Bruno Capurro1,2,3,4, Francisco Reina3, Anna Carrera3, Joan Carles Monllau1, Fernando Marqués-López1, Oliver Marín-Peña4,5, Raúl Torres-Eguía4,6, Marc Tey-Pons1,4. 1. Department of Orthopaedic Surgery and Traumatology, Hospital del Mar, Universitat Autònoma de Barcelona, Barcelona, Spain. 2. Department of Orthopaedic Surgery and Sport Traumatology, ReSport Clinic, Barcelona, Spain. 3. NEOMA Research Group, Department of Medical Sciences, Universitat de Girona, Girona, Spain. 4. GIPCA - Grupo Ibérico de cirugía Preservación de Cadera, Spain - Portugal. 5. Department of Orthopaedic Surgery and Traumatology, Hospital Infanta Leonor, Universidad Complutense de Madrid, Madrid, Spain. 6. Department of Orthopaedic Surgery and Traumatology, Clínica CEMTRO, Madrid, Spain.
Abstract
Background: Many advances have been made in hip labral repair and reconstruction and in the restoration of the suction seal. Purpose/Hypothesis: The purpose of this study was to evaluate the biomechanical effects of segmental labral reconstruction with a synthetic polyurethane scaffold (PS) in comparison with segmental labrectomy. Our hypothesis was that reconstruction with a icroporous polyurethane implant would normalize joint kinetics of the hip and restore the suction seal. Study Design: Controlled laboratory study. Methods: We used 10 hips from 5 fresh-frozen pelvises with an intact acetabular labrum without osteoarthritis. Using an intra-articular pressure measurement system, the contact area, contact pressure, and peak force were assessed for the following conditions: intact labrum, partial anterosuperior labrectomy, and PS reconstruction. For each condition, all specimens were analyzed in 4 positions (90° of flexion, 90° of flexion and internal rotation, 90° of flexion and external rotation, and 20° of extension) and underwent a labral seal test. The relative change from the intact condition was determined for all conditions and positions. Results: Compared with the intact labrum, labrectomy resulted in a significant decrease in the contact area (P < .001) and a significant increase in the peak force (P < .001) and contact pressure (P < .001) across all positions. Compared with labrectomy, PS reconstruction resulted in a significant increase in the contact area (P < .001) and a significant decrease in the contact pressure (P ≤ .02) and peak force (P < .001) across all positions. Compared with the intact labrum, PS reconstruction restored the contact area and peak force to normal values in all positions (P > .05), whereas the contact pressure was significantly decreased compared with labrectomy (P < .05) but did not return to normal values. The labral seal was lost in all specimens after labrectomy but was restored in 80% of the specimens after PS reconstruction. Conclusion: Femoroacetabular contact biomechanics significantly worsened after partial labrectomy; reconstruction using a PS restored the contact area and peak force to the intact state and improved the contact pressure increases seen after partial labrectomy. The contact area and peak force were normalized, and the labral seal was re-established in most cases. Clinical Relevance: This study provides biomechanical evidence for the use of a scaffold for labral reconstruction.
Background: Many advances have been made in hip labral repair and reconstruction and in the restoration of the suction seal. Purpose/Hypothesis: The purpose of this study was to evaluate the biomechanical effects of segmental labral reconstruction with a synthetic polyurethane scaffold (PS) in comparison with segmental labrectomy. Our hypothesis was that reconstruction with a icroporous polyurethane implant would normalize joint kinetics of the hip and restore the suction seal. Study Design: Controlled laboratory study. Methods: We used 10 hips from 5 fresh-frozen pelvises with an intact acetabular labrum without osteoarthritis. Using an intra-articular pressure measurement system, the contact area, contact pressure, and peak force were assessed for the following conditions: intact labrum, partial anterosuperior labrectomy, and PS reconstruction. For each condition, all specimens were analyzed in 4 positions (90° of flexion, 90° of flexion and internal rotation, 90° of flexion and external rotation, and 20° of extension) and underwent a labral seal test. The relative change from the intact condition was determined for all conditions and positions. Results: Compared with the intact labrum, labrectomy resulted in a significant decrease in the contact area (P < .001) and a significant increase in the peak force (P < .001) and contact pressure (P < .001) across all positions. Compared with labrectomy, PS reconstruction resulted in a significant increase in the contact area (P < .001) and a significant decrease in the contact pressure (P ≤ .02) and peak force (P < .001) across all positions. Compared with the intact labrum, PS reconstruction restored the contact area and peak force to normal values in all positions (P > .05), whereas the contact pressure was significantly decreased compared with labrectomy (P < .05) but did not return to normal values. The labral seal was lost in all specimens after labrectomy but was restored in 80% of the specimens after PS reconstruction. Conclusion: Femoroacetabular contact biomechanics significantly worsened after partial labrectomy; reconstruction using a PS restored the contact area and peak force to the intact state and improved the contact pressure increases seen after partial labrectomy. The contact area and peak force were normalized, and the labral seal was re-established in most cases. Clinical Relevance: This study provides biomechanical evidence for the use of a scaffold for labral reconstruction.
