Tony Tien1, Samuel C Saccomano1, Pilar A Martin1, Madeleine S Armstrong2, Robert K Prud'homme2, Kevin J Cash1,3. 1. Chemical and Biological Engineering, Colorado School of Mines, Golden, Colorado 80401, United States. 2. Chemical and Biological Engineering, Princeton University, Princeton, New Jersey 08544, United States. 3. Quantitative Biosciences and Engineering, Colorado School of Mines, Golden, Colorado 80401, United States.
Abstract
Flash nanoprecipitation (FNP) is an efficient and scalable nanoparticle synthesis method that has not previously been applied to nanosensor fabrication. Current nanosensor fabrication methods have traditionally exhibited poor replicability and consistency resulting in high batch-to-batch variability, highlighting the need for a more tunable and efficient method such as FNP. We used FNP to fabricate nanosensors to sense oxygen based on an oxygen-sensitive dye and a reference dye, as a tool for measuring microbial metabolism. We used fluorescence spectroscopy to optimize nanosensor formulations, calibrate the nanosensors for oxygen concentration determination, and measure oxygen concentrations through oxygen-sensitive dye luminescence. FNP provides an effective platform for making sensors capable of responding to oxygen concentration in gas-bubbled solutions as well as in microbial environments. The environments we tested the sensors in arePseudomonas aeruginosa biofilms andSaccharomyces cerevisiae liquid cultures─both settings where oxygen concentration is highly dependent on microbial activity. With FNP now applied to nanosensor fabrication, future nanosensor applications can take advantage of improved product quality through better replicability and consistency while maintaining the original function of the nanosensor.
Flash nanoprecipitation (FNP) is an efficient and scalable nanoparticle synthesis method that has not previously been applied to nanosensor fabrication. Current nanosensor fabrication methods have traditionally exhibited poor replicability and consistency resulting in high batch-to-batch variability, highlighting the need for a more tunable and efficient method such as FNP. We used FNP to fabricate nanosensors to sense oxygen based on an oxygen-sensitive dye and a reference dye, as a tool for measuring microbial metabolism. We used fluorescence spectroscopy to optimize nanosensor formulations, calibrate the nanosensors for oxygen concentration determination, and measure oxygen concentrations through oxygen-sensitive dye luminescence. FNP provides an effective platform for making sensors capable of responding to oxygen concentration in gas-bubbled solutions as well as in microbial environments. The environments we tested the sensors in arePseudomonas aeruginosa biofilms andSaccharomyces cerevisiae liquid cultures─both settings where oxygen concentration is highly dependent on microbial activity. With FNP now applied to nanosensor fabrication, future nanosensor applications can take advantage of improved product quality through better replicability and consistency while maintaining the original function of the nanosensor.
Flash nanoprecipitation (FNP)
is a technique for the efficient and consistent fabrication of polymer-based
nanoparticles. FNP is a highly tunable and scalable formulation method
that has produced nanoparticles from the mg to ton scale.[1] The process takes advantage of a rapid mixing
process to create a supersaturated solution of nanoparticle components,
which causes the precipitation of nanoparticles through the mixing
of a solvent and antisolvent.[2] Other nanoprecipitation
techniques have been developed for the creation of nanoparticles,
but issues with these approaches remain ever-present with inconsistent
replicability, size uniformity, and product yield.[1,3] During
FNP, when the component mixing streams flow rapidly and turbulently,
nanoparticle size and composition can be related directly to stream
composition with little effect from streamflow rate, while the stability
is affected by formulation choice rather than fabrication parameters.[4] Through process and formulation tuning, FNP can
create nanoparticles with a controlled diameter from approximately
40 to 400 nm—useful in applications where particle size is
related to function, such as biocompatible protein conjugation and
biodistribution of the particles.[5,6] With this approach,
researchers have used FNP for applications such as drug delivery[7−10] and optical imaging applications.[10−12]To date, FNP has
not been applied for the fabrication of nanosensors.
Nanosensors change their optical properties in response to analyte
concentration changes—enabling optical measurement of local
conditions. Optical polymeric nanosensors are particularly well suited
for the detection of analytes due to their small size and tunability
of the response based on formulation.[13,14] Nanosensors
are particularly of interest for use in situ due
to their small sizes, inert and nontoxic formulations, and potential
to function in biologically relevant pH ranges.[15] Additionally, nanoparticles provide a unique application
for imaging as they can be used to measure 3-dimensional spatial gradients
and temporal changes in complex, heterogeneous samples.[16] However, the current fabrication approach for
nanosensors is reliant on emulsification solvent evaporation (ESE)—yielding
minimal nanoparticle size control, large polydispersity, and poor
options to select specific nanoparticle sizes while retaining sensor
function.[17] While some alternate approaches
with different materials have shown some limited size tuning,[18] the current approaches do not offer the control
over nanoparticle fabrication seen with FNP.While there are
many approaches to designing quantitative optical
sensors, one common approach is to measure ratiometric luminescence
changes. In these approaches, there are two luminescent signals from
the sensor based on two separate dyes: one which changes with analyte
concentration, and one which does not change with analyte concentration
to be used as a reference, as shown in Figure . For more detail on ratiometric approaches,
please see Doussineau et al.[19] The incorporation
of these sensing components into a polymeric matrix creates what we
call a nanosensor. The ratio of these two dyes remains fixed once
encapsulated, which enables analyte measurements in complex biological
systems by adjusting for nanoparticle concentration.[16,20,21] Nanosensors have been developed
for a variety of biologically relevant parameters such as pH,[22] electrolytes,[23] glucose,[24] heavy metal ions,[25] and pesticides.[26] A wide variety of analyte
nanosensors can be made by simply changing sensing components while
tuning the formulation for the concentration range of interest. Consequently,
improving the fabrication method of nanosensors has the potential
to bolster nanosensor resilience and consistency for use across the
wide range of existing applications.
