Steroids play key roles in various biological processes and are characterized by many isomeric variants, which makes their unambiguous identification challenging. Ion mobility-mass spectrometry (IM-MS) has been proposed as a suitable platform for this application, particularly using collision cross section (CCS) databases obtained from different commercial IM-MS instruments. CCS is seen as an ideal additional identification parameter for steroids as long-term repeatability and interlaboratory reproducibility of this measurand are excellent and matrix effects are negligible. While excellent results were demonstrated for individual IM-MS technologies, a systematic comparison of CCS derived from all major commercial IM-MS technologies has not been performed. To address this gap, a comprehensive interlaboratory comparison of 142 CCS values derived from drift tube (DTIM-MS), traveling wave (TWIM-MS), and trapped ion mobility (TIM-MS) platforms using a set of 87 steroids was undertaken. Besides delivering three instrument-specific CCS databases, systematic comparisons revealed excellent interlaboratory performance for 95% of the ions with CCS biases within ±1% for TIM-MS and within ±2% for TWIM-MS with respect to DTIM-MS values. However, a small fraction of ions (<1.5%) showed larger biases of up to 7% indicating that differences in the ion conformation sampled on different instrument types need to be further investigated. Systematic differences between CCS derived from different IM-MS analyzers and implications on the applicability for nontargeted analysis are critically discussed. To the best of our knowledge, this is the most comprehensive interlaboratory study comparing CCS from three different IM-MS technologies for analysis of steroids and small molecules in general.
Steroids play key roles in various biological processes and are characterized by many isomeric variants, which makes their unambiguous identification challenging. Ion mobility-mass spectrometry (IM-MS) has been proposed as a suitable platform for this application, particularly using collision cross section (CCS) databases obtained from different commercial IM-MS instruments. CCS is seen as an ideal additional identification parameter for steroids as long-term repeatability and interlaboratory reproducibility of this measurand are excellent and matrix effects are negligible. While excellent results were demonstrated for individual IM-MS technologies, a systematic comparison of CCS derived from all major commercial IM-MS technologies has not been performed. To address this gap, a comprehensive interlaboratory comparison of 142 CCS values derived from drift tube (DTIM-MS), traveling wave (TWIM-MS), and trapped ion mobility (TIM-MS) platforms using a set of 87 steroids was undertaken. Besides delivering three instrument-specific CCS databases, systematic comparisons revealed excellent interlaboratory performance for 95% of the ions with CCS biases within ±1% for TIM-MS and within ±2% for TWIM-MS with respect to DTIM-MS values. However, a small fraction of ions (<1.5%) showed larger biases of up to 7% indicating that differences in the ion conformation sampled on different instrument types need to be further investigated. Systematic differences between CCS derived from different IM-MS analyzers and implications on the applicability for nontargeted analysis are critically discussed. To the best of our knowledge, this is the most comprehensive interlaboratory study comparing CCS from three different IM-MS technologies for analysis of steroids and small molecules in general.
Steroids are a class of cholesterol
derivatives and are characterized by a large diversity of isomers
that are of interest as even small structural changes can have a huge
impact on their biological activity.[1−3] Due to their role in
regulating metabolism, growth, reproductive function, and immune response,
steroids are analyzed across a wide range of research fields.[1,2] Furthermore, the use of exogenous anabolic androgenic steroids (AASs)
to enhance growth in livestock production was banned by the European
Union (EU),[4] and various natural and synthetic
steroids are prohibited doping agents in competitive sports.[5−7] Within this context, steroids are analyzed in diverse biological
matrices including serum, brain tissue, or urine.[1,6,8] Relatively lipophilic steroids are metabolized
via a two-phase reaction, and typically phase II metabolites (i.e.,
sulfates and glucuronides) are excreted due to their better solubility.[3] Steroid phase II metabolites are of major interest
when analyzing urine (e.g., for drug testing, in chemical food safety
or doping control) due to a high degree of metabolization and secretion.[5,6] The combination of complex matrices, low concentrations, and large
number of possible steroid isomers[8] demands
high-performance methods in terms of selectivity and sensitivity.[6] Therefore, mass spectrometry (MS) coupled to
front-end chromatographic separation such as liquid chromatography
(LC) has become one of the key technologies for the analysis of both
steroids and their phase II metabolites.[2]More recently, ion mobility-mass spectrometry (IM-MS) has
also
gained significant attention for potential application to separate
and identify isomeric and isobaric steroids.[5,9,10] As a standalone technology, IM-MS has been
