BACKGROUND: Cervical disc herniation is a common condition usually treated with anterior cervical discectomy and fusion (ACDF) or, more recently, with cervical disc arthroplasty (CDA). Both treatments offer similar clinical results. However, CDA has been found to offer fewer medium- to long-term complications as well as potential reduction of long-term adjacent disc degeneration. OBSERVATIONS: A 40-year-old man was treated with cervical discectomy and arthroplasty due to a C6-C7 disc herniation with left C7 radiculopathy. After the treatment, his postoperative follow-up appointments were uneventful for 9 months. However, after 9 months, he reported cervical pain and a right C7 radiculopathy after neck extension. Imaging confirmed a posterior intraprosthetic dislocation, the first case reported to date. The patient was received emergency surgery under neuromonitoring, and the prosthesis was replaced by an ACDF and anterior plate. The insert presented a rupture of the anterior horn. The patient presented no preoperative or postoperative neurological deficit, and his follow-up review revealed no issues. LESSONS: Posterior intraprosthetic dislocation is an extremely rare complication. It may occur with Mobi-C cervical arthroplasty in the case of rupture and oxidation of the polyethylene insert. Spine surgeons should be aware of this potential major complication.
BACKGROUND: Cervical disc herniation is a common condition usually treated with anterior cervical discectomy and fusion (ACDF) or, more recently, with cervical disc arthroplasty (CDA). Both treatments offer similar clinical results. However, CDA has been found to offer fewer medium- to long-term complications as well as potential reduction of long-term adjacent disc degeneration. OBSERVATIONS: A 40-year-old man was treated with cervical discectomy and arthroplasty due to a C6-C7 disc herniation with left C7 radiculopathy. After the treatment, his postoperative follow-up appointments were uneventful for 9 months. However, after 9 months, he reported cervical pain and a right C7 radiculopathy after neck extension. Imaging confirmed a posterior intraprosthetic dislocation, the first case reported to date. The patient was received emergency surgery under neuromonitoring, and the prosthesis was replaced by an ACDF and anterior plate. The insert presented a rupture of the anterior horn. The patient presented no preoperative or postoperative neurological deficit, and his follow-up review revealed no issues. LESSONS: Posterior intraprosthetic dislocation is an extremely rare complication. It may occur with Mobi-C cervical arthroplasty in the case of rupture and oxidation of the polyethylene insert. Spine surgeons should be aware of this potential major complication.
Cervical disc disease is a common condition that may lead to cervical disc herniation and radiculopathy of the upper extremities.[1] Cervical disc herniation was long treated by anterior cervical discectomy and fusion (ACDF) with good results and long-term efficacy and safety.[2] Over the last two decades, cervical disc arthroplasty (CDA) has become an alternative treatment of cervical disc disease. Some literature reviews found better medium- to long-term[3,4] clinical results, reoperation rates, and adjacent segmental disease[5-7] with CDA. Other reviews showed that CDA was not inferior to ACDF,[8] and a Cochrane review concluded clinical results in favor of CDA.[9] However, in the latter, the issues discussed were at such a low sample rate of studies with such small differences obtained, not blinded studies, that a bias potentially existed related to caregiver expectations.Several trademarked products are available, such as Bryan (Spinal Dynamics Corp. and Medtronic Sofamor Danek), ProDisc-C (DePuy Synthes), Prestige (Medtronic), and Mobi-C (Zimmer-Biomet), with different settings and shapes and a common objective of disc motion preservation. The Mobi-C is a semiconstrained cervical prosthesis containing two chrome-cobalt plates and a mobile polyethylene insert in between.[10]Complications related to CDA are primarily linked to the anterior cervical approach, such as dysphagia (2–70%),[11] recurrent laryngeal nerve compression (3–16.7%),[12,13] hematoma (suffocating or epidural, incidence 0.2% and 0.9%, respectively),[14] dural tear with pseudomeningocele (0.5–3%),[15] esophageal lesion (0.4–1.15%),[16,17] spinal cord compression (0.5%),[18] vertebral artery injury (0.4%),[18] and exceptional tracheal or thoracic duct lesion.[19]Postoperative neurological impairment is the most dreaded complication in cervical surgery for patients and surgeons. Some specific complications of CDA are well-known, including anterior bone loss (41.84%)[20] and material subsidence,[21] infection (<3.7%),[22] and heterotopic ossifications (7.7–94.1%).[23-25]Herein we describe a case of delayed posterior intraprosthetic dislocation of cervical arthroplasty.