Figure 1
We fabricated oxygen-sensitive nanosensors
using flash nanoprecipitation
(FNP) with sensing components used in other oxygen nanosensors. (a)
The oxygen sensing mechanism of the sensors is based on PtTPP—a
dye which luminesces most brightly in the absence of oxygen and decreases
in luminescence with increasing oxygen concentrations. DiA, used as
a reference dye, luminesces at a consistent level regardless of oxygen
concentration. The presence of DiA as a reference dye allows for oxygen
measurements, which are independent of many artifacts (e.g., sensor
concentration). (b) A confined-impinging jet mixer (CIJ) is used to
turbulently mix oxygen sensing components in THF with water as an
antisolvent, resulting in nanosensors fabricated with FNP. Created
with BioRender.com.
We fabricated oxygen-sensitive nanosensors
using flash nanoprecipitation
(FNP) with sensing components used in other oxygen nanosensors. (a)
The oxygen sensing mechanism of the sensors is based on PtTPP—a
dye which luminesces most brightly in the absence of oxygen and decreases
in luminescence with increasing oxygen concentrations. DiA, used as
a reference dye, luminesces at a consistent level regardless of oxygen
concentration. The presence of DiA as a reference dye allows for oxygen
measurements, which are independent of many artifacts (e.g., sensor
concentration). (b) A confined-impinging jet mixer (CIJ) is used to
turbulently mix oxygen sensing components in THF with water as an
antisolvent, resulting in nanosensors fabricated with FNP. Created
with BioRender.com.Given that oxygen-sensing dyes have already been well characterized
in nanosensors,[16,27,28] oxygen sensors make for an ideal system to study the FNP formulation
method. The biological relevance of these sensors would benefit from
improved replicability with a successful implementation in FNP.Nanoscale systems for measuring biological samples have become
increasingly important due to the need for methods which can capture
spatial and temporal dynamics. The metabolic activity of a microbial
community is important to understand the vitality and viability of
the consortia in response to the environmental conditions.[29] Under favorable growth conditions, such as high
nutrient availability, cell metabolism is more likely to be high while
unfavorable conditions are more likely to slow or cease metabolic
activity. For example, metabolic activity has become a common proxy
for determining antibiotic susceptibility of antimicrobial resistant
species.[20,30,31] Additionally,
metabolic activity has a widespread effect on the surrounding as the
depletion or production of metabolites can have a profound effect
on community.[32]Many systems have
been used to sense a variety of indicators related
to metabolic activity. Most of these sensors target key biomolecules
which are either consumed or produced by the microbial community of
interest. Yeor-Davidi et al. used silicon nanowires to detect redox
active metabolites such as glucose which could be oxidized by oxidase
enzymes. As metabolites are oxidized hydrogen peroxide is formed and
detected through a change in electrochemical potential.[33] Progress has been made to monitor NADP and NADPH
levels in biological systems through fluorescent probes. Goldbeck
et al. used a blue fluorescent protein (mBFP) to track NADPH levels
in Corynebacterium glutamicum and Escherichia coli.[34] Surger
et al. used an optical gas-phase carbon dioxide sensor for the detection
of microbial growth of multiple species in contaminated heating oils
and diesel fuels.[35] A variety of other
methods have been discussed by Braissant et al. which include the
detection of ATP, stable isotopic labeling, isothermal calorimetry,
and the monitoring of various electron acceptors such as nitrate,
sulfate, metal ions, and oxygen.[29]Oxygen is one of the most common indicators of metabolic activity
because of its role in aerobic respiration. Rapid oxygen consumption
within a system is evidence of a thriving microbial community. Oxygen
monitoring methods are most commonly categorized into electrochemical
methods, optode films, and nanoparticle/microparticle sensors. The
clark electrode is one of the most common methods of oxygen detection
due to its high sensitivity; however, oxygen is consumed in the process
and this method is often disruptive to the system under investigation.[36,37] Optode films are advantageous for measuring microbial growth on
a surface; however, due to geometric limitations optode films tend
to have slower response times and are limited to 2-dimensional imaging
applications.[38] Particle-based approaches
are ideal for oxygen monitoring due their fast response and uniform
dispersity within the system, allowing them to capture spatial and
temporal dynamics.In this work, we show that FNP provides a
platform for easily fabricating
nanosensors and can be controlled to produce nanosensors of replicable
size and composition. We demonstrate FNP’s potential with nanosensors
in measuring oxygen concentrations, showing that FNP-based nanosensors
respond similarly to prior oxygen sensors and function well in complex
biological settings. FNP is a new approach to fabricate polymer-based
nanosensors, opening the nanosensor field for the future potential
of producing nanosensors with lower barriers of accessibility and
new synthesis materials that have yet to be explored.
