Literature DB >> 36068805

Doubly resonant sub-ppt photoacoustic gas detection with eight decades dynamic range.

Zhen Wang1,2, Qiang Wang1,3, Hui Zhang1,3, Simone Borri4, Iacopo Galli4, Angelo Sampaolo5, Pietro Patimisco5, Vincenzo Luigi Spagnolo5, Paolo De Natale4, Wei Ren2.   

Abstract

Photoacoustic spectroscopy (PAS) based gas sensors with high sensitivity, wide dynamic range, low cost, and small footprint are desirable in energy, environment, safety, and public health. However, most works have focused on either acoustic resonator to enhance acoustic wave or optical resonator to enhance optical wave. Herein, we develop a gas sensor based on doubly resonant PAS in which the acoustic and optical waves are simultaneously enhanced using combined optical and acoustic resonators in a centimeter-long configuration. Not only the lower detection limit is enhanced by the double standing waves, but also the upper detection limit is expanded due to the short resonators. As an example, we developed a sensor by detecting acetylene (C2H2), achieving a noise equivalent absorption of 5.7 × 10-13 cm-1 and a dynamic range of eight orders. Compared to the state-of-the-art PAS gas sensors, the developed sensor achieves a record sensitivity and dynamic range.
© 2022 The Authors.

Entities:  

Keywords:  Acoustic resonator; Gas sensing; Optical cavity; Photoacoustic spectroscopy; Quartz tuning fork; Ultra-high sensitivity; Ultra-wide dynamic range

Year:  2022        PMID: 36068805      PMCID: PMC9441262          DOI: 10.1016/j.pacs.2022.100387

Source DB:  PubMed          Journal:  Photoacoustics        ISSN: 2213-5979


Introduction

Laser-based optical gas sensors are increasingly required in many fields such as environmental monitoring [1], [2], marine science [3], [4], biological studies [5], [6], and breath analysis [7], [8]. By measuring light attenuation through a gas sample, quantitative gas analysis can be performed. To enhance sensitivity, the absorption path length can be increased to hundreds of meters by a multipass cell or several kilometers by a high-finesse optical cavity. Among the most sensitive laser-based gas sensing techniques, noise-immune cavity-enhanced optical heterodyne molecular spectroscopy (NICE-OHMS) achieved 10−14 cm−1 in noise equivalent absorption (NEA) [9]. The minimum detectable concentration of a specific molecule has been achieved by saturated-absorption cavity ring-down (SCAR) with a limit of a few parts-per-quadrillion (ppq) [6], [10]. However, the relatively long cavity used in this class of gas sensors sets a limit on the dynamic range because the high finesse degrades even at low gas concentrations, which also leads to, in turn, the large overall footprint and weight. For example, with a cavity finesse of 30,000 and a cavity length of 37.8 cm used in NICE-OHMS, the dynamic range is limited to four orders of magnitude [11]. Dynamic range is a very important parameter of gas sensors. For example, the NH3 concentration changes by orders from tens of parts-per-million (ppm) to parts-per-billion (ppb) in applications including chemical hazards monitoring for occupational safety and health, contaminant control in hydrogen source for fuel cell, and contaminant requirement in lithography processing in semiconductor industry [12]. To get a combined high sensitivity and wide dynamic range, photoacoustic spectroscopy (PAS) is a promising candidate as the acoustic signal linearly increases with laser power, rather than a long absorption path. Numerous studies have been focused on applying various acoustic transducers and designing resonators to enhance the acoustic wave [13], [14], [15], [16]. An optical resonator can also be used to enhance the laser power by precisely matching the cavity mode and laser frequency using low-frequency dither locking [17], [18], optical feedback [19], [20], and Pound-Drever-Hall (PDH) locking method [21], [22]. However, compared to the well-assessed cavity ring-down spectroscopy (CRDS) and NICE-OHMS techniques (10−14–10−13 cm−1 in NEA), the NEA of state-of-the-art PAS gas sensors is limited to 10−11–10−8 cm−1, while the dynamic range is limited to five orders of magnitude [13], [14], [17], [21], [22], [23], [24]. It is worth mentioning that researchers recently tried to combine an acoustic resonator with an optical resonator to advance the performance [25]. However, the sensitivity and dynamic range cannot be simultaneously improved because of the limited optical and acoustic buildup factors, noise floor limited by the optical feedback strategy and the long optical cavity with a Brewster window [25]. A high-efficiency doubly resonant strategy, which can enhance the intracavity PAS signal by orders linearly, is worth pursuing. In this work, we report a PAS sensor with opto-acoustic resonance enhancement for gas detection with ultra-high sensitivity and ultra-wide dynamic range. The proposed photoacoustic design, leveraging on a double standing wave effect, achieves a combined acoustic amplification factor of 144 and laser power enhancement of almost three orders of magnitude. As a proof-of-principle, we show that our C2H2 sensor reaches a record sensitivity of 10−13 cm−1 (NEA) and a record dynamic range of eight orders of magnitude.

