Chenchen Wang1,2, Panpan Tian3, Hao Cao1, Bin Sun1, Jincan Yan1, Yuan Xue1, Hualin Lin1, Tianhui Ren2, Sheng Han1,3, Xin Zhao4. 1. School of Chemical and Environmental Engineering, Shanghai Institute of Technology, 201418 Shanghai, China. 2. School of Chemistry and Chemical Engineering, Shanghai Jiao Tong University, 200240 Shanghai, China. 3. School of Chemistry and Chemical Engineering, Shihezi University, 832003 Shihezi, China. 4. Shanghai Key Laboratory of Orthopaedic Implants, Department of Orthopaedic Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, 200041 Shanghai, China.
Abstract
The poor biotribological properties and bioinertness of Ti6Al4V have restricted its application in biomedical materials. In this study, microgrooves of different widths were prepared on the surface of a Ti6Al4V alloy by laser treatment. The tribological properties under dry lubrication and simulated body fluid (SBF) lubrication conditions, the electrochemical corrosion properties in SBF solution, and the bone marrow mesenchymal stem cell (BMSC) behavior on the surfaces were systematically tested. The corresponding mechanisms were discussed. The results showed that Ti6Al4V with a microgroove width of 45 μm (Ti64-45) exhibited excellent wear resistance with decreasing wear rates of 89.79 and 85.43% under dry friction and SBF lubrication compared to the Ti64 sample, which might be due to the increase of surface microhardness. Moreover, the excellent anticorrosion performance of Ti64-45 was attributed to the grain refinement on the titanium alloy surface with a lower volume fraction ratio of β phase to α phase. In addition, the microgrooves with a width of 45 μm are more conducive to BMSC proliferation and adhesion, related to promoting cell signal transduction due to cell extrusion. These studies imply that the microgroove structures are potential for application in the medical field.
The poor biotribological properties and bioinertness of Ti6Al4V have restricted its application in biomedical materials. In this study, microgrooves of different widths were prepared on the surface of a Ti6Al4V alloy by laser treatment. The tribological properties under dry lubrication and simulated body fluid (SBF) lubrication conditions, the electrochemical corrosion properties in SBF solution, and the bone marrow mesenchymal stem cell (BMSC) behavior on the surfaces were systematically tested. The corresponding mechanisms were discussed. The results showed that Ti6Al4V with a microgroove width of 45 μm (Ti64-45) exhibited excellent wear resistance with decreasing wear rates of 89.79 and 85.43% under dry friction and SBF lubrication compared to the Ti64 sample, which might be due to the increase of surface microhardness. Moreover, the excellent anticorrosion performance of Ti64-45 was attributed to the grain refinement on the titanium alloy surface with a lower volume fraction ratio of β phase to α phase. In addition, the microgrooves with a width of 45 μm are more conducive to BMSC proliferation and adhesion, related to promoting cell signal transduction due to cell extrusion. These studies imply that the microgroove structures are potential for application in the medical field.
Ti6Al4V is widely used
for surgical implants. However, its poor
wear resistance and biological inertia have a great effect on the
service life of titanium alloy implants, especially bone implants.[1−4] The debris from wear and corrosion might react with bone tissues,
leading to implant loosening.[5−7] Moreover, the weak interface bonding
between the implant and bone also accelerates implant failure.[8,9] Therefore, it is necessary to improve the antiwear and anticorrosion
properties and bioactivity of titanium alloys through surface modification.Generally, surface treatment methods include surface texturing
and surface coating techniques.[10−12] Among them, laser processing
technology, as a popular surface texturing method, has been concerned
due to its nontoxic, environmentally friendly, simple, feasible, accurate,
and controllable features. Several research studies reported that
the microstructures obtained by laser texturing technology have a
positive effect on the surface properties of materials, such as the
improvement of tribological behavior, electrochemical corrosion performance,
and biological activity.[13−15] Wu et al.[16] studied the influence of laser texture micropatterns on
the tribological properties of titanium alloys and pointed out that
the antiwear properties of titanium alloys were significantly improved
in the friction process. The reason might be attributed to the increasing
bearing capacity and the function of capturing debris from the structure.
Similar studies have been confirmed by other researchers.[17] Xu et al.[18] enhanced
the corrosion resistance of the titanium alloy surface by laser processing.
It was found that the effect of the treatment on the corrosion resistance
of the surface was mainly due to the phase transformation of the substrate.
Kumari et al.[19] also pointed out that the
significant improvement of anticorrosion on the laser-treated material
surface might be due to the surface grain refinement and homogenization
of the microstructure. Furthermore, according to the previous research,[15] the laser-textured microstructure can affect
the cell behavior and bone integration of implants. Some researchers
have pointed out that among these microstructures, laser-microgrooved
structures have a positive promotion effect on the tribological behavior,
anticorrosion properties, and bioactivity of the material surface.[20−23] Therefore, texturing microgroove structures on the titanium alloy’s
surface is a feasible method for improving implant surface properties.In addition, as previous reports described, the laser microgroove
parameters might significantly affect the surface performance.[5,24] Several studies[25−28] indicated that the widths of microgrooves had an important effect
on the wear resistance, corrosion resistance, and cell response of
the material surface, and they pointed out that the difference in
mechanical properties and the microstructure crystal phase structure
are the main factors affecting surface properties. Bhaduri et al.[29] reported that the groove arrays with a width
of 40 um effectively improved the tribological performance on tungsten
carbide (WC) blocks. Moreover, Raimbault et al.[30] pointed out that cellular behavior was sensitive to the
width of the microgroove, related to the arrangement and growth of
cells. Gui et al.[31] prepared the grooves
on the silicon substrate with a width of 5–20 μm and
implied that the groove with a 20 μm width increased the number
of osteoblast soars. However, there is no unified conclusion on the
optimistic microgroove width, beneficial to the improvements of antiwear,
anticorrosion, and bioactivity simultaneously. Therefore, it is necessary
to investigate the effect of textured microgroove patterns on the
enhancement of tribological properties, corrosion resistance, and
biological activity.In this study, microgrooves of different
widths were processed
on the surface of titanium alloy by adjusting UV nanosecond laser
parameters. The morphology and chemical composition of the textured
surface were characterized by scanning electron microscopy (SEM),
laser confocal microscopy (LSCM), and X-ray diffraction (XRD). The
effects of microgroove width on surface roughness, mechanical properties,
and wettability of the titanium alloy were investigated. The biotribological
properties in dry friction and SBF solution, corrosion resistance,
and bone marrow mesenchymal stem cell (BMSC) behavior of the titanium
alloy with different groove widths were evaluated. The corresponding
mechanisms of antiwear, anticorrosion, and bioactivity were analyzed
through surface characterization.