Experimental Section
Methods
For nanosensor fabrication,
platinum(II) 5,10,15,20-(tetraphenyl)porphyrin
(PtTPP) was purchased from Frontier Scientific (Logan, UT, USA), 4-(4-dihexadecylaminostyryl)-N-methylpyridinium iodide) (DiA) was purchased from ThermoFisher
Scientific (Waltham, MA, USA), and polystyrene-block-polyethylene glycol (PS–PEG) was from Polymer Source, Inc.
(Montreal, QC, CA); Vitamin E, Vitamin E acetate, and tetrahydrofuran
were purchased from MilliporeSigma (St. Louis, MO, USA).For
glucose oxidase assays, α-d-Glucose, Glucose Oxidase
from Aspergillus niger, and Dulbecco’s
Phosphate Buffered Saline (dry powder, DPBS Modified, without calcium,
without magnesium, suitable for cell culture) were purchased from
MilliporeSigma (St. Louis, MO, USA). 96-well black-walled optical
bottom plates from ThermoFisher Scientific (Waltham, MA, USA) were
used to contain the samples.For pH testing, a Fisherbrand accumet
AB150 pH meter and probe,
10 N HCl, and 10 N NaOH from ThermoFisher Scientific (Waltham, MA,
USA) were used.For antibiotic susceptibility testing, Pseudomonas
aeruginosa PAO1 was purchased from ATCC (Manassas,
VA, USA), colistin sulfate was purchased from MilliporeSigma (St.
Louis, MO, USA), Saccharomyces cerevisiae (brewing yeast) strains MIP-510 (Kolsch I) and MIP-354 (Kveik: Oslo)
were purchased from Propagate Laboratories (Golden, CO, USA), and
Campden tablets (potassium metabisulfite, PMB) obtained from Crosby
and Baker (Westport, MA, USA). 96-well black-walled optical bottom
plates from ThermoFisher Scientific (Waltham, MA, USA) were used to
contain the samples.For the gas bubbling setup, ultra-high-purity
nitrogen gas and
compressed air from Matheson (Denver, CO, USA), a 10 mm path length
quartz cuvette with rubber septa seal from Starna Cells (Atascadero,
CA, USA), Aalborg nitrogen
mass flow controller (Orangeburg, NY, USA), and an air mass flow controller
from Alicat Scientific (Tucson, AZ, USA) were used.
Nanosensor
Fabrication
Nanosensors were produced via
FNP using a confined-impinging jet mixer (CIJ) as described previously
and as shown in Figure b.[30] Briefly, PtTPP, DiA, PS–PEG,
and either Vitamin E or Vitamin E acetate were dissolved in a THF
stream and rapidly mixed against a DI water antisolvent stream into
a quench bath of DI water to drive controlled precipitation and produce
the stabilized nanosensors. Nanosensor size was measured by dynamic
light scattering (Malvern Nanosizer, Westborough, MA, USA). The concentrations
of components in the THF feed stream were adjusted to obtain nanosensors
with different compositions and explore the effect of various parameters
on the nanosensors’ performance. The varied parameters include
the ratio of the oxygen-responsive dye, PtTPP, to the reference dye,
DiA, the total mass concentration of the core of the nanosensor, the
concentration of the PS–PEG stabilizer, and the use of either
Vitamin E or Vitamin E acetate as the main core component. The compositions
are given in Table . Emulsification solvent evaporation sensor fabrication was adapted
from Jewell et al.[16] The dyes were replaced
with 1.25 mg of PtTPP and 0.5 mg DiA in the optode formulations.
Table 1
Formulations of FNP Oxygen Nanosensors
Tested
Sample Number
PtTPP:DiA
(mg:mg)
PS–PEG (mg/mL)
Core (mg/mL)
Core Component
1
5:0.2
5
5
Vitamin E acetate
2
5:0.2
1.25
5
Vitamin E acetate
3
5:0.2
5
5
Vitamin E
4
5:0.2
1.25
5
Vitamin E
5
5:2
5
5
Vitamin E acetate
6
5:2
1.25
5
Vitamin E acetate
7
5:0.2
5
10
Vitamin E acetate
8
5:0.2
1.25
10
Vitamin E acetate
9
5:2
5
10
Vitamin E acetate
10
5:2
1.25
10
Vitamin E acetate
Sample Selection via Glucose
Oxidase Assay
Ten variations
of FNP nanosensors (Table ) were tested on a BioTek Synergy H1 Hybrid Multi-Mode microplate
reader (Winooski, VT, USA). After samples were added to a 96-well
optical bottom well plate, they were excited at 488 nm to generate
luminescence spectra in intervals of 1 nm, and ratiometric data was
analyzed to find the best sample. Optimal samples had similar magnitudes
of PtTPP and DiA luminescence, greater signal intensity, and greater
signal contrast between deoxygenated and oxygenated states. A glucose
oxidase assay was performed using 100 mM glucose and 20 IU/mL glucose
oxidase stocks in PBS. Each well tested contained 60 μL of nanosensors.
Deoxygenated samples contained 18 μL each of the glucose and
glucose oxidase stocks with an additional 84 μL of PBS. Oxygenated
controls contained PBS only, glucose and PBS, or glucose oxidase and
PBS with the same total well volume of 180 μL.