The acetabular labrum plays a crucial role in the stability, lubrication, and kinematics
of the hip joint.
The purpose of hip preservation surgery in the presence of femoroacetabular
abnormalities is to maintain and preserve the anatomy, correcting morphological
alterations and repairing the labral lesion (when possible instead of performing labral
debridement). When this repair is not possible, reconstructing the acetabular labrum
should be attempted with the aim of regaining joint function, decreasing pain, and
allowing a return to activities of daily living.When labral ruptures occur, the vast majority can be addressed with primary repair
techniques. With the shift in recent decades from debridement to preservation, treatment
methods that are aimed at restoring the negative pressure gradient created by the labral
suction seal and improving contact mechanics have become popular.
Labral repair has been shown to have better clinical outcomes than labrectomy,
but when repair is not possible, labral reconstruction is indicated.
Performing allogeneic labral transplantation can achieve good results, but graft
availability from tissue banks may be limited, and sizing may be challenging
particularly when a more severe labral rupture than expected is diagnosed
intraoperatively. In this situation, autografts can be used from the fascia lata or
hamstring with satisfactory results.
The use of these autografts requires express authorization from the patient
through informed consent and can also lead to possible morbidity in the donor area and
an increase in surgical time.As an alternative to human tissue usage, biodegradable, synthetic, polyethylene scaffolds
were developed more than 15 years ago for restoration of the meniscus in the knee, with
successful results clinically up to 10 years.
Owing to the abundant vascular supply in the acetabulum,
integration of the scaffold can be expected. With the lack of compressive and
shear forces found in the knee, comparable or superior results may be obtained when
implanted in the hip.The objective of this biomechanical study was to evaluate the effect of reconstruction of
an acetabular labral injury, caused by partial anterosuperior labrectomy, using an
artificial macroporous polyurethane scaffold (PS). Our hypothesis was that partial
anterosuperior labrectomy would decrease the contact area and increase the contact
pressure and peak force compared with the intact acetabular labrum, while reconstruction
with a macroporous PS would normalize joint kinetics of the hip and restore the labral
seal.
Methods
Specimens
The protocol for this study was approved by the ethics committee of the local
university. A total of 22 consecutively donated deidentified cadaveric hips (11
complete pelvises and proximal femurs) were obtained from a voluntary human body
donation program of the local university, which complies with all legal and
ethical procedures for human donation.The complete pelvises were subjected to a fluoroscopic examination to confirm
that there were no dysplastic hips, femoroacetabular impingement, or impairment
of the femoral head (Wiberg angle of 25°-40°, alpha angle <55°) and that
there was conservation of the joint space (Tönnis grade <2). No specimens
with a severe osteochondral lesion were detected radiographically; however, 4
specimens (2 pelvises) presented with severe osteochondral damage on direct
inspection and were therefore excluded.The cadaveric hips were handled following a previously described protocol;
however, they were not separated, and the pelvises were kept intact.
Briefly, the specimens were thawed at room temperature for 12 hours, then
carefully dissected, and all the muscles and soft tissue were systematically
eliminated. They were then refrozen to be thawed on the day of biomechanical
testing. After testing, the pelvises were frozen a third time using the same
protocol to preserve them for revalidation if necessary. None of the specimens
required a third thawing.