Materials and Methods
Materials
Ti6Al4V alloy plates (medical
grade, referred to as Ti64) were purchased from Baoji Junhang Metal
Material Co., Ltd. and the chemical composition of the Ti6Al4V alloy
is listed in Table S1. They were machined
into dimensions of 10 mm × 10 mm × 2 mm (length × width
× thickness) by wire cutting, and polished to a roughness (Ra) of 0.2 ± 0.02 μm. They were ultrasonically
cleaned with deionized water and anhydrous ethanol three times, respectively.
Si3N4 ceramic ball was purchased from Shanghai
Wha Bearing Co., Ltd. Acetone, anhydrous ethanol, and other solvents
in this work are analytically pure, purchased from Shanghai Sinopharm
Group Co., Ltd. Simulated body fluid (SBF) was purchased from Xian
Hutt Biological Co., Ltd. Its solution composition is shown in Table S2. Fetal bovine serum was purchased from
Gibco. Paraformaldehyde, penicillin, streptomycin, 4′,6-diguinyl-2-phenyl
indole, and rhodamine ghost pen ring peptide were purchased from Sigma-Aldrich.
Cell counting kit-8 was purchased from Dojindo Laboratories in Kumamoto,
Japan.
Laser Equipment and Parameters
A
UV nanosecond laser (Shanghai Fermi Laser Processing Co., Ltd.) with
a wavelength of 355 nm and total power of 5 W was applied to process
a microgroove structure on the surface of a Ti6Al4V alloy plate. The
schematic diagram of the principle is shown in Figure . Microgroove patterns with different widths
were obtained by adjusting the laser processing parameters. The corresponding
laser parameters are shown in Table . The laser-processed titanium alloy samples were referred
to as Ti64-25, Ti64-35, Ti64-45, Ti64-55, and Ti64-65, according to
the width of the microgrooves. The textured parameters of the sample
can be marked as width (W), spacing (λ), and
depth (D). At the same time, the corresponding laser
energy densities (E) were calculated by formula , respectively.[32] The prepared samples were ultrasonically cleaned
with acetone and deionized water for 15 min.where E is the laser energy
density, P is the average power, Z is the frequency, and r is the spot radius.
Figure 1
Schematic of
the laser surface-microgrooved process on Ti6Al4V
alloy plates.
Table 1
Surface-Microgrooved Laser Parameters
parameters
pulse speed (mm/s)
frequency (kHz)
power (W)
distance
(cm)
scan times
spot diameter (μm)
laser energy density (J/cm2)
Ti64-25
400
35
1.5
11.7
37
10
54.6
Ti64-35
300
30
2.5
12.4
42
5
424.4
Ti64-45
600
32
3
12.7
55
5
477.5
Ti64-55
300
30
2.7
12.0
50
10
114.6
Ti64-65
700
30
3
11.4
124
10
127.3
Schematic of
the laser surface-microgrooved process on Ti6Al4V
alloy plates.
Surface Wettability Measurement
The
static contact angle (CA) of the sample surface was measured by the
droplet suspension method with a contact angle tester (DSA30, Kruss,
German). The wettability of the titanium alloy surface was further
evaluated by calculating the CA value. The detail test method was
as follows: the deionized water was dropped on the surface of Ti64,
Ti64-25, Ti64-35, Ti64-45, Ti64-55, and Ti64-65 sample plates. The
CA value of the sample surface is obtained by calculating the spherical
approximate value of the surface droplet. Each group of samples was
tested five times and the average value was taken.
Surface Microhardness Measurement
A Vickers hardness tester (MC010-HVS-1000, China) was used to measure
the surface hardness of Ti64, Ti64-25, Ti64-35, Ti64-45, Ti64-55,
and Ti64-65 sample plates. The set load was 100 g. Samples were tested
five times under each condition and the obtained average value was
the final microhardness.
Electrochemical Corrosion and Biotribological
Property Tests
To study the influence of the microgroove
structure on the corrosion resistance of the titanium alloy surface
in SBF solution, the electrochemical corrosion behavior of sample
plates (Ti64, Ti64-25, Ti64-35, Ti64-45, Ti64-55, Ti64-65) in SBF
solution at 37 °C was tested by an electrochemical workstation
(CHIT60E, Shanghai Chenhua Co., Ltd.). Moreover, the corrosion potential
and self-corrosion current density of the samples were analyzed through
the workstation.The tribological tests of titanium alloy samples
were recorded using a tribometer (UMT-3, Bruker). The test module
was a reciprocating ball–plate module. The test conditions
of the sample under dry friction and SBF solution lubrication were
as follows: room temperature, applied load 1 N, frequency 1 Hz, amplitude
2 mm, and test time 1800 s. As shown in the schematic diagram of tribological
measurements (Figure ),[8,9] the upper friction pair was a Si3N4 ball (Φ 6 mm), with a Ra of 0.02 μm and a Vickers hardness of 1500 HV. The titanium
alloy sample was the lower friction pair. The real-time friction coefficient
curve was automatically recorded by a computer. Each group of tribological
tests was tested three times, and the average value of the average
friction coefficient was taken as the final data.
Figure 2
Schematic diagram of
biotribological measurements using a tribometer.
Schematic diagram of
biotribological measurements using a tribometer.After the friction tests were completed, the wear
rate (k, mm3/(N·m)) of the surface
of those titanium
alloy samples was calculated according to formula .[8]where V is the wear volume, X is the sliding distance, and FN is the applied load.
In Vitro Bioactivity Test
The cell
evaluation experiment on the titanium alloy surface before and after
the microgroove treatment follows the principles of the Declaration
of Helsinki. Before the experiment, six groups of titanium alloy plates
(including Ti64, Ti64-25, Ti64-35, Ti64-45, Ti64-55, and Ti64-65)
were sterilized at high temperatures and high pressures. In addition,
the BMSCs were isolated from female C57BL/6 mice aged 4–6 weeks.