Working principle

Fig. 1 shows the basics of opto-acoustic resonance for PAS. The core gas sensing element consists of an optical resonator, an acoustic resonator, and an acoustic transducer, which are arranged in a coupled configuration. Indeed, when the optical frequency of the incident laser is in resonance with a longitudinal cavity mode of the optical resonator, a standing optical wave is formed between the resonator mirrors. A high-finesse optical resonator can significantly build up the laser power [26], by several orders of magnitude, directly enhancing the photoacoustic signal. The laser intensity is modulated at the same resonance frequency as the acoustic resonator and transducer. A specifically designed one-dimensional longitudinal tube can be used to amplify the acoustic signal by forming a standing acoustic wave inside it. Any types of acoustic transducers can be used to detect the amplified acoustic wave. In this work, we use a quartz tuning fork (QTF) with a high Q-factor for demonstration. The QTF locates nearby the antinode of the standing acoustic wave, which is generated by the acoustic resonator composed of two stainless-steel tubes placed at the opposite sides of the QTF [27].
Fig. 1

Working principle of doubly resonant PAS. ASW: acoustic standing wave; OSW: optical standing wave; QTF: quartz tuning fork. The amplitude of the OSW is modulated at the resonant frequency of the QTF. The geometry of the acoustic resonator is designed according to the acoustic frequency.

Working principle of doubly resonant PAS. ASW: acoustic standing wave; OSW: optical standing wave; QTF: quartz tuning fork. The amplitude of the OSW is modulated at the resonant frequency of the QTF. The geometry of the acoustic resonator is designed according to the acoustic frequency. The frequency-dependent photoacoustic signal (S) at its resonant frequency f is given by [28]:where b is the laser power buildup factor, g is the acoustic wave enhancement factor, K is the sensor constant, W is the incident laser power, λ is the laser wavelength, α(λ) is the effective absorbance by the analyte, τ(P) is the relaxation time at the gas pressure P, and ε is the radiation-to-sound conversion efficiency, which depends on f and τ. Note that the factor g is independent of laser power but is related to the geometry, to the material, and to the Q-factor of the acoustic resonator, as well as to the frequency of acoustic waves [29]. The power buildup factor b is determined by coupling efficiency and finesse of the optical resonator which needs to be properly selected so that a wide dynamic range and a high sensitivity can be simultaneously obtained (Appendix A).

Experimental setup

The schematic of the doubly resonant PAS sensor is shown in Fig. 2. An external cavity diode laser (ECDL, TOPTICA Photonics) is used to detect the P(11) transition of C2H2 at 1531.59 nm. The ECDL is phase modulated by an electro-optic modulator (EOM, iXblue Photonics) at 20 MHz and locked to the optical resonator using the PDH method [30]. The polarization beam splitter is used with the quarter-wave plate to pick up the reflected beam from the optical resonator, which is then detected by a photodetector (PD1). Details of the application of the PDH method to PAS can be found in [21]. In this work, the current and piezo transducer (PZT) feedback loops of the ECDL are both used to improve the locking performance. Each mirror (Layertec Inc.) of the optical resonator has a radius of curvature of 150 mm and reflectivity of 99.923% (finesse 4078) at the laser wavelength, as measured by cavity ring-down (see Supplementary Note 1). Compared with other cavity-enhanced absorption spectroscopic methods [6], [9], the optical resonator used here for the opto-acoustic resonance features a much shorter length (60 mm in this work). Two mode matching lenses (focus length: L1 =30 mm and L2 =50 mm) and a photodetector (PD2) are used to maximize the coupling efficiency (84%) between the laser and the optical resonator (see Supplementary Note 2). With a maximum incident laser power of 300 mW, the intracavity optical power is boosted to 264 W in this work (Appendix B).
Fig. 2

