than the untreated
tumors. (C) Scans were performed on tumor cores and compared between
treated (ATUX-792) and untreated (vehicle) tumors. A representative
scan of a core of the treated and core of the control tumor demonstrated
a left shift in the x axis (arrow) in the resonant
frequency (measured at trough, dagger symbol) of the treated tumor.
(D) Tissue cores from ATUX-792 treated tumors had a significantly
lower resonant frequency value than the cores of untreated tumors.
A minimum of three scans were obtained for each data point. Data are
reported as mean ± SEM. Statistical comparisons were completed
with Student’s t-test. **p ≤ 0.01.
Chemotherapy
alters resonant frequency. (A) Animals bearing SK-N-AS
neuroblastoma flank tumors were treated with a small-molecule protein
phosphatase 2A (PP2A) activator (ATUX-792). Once tumor volumes reached
100 mm3, animals were randomized to receive either vehicle
(n = 5) or ATUX-792 (n = 7) via
oral gavage. Prior to animal euthanasia, the flank tumors were scanned.
Representative scan of a treated (ATUX-792) and control (vehicle)
tumor demonstrated a left shift in the x axis (arrow)
of the resonant frequency (measured at trough, dagger symbol) of the
treated tumor. (B) Statistical analysis confirmed that the treated
tumors had a significantly lower resonant frequency than the untreated
tumors. (C) Scans were performed on tumor cores and compared between
treated (ATUX-792) and untreated (vehicle) tumors. A representative
scan of a core of the treated and core of the control tumor demonstrated
a left shift in the x axis (arrow) in the resonant
frequency (measured at trough, dagger symbol) of the treated tumor.
(D) Tissue cores from ATUX-792 treated tumors had a significantly
lower resonant frequency value than the cores of untreated tumors.
A minimum of three scans were obtained for each data point. Data are
reported as mean ± SEM. Statistical comparisons were completed
with Student’s t-test. **p ≤ 0.01.
Discussion
In the current study, we investigated whether
tissue permittivity
changes would allow detection between live and non-viable tissues.
This determination has potential to guide a wide range of clinical
decisions regarding tumor specimen viability and adequacy, the presence
of organ ischemia, or progression of wounds.Data from the current
study show that resonant frequency values
of live compared to non-viable tissues are significantly different.
A left shift in frequency was observed in non-viable tissues compared
to live tissues irrespective of location (abdomen or flank) or type
of tissue (tumor versus normal tissue). Readings for the non-viable
tissues were made following animal euthanasia. A potential explanation
for the study findings may be changes in temperature. Temperature
has been demonstrated to affect tissue permittivity such that the
dielectric properties of ex vivo bovine and porcine
tissue changed about 1% per °C of tissue cooling.[13] Since measurements in the current studies were
performed after only 5 min following animal euthanasia, it is unlikely
that the significant changes detected by the LC sensor could be contributed
entirely to changes in temperature.The ability to detect normal
tissues from tumor tissues using changes
in tissue permittivity has clinical relevance in the recognition of
complete tumor resection margins, detecting tumor metastasis without
invasive techniques, and even assisting in rapid tissue diagnostics.
In the current study, the resonant sensor detected a significant difference
in tissue permittivities between different types of tumors, between
normal tissues and their associated tumors, and even in the presence
of metastasis. Our findings corroborate findings from other researchers.