First, BMSCs were washed out of the tibia and femur of the mouse by
MEM Alpha modification (α-MEM) in a sterile environment. Subsequently,
the cells were centrifuged at 1000 rpm for 8 min, and cultured in
α-MEM containing 10% fetal bovine serum and 1% penicillin and
streptomycin, and the temperature was controlled to 37 °C. After
the cells were cultured for 3 days, the original medium was replaced
with a fresh medium. The cell confluence rate reaches 80–90%
for the following cell proliferation experiments.To evaluate
the cell viability of BMSCs on the titanium alloy sample plates, BMSCs
(2 × 104 cells/mL) were inoculated on 24-well culture
plates containing different samples. After 1, 3, and 7 days of culture,
according to the manufacturer’s instructions, the activity
of BMSCs on the samples was quantitatively tested with a cell counting
kit-8 (CCK-8). In addition, the adhesion performance of BMSCs on Ti64,
Ti64-25, Ti64-35, Ti64-45, Ti64-55, and Ti64-65 samples was studied.
First, BMSCs were inoculated and cultured on the sample plates. After
1 day, the cells on the sample plates were fixed with 4% paraformaldehyde
for 15 min. Subsequently, the fixed cells were washed three times
with a phosphate buffer solution. After that, the cells were cultured
in DAPI solution for 5 min to stain the nuclei. The cytoskeleton was
stained by immersing in rhodamine cyclopeptide solution for 30 min.
Finally, the cells that adhered to the sample surface were characterized
with a fluorescence microscope. In the fluorescence image, the nucleus
is marked in blue color and the cytoskeleton is marked in red color.
Analysis Methods
The microstructure
and chemical composition on the surfaces of these titanium alloy samples
before and after texturing were characterized by a scanning electron
microscope (SEM, FEI Nova 450) and an energy-dispersive X-ray spectrometer
(EDX) at 10 kV voltage. The three-dimensional topography of the sample
surface was characterized by a confocal laser microscope (CLM, VK-X100).
VK analysis software was used to measure the surface roughness of
titanium alloys with different microgroove widths.X-ray diffractometer
(XRD, D8 Advance, Germany) was used to analyze the crystal phase state
of Ti64, Ti64-25, Ti64-35, Ti64-45, Ti64-55, and Ti64-65 samples.
To obtain the ratio of the volume fraction of the α phase to
the β phase on the sample surface, Origin 9.1 software was applied
to fit the XRD spectra of each sample.[33] According to formulas and 4, the volume fraction of the α
phase and the volume fraction ratio of the α phase to the β
phase can be calculated, respectively.[18]where φα is the volume
fraction of the α phase on the titanium alloy, Sα and Sβ are,
respectively, the sum of the peak areas of the α phase and the
β phase (the 2θ angle range is 34–43°), and
φ is the volume ratio of the β phase to the α phase.A three-dimensional optical profiler (white light interferometer,
WLI, Contour GT-K0, Bruker, Germany) was conducted to test the 3D
profile and wear rate of the worn surface of the samples.
Experimental Results
Morphological Characterization
The
surface morphology of the titanium alloy treated with UV nanosecond
laser treatment was characterized by SEM and CLM, as shown in Figure . It can be seen
from Figure a–f
(SEM images) that the groove patterns with different groove widths
were obtained on the surface of the titanium alloy by controlling
the laser texturing parameters. Moreover, a small amount of deposition
was observed on the edge of the microgrooves of the titanium alloy.
This might be attributed to the formation of a remelting layer on
the titanium alloy surface after the thermal effect during laser processing. Figure a1–f1 shows
the three-dimensional topography of Ti64, Ti64-25, Ti64-35, Ti64-45,
Ti64-55, and Ti64-65 samples. The corresponding sample parameters
were measured and are listed in Table . It can be seen from Table that the groove width (W) and the groove spacing (λ) are precisely controlled during
the laser treatment. The depth (D) was in the range
of 9–13 μm.
Figure 3
Surface morphologies of titanium alloy specimens
before and after
texturing. (a–f) SEM images of Ti64, Ti64-25, Ti64-35, Ti64-45,
Ti64-55, and Ti64-65 plates; (a1–f1) 3D topography of Ti64,
Ti64-25, Ti64-35, Ti64-45, Ti64-55, and Ti64-65 plates.
Table 2
Parameters of Microgrooved Ti6Al4V
Alloy Platesa
samples
W (μm)
λ (μm)
D (μm)
Ti64-25
25
100
10.4
Ti64-35
35
100
9.8
Ti64-45
45
100
11.1
Ti64-55
55
100
10.9
Ti64-65
65
100
12.3
Note: W represents
the width of the textured microgroove, λ represents the spacing
of the textured microgroove, and D represents the
depth of the textured microgroove, with a range of 9–13 μm.
Surface morphologies of titanium alloy specimens
before and after
texturing. (a–f) SEM images of Ti64, Ti64-25, Ti64-35, Ti64-45,
Ti64-55, and Ti64-65 plates; (a1–f1) 3D topography of Ti64,
Ti64-25, Ti64-35, Ti64-45, Ti64-55, and Ti64-65 plates.Note: W represents
the width of the textured microgroove, λ represents the spacing
of the textured microgroove, and D represents the
depth of the textured microgroove, with a range of 9–13 μm.
Surface Roughness of Titanium Alloy Samples
The surface roughness of titanium alloy samples before and after
laser processing were analyzed and are listed in Table . This mainly includes the arithmetic
mean deviation of the contour (Ra), the
root-mean-square value of the deviation of the contour mean line (Rq), and the maximum height of the contour (Rz). Among them, Ra and Rq are the most important evaluation
parameters. It can be seen from Table that the Ra and Rq of the titanium alloy samples showed the same
sequence as follows: Ti64 < Ti64-25 < Ti64-35 < Ti64-55 <
Ti64-45 < Ti64-65. According to previous reports, the changes in
the surface roughness of laser-textured samples were mainly attributed
to two important laser parameters: laser intensity and pulse overlap
times.[34,35] According to the laser parameters listed
in Table , it can
be found that with the increasing groove width, the required laser
power and the number of pulse overlaps exhibited an increasing trend.
However, the fact that the surface roughness of Ti64-45 was slightly
larger than that of the Ti64-55 sample, might result from the higher
laser power (laser power of 3 W) and more pulse overlaps (55 times)
on Ti64-45 than that on Ti64-55 (laser power of 2.7 W) and the number
of pulse overlaps (50 times).