Illustrative Case
A 40-year-old male patient (smoker) received surgery by the team at Clinic La Source in Lausanne, Switzerland, because of a left C7 radiculopathy with motor deficit M4+/5 and pain-resisting medical treatment. Cervical magnetic resonance imaging (MRI) (Fig. 1) found a left C6–C7 disc herniation with C7 conflict and no sign of posterior facet or instability. The surgery consisted of a right cervicotomy, C6–C7 microdiscectomy, and CDA with a Mobi-C prosthesis (Zimmer-Biomet) (Fig. 2). The patient was discharged from the clinic after 24 hours. The follow-up review showed adequate wound healing, complete resolution of motor deficit, and radicular pain recovery without any complication.
FIG. 1.
Preoperative cervical MRI. A: Axial T2-weighted view showing a left C6–C7 disc herniation (arrows). B: Sagittal T2-weighted view showing C6–C7 discopathy. C and D: Right and left sagittal views showing no signs of posterior facet arthritis or instability. We noticed a sigmoid aspect of the cervical spine, considered antalgic.
FIG. 2.
Lateral radiographs of the cervical spine. A: Immediate postoperative view. B: At 9 months postoperatively; one can notice the loss of height of the intraprosthetic space and anterior closure aspect of the implant, supposing a posterior migration of the polyethylene insert.
Preoperative cervical MRI. A: Axial T2-weighted view showing a left C6–C7 disc herniation (arrows). B: Sagittal T2-weighted view showing C6–C7 discopathy. C and D: Right and left sagittal views showing no signs of posterior facet arthritis or instability. We noticed a sigmoid aspect of the cervical spine, considered antalgic.Lateral radiographs of the cervical spine. A: Immediate postoperative view. B: At 9 months postoperatively; one can notice the loss of height of the intraprosthetic space and anterior closure aspect of the implant, supposing a posterior migration of the polyethylene insert.After 9 months, the patient presented to the office after feeling neck pain for 5 days and right C7 radicular pain after neck extension. The pain was immediately unbearable (numerical rating scale [NRS] of 8/10), with transient right complete motor deficit of the lower limb in a few minutes, so he went to the emergency department in another location (Yverdon-les-Bains). Radiographs were obtained (Fig. 2) and were considered normal. The patient was discharged with analgesics, and he was recommended to consult his surgeon.Clinical examination showed a NRS 10/10, a Neck Disability Index of 82%, and a well-healed scar without any sign of inflammation. Cervical mobilization was painful and limited to 30° of rotation. The Spurling test was bilaterally positive, triggering a right C7 radicular pain. There was no motor or sensitive neurological deficit of the upper limbs. Lhermitte and Hoffmann signs were negative.Emergency cervical MRI showed a suspicion of intraprosthetic dislocation, so the imaging was completed by computed tomography (Fig. 3), which confirmed the diagnosis.
FIG. 3.
MRI and computed tomography (CT) of the cervical spine 9 months postoperatively. A: Axial T2-weighted MRI showing artifact of the prosthesis plate and potential anterior spinal cord compression (blue arrow). B: Frontal CT showing anterior contact of the metallic plates, suggesting an absence of the polyethylene insert. C: Sagittal CT demonstrating posterior dislocation of the polyethylene insert (green arrow) and osteolysis around the upper plate. D: Sagittal T1-weighted MRI showing a footprint on the anterior spinal cord (blue arrow) without any signs of myelopathy.
MRI and computed tomography (CT) of the cervical spine 9 months postoperatively. A: Axial T2-weighted MRI showing artifact of the prosthesis plate and potential anterior spinal cord compression (blue arrow). B: Frontal CT showing anterior contact of the metallic plates, suggesting an absence of the polyethylene insert. C: Sagittal CT demonstrating posterior dislocation of the polyethylene insert (green arrow) and osteolysis around the upper plate. D: Sagittal T1-weighted MRI showing a footprint on the anterior spinal cord (blue arrow) without any signs of myelopathy.The patient received emergency surgery the same day under C3–T1 neuromonitoring (NIM Eclipse Surgeon Directed, Medtronic). A left cervicotomy was performed using microscopic magnification. The superior plate of the prosthesis was not adhesive and was easily removed. The mobile part of the prosthesis in polyethylene was posteriorly dislocated and retained by the posterior longitudinal ligament. The anterior horn of the polyethylene insert was ruptured and the global aspect was partially supple, allowing a hook to be inserted within (Fig. 4). The inferior plate of the prosthesis adhered well to the vertebral endplate and was removed using a bone osteotome. A sample of deep membrane was taken for microbiological analysis. A polyether-ether ketone cage was implanted with 1 cm3 of bone substitute in the interbody space. An anterior C6–C7 plate completed the instrumentation (Fig. 5).