Calibration
Curves and Reversibility
The FNP oxygen
nanosensors were tested for response to changes in oxygen concentration
in an air-nitrogen gas bubbling system as described by Saccomano et
al.[21] Oxygen calibration curves and reversibility
tests were completed with an Avaspec 2408L spectrometer (Avantes Inc.,
Louisville, CO) and analyzed using Avantes AvaSoft 8. Two mL of nanosensors
were used in a septum sealed quartz cuvette and fitted with gas line
and vent using 22-gauge needles. Gas flow rates from an air and nitrogen
tank were controlled by mass flow controllers and mixed in a 25 mL
mixing chamber to form gas streams with various mole fractions of
oxygen from 0% to 21% or 0 to 6.65 mg/L dissolved oxygen at 5,780
ft elevation (Golden, CO). Gas was bubbled into the nanosensor solution
in the cuvette at a total flow rate of 20 mL/min. To test the luminescence
at each dissolved oxygen concentration the gas stream was allowed
to bubble for 20 min at which point the needle was removed and a fiber
optic 532 nm LED was shone on the cuvette. This was repeated for 0
to 6.65 mg/L dissolved oxygen concentrations in increments of 0.79
mg/L (2.5%). Sensor reversibility was tested by alternating between
0 mg/L dissolved oxygen (nitrogen bubbling) and 6.65 mg/L dissolved
oxygen (air bubbling) at 20 min each to observe the response of the
sensors to quenched and unquenched states. The samples were measured
over three cycles of oxygenation and deoxygenation states.
Pseudo-Stern–Volmer
Analysis
The calibration
curve data was used to perform a pseudo-Stern–Volmer analysis.
To measure the luminescence of the individual dyes, the maximum peak
intensities at 580 nm (DiA) and 660 nm (PtTPP) were taken. The ratiometric
signal was determined by dividing the luminescence of the of the oxygen
peak (PtTPP) to the reference peak (DiA). A pseudo-Stern–Volmer
plot was generated by dividing the ratiometric signal in the absence
of oxygen by the ratiometric signal at each of the tested oxygen concentrations
using the Stern–Volmer equation. These values are plotted with
respect to oxygen concentration, generating a linear correlation between
the luminescence ratio and oxygen concentration. A linear regression
of the plot was taken to determine the pseudo-Stern–Volmer
constant KpSV.
Functional Lifetime
We tested the viability of the
sensor response over an extended period of time with expected sensor
and dye degradation over time. The sensors were evaluated at 40 days
and 100 days from sensor fabrication using the glucose oxidase assay
as described above.
pH Response
PBS samples were prepared
in a pH range
of 5–8 in intervals of 1, as well as a stock pH of 7.40. The
pH of the various PBS samples was adjusted using 1 N HCl and 1 N NaOH
(diluted from their respective 10 N stocks) using live measurements
with a pH electrode. 160 μL of nanosensors were mixed with 40
μL of each PBS variation in a 96-well plate. The samples were
excited at 488 nm in intervals of 1 nm to generate emission spectra
from 500 to 700 nm analyze dye signal changes in an ambient oxygen
environment.
Antibiotic Susceptibility Test (AST) with Pseudomonas
aeruginosa
Nanosensors were concentrated
to 10× using MilliporeSigma Amicon Ultra-0.5 Centrifugal Filter
Units (St. Louis, MO, USA) and a microcentrifuge. Setup and execution
of the AST was adapted from Jewell et al.[20] The antibiotic challenge plate was set up as follows: column 1 contained
no antibiotic, column 2 was blank, and columns 3–10 contained
serial half dilutions of 250 μg/mL colistin sulfate in PBS.
Columns 11–12 contained PBS to observe the growth of the biofilm
in absence of nanosensors.
Plate Reader Analysis for AST
Luminescence
end point
readings were taken at 488 nm excitation. Emission wavelengths of
580 and 660 nm were used for DiA and PtTPP, respectively. The ratiometric
signal of the oxygenated peak (PtTPP) to deoxygenated peak (DiA) was
compared to antibiotic concentration. The luminescence readings were
normalized to the blank readings, after which we fitted a linear regression
to each column’s data from 0 to 5 h. The slope of the ratiometric
signal was graphed as a function of antibiotic concentration.
Tracking
Metabolic Changes with Saccharomyces
cerevisiae (Brewing Yeast) Strains MIP-510 (Kolsch
I) and MIP-354 (Kveik:Oslo)
Metabolic activity was measured
with the nanosensors using a protocol similar to that in Saccomano
et al.[21] The yeast strains were diluted
in their respective wort solutions at 1:3, 1:10, and 1:100 concentrations.
Samples were dispensed into a 96-well plate with all permutations
of the given yeast strains, yeast dilutions, and potassium metabisulfite
(PMB) condition in quadruplicate. In addition, controls of each strain
without nanosensors and nanosensors with/without PMB were performed
in sextuplicate. The samples were incubated at 30 °C and shaken
for 5 min before each reading, with readings taken every 15 min over
a 60 h period. Aluminum foil was used to cover the 96-well plate with
minute holes punched into the foil for potassium metabisulfite addition
at 42 h.
Results
Our nanosensors contain
two dyes, PtTPP and DiA, which are used
to sense changes in oxygen concentration. DiA was chosen as the oxygen-insensitive
reference dye, and PtTPP was chosen as the oxygen-responsive dye.