Table 3
Surface Roughness of Titanium Samples
samples
Ra (μm)
Rq (μm)
Rz (μm)
Ti64
0.19
0.23
1.17
Ti64-25
3.09
4.99
58.29
Ti64-35
3.39
4.66
31.29
Ti64-45
6.43
7.80
42.52
Ti64-55
6.37
7.38
39.96
Ti64-65
10.73
13.43
80.45
XRD Characterization of Titanium Alloy Samples
To analyze the effect of ultraviolet laser treatment on the crystalline
state of the titanium alloy surface, Figure shows the partial XRD spectrum of titanium
alloy specimens before and after texturing. The corresponding crystal
plane diffraction peaks are marked and two phases (α phase (hcp)
and β phase (bcc)) are observed in this figure. It can be clearly
seen from Figure that
after laser processing, the intensity of the 002 hcp peak on the surface
of the titanium alloy sample reduced, while the intensity of the 101
hcp peak increased. Moreover, compared with the Ti64 sample, a wider
diffraction peak was observed on the surface of the microgrooved titanium
alloy, which is consistent with previous reports.[18] In addition, the volume fraction ratio of the α phase
and β phase was calculated by fitting the peak areas of the
α and β phases (α phase is the yellow area and β
phase is the red area). The results are shown in Table . The volume ratio (φ)
of the β phase to the α phase (φ%) was sorted as
follows: Ti64-45 (5.51%) < Ti64-35 (11.16%) < Ti64-65 (11.54%)
Figure 4
XRD spectrum
magnifications of titanium alloy specimens before
and after texturing, including Ti64, Ti64-25, Ti64-35, Ti64-45, Ti64-55,
and Ti64-65 plates.
Table 4
Phase Volume Fraction of Titanium
Alloy Samples
samples
phase
φα (%)
φβ (%)
φ (%)
Ti64
α + β
87.15
12.85
14.74
Ti64-25
α + β
88.41
12.47
14.25
Ti64-35
α + β
89.65
10.04
11.16
Ti64-45
α + β
94.78
5.22
5.51
Ti64-55
α + β
87.53
11.59
13.11
Ti64-65
α + β
89.96
10.35
11.54
XRD spectrum
magnifications of titanium alloy specimens before
and after texturing, including Ti64, Ti64-25, Ti64-35, Ti64-45, Ti64-55,
and Ti64-65 plates.The results might be further explained by the Burgers
orientation
relationship.[33,36] The surface temperature of the
Ti6Al4V alloy would increase rapidly to above 882.5 °C during
the laser ablation process, resulting in the remelting of the heat-affected
zone on the surface of titanium alloys. After that, as the surface
temperature of the titanium alloy reduced rapidly, the body-centered
cubic structure (β phase) of the surface crystals would partially
transfer to the hexagonal lattice structure (α phase). Moreover,
the previous reports[37,38] indicated that the phase transformation
on these samples was related to the laser energy density during the
laser processing. Compared with the laser-treated titanium alloy samples,
it can be found from Tables and 4 that the laser-microgrooved
samples treated with a larger E value exhibited a
lower β/α value, and Ti-45 showed the highest E value. Zafari et al.[37] pointed
out that a larger laser energy density would lead to a large thermal
gradient and a faster cooling rate. It is known that due to the rapid
solidification, the microstructure of the laser-textured part is prone
to form very fine needles of α′ prime martensite rather
than forming the β phase during the remelting process.[26] Therefore, Ti-45 has the lowest β/α
content. Besides, with the increase of laser energy density, the lamella
width of a single α phase decreased, while the smaller columnar
crystal width increased.[38] Therefore, after
laser treatment, the grain microstructures were refined, and Ti-45
with a higher E value might exhibit a more refined
microstructure.
Surface Wettability of Titanium Alloy Samples
As shown in Figure , the effect of laser treatment on the wettability of the titanium
alloy surface was evaluated by calculating the contact angle of water
droplets on the surface of the sample. The wetting behavior of the
sample surface might be affected, due to the anisotropy of the surface
microgroove pattern.[39] According to the
observation directions, the contact angles were abbreviated as contact
angle perpendicular to the groove direction (VCA) and the contact
angle parallel to the groove direction (PCA). It can be seen from Figure a that the VCA value
of the titanium alloy surface decreased after laser treatment. Moreover,
as the groove width increased, the VCA value exhibited a decreasing
trend, suggesting the improving surface wettability. This might be
ascribed to the fact that the groove structure with a certain depth
has the function of storing water droplets, making it easier to spread
the water droplets on the surface of the titanium alloy and exhibiting
a lower contact angle. Moreover, as the width of the microgrooves
increased, the volume of the stored water droplets increased and the
contact angle decreased. The contact angle parallel to the groove
direction is shown in Figure b. It can be intuitively found that the PCA values of Ti64-25
and Ti64-35 were higher than those of the untreated titanium alloy
samples. The reason was ascribed to the anisotropic wettability of
the laser microgroove structure, which is consistent with the previous
report.[39] The squeezing effect on the water
droplets of the groove structure would lead to a larger PCA value.
Therefore, the above results indicated that the surface wettability
of Ti6Al4V was enhanced through texturing the groove structure with
a width greater than 45 μm.
Figure 5
Measurement of wetting properties on the
titanium alloy specimens
before and after texturing. (a) CA values in the vertical observation,
referred to as VCA and (b) CA values in the parallel observation,
referred to as PCA.
Measurement of wetting properties on the
titanium alloy specimens
before and after texturing. (a) CA values in the vertical observation,
referred to as VCA and (b) CA values in the parallel observation,
referred to as PCA.
Microhardness of Titanium Alloy Samples
Figure shows the
microhardnesss of titanium alloy samples before and after laser treatment.
It can be seen that compared with the Ti64 sample, the microhardness
of the grooved titanium alloy surface was significantly increased,
in the following sequence: Ti64-45 (599 HV) > Ti64-35 (437 HV)
> Ti64-25
(416 HV) > Ti64-55 (349 HV) > Ti64-65 (336 HV) > Ti64 (303
HV). It
can be calculated that the microhardness of Ti64-45 increased by 97.7%,
exhibiting the highest microhardness. This might be related to the
organization’s homogenization on the surface after laser texturing.
As described in the XRD spectrum analysis (Figure ), it was found that the obvious transfer
of the crystal phase from the β phase to the α phase appeared
on the Ti64-45 sample. This phase transformation would further refine
the grain of the titanium alloy and hinder the movement of dislocations,
thus increasing the surface microhardness.
Figure 6
Measurement of microhardness
on the titanium alloy specimens before
and after texturing.
Measurement of microhardness
on the titanium alloy specimens before
and after texturing.