Schematic of the photoacoustic sensor with opto-acoustic resonance. ECDL, external cavity diode laser; EOM, electro-optic modulator; OS, optical switch; PBS, polarization beam splitter; QWP, quarter-wave plate; PD, photodetector; PZT, piezo transducer; FG, function generator; PDH, phase demodulator and laser servo (PDD110/mFALC110, TOPTICA Photonics); TIA, trans-impedance amplifier; LIA, lock-in amplifier. The opto-acoustic resonance element inside the PAS cell has the same configuration as that shown in Fig. 1.

Schematic of the photoacoustic sensor with opto-acoustic resonance. ECDL, external cavity diode laser; EOM, electro-optic modulator; OS, optical switch; PBS, polarization beam splitter; QWP, quarter-wave plate; PD, photodetector; PZT, piezo transducer; FG, function generator; PDH, phase demodulator and laser servo (PDD110/mFALC110, TOPTICA Photonics); TIA, trans-impedance amplifier; LIA, lock-in amplifier. The opto-acoustic resonance element inside the PAS cell has the same configuration as that shown in Fig. 1. The intracavity laser beam passes through the acoustic resonator, which consists of two stainless-steel tubes (inner diameter 1.3 mm, length 23 mm), and does not touch any surface. The central axis of the acoustic resonator is placed 1.2 mm below the top of the QTF prongs to optimize the piezoelectrical conversion efficiency [15]. The two tubes are placed at a distance of ~60 µm from the QTF, so that it lies near the antinode of the acoustic wave and leaves its Q-factor unaffected. The beam waist (340 µm in diameter) is located between the two prongs of the QTF, which has a gap of 800 µm, a resonant frequency of 7.2 kHz and a Q-factor of ~8000 (gas pressure 760 Torr) [15]. Photoacoustic gas sensors using QTF as an acoustic transducer have been previously developed for detecting many different gas species [15], [16], [17]. The optical resonator, the acoustic resonator and the QTF are all enclosed inside a chamber (PAS cell in Fig. 2). A high-speed lithium niobate optical switch (NanoSpeed, Agiltron) is used to chop the laser beam at the same frequency as the resonant frequency of the QTF. The piezoelectric current from the QTF is collected and amplified by a trans-impedance amplifier. Finally, a lock-in amplifier (MFLI 5 MHz, Zurich Instruments) with a detection bandwidth of 1 Hz (integration time 80 ms, filter roll-off 18 dB/oct) is used to demodulate the first harmonic signal (1f) at the sensor output. To retrieve a complete spectrum, the optical cavity length is tuned by a PZT attached to one cavity mirror via the function generation.

Double standing wave enhancement

To evaluate the enhancement effects contributed by the integrated acoustic and optical resonators, the PAS-1f signal of the C2H2 line at 1531.59 nm is measured under three different configurations. Fig. 3 compares the typical PAS-1f signal measured using a bare QTF (2% C2H2), a QTF with the mere acoustic resonator (0.1% C2H2), and a QTF with the complete opto-acoustic resonator (1 ppm C2H2). Note that different C2H2 concentrations are used for these three configurations because of their quite different sensitivities. All the experiments are performed at the same incident laser power of 12 mW, lock-in detection bandwidth of 1 Hz, and gas pressure of 760 Torr. After normalization by the gas concentration, a comparison of Fig. 3(a) and (b) shows that the PAS-1f signal is enhanced by 175 times. However, with the high concentration C2H2 used in the bare QTF measurement, we must consider power attenuation for the incident power before reaching the QTF. With a power attenuation of 17.3%, the enhancement factor of acoustic resonator is evaluated as 144. Besides, the optical resonator provides another enhancement factor of 980, as emerging from a comparison of Fig. 3(b) and (c). Hence, the combined opto-acoustic amplification provides an overall enhancement of the PAS signal by a factor of 105 via the double standing wave effect.
Fig. 3

PAS-1f signal measured by three different configurations: (a) bare QTF, (b) QTF with acoustic resonator, and (c) QTF with acoustic and optical resonators. The amplitude of the single-pass PAS signal measured for 2% C2H2/N2 using the bare QTF is about 4.24 mV. Benefitting from the double standing wave enhancement, the sensor has a response of 36 mV to 1 ppm C2H2. An enhancement factor of up to five orders of magnitude for the PAS signal is achieved.