Oxygen functions as a quenching agent for PtTPP, where PtTPP luminescence
decreases with increasing oxygen concentrations. At low oxygen concentrations,
luminescence quenching is minimized resulting in greater signal from
the PtTPP dye. As oxygen concentrations increase, it quenches only
the PtTPP, leaving the DiA luminescence unchanged, which changes the
overall luminescence ratio between the two dyes.In fabricating
nanosensors using FNP, the formulation of the supporting
matrix is different from our prior sensors. These FNP nanosensors
contain PS–PEG and a core material of either Vitamin E or Vitamin
E acetate. However, this change in material properties is not a concern
for the sensing mechanism, as other work has already demonstrated
that for this class of sensors, the structural nanoparticle matrix
is easily adaptable to other materials assuming the hydrophobicity
remains similar, a result extended in this work.[14,18,39]The first step in the process of developing
these new FNP produced
nanosensors was to determine the impact of formulation parameters
on the nanosensor response. The key parameters we tested were as follows:
oxygen-responsive dye to reference dye ratio (PtTPP:DiA), structural
component amounts (PS–PEG, Core), and core component identity
(vitamin E or vitamin E acetate), as seen in Table . The designations in mg/mL in the table
represent the concentrations of the components in the solvent stream
introduced in the CIJ mixer. Each sample was measured to compare particle
size and size distribution as seen in Supporting Information (SI) Figures S1 and S2. Greater average particle
diameter was correlated to smaller PS–PEG amounts and larger
core amounts highlighting the ability to tune particle size. We also
assessed luminescence output from the nanosensors and their response
to oxygen concentration.Resulting luminescence spectra for
all formulations are shown in Figures S3 and S4 in SI. Formulations containing
the 5:2 ratio of PtTPP:DiA provided reference dye signals (DiA) similar
in magnitude to oxygen-responsive dye signals (PtTPP). This is desirable,
as ratios of different magnitudes can complicate sensor interrogation
with a single measurement. Nanosensor formulations containing a lower
concentration of PS–PEG (1.25 mg/mL) exhibited ∼20–50%
decreases in luminescence intensity compared to 5 mg/mL counterparts,
creating a preference for higher PS–PEG concentrations. Higher
quantities of core material (10 mg/mL) reduced the luminescence intensity
of the reference dye (DiA). For this reason, lower amounts of core
material are preferred to ensure similar peak sizes between the two
dyes. Furthermore, Samples 3 and 4, which contained Vitamin E in place
of Vitamin E acetate, exhibited lower overall luminescence when compared
to the other samples. Given that absorbance data between the formulations
tested were comparable in value, Vitamin E appears to adversely affect
sensor luminescence, potentially through quenching of the nanosensor
signal, since Vitamin E functions as an antioxidant. Therefore, we
used formulations containing Vitamin E acetate to achieve greater
signal intensities. Given the desirable characteristics and parameters
from this screening, sample 5’s formulation (5:2 PtTPP:DiA;
5 mg/mL PS–PEG, 5 mg Vitamin E acetate) exhibited the optimal
balance between ratiometric signal of the two dyes, luminescence signal
intensity, and response to oxygen. This formulation was used for all
the following characterization tests.The encapsulation efficiency
of the Flash Nanoprecipitation approach
for the optimized sensor composition was measured by comparing the
absorbance value of the dyes in the filtrate and the retentate of
samples filtered through a 30k MW amicon column. No signal was found
in the filtrate where any free dye molecules would be found indicating
that close to 100% of the dye was in the particles (see Figure S5). We evaluated the loading efficiency
of the FNP particles as compared to a standard emulsification solvent
evaporation method by comparing the absorbance of the two dyes in
solution of each sensor batch. Each method was loaded with equivalent
masses of the two dyes per 5 mL batch, 125 μg of PtTPP and 50
μg of DiA. The FNP batch clearly showed a greater absorbance
above the background than the ESE sensors (see Figure S5) which is supported by the observation that dye
is observed in large aggregates left over from the sonication process,
indicating that some amount of the dye did not make it into the particles.
We verified that dye leaching from the particle was very slow based
on filtering out any free dye through a 30k molecular weight filter
and measuring the change in absorbance over time. Over 2 weeks we
saw no increase in absorbance (Figure S5).Aggregation caused quenching (ACQ) was looked into as a
potential
phenomenon which may be present when we encapsulate our dyes in nanoparticles. Figure S6 shows that fluorescence was brighter
in the nanoparticle phase versus an organic solvent phase, indicating
that ACQ is unlikely as we would expect the brightness to decrease
relative to that in solution.Sensors made by FNP and ESE methods
were characterized by DLS to
obtain a comparison of nanoparticle size and zeta potential between
the two methods (Figure S7). We tested
stability of the two methods using the sample 5 formulation as our
test case. To measure stability of the nanoparticle itself we measured
the size of the sensor batch at 0 weeks, 1 week, and 2 weeks from
the initial synthesis. Results showed that there were no significant
changes in particle size over that time period (see Figure S8).When our nanosensor formulation was tested
for oxygen response
from 0 to 6.65 mg/L a notable response can be observed in the luminescence
of PtTPP at 660 nm, which decreases as oxygen concentration increases
(see Figure ). Minor
changes in the luminescence of the reference dye, DiA, can also be
observed at 580 nm; the intensity of the DiA peak decreased with increasing
oxygen concentration as well, as observed in Figure S9. To counteract this issue, a ratiometric approach was used
to normalize the spectra to the DiA peak allowing us to better evaluate
the oxygen response. The spectra shows a general decrease in signal
ratio with increasing oxygen concentration, as shown in Figure . A pseudo-Stern–Volmer
plot, as shown in Figure , demonstrates the relationship with changing oxygen concentration
in terms of the change in luminescence ratio compared to that ratio
at zero oxygen. Despite slight fluctuations in the luminescence of
DiA, a linear relationship is still achieved over the full oxygen
range, as shown in Figure S10.
