Literature DB >> 36032337

Validating Autoclave Cycles for Carcass Disposal in Animal Biosafety Level 2/3 Containment Laboratories.

Rebecca McGirr1, Christopher Sample2, Leslee Arwood3, James Burch3, Scott Alderman3.   

Abstract

Introduction: Animal carcasses differ in composition from other types of solid waste, and through prior testing it was determined that cycle parameters applied to general, solid biohazardous waste did not ensure proper sterilization of ferret carcasses.
Objectives: The goals of this study were to develop and validate an autoclave cycle that would ensure the decontamination of infectious animal carcasses before removal from an animal biosafety level 2/3 containment suite for downstream disposal and to test different ways to prepare and package animal carcasses for autoclaving.
Methods: Intact ferret carcasses were implanted with biological indicators, and the carcasses were placed in biohazard bags, then into metal pans. To test the efficacy of the autoclave cycle on larger biomasses, 1, 2, or 4 ferret carcasses were placed in a biohazard bag. A total of 4 carcasses were placed in each pan. An autoclave cycle was created to begin the study. After initial tests, minor modifications to the initial test cycle parameters were made, and a new cycle was validated for ferret carcasses up to 2 kg each. Parameters for the validated cycle were as follows: sterilization time 240 minutes, temperature 125°C, 5 prevacuum pulses, and chamber pressure 15 psi.
Results: The results of this study indicate that an extended sterilization time is required to successfully decontaminate animal carcasses compared with regular, solid, and biohazardous waste. Conclusions: This study demonstrates that it is possible to sterilize multiple intact ferret carcasses per load under validated autoclave cycle conditions. © ABSA International 2019.

Entities:  

Keywords:  animal biosafety; biocontainment; biosafety level 3; decontamination; sterilization

Year:  2019        PMID: 36032337      PMCID: PMC9134466          DOI: 10.1177/1535676019856799

Source DB:  PubMed          Journal:  Appl Biosaf        ISSN: 1535-6760


  2 in total

1.  Influence of packaging and processing conditions on the decontamination of laboratory biomedical waste by steam sterilization.

Authors:  G Ozanne; R Huot; C Montpetit
Journal:  Appl Environ Microbiol       Date:  1993-12       Impact factor: 4.792

2.  Measurement of Equilibration Time of Dry Goods Loads in Autoclaves.

Authors:  Soham Shah; Sara Alayli; Yiming Peng
Journal:  PDA J Pharm Sci Technol       Date:  2019-02-15
  2 in total