applied for rapid steroid analysis, but its application is limited
for the most demanding applications due to the high complexity of
biological matrices and by the limited resolution of current IM technology.[1,11] However, in combination with LC (i.e., LC-IM-MS), the benefits of
both analytical platforms can be exploited.[5] For example, signal-to-noise ratios were improved when IM-filtering
was used for mass spectra cleanup.[6] In
addition to increased peak capacity and cleaner fragment spectra,[12] IM-derived collision cross sections (CCSs) have been extensively discussed as a potential ion
species-specific descriptor for identity confirmation,[6,13,14] and some notable studies focused
on the potential analysis of steroids using (LC−)IM-MS have
been published recently.[5,6,9,10,14−16] Application of CCS as additional
identification parameter is of special interest for small molecules
that exhibit limited fragmentation or formation of only unspecific
fragments.[17,18] This is indeed the case for many
steroids, which leads to ambiguous assignment of isomer identities.[5,6] Since transitioning into the commercial arena in 2006, several different
IM-MS instrument types are now commercially available including high-field
asymmetric IM (FAIM-MS), drift tube (DTIM-MS), traveling wave (TWIM-MS),
and trapped IM (TIM-MS) technologies.[19−21] While FAIMS is typically
used as a selective IM filter (e.g., in drug screening[22]), time-dispersive techniques such as DTIM-MS
and TWIM-MS or confinement-and-release technology (i.e., TIM-MS) can
be used as generic IM separators for nontargeted full-scan analysis.[13] When comparing data derived from these instrument
types, the different principles of IM devices are important to highlight.
In DTIM, ions are accelerated in a low, uniform electric field and
separated in the drift tube containing a neutral buffer gas.[13,23] Ion separation is analogous in TWIM, but dynamic electric fields
are applied to drag the ions through the buffer gas.[20,24] Finally, TIM involves the spatial trapping of ions by opposing forces
of a moving buffer gas flow and an electric field gradient. In this
case, ions are released by incrementally lowering the applied potential
barrier and exit the TIM separator in inverse order compared to DTIM
or TWIM.[21,25] While the principles differ, these three
instrument types all offer a fast and partly orthogonal IM-separation
that can improve analytical performance of LC-MS methods,[9,26,27] allow the application of novel
acquisition modes for data independent acquisition[12] and data dependent acquisition,[27] and can be externally calibrated for determination of CCS values for all detected ions.[23,28,29]With a focus on database-driven identity confirmation of small
molecules, several experimental CCS data sets and
databases for small molecules have been published in recent years.
Besides extensive multiclass databases such as the CCS Compendium,[30] several data sets for certain
compound classes are publicly available and include metabolites,[31] lipids[32] and steroids.[5,9,10,14] In addition to experimental CCS databases, the
number of computationally (in silico) predicted CCS libraries, either based on structural predictions based
on density functional theory calculations[33−35] or machine
learning is also increasing.[8,36,37] However, a comprehensive evaluation of the comparability of CCS derived from different classes of IM-MS instruments
is still lacking and assessment of possible differences is mandatory
before such databases can be applied across different classes of IM-MS
instruments.[19,38] This evaluation is critically
important when considering the key question of “what
does a CCS actually represent?”—a question
that continues to be addressed in fundamental research,[39−41] despite its increasing use across diverse analytical applications.
From a fundamental perspective, CCS differs from m/z information as it is a conditional
value derived from an ion’s mobility (K) and
depends on properties of the ion, such as size or charge state, as
well as the buffer gas, temperature, and the field strength-to-pressure
ratio.[19] However, it is well-established
that ionization, ion transfer, and ion separation can also influence
the observed ion structure and hence the ion’s mobility, e.g.,
via formation of protomeric isomers,[19,42] other types
of open/closed conformers,[43] or metastable
solvent clusters.[44] In combination, this
influences the comparability of CCS values and the
question to which extent CCS reference values can
be established and used independently of the instrument type is still
under discussion.[45] Finally, we note that
the external calibration strategies for CCS determination
employed affect the comparability of CCS across laboratories.[13,19] Of the current major instrument technologies, low-field DTIM-MS
presents the closest relationship to fundamental ion mobility theory
and is used to generate primary data (i.e., stepped-field method)
for the establishment of reference CCS values for secondary
calibration approaches. As a consequence, this direct link to primary CCS values is maintained between a well-characterized reference
instrument[23] and other DTIM-MS instruments
using the same set of reference values for single-field calibration.