O’Rourke and colleagues demonstrated that the dielectric properties
of hepatocellular carcinoma, an adult primary liver tumor, and other
metastatic tumors in the liver were 16% higher than normal liver tissues.[10] We noted similar findings in our study where
the resonant frequency of the hepatoblastoma tissue, a pediatric liver
cancer, was 7% higher than normal liver tissues.There are many
potential explanations for the differences in tissue
permittivity of tumor tissue compared to normal tissues and variance
between tumor types. The water versus lipid content of tissues and
tumors is well documented and serves as a basis for imaging techniques
such as computerized tomography, magnetic resonance imaging, and diffuse
optical spectroscopy.[14] These differences
in water content affect the tissue permittivity, with the relative
permittivity of water being about 80 times that of lipids.[15] These differences in water content would allow
the resonant sensor to distinguish between different types of tumors
and also between tumor and normal tissues. The water content of the
tissues may also be affected by cell processes such as apoptosis and
necrosis. These changes in tissue water content would be one explanation
for the noted changes in tissue permittivity demonstrated in tumors
that were treated with chemotherapeutics since they demonstrated significant
necrosis after treatment.The water content of tissues is also
affected by blood flow, which
varies significantly between organ systems. For example, the liver
receives about 27% of the cardiac output compared to about 21% in
the skeletal muscle. For the current study, to mitigate this potential
confounder, tumor tissues and organs were scanned ex vivo soon after animal euthanasia before significant fluid losses through
dehydration occurred. Therefore, the differences noted between resonant
frequencies of different organs or between flank and abdomen alive
versus dead tissues were likely not entirely secondary to blood flow.
Additionally, electrical conductivity of human tissues varies depending
upon pH and ionic makeup of the tissues. Studies evaluating resonant
frequencies of ionic solutions demonstrated differing resonant frequencies
that could be predicted based upon the ionic content.[16] Based on these findings, we postulate that the differences
in resonant frequencies in the organs and tissues studied might be
attributed to variances in their ionic makeup, rather than blood flow.The current resonant sensor design is based on a passive sensor
platform.[3] The advantages to using this
LC sensor system for characterizing biologic tissues are several.
The design is inexpensive to fabricate, which improves accessibility
to research groups and to health care professionals who practice in
resource limited areas, broadening the spectrum of potential applications
of the system. It also lends itself well for single-use applications,
such as integration into surgical pads, for convenient use during
surgery to rapidly assess tissue quality. Further, the VNA employed
in the current design has Bluetooth capabilities and does not require
a direct power source, rendering the LC sensor versatile in widely
differing environments.Aside from the physical and financial
advantages of the resonant
sensor system presented in these studies, the data that the system
is capable of generating are valuable and have the potential for multiple
clinical applications. For example, based on the capability of the
resonant sensor system to detect differences between normal versus
tumor tissues and between alive (viable) and dead (non-viable) tissues,
we believe that the resonant sensor system would allow for rapid detection
of core needle sample adequacy, potentially abrogating the need for
a more invasive procedure for solid tumor biopsy. Further, the ability
to differentiate a normal from a tumor tissue could aid in the real-time
determination of tumor resection margins. Finally, the ability to
detect metastatic disease could lead to the elimination of invasive
lymph node sampling that is currently required for many tumor diagnoses
such as breast cancer and melanoma.As with any experimental
device, there are drawbacks to using LC
sensors to measure the biologic activity of a system. The resonant
frequency of the resonant sensor and reader system is dependent upon
the mutual inductance between the sensor and reader. Mutual inductance
is a function of a coupling constant, which depends upon positional
alignment between the two inductors: the sensor and the reader.[17] If the tissues are scanned in positions that
are significantly different from one another, then the corresponding
resonant frequency values obtained will not accurately represent the
tissue permittivity.[18] The experiments
in this study were designed to control for the sensor position. When
resonant frequency values were obtained for the live murine models,
the exact position of the resonant sensor and reader was kept constant
between the experimental groups. When the tissues were excised and
scanned by being placed on top of the resonant sensor system, the
tissues were always placed in the exact same location on the sensor,
and the sensor itself was also placed in a constant location on the
reader. The LC sensor is also subject to pressure variations that
deform the resonant sensor leading to changes in inductance and capacitance.[19,20] If the pressure applied to the sensor by the experimental tissue
is not constant, then the data obtained may not accurately represent
the tissue permittivity and may not be reproducible. The current experimental
design factored in the variable of pressure. When the resonant frequency
was obtained for the live murine models, the distance between the
reader and the sensor was kept constant to ensure that a constant
pressure was applied between the two experimental devices across the
study. For the ex vivo studies, a sterile 2 mm trephine
was used to sample every tissue examined. Excising all the tissues
to the same volume significantly reduced the amount of pressure variation
between the tissues. Recent studies have been performed to overcome
the positional alignment sensitivity issues of the LC sensor.[21,22] By understanding the potential implications of misaligning the resonant
sensor system and maintaining consistent pressure on the system for
the various experiments, we were able to successfully design the experiments
to control for potential frequency variation and confidently report
the data.
