| Literature DB >> 36068509 |
Muge Fan1, Qiuli Wang2, Jiaqi Liu1, Lingyun Su1, Bingjian Wang3, Hai Xu3, Qing Li4, Zhi Zhang5.
Abstract
BACKGROUND: With the spread of COVID-19, telemedicine has played an important role, but tele-auscultation is still unavailable in most countries. This study introduces and tests a tele-auscultation system (Stemoscope) and compares the concordance of the Stemoscope with the traditional stethoscope in the evaluation of heart murmurs.Entities:
Keywords: Auscultation; Electronic stethoscope; Heart murmur; Tele-auscultation; Valvular heart diseases
Mesh:
Year: 2022 PMID: 36068509 PMCID: PMC9446597 DOI: 10.1186/s12938-022-01032-4
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 3.903
Baseline characteristics of patients
| Variable | Value |
|---|---|
| Age, mean ± SD, years | 61.58 ± 18.30 |
| Male, n (%) | 25 (43.86%) |
| Aortic regurgitation | 12 (11.32%) |
| Mitral regurgitation | 38 (35.85%) |
| Mitral valve prolapse | 3 (2.83%) |
| Tricuspid regurgitation | 15 (14.15%) |
| Aortic stenosis | 9 (8.49%) |
| Mitral stenosis | 4 (3.77%) |
| Ventricular septal defect | 3 (2.83%) |
| Atrial septal defect | 1 (0.94%) |
| Mitral valve replacement | 14 (13.21%) |
| Tricuspid valve replacement | 2 (1.89%) |
| Aortic valve replacement | 5 (4.72%) |
Agreement of in-person auscultation results between the Stemoscope and an acoustic stethoscope among three cardiologists
| Cardiologist 1 | Cardiologist 2 | Cardiologist 3 | ||||
|---|---|---|---|---|---|---|
| Acoustic | Acoustic | Acoustic | ||||
| Abnormal | Normal | Abnormal | Normal | Abnormal | Normal | |
| Abnormal | 150 | 9 | 148 | 11 | 144 | 15 |
| Normal | 9 | 90 | 15 | 84 | 13 | 86 |
| Agreement | 93% | 90% | 89% | |||
| P value | 1 | 0.44 | 0.57 | |||
| Kappa value | 0.85 | 0.78 | 0.78 | |||
Fig. 1Distribution of in-person auscultation findings. A Total abnormal auscultatory findings including heart arrhythmias. B Abnormal in-person auscultatory findings
Agreement of in-person auscultation and Stemoscope tele-auscultation results among three cardiologists
| Cardiologist 1 | Cardiologist 2 | Cardiologist 3 | |||||||
|---|---|---|---|---|---|---|---|---|---|
| In-person | In-person | In-person | |||||||
| Abnormal | Normal | Abnormal | Normal | Abnormal | Normal | ||||
| Abnormal | 153 | 14 | 150 | 16 | 141 | 18 | |||
| Normal | 6 | 85 | 12 | 80 | 11 | 88 | |||
| Agreement | 92% | 89% | 89% | ||||||
| P value | 0.27 | 0.57 | 0.26 | ||||||
| Kappa value | 0.77 | 0.77 | 0.83 | ||||||
Comparisons between echocardiographic findings and Stemoscope tele-auscultation/acoustic in-person auscultation findings (Stemoscope tele-auscultation results of three cardiologists were combined)
| Stemoscope tele-auscultation vs. acoustic in-person auscultation | ||||
|---|---|---|---|---|
| Systolic murmurs | Diastolic murmurs | |||
| Abnormal | Normal | Abnormal | Normal | |
| Abnormal | 50 vs. 48 | 3 vs. 4 | 7 vs. 6 | 5 vs. 5 |
| Normal | 3 vs. 4 | 0 vs. 0 | 3 vs. 4 | 0 vs. 0 |
| Agreement | 89% vs. 86% | 47% vs. 40% | ||
| Sensitivity | 94% vs. 92% | 58% vs. 55% | ||
Summarized results of different electronic stethoscopes
| References | Product name | Tele-auscultation | Smartphone app | Agreement with echocardiography | Comparison with acoustic stethoscope |
|---|---|---|---|---|---|
| Our study | Stemoscope | Y | Y | 89% for systolic, 58% for diastolic murmurs | 90.6% |
| Kalinauskiene E | 3 M Littmann 3200 | NA | NA | 64.17% by cardiologists, 65% by residents | NA |
| Chowdhury MEH | Eko Core Digital Stethoscope | Y | Y | NA | 87.6% |
| Ghanayim T | VoqX | NM | NM | 86% for aortic stenosis | NA |
| Hirosawa T | JPES-01 | NA | NA | NA | 83.3% (vs. stimulator) |
| Belmont JM | ATI | Y | NA | Kappa = 0.55 | 88% |
NA: not available; NM: not mentioned; Y: yes
Fig. 2Results of the qualitative survey concerning the experience of using the Stemoscope
Fig. 3Study flowchart
Fig. 4A Stemoscope. B Stemoscope app to record heart sounds. The red dot on the screen is tapped to start recording sounds. Then, the investigators can play the recorded sounds. The app provides several options to share recorded sounds, such as via Twitter, Facebook, and WeChat
Definition for the correct identification of an echocardiography finding by auscultation
| Echocardiography | ||
|---|---|---|
| Not observed | Observed | |
| Not heard | Correct | False-negative |
| Heard | False-positive | Correct |