Biomechanical Testing
There were 3 complete pelvises (6 specimens) used for a pilot study to test
the calibration and sensor positioning and to refine the reconstruction
technique and biomechanical workflow. One pelvis (2 specimens) did not allow
for correct positioning of the sensor because of the small size of the
acetabulum. These cases were not included in the study. This left 5 pelvises
(10 hips) remaining for testing.The complete pelvis specimens were fixed in an anatomic position on a
biomechanical testing bench (Figure 1). This bench allows for
angular control in the 3 axes of joint mobility. In the absence of
servomotors for movement control, maximum mobility for each specimen can be
achieved. The design and methodology have been previously described.
Figure 1.
Schematic diagram of the bench used to secure the specimens.
Schematic diagram of the bench used to secure the specimens.Testing was carried out in 3 different clinical situations for each specimen:
(1) with the acetabular labrum intact, (2) after 3 cm–long anterior
labrectomy, and (3) after labral reconstruction with a PS.
Intact Labrum
Although normally carried out arthroscopically, for the purposes of this
study, we performed an open procedure. Wide 360° capsulotomy was
performed on each specimen to 2 mm of the acetabular rim, and the round
ligament was sectioned (Figure 2). Visual inspection
confirmed the integrity of the acetabular labrum and the state of the
acetabular cartilage. Capsulotomy was carried out to manipulate the
acetabular labrum to insert dynamic pressure sensors with a semicircle
design (model 4400; Tekscan). The sensors were calibrated according to
manufacturer guidelines and attached to the capsule and labrum with
3-point capsular sutures (silk 2/0) (Figure 2), achieving a static
position during testing and allowing for reproducible measurements.
Figure 2.
Tekscan sensor’s cotyloid cavity attachment after joint capsule
resection and round ligament sectioning to remove the femoral
head. Articular cartilage (crescent shape) (a);
acetabular labrum (b); distally sectioned round
ligament (c); acetabular fossa
(d); transverse ligament
(e); and ischial tuberosity
(f). The numbers show the distribution in
zones according to the clockface position.
Tekscan sensor’s cotyloid cavity attachment after joint capsule
resection and round ligament sectioning to remove the femoral
head. Articular cartilage (crescent shape) (a);
acetabular labrum (b); distally sectioned round
ligament (c); acetabular fossa
(d); transverse ligament
(e); and ischial tuberosity
(f). The numbers show the distribution in
zones according to the clockface position.A precision dynamometer (Digital Force Gauge SF-500; Beslands) was used
to ensure that stable pressure was maintained throughout the study for
all specimens (maximum load: 500 N/50 kg; load division value: 0.1
N/0.01 kg; inner sensor accuracy: ±0.5%; operating temperature: –15° to
35°C; transport temperature: 23° to 60°C; relative humidity: 15% to
80%).The peak force, contact pressure, and contact area of the acetabular
cartilage were measured after the introduction of the femoral head into
the acetabular cavity. The same 4 sequences of movement were reproduced
for each specimen. Dynamic testing began from the anatomic position in
extension, neutral rotation, and neutral abduction, and progressive
flexion was performed to 90°. The joint was then brought to maximum
internal rotation (flexion + IR), followed by maximum external rotation
(flexion + ER) and back to neutral, finishing with an excursion at 20°
of extension. The mean of 3 consecutive examinations was used for each
measurement. Movement was carried out manually, timing and documenting
each position. Throughout the study, a constant compressive force of 50
N was applied with the dynamometer, with a variation of ±5 N. Each cycle
lasted 30 seconds (±10%). If greater variations in strength or duration
occurred, they were discarded and the cycle repeated. Each test was
repeated 3 times, and the mean of the results was used (Figure 3).
Figure 3.
Cycle performed on a biomechanical bench with a 50-N dynamometer
fixed intramedullary in the distal femur with K-wire to
manipulate rotations: (A) 90° of flexion, (B) 90° of flexion
plus maximum internal rotation (arrow), (C) 90° of flexion plus
maximum external rotation (arrow), and (D) 20° of extension
(arrow).
Cycle performed on a biomechanical bench with a 50-N dynamometer
fixed intramedullary in the distal femur with K-wire to
manipulate rotations: (A) 90° of flexion, (B) 90° of flexion
plus maximum internal rotation (arrow), (C) 90° of flexion plus
maximum external rotation (arrow), and (D) 20° of extension
(arrow).