Conclusions
The findings of this study demonstrate
that a portable, wireless,
LC resonant sensor system was capable of accurately characterizing
pediatric solid tumors both in vivo and ex
vivo. By detecting the permittivity of the tissue, the LC
sensor differentiated between live and dead tissues and between an
array of murine organs including the brain, spleen, liver, pancreas,
lung, kidney, and heart. The ability to differentiate these tissues
led to characterization between (i) pediatric solid tumors of differing
histology, (ii) normal tissues and tumor tissues, (iii) primary and
metastatic tissues, and (iv) treated and untreated tumors. Based on
these findings, we postulate that the LC sensor system could be translated
for novel clinical cancer applications that may vastly improve the
quality of care for patients. For example, the resonant sensor technology
could assist in the rapid determination of adequacy of tumor biopsy
specimens from the standpoint of obtaining a viable tissue for diagnosis.
The ability to accurately differentiate between cancerous and healthy
tissue demonstrates the potential role of the LC sensor in the rapid
determination of adequacy of tumor resection margins. Other clinical
applications include the early detection of metastasis, based on changes
in tissue permittivity. We will continue exploring the potential for
clinical applications of the LC sensor system as we believe that the
technology will have a pivotal role in future pediatric cancer research
and therapy.
Methods
Resonator Fabrication
A Silhouette Curio X-Y plotter
(Silhouette America, Inc., Lindon, UT) was used in combination with
an ultrafine point permanent sharpie marker to mask an Archimedean
spiral on the copper surface of copper-coated polyimide (DuPont Pyralux,
DuPont, Torrance, CA). The Archimedean spirals were designed using
the CAD software Rhinoceros5 (https://www.rhino3d.com) to have a 40 mm outer diameter with
a pitch size of 1.2 mm. After the spiral was masked onto the polyimide,
the unmasked copper was etched from the surface by submerging it into
a 2:1 volume ratio solution of hydrogen peroxide/hydrochloric acid
(H2O2/HCl). Acetone was then used to remove
the mask from the resonator, and the resonator was air-dried and encapsulated
in a Tegaderm dressing (3M, Maplewood, MN).
Signal Acquisition
A MetroVNA Deluxe 250 MHz (Metropwr,
Italy) acquired the data. This portable vector network analyzer (VNA)
was used in combination with a three-dimensionally (3D) printed antenna
holder to create the resonant reader. The printed holder contained
a dual-loop coil antenna with 54 mm diameter copper loops, with an
overlap of 26.7 mm3.[3] The VNA
in conjunction with JAVA-based software, vna/J (https://vnaj.dl2sba.com/), generated
the tissue resonant sensor data. The VNA monitored the S21 parameter
in the scattering parameter matrix.[23] The
scans had a start frequency of 0.1 MHz and an end frequency of 1.8
MHz, and the number of points (or frequency step size) was 10,000
kHz. The basic scan setup is diagrammed in Figure S4.
Data Analysis
The VNA used in the experiments determined
the transmission loss of the resonant sensor system. The transmission
loss of the resonant sensor system changes in both amplitude and peak
frequency based on the permittivity of the medium exposed to the resonant
sensor.[8] Once the data were collected for
each experiment via the MetroVNA, they were imported into the numeric
computing platform MATLAB for analysis (https://www.mathworks.com/products/matlab.html) (Figure S5). Several features of the
S21 parameter were quantified for the various experiments including
transmission loss and peak frequency of the resonant sensor system.