PAS-1f signal measured by three different configurations: (a) bare QTF, (b) QTF with acoustic resonator, and (c) QTF with acoustic and optical resonators. The amplitude of the single-pass PAS signal measured for 2% C2H2/N2 using the bare QTF is about 4.24 mV. Benefitting from the double standing wave enhancement, the sensor has a response of 36 mV to 1 ppm C2H2. An enhancement factor of up to five orders of magnitude for the PAS signal is achieved.

Ultrasensitive gas detection

Fig. 4 shows the results of C2H2 detection with doubly resonant PAS. The measurement of PAS-1f signal is performed with an incident optical power of 12 mW at 760 Torr for different C2H2 concentrations (1 ppm, 100 ppb, 10 ppb and 1 ppb in nitrogen balance) and high-purity nitrogen (99.999% purity) when the laser wavelength is tuned from 1531.32 nm to 1531.75 nm, as shown from Fig. 4(a) to (e). Note that the background signal is caused by unwanted absorption at the optical windows and resonator mirrors [31]. This background signal is not subtracted from the PAS-1f signal shown in Fig. 4. Interestingly, due to the excellent signal-to-noise ratio, a neighboring water line with quite low intensity of 2.896 × 10−24 cm−1/(molec·cm−2) near 1531.37 nm is detected, as shown in Fig. 4. This is probably due to residual water in the gas chamber after the desiccation process (Appendix C). Our hypothesis is confirmed by repeating the measurement of pure nitrogen (water vapor < 0.3 ppm) shown in Fig. 4(d). To reduce the measurement error, a multi-spectral fitting method, with prior knowledge of infrared spectra from the HITRAN database, is implemented [32], [33]. Hence, the background signal is automatically eliminated during the fitting procedure.
Fig. 4

Results of C2H2 detection with doubly resonant PAS. Representative PAS-1f signal of C2H2/N2 mixtures and N2 balance gas: (a) 1 ppm, (b) 100 ppb, (c) 10 ppb and (d) N2 cylinder. The target C2H2 peak is observed at 1531.59 nm, whereas another H2O line is also evident at 1531.37 nm. (e) Representative PAS-1f signal of 1 ppb C2H2 measured at two different incident optical power levels: 12 mW (top panel) and 300 mW (bottom panel). (f) PAS-1f amplitude for 100 ppb C2H2 and 1-σ standard deviation (STD) of the noise (a measurement of nitrogen over 120 s) versus incident power. The PAS-1f amplitudes of 100 ppb C2H2 at different incident power levels are inferred from the fitting results. All these measurements are performed at 760 Torr and a detection bandwidth of 1 Hz. The error bars (1-σ STD) are calculated from the raw data taken in a time interval of 120 s.

Results of C2H2 detection with doubly resonant PAS. Representative PAS-1f signal of C2H2/N2 mixtures and N2 balance gas: (a) 1 ppm, (b) 100 ppb, (c) 10 ppb and (d) N2 cylinder. The target C2H2 peak is observed at 1531.59 nm, whereas another H2O line is also evident at 1531.37 nm. (e) Representative PAS-1f signal of 1 ppb C2H2 measured at two different incident optical power levels: 12 mW (top panel) and 300 mW (bottom panel). (f) PAS-1f amplitude for 100 ppb C2H2 and 1-σ standard deviation (STD) of the noise (a measurement of nitrogen over 120 s) versus incident power. The PAS-1f amplitudes of 100 ppb C2H2 at different incident power levels are inferred from the fitting results. All these measurements are performed at 760 Torr and a detection bandwidth of 1 Hz. The error bars (1-σ STD) are calculated from the raw data taken in a time interval of 120 s. The relationship between the PAS-1f signal and incident laser power is also investigated. Fig. 4(e) compares the PAS-1f signal at 1 ppb C2H2 under two different incident power levels (12 mW and 300 mW). The peak value of the PAS-1f signal is increased by a factor of about 22.5 when the incident laser power is increased from 12 mW to 300 mW (a factor of 25). The slight deviation of the enhancement factor between laser power and PAS signal is caused by the variation of the optical coupling efficiency. By repeating the measurement at 100 ppb C2H2, Fig. 4(f) shows the variation of the PAS-1f amplitude with the incident optical power. The sensor signal (red circles) increases almost linearly (0.25 mV/mW) with the incident power. In contrast, the noise level (black rectangles) remains almost unchanged over the entire power range, as shown in Fig. 4(f) (1-σ standard deviation of N2 over 120 s). This makes our set-up very promising to further enhance the minimum detection limit by simply increasing the incident laser power.