Labrectomy
After testing with the intact native labrum was completed, the pressure
sensor was detached from the capsule, and the capsule was retracted
laterally. A longitudinal labral cut of 3 cm in length in the
anterolateral zone was performed with a No. 10 scalpel, leaving a 2-mm
lateral margin for later attachment of the scaffold (Figure 4). The
sensor was reattached to the capsule and the specimen repositioned on
the bench. Testing was performed with the routine described above.
Figure 4.
Anterosuperior labrectomy of 3 cm performed on a right hip, with
resected labrum (a) and native labrum
(b). The number represents the 12-o’clock
position.
Anterosuperior labrectomy of 3 cm performed on a right hip, with
resected labrum (a) and native labrum
(b). The number represents the 12-o’clock
position.
PS Reconstruction
After testing in the labrectomy condition, the sensor was removed as
described above. Reconstruction was carried out with the insertion of a
biodegradable, synthetic PS (Actifit; Orteq). This scaffold is highly
porous and composed of aliphatic polyurethane (Figure 5). There were three 5-mm
anchors (Wedge Anchor II with No. 2 Force Fiber; Stryker) placed every 7
mm in the middle to secure the implant, and 1 side-to-side suture was
used to attach the labral remnant at each end (Figure 6). Because of the
quality of the bone for specimen 8, an extra anchor was used in the
middle of the scaffold. Testing was completed as described above.
Figure 5.
Macroporous polyurethane implant (Actifit).
Figure 6.
(A) Reconstruction with a polyurethane implant (specimen 4, left
hip) using 3 anchors and 2 side-to-side translabral sutures. (B)
Superior and side view of the sensor position fixed with silk
sutures. Sup, superior; Med, medial; Lat, lateral, Inf,
inferior.
Macroporous polyurethane implant (Actifit).(A) Reconstruction with a polyurethane implant (specimen 4, left
hip) using 3 anchors and 2 side-to-side translabral sutures. (B)
Superior and side view of the sensor position fixed with silk
sutures. Sup, superior; Med, medial; Lat, lateral, Inf,
inferior.
Labral Seal Testing
The suction seal test
consisted of a qualitative type of analysis that was performed with the
hip in a neutral anatomic position. After capsulotomy, a section of the round
ligament of the femoral head was rearticulated, and an axial distraction force
was applied according to Suppauksorn et al
and rated qualitatively to see if the seal was maintained. Losing the
seal or vacuum results in a negative sealing pressure that produces an audible
sound and dislocation of the hip. If the seal is not maintained, no resistance
to the distracting force is perceived, nor is the sound heard. This test was
performed on all specimens to determine if it was present in the intact labrum,
after labrectomy, and again after reconstruction with the PS. All labral seal
tests were performed by the same investigator (B.C.).
Statistical Analysis
We measured quantitative variables: maximum pressure determined by the peak force
and contact area on the acetabular surface. These 2 parameters were measured in
each specimen in 3 predetermined clinical situations. When measuring the contact
area, normalization of the values was necessary, as each hip was of a different
size and adaptation of the sensor may have been influenced by that difference.
This was achieved by dividing the raw data for each value of stage 1 (intact
labrum).PS values were normalized to analyze differences in the peak force, contact area,
and contact pressure and were compared with the intact state using measurements
from the 4 different positions (90° of flexion in neutral, flexion + IR, flexion
+ ER, and 20° of extension). Because multiple measurements were taken per
specimen, normalized values were modeled using repeated-measures analysis of
variance to account for potential correlations within each specimen. The least
squares means and 95% CIs are reported, along with a P value
that reflects the test: H0: mean = 1 versus H1: mean ≠ 1.
Statistical significance was set at P < .05 with an alpha
error of .05 and an effect size of 0.8. Statistical analyses were performed
using SAS (Version 9.4; SAS Institute).To assess each specific area of the acetabular surface, an analysis of the data
was carried out for each of the 4 positions. The peak force, contact area, and
contact pressure on the acetabular surface were compared in the following
situations: (1) intact labrum versus labrectomy, (2) intact labrum versus
PS-reconstructed labrum, and (3) labrectomy versus PS-reconstructed labrum. The
relative change from the intact condition (value of 1) was determined for all
conditions and positions.