Nevertheless, uncertainties on these reference CCS values have to be considered as these are propagated into secondary
calibration strategies for determination of CCS across
all three major instrument types.[23,46] For these
reasons, the true merit of applying CCS in steroid
analysis demands for a comprehensive study of all major commercial
instruments including their prescribed external calibration procedures.To address these open questions, we undertook a large-scale comparison
of three major commercial IM-QTOF instrument classes for application
to the analysis of steroids. For this purpose, new CCS and CCS reference data sets were established to complement existing
single laboratory and interlaboratory reference CCS values.[9,14] To the best of our knowledge, this is the only comprehensive study
comparing experimental CCS values for steroids using all
three commercially available technologies relevant to this application
(i.e., DTIM-MS, TWIM-MS, and TIM-MS).
Materials and Methods
Chemicals and Reagents
Stock solutions of steroid standards
(100 μg/mL or 1 mg/mL) from Steraloids (Newport, RI, USA), Sigma-Aldrich
(St. Louis, MO, USA), and National Measurement Institute (NMI, Pymble,
Australia) were stored in ethanol at −20 °C, and several
mixed solutions (10 μg/mL) were prepared for LC-IM-MS measurements
(see the Supporting Information). For LC-DTIM-MS
analysis, LC-MS-grade water from a Milli-Q IQ 7000 purification system
equipped with an LC-Pak polisher cartridge (Merck Chemicals and Life
Science GmbH, Vienna), and LC-MS-grade acetonitrile (CAN) and formic
acid (FA) from Sigma-Aldrich were used for dilutions of standards
and preparation of mobile phase. ESI Tune Mix (ESI-L, G1969-85000,
Agilent Technologies) along with 0.1 mmol/L HP-0321 from Agilent Biopolymer
Reference Kit was used to tune and calibrate the DTIM-MS instrument
according to the manufacturer instructions. For LC-TIM-MS measurements,
ultrapure water from an ELGA LabWater–water purification system
was used along with MS-grade acetonitrile (Biosolve, Netherlands)
and MS-grade formic acid (Honeywell Fluka).
Standards and Sample Preparation
A set of 87 steroids
based on a previous interlaboratory comparison of different TWIM-MS
systems was adapted for this study.[14] Pure
standard mixtures were prepared with concentrations of 1 or 5 μg/mL
for direct infusion and 1 μg/mL for LC-DTIM-MS analysis. Standard
mixtures of water-soluble steroids were prepared in 95:5 of 0.1% (v/v)
formic acid/acetonitrile, while hydrophobic steroids (e.g., sterol
esters) were prepared in 50:50 of 0.1% (v/v) formic acid/acetonitrile.
To obtain a well-defined solvent composition, spiked samples and standard
mixtures were dried under nitrogen and redissolved in an appropriate
final composition ready for analysis. The exact composition of the
steroid mixtures is presented in the Supporting Information. Bovine urine samples from adult animals and calves
used to study matrix effects were stored in the LABERCA biobank. Urine
samples were thawed, diluted 10-fold, and spiked with a mixture containing
69 steroids to a final concentration of 0.5 μg/mL prior to analysis.
IM-MS Instrumentation, Calibration, and Acquisition
Across the three commercial instrument technologies assessed in this
study, the ion transport mechanisms and routine analytical procedures
for external CCS calibration are known to be inherently
different.[13,19] Recommended acquisition settings
and application of routine external CCS calibration
were used. Details of the underlying theory for DTIM-MS, TWIM-MS and
TIM-MS are presented in the Supporting Information. TWIM-MS data sets were reported in two of our previous publications
and publicly available data was used for all comparisons.[9,14] Data from these studies are referred to as the single laboratory[9] and interlaboratory[18]CCS libraries, respectively. DTIM-MS measurements were performed
using an Agilent 6560 IM-QTOFMS instrument equipped with an Agilent
Dual Jet Stream ESI source using various acquisition methods and conditions.
Prior to measurements, the instrument was tuned and calibrated using
manufacturer recommendations (ESI-L, G1969-85000, Agilent Technologies).