Figure 2
Normalized
luminescence spectra for ratiometric nanosensors containing
an oxygen-sensitive PtTPP dye and an oxygen-insensitive DiA dye at
varying dissolved oxygen concentrations. Oxygen nanosensors change
in luminescence according to ambient oxygen changes by bubbling air/nitrogen
mixtures into the sample within a cuvette system. Dashed lines indicate
the peak wavelengths for the reference dye (DiA) at 580 nm and the
oxygen-responsive dye (PtTPP) at 660 nm. As oxygen concentration is
increased from 0 mg/L to 6.65 mg/L, PtTPP luminescence is quenched,
while DiA luminescence remains relatively constant. Data was normalized
to peak intensity at 580 nm.
Figure 3
Pseudo-Stern–Volmer
(pSV) plot of the ratiometric in response
to oxygen concentration change from 0 mg/L to 6.65 mg/L, where 6.65
mg/L represents the dissolved oxygen mixture at 5780 ft elevation.
A relatively linear slope indicates optimal ratiometric behavior of
the sensor response, allowing for a pseudo-Stern–Volmer constant
(KpSV) to be extrapolated to determine
sensor response at oxygen concentrations not included in initial testing.
The error bars represent standard deviation for n = 3 replicates. Dotted lines indicate 95% confidence interval of
the linear fit.
Normalized
luminescence spectra for ratiometric nanosensors containing
an oxygen-sensitive PtTPP dye and an oxygen-insensitive DiA dye at
varying dissolved oxygen concentrations. Oxygen nanosensors change
in luminescence according to ambient oxygen changes by bubbling air/nitrogen
mixtures into the sample within a cuvette system. Dashed lines indicate
the peak wavelengths for the reference dye (DiA) at 580 nm and the
oxygen-responsive dye (PtTPP) at 660 nm. As oxygen concentration is
increased from 0 mg/L to 6.65 mg/L, PtTPP luminescence is quenched,
while DiA luminescence remains relatively constant. Data was normalized
to peak intensity at 580 nm.Pseudo-Stern–Volmer
(pSV) plot of the ratiometric in response
to oxygen concentration change from 0 mg/L to 6.65 mg/L, where 6.65
mg/L represents the dissolved oxygen mixture at 5780 ft elevation.
A relatively linear slope indicates optimal ratiometric behavior of
the sensor response, allowing for a pseudo-Stern–Volmer constant
(KpSV) to be extrapolated to determine
sensor response at oxygen concentrations not included in initial testing.
The error bars represent standard deviation for n = 3 replicates. Dotted lines indicate 95% confidence interval of
the linear fit.When tested for reversibility,
the luminescence ratio increased
by 67.4% over three reversibility cycles for oxygenated environments
and increased by 54.2% for deoxygenated environments (see Figure S11a). The increasing luminescence ratio
over an increased number of reversibility cycles is driven by decreases
in DiA signal over time with increasing PtTPP signal over time, as
also observed in Figure S3b. When analyzed
separately, PtTPP increases in deoxygenated luminescence by approximately
17.5% after three cycles, with the DiA signal being reduced by 39.0%
in the same time frame (see Figure S11b,c).The nanosensors were also periodically tested for signal
stability
over a period of four months. Between 40 and 100 days after fabrication,
the reference dye (DiA) was less stable when treated with glucose
and glucose oxidase, producing a noticeable signal drop from the oxygenated
control (see Figure S12). Even though the
overall signal of the sensors reduced by approximately 50%, the sensors
remained responsive to oxygen—though with a different calibration
curve—necessitating recalibration for quantitative application.
The nanosensor response was tested in a variety of pH conditions from
pH 5 to 8. The pH had little effect on the raw intensity or the response
of the sensors to oxygen (see Figure S13).To test the function of these nanosensors in biological
systems,
we applied them to monitor metabolic oxygen consumption in Pseudomonas aeruginosa biofilms in response to antibiotic
concentration log(2) dilution series using a previously developed
assay.[20] Under natural conditions these
biofilm-forming microbes rapidly deplete oxygen in the extracellular
matrix; however, when exposed to high colistin sulfate (antibiotic)
doses, metabolic oxygen consumption rates are slowed resulting in
increased oxygen levels in the sample with respect to antibiotic concentration.[16] This approach allows for the determination of
a minimal biofilm inhibitory concentration (MBIC).[20] When tested in P. aeruginosaPAO1 biofilms with 1 to 250 μg/mL antibiotic
concentrations, our nanosensor assay measurement (as a proxy for metabolic
rate) decreases with increasing antibiotic concentration as expected
(Figure ). The smaller
assay response indicates that oxygen concentration is closer to atmospheric
values—meaning that oxygen consumption within the biofilm was
lower in samples with higher concentrations of antibiotic as expected.