Corrosion Resistance Performance
To evaluate the effect of laser-grooved treatment on the corrosion
resistance of the titanium alloy surface, the electrochemical corrosion
performance of the samples in the SBF solution was tested. The results
are shown in Figure and Table . Figure shows the Nyquist
impedance diagram and Tafel curve of the Ti6Al4V samples before and
after the surface microgrooves. According to the previous report,[8] the larger the radius of curvature in the Nyquist
impedance diagram, the better the corrosion resistance of the sample
surface. It can be seen from Figure a1 that the radius of curvature of the microgrooved
titanium alloy sample was significantly larger than that of the blank
titanium alloy plate, in the following order: Ti64-45 > Ti64-35
>
Ti64-65 > Ti64-55 > Ti64-25 > Ti64. This implied that laser
texturing
treatment improved the corrosion resistance of the titanium alloy
surface. Figure b
is the Tafel curve of these samples. The fitting corrosion current
density (icorr), corrosion potential,
and inhibition efficiency (refer to formula (28)) were calculated
and shown in Table . It was found that the corrosion current of the samples has a negative
correlation with the radius of curvature. Among them, the corrosion
current density of the Ti64-45 sample was the smallest (7.189 ×
10–7) compared with the Ti64 sample (1.902 ×
10–6). Moreover, Ti64-45 exhibited a higher inhibition
efficiency (62.20%), further suggesting the excellent corrosion resistance
of Ti64-45 samples.where icorr (untreated)
and icorr (grooved) are the corrosion
current density values without and with the coating, respectively.
Figure 7
Anticorrosion
property measurements on the titanium alloy specimens
before and after texturing. (a) Nyquist plots, including the magnification
(a1) of the six different plates and (b) Tafel plots of the six different
plates.
Table 5
Electrochemical Corrosion Parameters
of Titanium Alloy Plates in SBF Solution
samples
corrosion potential (mV)
corrosion current density (A/cm2)
inhibition efficiency (%)
Ti64
–576
1.902 × 10–6
Ti64-25
–421
1.710 × 10–6
10.09
Ti64-35
–475
9.525 × 10–7
49.92
Ti64-45
–495
7.189 × 10–7
62.20
Ti64-55
–527
1.440 × 10–6
24.29
Ti64-65
–612
1.169 × 10–6
38.54
Anticorrosion
property measurements on the titanium alloy specimens
before and after texturing. (a) Nyquist plots, including the magnification
(a1) of the six different plates and (b) Tafel plots of the six different
plates.
Biotribological Properties
Figure shows the tribological
properties of the titanium alloy before and after the surface texturing
in dry friction and SBF lubrication conditions. Among them, the variation
curves of friction coefficients (COF) with the time of the six group
samples during the dry friction and SBF lubrication process are, respectively,
displayed in Figure a,b. It can be seen from these figures that after a running-in period
during the friction process, the COF curve trend stabilizes with the
extension of time under both lubrication environments. The corresponding
average friction coefficient (referred to as ACOF) is shown in Figure c and Table . It can be found that under
dry friction conditions, the friction coefficient of the microgrooved
titanium alloy surface was lower than that of the untreated titanium
alloy sample. Moreover, the groove width has a significant effect
on the reduction of COF. Among these microgrooved samples, the Ti64-45
sample showed the lowest ACOF value (0.215), implying excellent antifriction
properties in dry conditions. While under SBF lubrication, the friction
coefficient of the titanium alloy has not been effectively reduced
after laser-grooved treatment. The above results indicated that the
microgroove structure, especially with the groove width of 45 μm,
can effectively reduce the friction coefficient on the Ti6Al4V substrate
in dry sliding friction.
Figure 8
Coefficient of friction (COF) and wear rates
of titanium alloy
specimens before and after texturing under dry friction and SBF lubrication
conditions. (a) Change of COF with sliding time under dry conditions;
(b) change of COF with sliding time in SBF lubrication; (c) average
coefficient of friction (ACOF) under both conditions; and (d) wear
rates of the five plates.
Table 6
Frictional Performance Parameters
of Samples under Dry Friction Conditions
samples
average coefficient of friction
wear rate, 10–3 mm3/(N·m)
ratio
of the reducing wear rate (%)
Ti64
0.435
0.235
Ti64-25
0.469
0.078
66.80
Ti64-35
0.289
0.053
77.45
Ti64-45
0.215
0.024
89.79
Ti64-55
0.359
0.034
85.53
Ti64-65
0.290
0.047
80.00
Coefficient of friction (COF) and wear rates
of titanium alloy
specimens before and after texturing under dry friction and SBF lubrication
conditions. (a) Change of COF with sliding time under dry conditions;
(b) change of COF with sliding time in SBF lubrication; (c) average
coefficient of friction (ACOF) under both conditions; and (d) wear
rates of the five plates.The surface wear rates on these titanium alloys during
the friction
process were calculated and displayed in Figure d and Table . It can be found that the wear rates of the six groups
of titanium alloy samples in SBF lubricants were higher than those
in the dry friction process. Moreover, a similar change trend in the
wear rate on the titanium alloy sample was observed (see Figure a). The wear rates
presented a trend of first increasing and then decreasing with the
increase of groove width. Compared with other samples, the Ti64-45
sample showed the lowest wear rate (0.042 × 10–3 mm3/(N·m) in dry sliding and 0.065 × 10–3 mm3/(N·m) in SBF sliding), with the
reduction of 89.79% under dry friction and 85.43% in SBF lubrication
compared to the Ti64 sample. Therefore, it can be concluded that a
microgroove structure with a suitable groove width (45 μm) can
effectively reduce the COF in dry sliding and improve the wear resistance
of the titanium alloy substrate.
Table 7
Frictional Performance Parameters
of Samples under SBF Lubrication Conditions
samples
average coefficient of friction
wear rate, 10–3 mm3/(N·m)
ratio
of the reducing wear rate (%)
Ti64
0.526
0.446
Ti64-25
0.901
0.252
43.50
Ti64-35
0.897
0.083
81.39
Ti64-45
0.872
0.065
85.43
Ti64-55
0.969
0.081
81.84
Ti64-65
1.011
0.101
77.35
Surface Analysis of Wear Scars
The
surface morphology of the wear scars was characterized by SEM and
white light interferometry, and the chemical composition of the surface
was analyzed by EDX. The characterization results are shown in Figures –11.
Figure 9
3D topography of worn scars on the Ti–6Al–4V
plates
before and after texturing under dry friction conditions (a–f)
and under SBF lubrication conditions (a1–f1). (a, a1) Ti64;
(b, b1) Ti64-25; (c, c1) Ti64-35; (d, d1) Ti64-45; (e, e1) Ti64-55;
and (f, f1) Ti64-65.