Results
Contact Area
In comparison with the intact labrum, labrectomy resulted in a significant
decrease in the contact area. In flexion, it decreased to 82% of normal values
(95% CI, 0.77-0.87), in flexion + IR to 83% (95% CI, 0.78-0.88), in flexion + ER
to 83% (95% CI, 0.78-0.88), and in extension to 82% (95% CI, 0.76-0.86)
(P < .001 for all).After labrectomy, which decreased the contact area, PS reconstruction restored
the contact area to more than 96% of the intact labrum in all positions
(P > .05), with no statistically significant differences
found between PS and native values. In flexion, the contact area was restored to
96% (95% CI, 0.93-1.00; P = .061), in flexion + IR to 98% (95%
CI, 0.94-1.01; P = .187), in flexion + ER to 98% (95% CI,
0.94-1.02; P = .243), and in extension to 97% (95% CI,
0.93-1.00; P = .076). When the contact area of PS
reconstruction was compared with labrectomy, the values were significantly
increased in all positions (P < .001).
Contact Pressure
The contact pressure increased significantly after partial labrectomy compared
with the intact labrum in all positions (P < .001 for all).
In flexion, it increased 1.28% (95% CI, 1.19-1.38); in flexion + IR, 1.28% (95%
CI, 1.19-1.38); in flexion + ER, 1.32% (95% CI, 1.23-1.42); and in extension,
1.28% (95% CI, 1.19-1.38).Compared with the partial labrectomy state, PS reconstruction significantly
decreased the contact pressure, but the values still remained significantly
higher than the intact state in all positions (flexion: 11% increase [95% CI,
1.05-1.17]; P = .003) (flexion + IR: 8% [95% CI, 1.03-1.13];
P = .008) (flexion + ER: 11% [95% CI, 1.04-1.16];
P = .0004) (extension: 12% [95% CI, 1.03-1.21];
P = .02). Figure 7 represents a 3-dimensional visualization of the results.
When comparing the contact pressure of PS reconstruction with labrectomy, all
P values were statistically significant for all positions
(P = .02).
Figure 7.
A 3-dimensional (3D) visualization of contact pressure across all
positions. Interpretation of a Tekscan software image, scaled from blue
to red, where the areas of least pressure are dark blue and of the
highest pressure are red. ER, external rotation; IR, internal
rotation.
A 3-dimensional (3D) visualization of contact pressure across all
positions. Interpretation of a Tekscan software image, scaled from blue
to red, where the areas of least pressure are dark blue and of the
highest pressure are red. ER, external rotation; IR, internal
rotation.
Peak Force
When comparing the intact labrum with partial labrectomy, the peak force
increased significantly in all positions: 27% (95% CI, 1.19-1.35) in flexion,
25% (95% CI, 1.17-1.33) in flexion + IR, 22% (95% CI, 1.14-1.29) in flexion +
ER, and 26% (95% CI, 1.19-1.34) in extension (P < .001 for
all).When comparing the peak force between the intact labrum and PS reconstruction, we
observed a recovery of more than 95% of the peak force, with no statistically
significant differences found. In flexion, the peak force decreased to 5% above
normal values (95% CI, 0.99-1.11; P = .07) from the labrectomy
state; in flexion + IR, it decreased to 3% (95% CI, 0.99-1.07;
P = .20), in flexion + ER to 2% (95% CI, 0.99-1.05;
P = .26), and in extension to 4% (95% CI, 1.00-1.08;
P = .07). When comparing the peak force of PS
reconstruction with labrectomy, the differences were statistically significant
for all positions (P < .001).In summary, reconstruction with a synthetic PS significantly increased the
contact area (P < .001) and significantly decreased both the
contact pressure (P ≤ .02) and peak force (P
< .001) compared with the effects caused by partial labrectomy, bringing the
alteration in joint kinematics closer to the intact condition. Box plots for all
procedures and conditions can be found in Figure 8.
Figure 8.