Figure 4
Metabolic
response of P. aeruginosa biofilms
with varying concentration of colistin sulfate as measured
by oxygen nanosensors. Nanosensors were grown in P.
aeruginosaPAO1 biofilms for antibiotic
susceptibility testing (AST). The assay response represents the slope
of the ratiometric signal (oxygen consumption) within the biofilms
from over the first 5 h of antibiotic exposure; higher values indicate
greater oxygen consumption over time, whereas values close to 0 represent
minimal oxygen consumption change over time. Error bars represent
standard deviation of the assay response for n =
4 replicates. Raw luminescence plots for individual dyes and ratiometric
signals are provided in Figures S14–S15 in SI.
Metabolic
response of P. aeruginosa biofilms
with varying concentration of colistin sulfate as measured
by oxygen nanosensors. Nanosensors were grown in P.
aeruginosaPAO1 biofilms for antibiotic
susceptibility testing (AST). The assay response represents the slope
of the ratiometric signal (oxygen consumption) within the biofilms
from over the first 5 h of antibiotic exposure; higher values indicate
greater oxygen consumption over time, whereas values close to 0 represent
minimal oxygen consumption change over time. Error bars represent
standard deviation of the assay response for n =
4 replicates. Raw luminescence plots for individual dyes and ratiometric
signals are provided in Figures S14–S15 in SI.The nanosensors were also tested
to monitor oxygen consumption
in Saccharomyces cerevisiae fermentations
with potassium metabisulfite (PMB) as an antimicrobial agent to halt
metabolic activity—an approach described by Saccomano et al.[21] As seen in Figure , once PMB is added 42 h after yeast incubation,
a noticeable decline in the ratiometric signal is observed with both
yeast strains, though the Kolsch strain showed a more dramatic initial
drop post-PMB addition. The decrease in ratiometric signal indicates
a decrease in yeast oxygen consumption and an increase in ambient
oxygen concentration indicating the efficacy of the antimicrobial
agent as expected. Thus, continuous monitoring of oxygen levels is
possible for measuring metabolic activity with these nanosensors.
Variations between the two oxygen consumption curves in Figure match previous oxygen consumption
trends between the two yeast strains as previously observed.[21] Thus, the nanosensors can continuously monitor
oxygen consumption with adequate reversibility capabilities.
Figure 5
Oxygen consumption
in S. cerevisiae for (a) Kolsch and
(b) Kveik strains was monitored using oxygen
nanosensors. Before PMB was added, the oxygen nanosensors continuously
monitored oxygen consumption levels in both the Kolsch and Kveik strains.
At 42 h after initial incubation, potassium metabisulfite (PMB) was
added to inhibit yeast growth and test nanosensor response to yeast
oxygen consumption, as indicated by the dotted line. Once PMB was
added, the ratiometric signal of the sensors decreased, indicating
decreased yeast oxygen consumption in both strains. The Kolsch PMB
sample experienced a sharper decrease in oxygen consumption than the
Kveik strain, as measured by the ratiometric signal. Error bars have
been removed for clarity; figures with error bars can be found in Figures S10–S12 in SI.
Oxygen consumption
in S. cerevisiae for (a) Kolsch and
(b) Kveik strains was monitored using oxygen
nanosensors. Before PMB was added, the oxygen nanosensors continuously
monitored oxygen consumption levels in both the Kolsch and Kveik strains.
At 42 h after initial incubation, potassium metabisulfite (PMB) was
added to inhibit yeast growth and test nanosensor response to yeast
oxygen consumption, as indicated by the dotted line. Once PMB was
added, the ratiometric signal of the sensors decreased, indicating
decreased yeast oxygen consumption in both strains. The Kolsch PMB
sample experienced a sharper decrease in oxygen consumption than the
Kveik strain, as measured by the ratiometric signal. Error bars have
been removed for clarity; figures with error bars can be found in Figures S10–S12 in SI.
Discussion
The fabrication of nanosensors with the FNP method
provides new
potential for the sensing field by coupling previously characterized
sensing mechanisms with the scalability and repeatability of FNP nanoparticles.
Oxygen sensing was used for testing, as it has been well characterized
with other nanosensor formulation methods, and it has applications
in tracking metabolic activity from various organisms, includingP. aeruginosaandS. cerevisiae (brewing yeast).[16,21,40] Because FNP has applications in scalable manufacturing, FNP can
be used to improve the scalability and replicability of nanosensors
for more widespread diagnostic and exploratory applications.The nanoparticle self-assembly step requires an amphiphilic block
copolymer to stabilize the nanoparticles. We have demonstrated that
a range of block copolymers can be used.[3,41] For this application
we used the polystyrene-b-polyethylene glycol (PS-b-PEG). The reason for this is that the block copolymer
is not degradable, and the nanoparticles are stable in aqueous solution
for indefinite periods of time. If this sensor technology were translated
into clinical applications, then the same nanoparticles could be made
from the block copolymer poly(lactic acid)-b-polyethylene
glycol (PLA-b-PEG), which is approved for parenteral
drug administration. The nanoparticles made with the same internal
components as described in this paper can be made using PLA-b-PEG as are made using PS-b-PEG. We demonstrated
this in the paper by Pagels.[3]The
FNP sensors showed that they have several advantageous properties
over traditional nanoparticle synthesis methods such as emulsification
solvent evaporation (ESE). The ability to tune the size of the nanoparticles
was demonstrated by altering the ratio of the core and structural
components. It is possible that we could achieve a wider range of
nanoparticle sizes by further optimization of the synthesis composition.