Dynamic range and detection limit

The linear response of the sensor is tested at the pressure of 760 Torr by filling the gas chamber with different C2H2/N2 mixtures. Fig. 5(a) shows the background-subtracted PAS-1f amplitude as a function of C2H2 concentration, varying in a 1 ppb–500 ppm interval. The sensor shows a very good linear response with a slope of 36.4 μV/ppb and an R-square value of 0.99 from 1 ppb to 50 ppm. Considering the incident optical power of 12 mW, the normalized linear response coefficient of the sensor is 3 μV/(mW·ppb). Due to the linear relationship between incident optical power and PAS-1f amplitude verified in Fig. 4(f), the response coefficient remains unchanged at 300 mW. However, the sensor deviates from the linear response at higher C2H2 concentrations due to the apparent degradation of finesse of the optical resonator. With an incident optical power of 300 mW, the sensor precision has also been evaluated by measuring each concentration for 120 s; here the relative precision is defined by the ratio of the standard deviation (1-σ) to the average. The sensor shows a precision of 0.1–1.1% for the C2H2 concentration range from 10 ppb to 1 ppb. The relative precision remains below 0.1% for gas concentrations over 10 ppb.
Fig. 5

Evaluation of the sensor performance. (a) Background-subtracted PAS-1f amplitude as a function of gas concentration for an incident optical power of 12 mW. Error bars show the 1-σ standard deviation from 120 s measurements. (b) The Allan–Werle deviation analysis. The laser wavelength is tuned to the peak of the C2H2 absorption line, with the chamber filled with nitrogen. The dashed line in the top panel represents the 1/t0.44 slope. The bottom panel depicts the raw data of noise measured for nearly 2 h. The detection bandwidth is the same as the signal measurement.

Evaluation of the sensor performance. (a) Background-subtracted PAS-1f amplitude as a function of gas concentration for an incident optical power of 12 mW. Error bars show the 1-σ standard deviation from 120 s measurements. (b) The Allan–Werle deviation analysis. The laser wavelength is tuned to the peak of the C2H2 absorption line, with the chamber filled with nitrogen. The dashed line in the top panel represents the 1/t0.44 slope. The bottom panel depicts the raw data of noise measured for nearly 2 h. The detection bandwidth is the same as the signal measurement. To evaluate the long-term stability and the minimum detection limit, the Allan–Werle deviation analysis is conducted by measuring nitrogen with the results shown in Fig. 5(b). The noise equivalent concentration (NEC) is determined to be 5.1 ppt (unity for signal-to-noise ratio) at an averaging time of 1 s. At an incident optical power of 300 mW and a detection bandwidth of 1 Hz, we obtain a NNEA coefficient of 1.7 × 10−12 W·cm−1·Hz−1/2 (Appendix D). The NEC can be improved to 0.5 ppt at a longer averaging time of 300 s, leading to a NEA coefficient of 5.7 × 10−13 cm−1. As a result, the proposed photoacoustic gas sensor achieves a dynamic range of 1.0 × 108 (Appendix E). The dynamic stability of the sensor is evaluated by operating the gas sensor when continuously filling C2H2/N2 gas samples into the gas chamber (see Supplementary Note 3).