Both stepped-field and single-field measurements were carried out
following the method of Stow et al.[23] Briefly, CCS determined using the stepped-field method as well as single-field
operation with either standard or 4-bit multiplexing settings. Long-term
repeatability of CCS was evaluated by reanalyzing steroids
9 months after the first measurements using 4m acquisition.For generation of a CCS data set, an Elute UHPLC was coupled to a timsTOF
Pro (Bruker Daltonics, Bremen Germany) using an ESI source and the
same LC conditions, and columns were used as described for DTIM-MS
analysis except for the LC flow rate (600 μL/min) and injection
volume (5 μL) applied. This instrument platform yielded an IM-resolving
power of R ∼
60–80 under the conditions applied in this study. The timsTOF
Pro was operated using the Bruker OTOFcontrol (6.2) software along
with HyStar (5.1) software. Prior to analysis, the instrument was
mass calibrated with sodium formate clusters (10 mM in 50:50 2-propanol/water)
and CCS was calibrated using ions from Agilent ESI-L Tune
Mix via a linear function. In addition to external calibration before
analysis, automatic postrun recalibration was used. Spiked urine samples
(see above) were analyzed using the same settings as used for standard
mixtures. Matrix effects of urine on determined CCS and CCS were determined by analyzing 69 steroids spiked into urine
samples with final concentrations of 0.5 mg/mL. Full method parameters
for all applied methods are found in the Supporting Information.
Data Processing and Evaluation
For DTIM-MS, Agilent
IM-MS Browser 10.0 was used for single-field calibration, evaluation
of stepped-field CCS, and manual inspection when required.
Agilent MassHunter Mass Profiler 10.0 was used for feature extraction
(peak picking) of triplicate LC-DTIM-MS measurements following single-field
calibration and direct infusion. PNNL Preprocessor 3.0 (2021.04.21)
was used for demultiplexing and data preprocessing steps.[47]For TIM-MS, Bruker TASQ software (version
2021) was used to analyze all data acquired using the timsTOF Pro.
This included automatic recalibration, generation of high-resolution
extracted ion chromatograms (EICs), feature detection (peak picking),
and CCS calculation. Full details of used data processing
can be found in the Supporting Information. Exported data was restructured, analyzed, and visualized using
Microsoft Office (Excel and Powerpoint) and R (4.1.2)[48] together with RStudio (2021.9.1.372)[49] (see the Supporting Information).
Results and Discussion
For a comprehensive evaluation
of reproducibility for CCS determination using different instrument
classes and methods, reference CCS values were generated.
For this purpose, stepped-field and secondary single-field methods
with either standard operation mode or 4-bit multiplexing were applied.
In the used DTIM-QTOFMS instrument, ions are accumulated in a trapping
funnel followed by release and DTIM-separation once per measurement
cycle (standard operation). The multiplexed operation increases ion
utilization efficiency, increases working range and reduces instrument
saturation including the minimization of space-charge effects in comparison
to standard operation.[50] In total, 135
single-field CCS values and 102 stepped-field CCS values were determined with the DTIM-MS platform and were used as
the basis of interplatform CCS comparisons (see the Supporting Information). Precision under conditions
of repeatability for measurements of CCS and CCS were both excellent (i.e., average RSD < 0.2%), and effects of
bovine urine matrix were negligible. In addition, determined CCS values were in good agreement with recently published data
sets by Velosa et al.[11] and Davis et al.[5] with average differences <0.5% for a small
number of steroids determined (see Table S2). Similar figures of merit for CCS determination of steroids
have been previously reported.[6,14]
Comparisons of CCS, CCS, and CCS Data Sets
In addition to single IM-MS technology interlaboratory studies,
some comparisons across two instrument technology classes for small
molecules were performed previously. For example, differences between CCS and CCS of 0.53% to 2.1% have been determined for
plant metabolite data sets,[29] and mean
percentage errors of CCS compared to CCS were 1.0%
and 1.1% for [M+H]+ and [M+Na]+ ions, but deviations
of up to 6.2% were reported for some ions in the first study that
compared the commercial Agilent DTIM-MS with Waters TWIM-MS for small
molecule applications.[38] Based on the new
data sets and the existing CCS database,[14] a new comprehensive interplatform and interlaboratory CCS database for steroids and phase II
metabolites containing 142 ions was created (see the Supporting Information). The correlation between individual
data sets from the three different instruments was investigated, and
the appearance of outliers and systematic differences observed for
[M–H]−, [M+H]+, and [M+Na]+ species of steroids was further studied. To compare all data
sets, Pearson correlation coefficients (r) were determined.