Figure 11
SEM
micrographs (a–f and a1–f1) and EDX composition
analysis (a2–f2) of worn scars on the Ti–6Al–4V
plates before and after texturing under SBF lubrication conditions.
(a–a2) Ti64; (b–b2) Ti64-25; (c–c2) Ti64-35;
(d–d2) Ti64-45; (e–e2) Ti64-65; and (f–f2) Ti64-65.
3D topography of worn scars on the Ti–6Al–4V
plates
before and after texturing under dry friction conditions (a–f)
and under SBF lubrication conditions (a1–f1). (a, a1) Ti64;
(b, b1) Ti64-25; (c, c1) Ti64-35; (d, d1) Ti64-45; (e, e1) Ti64-55;
and (f, f1) Ti64-65.The three-dimensional morphologies of the surface
wear scar on
the titanium alloy sample under dry friction and SBF lubrication friction
are exhibited in Figure . It can be seen from Figure a–f that serious wear with a maximum wear scar depth
of ∼5.5 μm appeared on the surface of the titanium alloy
plate, after the sliding friction in a dry environment. After laser
processing, with the increase of groove width, the wear depth decreased
preferentially and then increased. Ti64-45 had a smaller wear scar
with a slighter wear depth of ∼1.5 μm. Under SBF lubrication
conditions, the wear depth of the titanium alloy samples has a similar
trend to that in dry friction, as shown in Figure a1–f1. It can be clearly seen from
the figure that the wear trace of the blank titanium alloy plate (∼6
μm) is deeper than that of other samples. The grooved samples
have relatively shallow wear marks, specially Ti64-45 with a wear
depth of ∼2.5 μm. In addition, it can be observed that
the depth of wear scars of each sample under SBF friction conditions
was greater than that under dry sliding friction conditions. These
characterization results were consistent with the change trend of
the wear rates (Figure d).Figure shows
the microscopic morphology and the corresponding chemical composition
analysis results on the six groups of titanium alloy samples in dry
friction. It can be seen from the SEM images (Figure a–f) that the blank titanium alloy
plate has the widest wear scar width (∼398 μm) with serious
material loss and obvious scratches, debris, and plows on the surface.
This indicated that the abrasive wear occurred on the titanium alloy
samples during the dry friction process. After the surface of the
titanium alloy was grooved by laser treatment, the wear scar width
was narrowed with reduced wear damage. Furthermore, as the groove
width increased, the wear scar width showed a trend of first decreasing
and then increasing. This corresponded to the change trend of the
sample wear rate under dry friction (Figure d). In addition, the corresponding component
analysis results of the surface of the titanium alloy samples are
shown in Figure a2–f2. Si and N were detected on the wear surfaces of the
six sets of titanium alloy samples, indicating the existence of Si3N4 transfer during the dry friction process and
further implying the adhesive wear mechanism. Moreover, the contents
of Si and N on the wear surface of the grooved titanium alloy were
less than those on the Ti64 sample, demonstrating that the surface
groove treatment might effectively reduce the adhesive wear. Among
them, the Si content of the worn surface of the Ti64-45 sample exhibited
the lowest value (0.82 wt %), suggesting a slight adhesive wear behavior.
Figure 10
SEM
micrographs (a–f and a1–f1) and EDX composition
analysis (a2–f2) of worn scars on the Ti–6Al–4V
plates before and after texturing under dry friction conditions. (a–a2)
Ti64; (b–b2) Ti64-25; (c–c2) Ti64-35; (d–d2)
Ti64-45; (e–e2) Ti64-65; and (f–f2) Ti64-65.
SEM
micrographs (a–f and a1–f1) and EDX composition
analysis (a2–f2) of worn scars on the Ti–6Al–4V
plates before and after texturing under dry friction conditions. (a–a2)
Ti64; (b–b2) Ti64-25; (c–c2) Ti64-35; (d–d2)
Ti64-45; (e–e2) Ti64-65; and (f–f2) Ti64-65.Figure exhibits the SEM images and the corresponding
element
content analysis of the titanium alloy samples in SBF lubrication.
According to the morphological characterizations (Figure a–f,a1–f1),
it can be observed that the width of the wear scar on the surface
of the Ti64 sample was the largest (∼436 μm), with severe
scratches and deep plows and more wear debris. This indicated that
the abrasive wear on the surface of the blank titanium alloy sample
was serious, while the wear widths of other grooved titanium alloys
were relatively narrow, implying that the surface microgroove structure
has a positive effect on improving the wear resistance of titanium
alloys. This is similar to that in dry friction. Additionally, the
element compositions on the worn surface of the titanium alloy samples
were detected and analyzed, as seen in Figure a2–f2. Si and N elements were found
on the surface of the six groups of titanium alloy samples, illustrating
the adhesion wear mechanism in the SBF friction process. The surface
Si content on the samples were sorted as follows: Ti64 (1.78 wt %)
> Ti64-25 (1.35 wt %) > Ti64-35 (1.22 wt %) > Ti64-45 (0.84
wt %),
and the Si content of Ti64-55 and Ti64-65 was approximative. The results
showed that the adhesion wear resistance of the Ti64 surface was significantly
improved through the microgroove structure under SBF lubrication,
especially the width of 45 μm. Besides, the occurrence of Ca
and P elements on the worn surface of the grooved titanium alloy may
be due to the formation of calcium phosphate, induced by the grooved
structure in SBF solution. In addition, compared with the wear surface
of these samples under dry friction, wider wear traces were observed
in the wet friction process. This also further confirmed the trend
of wear rates under both lubrications (Figure d). The above analysis results indicated
that the abrasive wear and adhesion wear appeared during both lubrication
conditions, and the microgroove structure especially with the 45 μm
width might have a positive effect on enhancing the wear behavior.SEM
micrographs (a–f and a1–f1) and EDX composition
analysis (a2–f2) of worn scars on the Ti–6Al–4V
plates before and after texturing under SBF lubrication conditions.
(a–a2) Ti64; (b–b2) Ti64-25; (c–c2) Ti64-35;
(d–d2) Ti64-45; (e–e2) Ti64-65; and (f–f2) Ti64-65.