Resonant Sensor Data Collection
All tissues assessed
by the resonant sensors were human in origin propagated in athymic
nude mice (Fredricks, Charles River, Wilmington, MA). For external
live tissue scans, animals were anesthetized with 3% isoflurane and
placed in the supine position, and the LC sensor was placed on the
center of the area of interest. The reader was then placed over the
center of the LC sensor, and the vna/J software was used to obtain
the transmission loss data from the reader. The University of Alabama
at Birmingham Institutional Animal Care and Use Committee (UAB IACUC)
approved all animal experiments (IACUC-022020), and experiments were
conducted within institutional, national, and NIH guidelines. After
resonant frequency readings were recorded, the animals were humanely
euthanized per IACUC protocols with CO2 in their home cages
followed by cervical dislocation. Five minutes after euthanasia, the
resonant sensor and reader were used in a similar fashion to obtain
scans of an external dead tissue. Comparisons were completed between
the relative permittivities of a live versus dead tissue and between
external tissue types and conditions. External tissue areas scanned
included the abdomen, flank, and established flank pediatric solid
tumors.To obtain scans of internal tissues, mice were euthanized
as described above. The tissue of interest was extracted from the
mouse following euthanasia under sterile conditions outlined by IACUC
protocols (IACUC-022020). To minimize water loss from the specimen,
immediately following tissue harvest, a 2 mm trephine was used to
cut a standard, uniform core of the selected tissue, which was immediately
placed on the resonant sensor. The core of the tissue was placed on
the furthest edge of the resonant sensor, with the center of the tissue
oriented on the outside end of the Archimedean spiral to control for
any variation between experimental samples. This attention to positioning
was important as the outside of the resonator is more sensitive than
the center.[24] The reason for this difference
in sensitivity is because the electromagnetic field inner arms of
the Archimedean LC resonator have greater destructive interference
than the outer arms.[8] Samples were placed
on the sensor in the same location for each reading. The resonant
sensor was then positioned in the center of the reader, and scans
were obtained as previously described. Internal tissues included the
brain, heart, kidney, liver, lung, pancreas, spleen, bone, and established
pediatric solid tumors. The resonant frequencies were compared across
tissue types. All scanning procedures were conducted under similar
conditions, and prior to each scan, the MetroVNA was calibrated to
reduce the background signal from the resonant sensor. Each specimen
was scanned, removed, replaced, and scanned again at least three times
to account for variation caused by positioning.
Tumor Models
Flank tumors were established in 6 week
old female athymic nude mice (Fredrick, Charles River, Fredrick, MD).
The UAB IACUC (IACUC-022020) approved all experiments. Experiments
were conducted within institutional, national, and NIH guidelines.
Animals were maintained in the specific pathogen-free facility with
standard 12 h light/dark cycles and access to chow and water ad libitum. Long-term passage human neuroblastoma SK-N-AS
cells (1.8 × 106) in 25% Matrigel (BD Biosciences,
San Jose, CA) were injected into the right flank of the animals. To
propagate patient-derived xenograft tumors, a piece of tumor (100
mm3) (neuroblastoma, hepatoblastoma, osteosarcoma, medulloblastoma,
and neuroendocrine tumors) was injected into the right flank of mice
in a 50% Roswell Park Memorial Institute 1640 Medium (RPMI) and 50%
Matrigel (BD Biosciences). We developed these PDXs at our institution
under IACUC (IACUC-09186) and UAB Institutional Review Board (IRB,
IRB-130627006) approval. Tumors were monitored weekly with caliper
measurements and calculation of tumor volume accomplished using the
formula, where width is the shortest measurement.