Coefficients of r = 0.9949 and r = 0.9953 were obtained when comparing CCS and CCS with CCS as the reference and r = 0.9989
when directly comparing CCS with CCS (Figure S1) indicating the good agreement between CCS data sets from all three instrument types. Single-field CCS data from multiplexed operation was used as reference to calculate
bias against results from TIM-MS and TWIM-MS instruments. The histograms
of the absolute bias (in %) of CCS and CCS and bias
distribution visualized as violin plots for comparisons of CCS and CCS data sets with CCS (4m) values are shown
in Figure S2. These comparisons illustrate
at first glance the lower absolute bias observed between CCS and CCS (mean = 0.47% ± 0.70%, 95th percentile = 1.03%)
than that between CCS and CCS (mean = 0.82%
± 0.76%, 95th percentile = 1.92%). Two further observations can
be made from these comparisons: (1) a systematic negative bias irrespective
of the ion species was observed between CCS and CCS; and (2) a small positive bias was apparent for the CCS [M+Na]+ data, which is notably different to the corresponding
comparisons for protonated and deprotonated ions. Detailed assessment
of individual compounds revealed that structural effects for some
steroids may play an important role for comparability of measurements
on different IM-MS instruments. For example, comparisons of CCS with CCS and for CCS with CCS revealed
that the ion with largest bias was [M–H]− of estradiol diglucuronide (EDG) which was the only analyzed diglucuronide
with similar bias of 6.8% and 6.6%, respectively. In addition, [M+Na]+ ion of boldenone undecylenate (BU) had a bias of 4.2% between CCS and CCS and a bias of 2.3% between CCS and CCS data. Furthermore, results for [M+H]+ of androstanolone and androstanedione with a bias of −1.9%,
−1.8% between CCS and CCS and a bias of
−2.7%, −2.5% between CCS and CCS indicate
similar differences between DTIM-MS and the two other instruments.
In a complementary comparison of CCS and CCS data (CCS used as reference) the average absolute bias was found to
be 0.66% ± 0.39% with a maximum of 1.81% and 95th percentile
of 1.34%. Overall, fewer ions with extremely large biases were observed
in this comparison than in the case of using CCS values as
the reference.To assess trends in the data sets with respect
to the transport mechanism of IM, bias data was also plotted using
the modified CCS′ as an independent variable
(CCS′ = CCS with μ being the reduced mass of
the ion-gas pair and z the charge number). This comparison
reflects the separation order and reveals a moderate positive correlation
(Pearson correlation r = 0.535) for bias of individual
ions’ CCS against CCS with respect to CCS′, while a weak positive correlation (r = 0.296) was observed in the corresponding comparison
of CCS and CCS (Figure ). Assessment of the correlation of bias
between CCS and CCS with CCS′ revealed
broadly similar results (r = 0.550) as when CCS was used as the reference.
Figure 1
Bias (%, ref = CCS) of (a) CCS and (b) CCS as a function of CCS′ for ion species [M+H]+, [M+Na]+, and [M–H]−.
Panel (c) shows bias (%, ref = CCS) between CCS and CCS as a function of CCS′ for
ion species [M+H]+, [M+Na]+, and [M–H]−. Pearson correlations (r) were calculated
for the entire data sets.
Bias (%, ref = CCS) of (a) CCS and (b) CCS as a function of CCS′ for ion species [M+H]+, [M+Na]+, and [M–H]−.
Panel (c) shows bias (%, ref = CCS) between CCS and CCS as a function of CCS′ for
ion species [M+H]+, [M+Na]+, and [M–H]−. Pearson correlations (r) were calculated
for the entire data sets.Furthermore, linear regressions were used to investigate
systematic
bias and possible calibration-related contribution to these observations.
A set of linear regressions comparing CCS, CCS, and CCS was built for this purpose, and residuals were
analyzed and ions with residuals outside of upper or lower whiskers
were excluded from the linear models (Figure ). The goodness of fit was excellent for
all linear models with coefficients of determination (R2) ≥ 0.9956, but systematic differences were observed
for the comparisons of CCS with CCS and for CCS with CCS. An intercept magnitude of <1 Å2 and a slope of 1.0008 in the linear model comparing CCS and CCS were obtained, whereas intercepts of >5 Å2 in combination with steeper slopes (>1.02) remained for
the
linear models comparing CCS with CCS and CCS. It is noteworthy that CCS and CCS are routinely
calibrated with the same commercially available compound mixture (i.e.,
reference ions and reference values) established by Stow et al.,[23] while CCS systems were calibrated
using a different commercial calibrant mix. Our findings further evidence
that an additional systematic difference is brought in by the external
calibration strategy for TWIM-MS, here in the case of small molecules.
Figure 2
Linear
regression models and residuals for comparisons of (a) CCS with CCS, (b) CCS with CCS, and (c) CCS with CCS. Boxplots were used to determined outliers
based on residuals and data points not included in the linear models
are indicated as red crosses in the corresponding scatterplot. Residuals
of outliers are not included in the corresponding scatterplots and
histograms. Colors for ion species shown correspond to those in Figure .