Additionally, the sensors showed better stability and similar levels
of leaching compared their ESE counterparts with the same dye loading
ratio making it more practical to scale and store FNP sensor batches
long-term. FNP also demonstrated a higher loading because the ESE
tended to form more dye-containing aggregates which get filtered out
in the sensor processing protocol.When the sensors were tested
for oxygen responsiveness, the signal
of the oxygen-responsive dye (PtTPP) decreased with increasing oxygen
concentrations, allowing for a relationship to be drawn between PtTPP
signal intensity and oxygen concentration. However, the reference
dye (DiA) signal also exhibited slight changes before normalization,
decreasing in luminescence with increasing oxygen concentrations,
although with a different slope. Fortunately, the normalized pseudo-Stern–Volmer
plot (Figure ) shows
that the ratiometric measurement can still be used to correlate luminescence
to dissolved oxygen even as the DiA signal is changing, highlighting
the value of ratiometric measurements. While the fluctuations in reference
dye signal provide an impetus to investigate other reference dye candidates
for future formulations, ratiometric changes between the oxygen-responsive
dye and reference dye used provide initial indications that the sensors
are capable of effectively responding to changes in oxygen concentrations
at least 40 days after fabrication and with recalibration up to 100
days after fabrication. Thus, FNP nanosensors can be used for longer
term sensing applications.We also found that the FNP sensors
exhibited reversibility when
purged with 0% and 21% oxygen environments. However, a gradual decrease
in reference dye (DiA) signal led to an increasing luminescence ratio
over multiple cycles. The excitation LED was powered for no longer
than 5 s per reading, which provided enough time to generate the reversibility
emission spectra. Thus, decreasing DiA signals may be attributed to
the degradation of DiA with prolonged excitation light exposure over
many cycles. Additional testing would be required to determine the
effect on full-range oxygen sensing (0 to 6.65 mg/L) that age would
contribute to.Furthermore, determining oxygen concentration
in biological samples
is of interest, particularly as nanosensors fabricated with alternate
approaches functioned in bacterial biofilms and yeast cultures. Preliminary
testing with P. aeruginosa biofilms
indicated that the FNP sensors are capable of sensing oxygen changes
in biofilms in response to antibiotic administration. The sensors
were also capable of measuring distinct, real-time changes in metabolic
activity between different S. cerevisiae brewing strains (Kolsch and Kveik) after antibiotic exposure.In both cases, the decreasing sensor signal with increased antibiotic
concentration is indicative of decreasing oxygen utilization caused
by cell death, where the fluorescence ratio at high antibiotic doses
(where we expect no metabolism) matches that of our atmospheric calibration—implying
no metabolic consumption of oxygen. The trends in sensor response
that we have observed in these experiments generally match other oxygen
sensor studies that have been done previously in similar microbial
environments showing a lack of significant change in function of these
nanosensors due to the FNP fabrication process. A future application
may include investigating the ability of FNP nanosensors to provide
a platform for 3D luminescence imaging of biofilms as previously done
with PtTFPP/DiA sensors fabricated by an emulsification–solvent
evaporation technique.[16]Our oxygen
sensors show that FNP is a viable method for fabricating
nanoparticles for analyte measurements. The FNP method for developing
nanosensors does not impede the use of dyes of interest for sensing,
providing potential for adaptations of existing sensor dyes to measure
various other analytes in biological samples.As nanosensors
are developed in the future for more strictly controlled
assays, the ability to scale up and produce a consistent product will
become very important. Even in exploratory bench-scale experiments,
the need for lower batch to batch variabilities makes FNP a desirable
method for nanosensor fabrication. Current nanoparticle synthesis
methods produce several milliliters of nanosensor solution per batch,[14,39] where FNP methods could potentially produce liter scale batches
of sensor for high-throughput or large volume testing.[1] The ability to tune nanoparticle size enables better control
over parameters such as sensor response time and diffusive properties in vivo.[42,43] Given the ability to implement
oxygen sensing into FNP products, future exploration of applicability
to ionophore sensors could allow for ion-sensing in various environments.
Conclusion
FNP is a valuable technique that enables fast and reliable fabrication
of nanosensors. When a typical nanoprecipitation formulation of oxygen
nanosensors was fabricated with the FNP process, the sensors exhibited
sensing behavior that was similar to other fabrication methods, allowing
for the sensor components to function as originally intended. The
nanosensors function in biological systems, measuring the antibiotic
and antimicrobial agent mediated changes in P. aeruginosa and S. cerevisiae metabolism. In
all of the tested conditions, the sensors were able to measure oxygen
in real time with adequate reversibility, as FNP allows the existing
sensor mechanisms to work effectively. Nanosensor technology can benefit
from the uniformity of FNP products to ensure uniform population for
a given application, and while FNP had not been previously applied
to nanosensors, this work opens up possibilities for future FNP-based
nanosensors for various analytes. Future applications can include
adaptations to other oxygen nanosensor formulations, ion sensors,
and testing in other biological systems.