Biological Activity Evaluation
To
study the effect of laser treatment on the biological activity of
the titanium alloy surface, the BMSC proliferation and adhesive behavior
were analyzed by applying the CCK-8 detection method and a fluorescence
microscope. The cell viability results are shown in Figure . It can be found that the
absorbance of the cells on the six group titanium alloy specimens
has increased with increasing culture time, indicating that these
titanium alloy materials showed ideal biocompatibility before and
after laser treatment. The calculated cell proliferation rates of
these six group samples are listed in Table , based on the absorbance. It is significant
that after 1 day of incubation, the surface cell activity of titanium
alloy samples with a groove width greater than 45 μm was better
than that of Ti64 samples. After culturing for 3 and 7 days, compared
with the Ti64 group, the surface of the titanium alloy sample with
the groove structure showed a higher cell survival rate, suggesting
that the microgroove could improve the BMSC proliferation of the titanium
alloy surface. Moreover, the Ti64-45 sample showed the highest cell
viability with a cell proliferation rate of 84.3%, indicating that
the groove structure with a 45 μm width was more conducive to
the BMSC proliferation on the surface.
Figure 12
Cell viability of BMSCs
by the CCK-8 assay after 1, 3, and 7 days
of culture on six different specimens.
Table 8
Increasing Cell Proliferation Rates
of Samples
increasing
cell proliferation rate (%)
samples
after 1
day
after 3 days
after 7 days
Ti64
27.8
38.6
68.9
Ti64-25
22.1
51.6
79.6
Ti64-35
22.3
52.0
80.0
Ti64-45
31.7
62.0
84.3
Ti64-55
29.8
59.9
82.5
Ti64-65
28.9
58.4
60.0
Cell viability of BMSCs
by the CCK-8 assay after 1, 3, and 7 days
of culture on six different specimens.The adhesion performance of BMSCs on the surface of
different titanium
alloy samples was observed and shown in Figure . Among them, the nucleus and cytoskeleton
of BMSCs were, respectively, marked in blue and red colors. It can
be clearly seen that the fluorescent area of the grooved titanium
alloy surface was more than that of the Ti64 group, indicating that
the BMSCs were prone to adhere to the grooved surface. Moreover, the
cells on the surface of Ti64-45 samples were denser than those on
the surface of other samples, with a large number of filopodia, further
indicating that the groove width of 45 μm was beneficial to
the BMSC adhesion. The above bioactivity results demonstrated that
the laser treatment with the microgroove structure improved the BMSC
adhesion behavior and cell proliferation on the titanium alloy surface,
and the optimal groove width was ∼45 μm.
Figure 13
Fluorescence images
of BMSCs cultured on six different specimens
for 1 day, including blue nucleus images, red cytoskeleton images,
and merged fluorescence images. (a) Ti64; (b) Ti64-25; (c) Ti64-35;
(d) Ti64-45; (e) Ti64-55; and (f) Ti64-65.
Fluorescence images
of BMSCs cultured on six different specimens
for 1 day, including blue nucleus images, red cytoskeleton images,
and merged fluorescence images. (a) Ti64; (b) Ti64-25; (c) Ti64-35;
(d) Ti64-45; (e) Ti64-55; and (f) Ti64-65.
Discussion
Electrochemical Corrosion Resistance Mechanism
According to our previous reports,[40] the blank Ti6Al4V alloy might readily corrode in SBF solution, due
to the following chemical reactions (Schemes –8). During the
immersion process of the titanium alloy in the SBF aqueous solution,
the hydroxylated titanium compound was first formed and then reacted
with Cl– and SO42– in
the SBF solution, further resulting in the corrosion of the titanium
alloy surface.In this study, the corrosion resistance results
(Figure and Table ) showed that the
laser microgroove treatment effectively improved the corrosion resistance
of the titanium alloy samples in the SBF solution. This result is
similar to previous studies.[41] According
to this report, laser processing can affect the corrosion rate of
the titanium alloy matrix, mainly for the reason that the laser texturing
treatment changes the crystal component on the titanium alloy surface.[42] By XRD analysis (Figure ) and calculated results (Table ) of the crystal ratio (φ%)
of the titanium alloy surface before and after laser treatment, it
was found that the φ% value of the grooved titanium alloy surface
was reduced compared to the Ti64 sample, indicating that the β
phase of the titanium alloy surface has transformed to the α
phase during laser ablation. This further implied that the formed
remelting layer due to the laser thermal effect might have a refined
and uniform tissue structure. As described in the research, the grain
refinement on the material surface has a positive effect on reducing
the surface defects, thus enhancing its corrosion resistance.[43] In addition, as these reports described,[26,37,44] the galvanic corrosion behavior
is the main corrosion form on the surface of the Ti6Al4V alloy, and
the density of grain boundaries and grain refinement are the important
factors affecting the galvanic corrosion behavior.[44] In this work, according to the microstructure analysis
(XRD analysis), it can be obtained that the density of grain boundaries
was increased and the microstructures of laser-microgrooved Ti6Al4V
samples were refined due to the phase transformation after laser treatment.[37] In the simulated body fluid, the grain refinement
improved the passive film formation, which prevented ions from entering
the interior of Ti6Al4V, thereby reducing the galvanic corrosion.[26] Therefore, a laser-grooved structure exhibited
the improvement of anticorrosion properties on the titanium alloy
surface. Compared with the grooved samples of different widths, the
Ti64-45 sample has the lowest crystal ratio on the surface, suggesting
a more uniform tissue microstructure, thus exhibiting excellent corrosion
resistance.
Wear Resistance Mechanisms
The tribological
properties of laser-microgrooved titanium alloys were studied. The
experimental results showed that the introduction of the microgroove
structure can significantly reduce the friction coefficient between
the titanium alloy and Si3N4 in dry friction.
Moreover, the wear resistance of the titanium alloy surface under
dry and SBF lubrications was significantly improved by the laser-grooved
structure.According to Coulomb’s law, the friction coefficient
(μ) can be calculated by formula .[9] In the friction process,
the friction coefficient changes with the friction force (Ff), after fixing the normal load (Fn). Johnson et al. mentioned that friction force was mainly
related to the adhesion force (Ff,adh)
and surface furrowing effect (Ff,pl),
as shown in formula .[45] In these formulas, Ff,adh can be determined by formula (τ is the shear strength and Ar is the actual contact area).In the dry friction process, the friction
coefficients of the titanium alloy sample with groove structures were
lower than those of the blank sample. Based on the surface analysis
results (Figure ), the groove structure might store the wear debris generated during
the friction process, resulting in the reduction of wear debris on
the surface of the grooved titanium alloy compared to the blank samples.