Box plots showing the distribution of (A) contact area, (B) contact
pressure, and (C) peak force by procedure and position compared with
normal (1.0). Lab, labrectomy; Nor, normal; PS, polyurethane
scaffold.
Box plots showing the distribution of (A) contact area, (B) contact
pressure, and (C) peak force by procedure and position compared with
normal (1.0). Lab, labrectomy; Nor, normal; PS, polyurethane
scaffold.
Labral Seal
The labral seal was preserved in 100% of the intact labrum specimens but was lost
in 100% of cases after labrectomy. The seal was restored in 80% of the specimens
after PS reconstruction.
Discussion
The most important finding of this study was that partial reconstruction of the
labrum with a macroporous PS restored joint kinematics to that of the intact labrum,
restoring the contact area and peak force back to normal and improving the contact
pressure closer to normal than the partial labrectomy state. Labral reconstruction
with a PS restored the labral seal in the majority of cases.In anatomic studies, the acetabular labrum increased the contact area of the cotyloid
cavity by 28%.
This increases congruence and joint stability, allowing the correct
distribution of loads while reducing contact pressure and stress in the joint cartilage.
In all measurements made in this biomechanical study, partial anterosuperior
labrectomy decreased the contact area from normal, on average, to 82%. It is
interesting to note that despite being partial anterosuperior labrectomy of 3 cm, it
produced a decrease in the contact area similar to that of total labrectomy that was
reported in a previous study.
PS reconstruction managed to increase the contact area after labrectomy by
normalizing that parameter relative to the intact labrum. Our results showed that
reconstruction with a PS re-established the contact area and normalized
intra-articular stress, which was consistent with the cadaveric study by Philippon
et al,
demonstrating that labral reconstruction produced improvements in the hip
joint’s contact area and contact pressure compared with labral resection.One of the main functions of the labrum is to distribute pressures in the
femoroacetabular joint.
In vivo and in vitro biomechanical studies have shown that the intact labrum
allows direct pressure loads from solid-to-solid and interstitial fluid to be evenly
distributed within the femoroacetabular joint.
This is corroborated by the results of this study, which showed that the
intra-articular contact pressure increased significantly after partial
anterosuperior labrectomy compared with the intact labrum and that the contact
pressure decreased significantly with PS reconstruction. Although reconstruction
decreased the contact pressure, it did not return to normal values of the intact
labrum.There is a relationship between increased stress forces and damage to the chondral
surface, which has been described as an independent risk factor of osteoarthritis.
Areas of cartilage exposed to higher loads may have an increased risk of
damage and a reduced ability to effectively distribute loads during movement.
Shi et al
reported that labral reconstruction with an autograft in a porcine model
provides benefits in terms of reducing the development of osteoarthritis compared
with labral resection. They concluded that the development of osteoarthritis can be
attenuated by labral reconstruction, contributing to homeostasis of the hip joint.
In this study, the peak force increased significantly after partial
labrectomy but decreased significantly after reconstruction with the PS to
approximately 95% of normal values. While not normalizing the contact pressure to
the intact state, if it manages to restore the peak force close to normal, a
decrease in elevated tissue deformation and shear forces that produce fibrillation
and chondral delamination could occur,
which may attenuate the progression of osteoarthritis.The positions of 90° of flexion, 20° of extension, and maximum internal and external
rotation at 90° of flexion were chosen with the aim of analyzing whether joint
kinetics could be reconstructed throughout the cone of circumduction of the enarthrosis.
In the range of motion that we studied, reconstruction with a PS after
labrectomy normalized the contact area, contact pressure, and peak force. We also
observed that labral reconstruction restored joint kinetics in internal and external
rotation at 90°.Biomechanical cadaveric studies have shown that simulated labral injury or resection
conditions decrease the pressurization of intra-articular fluid
and decrease distraction resistance
but that repair and reconstruction improve the distracting stability of the
hip fluid seal. Ejnisman et al
demonstrated that incorporation of the allograft and maintenance of the
suction seal in the joint were obtained. To analyze this scenario in depth,
Suppauksorn et al
compared the results of primary labral repair of an anterosuperior lesion and
270° allograft reconstruction in 8 cadaveric specimens. They found that when
reconstruction was performed at 270°, there was a decrease in the intra-articular
contact area and a greater loss of the labral seal compared with labral repair. With
the “suction seal test” that they described, 100% of the specimens with labral
repair recovered the seal compared with only 22.5% with 270° reconstruction. Our
results using the same labral seal test showed that the labral seal after wide
capsulotomy was maintained in 100% of the intact labrum specimens, that 100% of
specimens after partial labrectomy lost their labral seal, but that it was
re-established in 80% of specimens after reconstruction with the PS. Unlike the
study of Suppauksorn et al, our study analyzed reconstruction after 3-cm partial
anterosuperior labrectomy, not total labrectomy after 270° reconstruction. This
could explain the difference between our 80% recovery of the seal and their 22.5%
rate with reconstruction at 270°. Based on these results, it could be suggested that
it is better to perform partial reconstruction rather than 270° reconstruction to
recover the labral seal function.