Linear
regression models and residuals for comparisons of (a) CCS with CCS, (b) CCS with CCS, and (c) CCS with CCS. Boxplots were used to determined outliers
based on residuals and data points not included in the linear models
are indicated as red crosses in the corresponding scatterplot. Residuals
of outliers are not included in the corresponding scatterplots and
histograms. Colors for ion species shown correspond to those in Figure .Residuals remaining within the whiskers of boxplots
were all below
±3 Å2 and the majority within ±2.0 Å2 of fitted values. Moreover, the distribution of residuals
for [M+Na]+ ions was observed to be positively shifted
for the comparisons of both CCS with CCS and CCS with CCS (blue histograms, Figure a,b), which is in agreement with the positive
bias for those ions reported in Figure S2. Furthermore, even after removal of outliers, the spread of values
for CCS vs CCS remained large in comparison to CCS vs CCS or CCS vs CCS.
Feasibility of Applying Single or Interplatform CCS Databases
for Identity Confirmation of Steroids
The value of CCS from instrument-specific, crowd-sourced, or in silico databases remains a topic of great interest for
a wide range of analytical applications where standards-free identity
confirmation is demanded.[14,30,36] Within our study, the generation of a large data set encompassing CCS values from the three major IM-MS instrument types enabled
a critical view on the feasibility of either instrument-specific or
interplatform CCS databases for the first time. Although
several studies have demonstrated the excellent repeatability and
minimal matrix effect for CCS in discriminating different steroid isomers,[5,10] uncertainty estimates for results and reference values must be considered
when CCS is employed
as an identification parameter on a routine basis.[46] However, reporting of accepted tolerance levels is usually
pragmatic and oriented around observed interlaboratory reproducibility
leading to precision estimates in the region of ±1% for DTIM-MS,[12] while ±2% is often considered for TWIM-MS
and TIM-MS applications.[14,29,45] To assess the merits of these limits in a clear way, average CCS for each instrument type
along tolerances of ±1% (boxes) and ±2% (whiskers) for a
series of isomeric ions from the new data sets are plotted in Figure . In addition to
these thresholds, uncertainty estimates (U, coverage
factor k = 1) for CCS are plotted to illustrate
the challenges using CCS for assigning the correct identity from several possible isomers
within a given database. For the selected subset of isomers, all average CCS fall into the uncertainty
estimates for single-field CCS and similar trends regarding isomer
differentiation were observed on all instruments. Except for [M+H]+ ions of testosterone glucuronide, epitestosterone glucuronide
and [M+Na]+ ion of epitestosterone glucuronide, unambiguous
differentiation of these isomers across platforms is already impossible
when a ± 1% tolerance is accepted without additional use of other
identification criteria such as retention time information. Furthermore,
the systematic bias observed in the CCS data would influence
false positive and false negative candidates if databases of other
instrument classes are used as reference. While only representing
a small subset of a particular class of small molecules, this result
illustrates the necessity of further collaborative efforts to investigate
the merits of consolidating external calibration ions, reference CCS values, and strategies for CCS determination
as well as the establishment and implementation of reference materials
with a view toward applying CCS as reliable parameter
within standards-free identity confirmation workflows. In considering
the physical interpretation of CCS data, it can be
stated that the ability to unambiguously diagnose outliers as being
representative of true differences in ion conformation is currently
hampered. Within the present study, several outliers for CCS and CCS were suspected due to the large differences between
−2.7% and +6.8% to the reference DTIM-MS values. However, it
is noteworthy that only sodiated boldenone undecylenate (BU) and deprotonated
estradiol diglucuronide (EDG) are outside the uncertainty estimates
(k = 1) for CCS. While representing only
a small fraction (<1.5%) of the present data set, these large differences
between CCS, CCS, and CCS are a clear issue
for the broad applicability of CCS as an identification
parameter across different IM-MS platforms. Figure a–c shows IM data for [M+H]+ and [M+Na]+ ions of a sterol ester (BU). The DTIM data
for [M+H]+ of BU (4-bit multiplexing) revealed two partly
separated peaks with CCS of ∼213 and ∼222 Å2 as well as an additional shoulder indicating multiple gas
phase conformations that are partly separated by DTIM (Figure b, solid blue line). Although
no separation was observed in the TWIM-MS data at least in part due
to its lower IM resolution, a broad arrival time distribution was
observed hinting toward the presence of additional unresolved conformers
(Figure a). Use of
TIM-MS allowed partial resolution of three peaks with CCS of 212.7, 222.9, and 227.5 Å2 (Figure c) while recently introduced
high-resolution demultiplexing (Hrdm)[51] for DTIM-MS data revealed a qualitatively similar result (Figure b, dashed line).