Discussion and conclusions

The key performance parameters (e.g., dynamic range, NEC, and NEA) of some state-of-the-art gas sensors are summarized in Table 1. The photoacoustic gas sensor developed in this work shows a dynamic range and NEA, superior to the state-of-the-art photoacoustic sensors using different types of laser sources and acoustic transducers. Most remarkably, the current dynamic range is three orders of magnitude wider than the state-of-the-art (105) [23], and the current NEA is two orders of magnitude better than the state-of-the-art (10−11 cm−1) [22]. Compared to other state-of-the-art gas sensors based on various spectroscopic methods such as NICE-OHMS, SCAR, CRDS and off-axis integrated cavity output spectroscopy (OA-ICOS) [6], [9], [34], [35], NICE-OHMS and CRDS exhibit 57 times and 4 times better sensitivities, respectively. But our sensor shows evident merits of size (5–8 times shorter cavity) and dynamic range (2–3 orders wider). It is worth mentioning that, among the state-of-the-art PAS studies, the cantilever with an interferometer readout has been proven to be the most sensitive acoustic transducer [22].
Table 1

Main parameters of state-of-the-art laser-based gas sensors.

GasMethodλ (µm)L (mm)DNEC (ppt)NEA (cm−1)t (s)P (Torr)
NO2PAS [23]PAS0.45NA1 × 105542.3 × 10−101640
SF6QEPAS [24]10.5NA3 × 103506.1 × 10−9175
CO2I-QEPAS [17]4.331743 × 1033001.4 × 10−82038
C2H2CECEPAS [22]1.531477 × 103242.4 × 10−11100150
C2H2CE-PAS [21]1.531301 × 1043001.9 × 10−1030050
HFCEPAS [13]2.47NA4 × 1040.651.3 × 10−101920150
SF6α-BiB3O6-PAS [14]10.6NA1 × 1050.758.1 × 10−11100760
C2H2This work1.53601 × 1080.55.7 × 1013300760
14CO2SCAR [6]CEAS4.51000NS0.0051.1 × 10−1272009.12
CO2CRDS [34]1.63307 × 10540,0001.4 × 10−1340.075
C2HDNICE-OHMS [9]1.06469NSNS1 × 10−1410.0018
COOA-ICOS [35]1.571100NS10001.9 × 10−121197

λ: laser wavelength. L: cavity length. D: dynamic range. t: averaging time. P: pressure. CEAS: cavity-enhanced absorption spectroscopy. NA: not applicable. NS: not stated.

Note that the dynamic range values are calculated using the same definition in the Appendix E. The two PAS sensors that can achieve sub-ppt sensitivity use mid-infrared lasers [13], [14]. The NEAs with different averaging times are listed here because these sensors have different stabilities. A sensor may have a sensitivity improvement by averaging if it is stable enough with white noise dominant. The averaging time is not normalized here as it cannot reflect the ultimate performance of sensors.

Main parameters of state-of-the-art laser-based gas sensors. λ: laser wavelength. L: cavity length. D: dynamic range. t: averaging time. P: pressure. CEAS: cavity-enhanced absorption spectroscopy. NA: not applicable. NS: not stated. Note that the dynamic range values are calculated using the same definition in the Appendix E. The two PAS sensors that can achieve sub-ppt sensitivity use mid-infrared lasers [13], [14]. The NEAs with different averaging times are listed here because these sensors have different stabilities. A sensor may have a sensitivity improvement by averaging if it is stable enough with white noise dominant. The averaging time is not normalized here as it cannot reflect the ultimate performance of sensors. In summary, we have developed a photoacoustic gas sensor using an integrated optical-acoustic resonator. Taking advantage of a double standing wave enhancement, the photoacoustic signal has been improved by five orders of magnitude compared to a bare QTF, thus providing ultrasensitive gas detection with low-power semiconductor lasers. The moderate cavity finesse maintains the high coupling efficiency of the incident laser and enables a wide dynamic range by using a short optical resonator. As a result, we have demonstrated a photoacoustic sensor for C2H2 (balanced in nitrogen) with a NEC of 0.5 ppt, a NEA of 5.7 × 10−13 cm−1, and a dynamic range of eight orders of magnitude. Although the demonstration is performed in the near-infrared region, the concept can be extended to the mid-infrared range to take advantage of the much stronger fundamental absorption bands. One may expect the possible challenges of obtaining high-quality mid-infrared laser beams and suitable acoustic resonators.