Limitations
There are limitations to our study. Although the initial sample size was 22
specimens, only 10 specimens were included in the biomechanical study. This was
a relative limitation, as most sample sizes in biomechanical cadaveric studies
are 5 to 10 specimens.
The Tekscan sensors utilized in this study were specifically designed for
use in the hip. They are arranged in a semicircle and connected by USB to a
computer. Their 5 connectors are approximately 1 cm wide and are separated by
1.5 cm. This requires wide capsulotomy with a section of the ligamentum teres to
be able to insert and fix the sensor in the required position, which may affect
the study in terms of the distribution of loads and the subjectivity of the
suction seal test. However, this was not considered a significant limitation, as
we compared 3 clinical conditions with the same instability that is produced by
total capsulotomy. The hip joint is a system that requires the capsule to
maintain joint liquid. Despite having lubricated the cartilage with saline and
cadaveric fat, it may not have remained in the same condition as normal.
Pressures measured in the study may differ from those measured in a surgical
setting, making the raw value of the data not transferable. However, this bias
is understood in all cadaveric biomechanical studies with a similar study protocol.
The biomechanical bench used in this study permitted the specimens to be
placed in an anatomic position and allowed for unlimited hip range of motion,
with analog control of all angles of motion. The angular measurements were made
on that bench with a visual analog system; however, angular measurements can
sustain a bias. A goal in this study was to standardize range of motion for each
specimen, limiting the differences among the specimens. Finally, the PS used in
this study was developed for meniscal replacement in the knee but was adapted to
the hip. Although there are clinical and histological studies of this implant
that show that it integrates well in the knee and no inflammatory response has
been reported up to 10 years,
it is largely unknown how this implant will act in the hip. We are
limited, at present, by the study of Tey-Pons et al,
who observed good clinical outcomes at 4 years’ follow-up and no
progression to osteoarthritis using the same scaffold as in this study.
Conclusion
The biomechanical effects of a synthetic PS for labral reconstruction are described
for the first time. Femoroacetabular contact biomechanics significantly worsened
after partial labrectomy; reconstruction with a PS restored the contact area and
peak force to the intact state and improved the contact pressure increases seen
after partial labrectomy. The contact area and peak force were normalized. The
labral seal was preserved after complete capsulotomy but was lost after partial
labrectomy and re-established in 80% of cases after reconstruction with a PS.
Further studies are needed to compare the results with other autograft
reconstruction techniques.
Authors: Daniel K Moon; Savio L-Y Woo; Yoshiyuki Takakura; Mary T Gabriel; Steven D Abramowitch Journal: J Biomech Date: 2005-04-05 Impact factor: 2.712
Authors: Peter Verdonk; Philippe Beaufils; Johan Bellemans; Patrick Djian; Eva-Lisa Heinrichs; Wouter Huysse; Heinz Laprell; Rainer Siebold; René Verdonk Journal: Am J Sports Med Date: 2012-02-09 Impact factor: 6.202
Authors: Cecile Toanen; Aad Dhollander; Paolo Bulgheroni; Giuseppe Filardo; Stefano Zaffagnini; Tim Spalding; Joan Carles Monllau; Pablo Gelber; Rene Verdonk; Philippe Beaufils; Nicolas Pujol; Erica Bulgheroni; Laura Asplin; Peter Verdonk Journal: Am J Sports Med Date: 2020-04-08 Impact factor: 6.202