While only one major IM peak was observed for the BU [M+Na]+ ion using DTIM-MS or TWIM-MS, a complex IM spectrum was obtained
on the TIM-MS instrument including a dominant larger conformation
(CCS = 260.1 Å2) and a peak with a similar conformation
as on the DTIM-MS with CCS = 248.3 Å2. One explanation
is that the high degree of flexibility of the fatty acid chain of
BU supports the formation of multiple different stable gas-phase conformations
regardless of the used analyzer. However, further theoretical calculations
would be required to fully answer this question. These observations
limit the use of a simple platform-independent application of CCS for identification purposes for such compounds. In the
second example (Figure d–f), IM spectra of EDG [M–H]− are
investigated in more detail. For this compound, several additional
ion species were observed including [2M-2H]2– (m/z 311.1136 with CCS ∼
301 Å2), a doubly charged dimer (623.2345 m/z with CCS ∼ 388 Å2), and the [M+Na-2H]− ion. Interestingly,
the [M–H]− ion exhibited a more compact conformation
in DTIM-MS compared to TIM-MS and TWIM-MS, whereas the CCS for [M+Na-2H]− was
∼240 Å2 across all data sets. The large CCS difference of >6.5% indicates that a different conformation
of the [M–H]− ion is sampled in DTIM-MS compared
to the other instruments. However, this might be related to fundamental
differences of the ion transport mechanism, the ion structure or due
to the influence of ESI source conditions used on the different instrument
classes. To fully elucidate such results would require measurements
with the same ESI source and conditions applied on different IM-MS
platforms to be performed, which was not feasible within the context
of this study. Such examples finally underscore the value of carefully
curated and instrument class-specific experimental CCS databases targeting specific analytical applications although the
agreement of CCS values derived from different instrument
classes is very good for most of the investigated ions. Furthermore,
although high IM resolving power is clearly beneficial for overcoming
some of the current limitations of IM-MS analysis for isomeric small
molecules, the issues of external calibration and the influence of
measurement conditions affecting ion conformation must be considered
separately. Recently commercialized IM-technologies such as the Structures
for Lossless Ion Manipulations (SLIM)-MS,[52] cyclic ion mobility (cIM),[53] and TIM-MS[54] can be operated with Rp > 200, while HRdm for DTIM-MS allows a comparable increase
via data processing.[51] Thus, as the range
of commercial IM-MS technologies with expanded resolution capabilities
increases, external calibration will remain as a critical issue for
derivation and comparison of CCS values.
Figure 3
Comparison
of CCS, CCS, and CCS data for
isomer examples assessed in this study. Boxes and whiskers indicate
typically applied tolerance limits of ±1% and ±2%, respectively.
Uncertainty estimate boundaries (coverage factor k = 1)[46] calculated for CCS of respective
ions are represented by the shaded areas bounded by dashed lines.
Figure 4
IM spectra for [M+H]+ and [M+Na]+ ions of
boldenone undecylenate (BU) in (a) TWIM-MS, (b) DTIM-MS data, and
(c) TIM-MS and arrival time spectra for [M–H]− and [M+Na–2H]− ions of estradiol diglucurondide
(EDG) in (d) TWIM-MS, (e) DTIM-MS data, and (f) TIM-MS data. DTIM-MS
Hrdm spectra (R ∼
110) are represented by dashed black lines in panel (b). Intensity
is normalized to the intensity of the largest peak in the IM spectrum.
Comparison
of CCS, CCS, and CCS data for
isomer examples assessed in this study. Boxes and whiskers indicate
typically applied tolerance limits of ±1% and ±2%, respectively.
Uncertainty estimate boundaries (coverage factor k = 1)[46] calculated for CCS of respective
ions are represented by the shaded areas bounded by dashed lines.IM spectra for [M+H]+ and [M+Na]+ ions of
boldenone undecylenate (BU) in (a) TWIM-MS, (b) DTIM-MS data, and
(c) TIM-MS and arrival time spectra for [M–H]− and [M+Na–2H]− ions of estradiol diglucurondide
(EDG) in (d) TWIM-MS, (e) DTIM-MS data, and (f) TIM-MS data. DTIM-MS
Hrdm spectra (R ∼
110) are represented by dashed black lines in panel (b). Intensity
is normalized to the intensity of the largest peak in the IM spectrum.