When tumors reached 2000 mm3, animals were utilized for
scanning experiments.A second experiment was completed to monitor
changes in tissue permittivity in tumors over time. Established human
neuroblastoma SK-N-AS cells (1.8 × 106) in 25% Matrigel
(BD Biosciences, San Jose, CA) were injected into the right flank
of the animals (n = 20). These animals were followed
for 21 days for tumor growth, and scans were completed at the time
of volume measurements.A unique metastatic model was established
using the neuroendocrine
PDX. A neuroendocrine PDX flank tumor was harvested and dissociated
following our previously published protocol,[9] and cells were kept in culture for 24 h at 5% CO2 and
37 °C. PDX cells (2 × 106) were then resuspended
in 100 μL of sterile phosphate buffered saline (PBS) and injected
into the abdominal cavity of the animals via a 27-gauge needle. Animals
were weighed weekly, and the overall body condition was monitored
daily. Abdomens were palpated routinely, and upon detection of tumors,
mice were used for external and internal scans as described above.
Established tumor cell lines and PDX tumor cells were validated regularly
using short tandem repeat analysis (UAB, Genetics Core). Polymerase
chain reaction studies confirmed no murine contamination of the human
PDXs (UAB, Genetics Core), and the cells were deemed free of mycoplasma.
In Vivo Treatment Studies
Once SK-N-AS
neuroblastoma flank tumors reached a volume of 100 mm3,
animals were randomized to three groups to receive 100 μL of
either vehicle [N,N-dimethylacetamide
(DMA, 271012, Sigma-Aldrich) and Kolliphor HS 15 (Solutol, 42996,
Sigma-Aldrich), n = 5] or ATUX-792 (50 mg kg–1 in DMA and Solutol, n = 7) twice
daily by oral gavage. Flank tumors were scanned prior to euthanasia
and 5 min after euthanizing the animals. We then resected the tumors
and used a trephine core as described, placing each core on the same
location on the resonance sensor. The resonant frequency of tumors
treated with vehicle was compared to that of tumors treated with ATUX-792.In the second study, trametinib (Selleckchem, Houston, TX), a small
molecule inhibitor of MEK1/2, was tested as a targeted therapeutic
against the metastatic neuroendocrine PDX. Metastatic tumors were
established as previously described,[9] and
drug treatment was initiated 24 h after injection of cells. An external
tissue scan of the abdomen was obtained prior to treatment initiation.
Mice received 2 mg kg–1 trametinib (n = 5) or 4% dimethyl sulfoxide (DMSO, vehicle, n = 5) in corn oil via oral gavage daily. External abdominal tissue
scans were obtained three times a week for a total of 6 weeks. Due
to the limited tumor specimen in the treatment group, tissue core
scans of the tumors were not obtained. Analysis of the relative permittivities
was compared over time and between experimental groups.
Statistical Analysis
All scans were performed in at
least triplicate, and data were represented with at least three biologic
replicates. Data were reported as mean ± SEM. Statistical analysis
was completed with GraphPad Prism 9. Student’s t-test or analysis of variance (ANOVA) was used as appropriate to
determine statistical significance. A p-value ≤0.05
was considered significant.
Authors: S Merritt; G Gulsen; G Chiou; Y Chu; C Deng; A E Cerussi; A J Durkin; B J Tromberg; O Nalcioglu Journal: Technol Cancer Res Treat Date: 2003-12
Authors: Ann P O'Rourke; Mariya Lazebnik; John M Bertram; Mark C Converse; Susan C Hagness; John G Webster; David M Mahvi Journal: Phys Med Biol Date: 2007-07-18 Impact factor: 3.609
Authors: Sadaf Charkhabi; Kyle J Jackson; Andee M Beierle; Adam R Carr; Eric M Zellner; Nigel F Reuel Journal: ACS Sens Date: 2020-12-31 Impact factor: 7.711
Authors: Colin H Quinn; Andee M Beierle; Adele P Williams; Raoud Marayati; Laura V Bownes; Hooper R Markert; Jamie M Aye; Jerry E Stewart; Elizabeth Mroczek-Musulman; David K Crossman; Karina J Yoon; Elizabeth A Beierle Journal: Transl Oncol Date: 2021-04-19 Impact factor: 4.243
Authors: Adam R Carr; Yash H Patel; Charles R Neff; Sadaf Charkhabi; Nathaniel E Kallmyer; Hector F Angus; Nigel F Reuel Journal: NPJ Digit Med Date: 2020-04-30