Furthermore, the higher microhardness on the grooved surface would
also reduce the shedding of debris, owing to the reduction of residual
stress at the interface of the titanium alloy and Si3N4 ball during the friction process.[46,47] This might lead to a decrease in deformation friction force. Additionally,
after the titanium alloy was treated with laser-microgrooved texturing,
the actual contact area at the friction interface became smaller than
that on the Ti64 surface, and the shear strength was decreased due
to the increase of microhardness. This implied the reduction of adhesive
friction force. Hence, the microgrooved titanium alloys showed a lower
COF value than the blank Ti64 plate. In addition, compared with the
microgrooved samples, the Ti64-45 sample has the largest microhardness
(599 HV), indicating a remarkably reduced deformation force. Furthermore,
almost no debris was found on the surface of Ti64-45 samples, suggesting
the adequacy of entrapment of wear debris with the groove of 45 μm,
further reducing the adhesion force. Therefore, Ti64-45 exhibited
a smaller friction coefficient (0.215) than other samples in dry sliding.
Under SBF lubrication conditions, the microgroove structures can not
only be used to store wear debris but also as a lubricant storage
tank. However, the existence of the groove structure during the contact
process may affect the supply of the lubricant at the contact interface,
resulting in the inability to form a sufficiently thick liquid film
on the surface. Therefore, under the conditions of SBF lubrication,
the friction coefficients on the grooved titanium alloys were higher
than that of the blank plate.In this work, the wear rates and
the corresponding ratio of the
reducing wear rate after the surface groove texture of the titanium
alloy were calculated. It can be found that the laser-grooved treatment
can significantly reduce the wear rate on the titanium alloy surface
with a reduction of 89.79% under dry friction and 85.43% in SBF lubrication.
First, the increase of surface microhardness on the titanium alloys
after laser treatment may be an important reason for the improvement
of the wear resistance of titanium alloys. This is similar to the
report by Kumari et al.[19] The increase
of surface hardness would reduce the generation of wear debris on
the friction interface, thereby reducing abrasive wear in the friction
process. Moreover, according to the analysis results of the chemical
composition on the worn surface, the reduction of the Si element on
the surface of the grooved titanium alloy also implies that the groove
structure can reduce the adhesive wear during the friction process.[48] Therefore, the microgroove structure improved
the antiwear properties of the titanium alloy surface significantly.
The highest microhardness (599 HV) and the smaller Si (0.98) contents
appeared on the Ti64-45 samples, and thus it exhibited superior wear
resistance.
Bioactivity Mechanism Analysis
The
excellent bioactivity on the microgrooved titanium alloys was evaluated,
by analyzing the BMSC adhesion behavior and proliferation. The results
can be explained from the following aspects. First, compared with
the Ti64 group, the surface roughness of the titanium alloy sample
with the microgroove structure is significantly increased (Table ), resulting in the
increase of the contact area of BMSCs on the surface, thereby improving
the adhesion of the cells.[49,50] Secondly, the surface
microgroove structure may affect the contact guidance of cells in
the diffusion process.[51,52] As the cells are in contact with
the surface, the actin from BMSCs tends to reach an equilibrium state
for promoting its self-differentiation on the surface. Due to the
influence of the groove structure, the actin filamentous pseudopodia
of BMSCs would extend along the direction of the groove, thereby increasing
the contact and communication of the cells.[53] This may be an important factor in improving surface cell proliferation.
Furthermore, the induced bonelike apatite was found on the microgrooved
surface (seen in Figure ), enhancing the surface bioactivity of the titanium alloy.In addition, the influence of the groove width on cell behavior
might be ascribed to the following two aspects: the BMSCs’
extensions and the improved wettability on the groove surface.[54] First, according to our previous reports,[9] the size of BMSCs is ∼50 μm. When
the groove width is less than 50 μm, the BMSCs’ extensions
might appear along the direction of the grooves on the surface, thus
promoting cell growth by increasing the possibility of intercellular
signaling. Besides, the wettability of the titanium alloy surface
is also considered to be the key performance for affecting BMSC behavior.
Research in the past reported that the hydrophilic surfaces were prone
to induce cell adhesion, proliferation, differentiation, and mineralization.[55] As shown in Figure , it is observed that the groove width and
surface wettability are positively correlated, implying the improvement
of cell behavior on the titanium alloy surface with a greater groove
width. Therefore, considering the effect of surface wettability and
the microgrooved structure on BMSCs’ behavior, Ti64-45 exhibited
excellent biological activity due to the improved wettability and
the function of promoting BMSC extensions.
Conclusions
In this study, microgroove
structures with different widths were
prepared on the surface of the titanium alloy by laser processing.
The surface morphology, roughness, and the surface properties such
as microhardness and wettability were characterized. Furthermore,
the electrochemical corrosion performance, tribological properties,
and biological activity were systematically investigated. The following
conclusions are drawn:The anticorrosion analysis results
showed that the laser-grooved structure improved the corrosion resistance
on the titanium alloy surface in SBF solution. Moreover, Ti64-45 showed
outstanding anticorrosion properties, resulting from the reduction
of the tissue defect. The lowest crystal ratio of the β phase
to the α phase on the surface might imply its uniform and dense
tissue microstructure and thus exhibit better excellent corrosion
resistance.The tribological
performance results
showed that the introduction of the microgroove structure can significantly
reduce the friction coefficient of the titanium alloy in dry friction.
Besides, the wear resistance under both lubrications on the titanium
alloy surface was significantly improved. Among them, the wear rates
of titanium alloys were reduced by 87% under dry friction and 85%
in SBF lubrication, after texturing the microgroove structure with
a 45 μm width. This might be mainly related to the increase
of the microhardness due to the laser-grooved treatment and the function
of storing abrasive particles of the grooved structure.The in vitro bioactivity results showed
that the surface of the titanium alloy samples with microgroove structures
exhibited higher BMSC proliferation and adhesion. Moreover, the Ti64-45
sample showed the highest cell viability with a cell proliferation
rate of 84.3% and cell adhesion number, which might be ascribed to
the improved wettability and the function of promoting BMSC extensions.The above results imply that the microgroove structure
with 45
μm has effectively improved the anticorrosion, antiwear performances,
and bioactivity in vitro of the titanium alloy surface. This textured
microgroove structure has potential application prospects in the field
of titanium alloy orthopedic implants.
Authors: Xiongcheng Xu; Yanjin Lu; Shuman Li; Sai Guo; Mengjiao He; Kai Luo; Jinxin Lin Journal: Mater Sci Eng C Mater Biol Appl Date: 2018-04-17 Impact factor: 7.328
Authors: Man Tik Choy; Chak Yin Tang; Ling Chen; Chi Tak Wong; Chi Pong Tsui Journal: Mater Sci Eng C Mater Biol Appl Date: 2014-06-20 Impact factor: 7.328