Literature DB >> 3605751

Flammability of endotracheal tubes in oxygen and nitrous oxide enriched atmosphere.

G L Wolf, J I Simpson.   

Abstract

Endotracheal tube (ETT) fire has been reported secondary to laser and electrocautery ignition. The flammability of polyvinylchloride (PVC), silicone (Si), and red rubber (RR) ETTs in oxygen (O2) and/or nitrous oxide (N2O) in nitrogen was determined and compared by means of the O2 and N2O indices of flammability. The O2 index of flammability is the minimum O2 fraction in nitrogen that will support candle-like flame using a standard ignition source. The O2 index of flammability for PVC ETTs is 0.263, for Si 0.189, and for RR 0.176. The N2O index of flammability is the minimum N2O fraction in N2 that will support candle-like flame using a standard ignition source. The N2O index of flammability for PVC ETTs is 0.456, for Si ETTs 0.414, and for RR ETTs 0.374. The indices are additive. Flammability is a valid method of comparing safety of various endotracheal tube materials. There is a need for new endotracheal tube material with a higher index of flammability. The significance of these findings and the clinical applications are discussed.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3605751     DOI: 10.1097/00000542-198708000-00014

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  8 in total

Review 1.  Airway fire during tracheostomy: prevention strategies for surgeons and anaesthetists.

Authors:  M L Rogers; R W Nickalls; E T Brackenbury; F D Salama; M G Beattie; A G Perks
Journal:  Ann R Coll Surg Engl       Date:  2001-11       Impact factor: 1.891

Review 2.  [Specific aspects of anesthesiological management of laser surgery in otorhinolaryngology].

Authors:  W A Wetsch; D Beutner; K J Stuermer; S A Padosch
Journal:  Anaesthesist       Date:  2014-06       Impact factor: 1.041

3.  [Specific aspects of anesthesiological management of laser surgery in otorhinolaryngology].

Authors:  W A Wetsch; D Beutner; K J Stuermer; S A Padosch
Journal:  HNO       Date:  2014-03       Impact factor: 1.284

4.  Abstracts: annual meeting of the Canadian Anesthetists' Society. June 26-29, 1988, Halifax, Nova Scotia.

Authors: 
Journal:  Can J Anaesth       Date:  1988-05       Impact factor: 5.063

5.  Use of i-gel for laser ablation of a bronchial lesion.

Authors:  Rakesh Garg; Sakshi Thakore; Karan Madan; Anant Mohan
Journal:  BMJ Case Rep       Date:  2017-10-24

Review 6.  Management of a fire in the operating room.

Authors:  Alan David Kaye; Daniel Kolinsky; Richard D Urman
Journal:  J Anesth       Date:  2013-08-30       Impact factor: 2.078

7.  Reducing the oxygen concentration of gases delivered from anaesthetic machines unadapted for medical air.

Authors:  R E Clutton; G Schoeffmann; M Chesnil; R Gregson; F Reed; H Lawson; M Eddleston
Journal:  Vet Rec       Date:  2011-08-22       Impact factor: 2.695

8.  Endotracheal tube fire during tracheostomy.

Authors:  Eunju Lee; Su-Nam Lee; Jong-Il Kim; Youbin Son
Journal:  Korean J Anesthesiol       Date:  2012-06-19