FIG. 4.
Intraoperative views of the polyethylene insert. Orientation of A to C: left side is caudal; right side is cranial. A: Left anterior rupture fragment. B: Membranes reformed anteriorly to the insert and were sampled for microbiological analysis. C and D: One can notice the ridges on the surface of the implant performed with the microsurgical hook during its extraction and the anterior rupture of the left horn of the insert.
FIG. 5.
Postoperative radiographs. A: Anterior-posterior view showing a good position of the plate and the polyether-ether ketone implant for C6–C7 fusion. B: Lateral view showing adequate position of the implants and length of the screws.
Intraoperative views of the polyethylene insert. Orientation of A to C: left side is caudal; right side is cranial. A: Left anterior rupture fragment. B: Membranes reformed anteriorly to the insert and were sampled for microbiological analysis. C and D: One can notice the ridges on the surface of the implant performed with the microsurgical hook during its extraction and the anterior rupture of the left horn of the insert.Postoperative radiographs. A: Anterior-posterior view showing a good position of the plate and the polyether-ether ketone implant for C6–C7 fusion. B: Lateral view showing adequate position of the implants and length of the screws.Postoperatively, the patient showed no complications, and the radicular pain was relieved. The patient was discharged from the clinic after 2 days. The sample taken remained negative for microbial culture. The follow-up review was without issue, with adequate wound healing and normal cervical motion after 2 months. The NRS reached 4/10 and Neck Disability Index was 34%.
Discussion
Observations
To our knowledge, this is the first case of posterior intraprosthetic dislocation of a CDA reported in the literature so far. Tsermoulas and Bhattathiri[26] reported the first case of anterior dislocation of a C5–C6 Mobi-C arthroplasty. They found a fixed expulsion of the inferior plate and insert from the prosthesis after 8 weeks and performed emergency ACDF with good results. They hypothesized that the lack of prosthesis restraint may have been the main cause of the implant migration. In that case, the 54-year-old female patient had severe vomiting and coughing episodes that may have increased stress on the implant and triggered its dislodgment.Pelletier et al. described a case of an early anterior intraprosthetic dislocation of a C4–C5 Mobi-C CDA[27] related to excessive motion of the mobile segment adjacent to a two-level C5–C6 and C6–C7 ACDF. The dislocated CDA was removed and ACDF was performed with an anterior plate placed from C4 to C7.Several causes of CDA failure have been discussed, such as inappropriate patient selection, under- or oversized implants, and technical error.[28] In the current case, the polyethylene insert showed signs of wear and partial rupture. The delay of occurrence and motion of the upper plate suggest a slow mechanism, with plate osteolysis and polyethylene oxidation, which might have been favored by tobacco exposure.[29] In addition to the lack of restraint of the prosthesis, these mechanisms may have resulted in intraprosthetic dislocation.
Lessons
Posterior intraprosthetic dislocation is an extremely rare complication that may occur with Mobi-C cervical arthroplasty in cases of rupture and oxidation of the polyethylene insert. Spine surgeons should be aware of this potential major complication. Further studies and investigations are needed to understand the exact causes of CDA failure.
Acknowledgments
We thank Mrs. Fiona Tweedie for her English language corrections.
Disclosures
The authors report no conflict of interest concerning the materials or methods used in this study or the findings specified in this paper.
Author Contributions
Conception and design: Prod’homme, Boscherini. Acquisition of data: Prod’homme, Grasset. Analysis and interpretation of data: Prod’homme. Drafting the article: Prod’homme. Critically revising the article: Prod’homme, Grasset. Reviewed submitted version of manuscript: Prod’homme, Grasset. Approved the final version of the manuscript on behalf of all authors: Prod’homme. Administrative/technical/material support: Grasset. Study supervision: Boscherini.