In vitro and in vivo evaluation of a novel wired transmission magnetically controlled capsule endoscopy system for upper gastrointestinal examination.

Yaoping Zhang1,2, Lina Qu1,2, Yani Gou1,2, Jinyong Hao1,2, Yanglin Pan3, Xiaojun Huang4,5.   

Abstract

BACKGROUND: Magnetically controlled capsule endoscopy (MCCE) has recently increasingly been used for gastric examination. However, the image quality and esophageal observation is suboptimal. We developed a novel wired transmission magnetically controlled capsule endoscopy (WT-MCCE) system and evaluated its feasibility through in vitro and in vivo experiments.
METHODS: A plastic stomach model and a pathological upper gastrointestinal model were used to evaluate the performance of WT-MCCE in vitro experiments. Twice of examination in the two in vitro models by WT-MCCE were performed by 5 endoscopists who were experienced in performing wireless capsule endoscopy. The examination of traditional gastroscopy (Olympus, GIF-HQ290) in the pathological upper gastrointestinal model was set as the control. In vivo experiments were performed in a live canine model by 3 endoscopists, in which WT-MCCE was inserted with the assistance of gastroscopy. Measurements included maneuverability, examination time, visualization of gastric mucosa, image quality and diagnostic accuracy.
RESULTS: WT-MCCE showed good performance in both in vitro and in vivo experiments with excellent visualization of mucosa (75-100%). The mean operation time is 17.6 ± 2.7 min, 22.3 ± 1.9 min and 29.3 ± 3.4 min in three models, respectively. In pathological upper gastrointestinal model, all lesions, including esophageal varices, one polyp, one foreign body, two gastric ulcers and one duodenal ulcer, were detected by both WT-MCCE and traditional gastroscopy by all endoscopists. For the observation of esophagus and stomach in the canine model, WT-MCCE also showed excellent maneuverability and good image quality.
CONCLUSIONS: The novel WT-MCCE system performed well in evaluating upper gastrointestinal landmarks and lesions in two in vitro models, and showed good performance in a canine model. WT-MCCE may be potentially useful for diagnosis of esophageal and gastric diseases.
© 2022. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.

Entities:  

Keywords:  Magnetically controlled capsule endoscopy; Upper gastrointestinal examination; Wired transmission

Year:  2022        PMID: 36112221     DOI: 10.1007/s00464-022-09603-1

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   3.453


  2 in total

1.  Sedation for gastrointestinal endoscopy. Analysis of tolerance and complications.

Authors:  C Ciriza; L García; A Fernández; A Díez; M Delgado; A I San Sebastián
Journal:  Rev Esp Enferm Dig       Date:  2001-09       Impact factor: 2.086

Review 2.  The endoscopic diagnosis of early gastric cancer.

Authors:  Kenshi Yao
Journal:  Ann Gastroenterol       Date:  2013
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.