Conclusion
This study systematically investigated the
comparability of CCS, CCS and CCS for the
analysis of steroids and phase
II metabolites for the first time. Most of the investigated ions fell
within ±1% of reference CCS for TIM-MS and within
±2% for TWIM-MS. Additionally, 95% of the CCS values were
found within ±1.34% of reported CCS values despite
a systematic negative and CCS′-dependent bias
in CCS data compared to DTIM-MS and TIM-MS data.Our findings also revealed the presence a calibration-dependent
bias for TWIM-MS that is not apparent between DTIM-MS and TIM-MS data
sets which are routinely calibrated with the same set of tune ions.
While overall agreement was found to be good across all three platforms,
these observed systematic differences for TWIM-MS hamper the applicability
of CCS databases across different types of IM-MS
technologies and increase the risk of false positive and false negative
identifications.Furthermore, by considering uncertainty estimates
associated with CCS reference values, a new approach toward
the unambiguous identification
of outliers was presented. Only a small number of CCS and CCS values (i.e., <1.5% of the ions) were found to be substantially
different from CCS values and their uncertainty estimates.
However, whether these experimental differences are due to the different
IM-separation mechanisms or originating from the influence of different
ESI parameters cannot be ascertained on a uniform basis.From
a broad analytical perspective, while the unavailability of
true CCS values remains as the major hindrance in
evaluating the use of CCS for identification workflows,
harmonization of calibrant ions and their reference CCS employed
for external instrument calibration demands for further investigation.
Good analytical practices including validation of external calibration
and better modeling of measurement uncertainty remain at the heart
of IM-MS research if this technology is to make the transition from
research into routine analytical laboratories.To the best of
our knowledge, this work encompasses the first and
most comprehensive comparison of CCS values obtained
from three major classes of IM-MS instruments for the analysis of
steroids so far. While limited to steroids within the present study,
these approaches can be viewed as a model that can be applied to a
broader range of small molecule databases whereby any of the three
instrument classes might be utilized and CCS values
employed for supporting standards-free identity confirmation.
Authors: Nicholas R Oranzi; Robin H J Kemperman; Michael S Wei; Violeta I Petkovska; Scott W Granato; Benjamin Rochon; Julia Kaszycki; Aurelio La Rotta; Kevin Jeanne Dit Fouque; Francisco Fernandez-Lima; Richard A Yost Journal: Anal Chem Date: 2019-03-08 Impact factor: 6.986
Authors: Diana C Velosa; Marcus E Rivera; Shon P Neal; Stine S H Olsen; Aurora Burkus-Matesevac; Christopher D Chouinard Journal: J Am Soc Mass Spectrom Date: 2021-12-22 Impact factor: 3.109
Authors: Susanta Das; Kiyoto Aramis Tanemura; Laleh Dinpazhoh; Mithony Keng; Christina Schumm; Lydia Leahy; Carter K Asef; Markace Rainey; Arthur S Edison; Facundo M Fernández; Kenneth M Merz Journal: J Am Soc Mass Spectrom Date: 2022-04-04 Impact factor: 3.262
Authors: Don E Davis; Katrina L Leaptrot; David C Koomen; Jody C May; Gustavo de A Cavalcanti; Monica C Padilha; Henrique M G Pereira; John A McLean Journal: Anal Chem Date: 2021-07-28 Impact factor: 8.008
Authors: Valérie Gabelica; Alexandre A Shvartsburg; Carlos Afonso; Perdita Barran; Justin L P Benesch; Christian Bleiholder; Michael T Bowers; Aivett Bilbao; Matthew F Bush; J Larry Campbell; Iain D G Campuzano; Tim Causon; Brian H Clowers; Colin S Creaser; Edwin De Pauw; Johann Far; Francisco Fernandez-Lima; John C Fjeldsted; Kevin Giles; Michael Groessl; Christopher J Hogan; Stephan Hann; Hugh I Kim; Ruwan T Kurulugama; Jody C May; John A McLean; Kevin Pagel; Keith Richardson; Mark E Ridgeway; Frédéric Rosu; Frank Sobott; Konstantinos Thalassinos; Stephen J Valentine; Thomas Wyttenbach Journal: Mass Spectrom Rev Date: 2019-02-01 Impact factor: 10.946
Authors: Max L Feuerstein; Maykel Hernández-Mesa; Younes Valadbeigi; Bruno Le Bizec; Stephan Hann; Gaud Dervilly; Tim Causon Journal: Anal Bioanal Chem Date: 2022-08-12 Impact factor: 4.478