| Literature DB >> 36046668 |
Hiroki Tanikawa1, Masahiko Mukaino2, Shota Itoh3, Hikaru Kondoh3, Kenta Fujimura1, Toshio Teranishi1, Kei Ohtsuka1, Satoshi Hirano2, Hitoshi Kagaya2, Eiichi Saitoh2, Yohei Otaka2.
Abstract
Background: Despite recent developments in the methodology for measuring spasticity, the discriminative capacity of clinically diagnosed spasticity has not been well established. This study aimed to develop a simple device for measuring velocity-dependent spasticity with improved discriminative capacity based on an analysis of clinical maneuver and to examine its reliability and validity.Entities:
Keywords: ankle joint; modified ashworth scale; quantification; resistance; spasticity
Year: 2022 PMID: 36046668 PMCID: PMC9420860 DOI: 10.3389/fbioe.2022.911249
Source DB: PubMed Journal: Front Bioeng Biotechnol ISSN: 2296-4185
Patient demographics for Experiment 2.
| Case | Sex | Age (years) | Height (cm) | Weight (kg) | Time after onset (months) | Diagnosis | Passive ankle D/F (°) | MAS of ankle plantar flexors | Achilles tendon reflex |
|---|---|---|---|---|---|---|---|---|---|
| 1 | M | 61 | 167.0 | 62.0 | 10 | CH | 10 | 3 | 3+ |
| 2 | M | 52 | 170.0 | 81.0 | 60 | CI | 10 | 2 | 2+ |
| 3 | M | 73 | 172.0 | 60.0 | 49 | CI | 25 | 2 | 3+ |
| 4 | M | 65 | 165.0 | 69.0 | 71 | CH | 20 | 2 | 2+ |
| 5 | M | 26 | 179.0 | 90.0 | 5 | CH | 15 | 1+ | 2+ |
| 6 | M | 47 | 172.0 | 73.0 | 40 | CH | 25 | 1+ | 2+ |
| 7 | M | 76 | 168.0 | 65.0 | 109 | CH | 20 | 1+ | 1+ |
| 8 | F | 75 | 157.0 | 57.0 | 4 | CI | 20 | 1 | 2+ |
| 9 | F | 61 | 152.0 | 40.0 | 85 | SAH | 15 | 1 | 2+ |
| 10 | F | 49 | 158.0 | 59.0 | 79 | CH | 15 | 1 | 1+ |
F female, M male, CH, cerebral hemorrhage; CI, cerebral infarction; SAH subarachnoid hemorrhage, D/F dorsiflexion, MAS, modified ashworth scale.
FIGURE 1The main body of the developed device and the standard position for measurement.
Angular velocity of the ankle joint (°/s) according to the raters’ manual assessment.
| Case | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 | Mean | |
| Rater A | 189 | − | 269 | 182 | 169 | − | − | − | − | − | − | − | − | − | 306 | 223 |
| B | 215 | − | − | − | − | 266 | − | − | 274 | − | − | − | 255 | 313 | 265 | |
| C | − | 162 | − | − | − | 229 | 409 | 296 | − | − | − | − | − | − | 408 | 301 |
| D | − | − | 365 | − | − | − | 332 | − | − | 414 | 352 | − | − | 361 | − | 365 |
| E | − | − | − | − | 304 | − | − | 202 | − | 243 | − | 205 | 150 | − | − | 221 |
| F | − | 231 | − | 182 | − | − | − | − | 208 | − | 244 | 301 | − | − | − | 233 |
| Mean | 202 | 196 | 317 | 182 | 236 | 248 | 370 | 249 | 241 | 328 | 298 | 253 | 202 | 337 | 357 | 268 |
FIGURE 2Peak resistance force values at angular velocities of 300°/s, 150°/s, 100°/s, and 5°/s are shown for the control and the patient subgroups, classified according to the Modified Ashworth Scale (MAS) score in Experiment 2.
FIGURE 3Peak resistance force values at angular velocities of 300°/s and 5°/s and their differences are shown for the control and the patient subgroups, classified according to the Modified Ashworth Scale (MAS) score, in Experiment 3.
FIGURE 4Receiver operating characteristic (ROC) curve analyses of peak resistance force values at angular velocities of 300°/s (blue, dashed line) and 5°/s (red, solid line) and their differences (orange, dotted line) for discriminating patients with spasticity from controls.
FIGURE 5Bland–Altman plots for the 1st and 2nd measurements. The solid black line represents the average difference between the two measurements. The solid gray lines represent the limits of agreement (LOA). The dotted lines represent 95% confidence intervals.