Declaration of Competing Interest

The authors declare no conflicts of interest.
  23 in total

1.  Part-per-trillion level SF6 detection using a quartz enhanced photoacoustic spectroscopy-based sensor with single-mode fiber-coupled quantum cascade laser excitation.

Authors:  Vincenzo Spagnolo; Pietro Patimisco; Simone Borri; Gaetano Scamarcio; Bruce E Bernacki; Jason Kriesel
Journal:  Opt Lett       Date:  2012-11-01       Impact factor: 3.776

2.  Optical feedback frequency stabilized cavity ring-down spectroscopy.

Authors:  Johannes Burkart; Daniele Romanini; Samir Kassi
Journal:  Opt Lett       Date:  2014-08-15       Impact factor: 3.776

3.  Cavity-enhanced rotational Raman scattering in gases using a 20  mW near-infrared fiber laser.

Authors:  Adam J Friss; Christopher M Limbach; Azer P Yalin
Journal:  Opt Lett       Date:  2016-07-15       Impact factor: 3.776

4.  Ultrasensitive photoacoustic detection in a high-finesse cavity with Pound-Drever-Hall locking.

Authors:  Zhen Wang; Qiang Wang; Weipeng Zhang; Haoyun Wei; Yan Li; Wei Ren
Journal:  Opt Lett       Date:  2019-04-15       Impact factor: 3.776

5.  Ultrasensitive multispecies spectroscopic breath analysis for real-time health monitoring and diagnostics.

Authors:  Qizhong Liang; Ya-Chu Chan; P Bryan Changala; David J Nesbitt; Jun Ye; Jutta Toscano
Journal:  Proc Natl Acad Sci U S A       Date:  2021-10-05       Impact factor: 11.205

6.  Highly Sensitive and Exceptionally Wide Dynamic Range Detection of Ammonia Gas by Indium Hexacyanoferrate Nanoparticles Using FTIR Spectroscopy.

Authors:  Supone Manakasettharn; Akira Takahashi; Tohru Kawamoto; Keiko Noda; Yutaka Sugiyama; Tohru Nakamura
Journal:  Anal Chem       Date:  2018-03-23       Impact factor: 6.986

7.  Mode-phase-difference photothermal spectroscopy for gas detection with an anti-resonant hollow-core optical fiber.

Authors:  Pengcheng Zhao; Yan Zhao; Haihong Bao; Hoi Lut Ho; Wei Jin; Shangchun Fan; Shoufei Gao; Yingying Wang; Pu Wang
Journal:  Nat Commun       Date:  2020-02-12       Impact factor: 14.919

8.  Quartz-enhanced photoacoustic spectroscopy exploiting low-frequency tuning forks as a tool to measure the vibrational relaxation rate in gas species.

Authors:  Stefano Dello Russo; Angelo Sampaolo; Pietro Patimisco; Giansergio Menduni; Marilena Giglio; Christine Hoelzl; Vittorio M N Passaro; Hongpeng Wu; Lei Dong; Vincenzo Spagnolo
Journal:  Photoacoustics       Date:  2020-12-09

9.  Mid-infrared intracavity quartz-enhanced photoacoustic spectroscopy with pptv - Level sensitivity using a T-shaped custom tuning fork.

Authors:  Jakob Hayden; Marilena Giglio; Angelo Sampaolo; Vincenzo Spagnolo; Bernhard Lendl
Journal:  Photoacoustics       Date:  2022-01-11

10.  Shipborne eddy covariance observations of methane fluxes constrain Arctic sea emissions.

Authors:  Brett F Thornton; John Prytherch; Kristian Andersson; Ian M Brooks; Dominic Salisbury; Michael Tjernström; Patrick M Crill
Journal:  Sci Adv       Date:  2020-01-29       Impact factor: 14.136

View more
  1 in total

1.  Dual-comb quartz-enhanced photoacoustic spectroscopy.

Authors:  Xinyi Ren; Ming Yan; Zhaoyang Wen; Hui Ma; Ran Li; Kun Huang; Heping Zeng
Journal:  Photoacoustics       Date:  2022-09-21
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.