| Literature DB >> 36032318 |
John R Henneman1, Elizabeth A McQuade1, Rachael R Sullivan1, Jen Downard1, Ashley Thackrah1, Meaghan Hislop2.
Abstract
Introduction: The applications of fumigation and the challenges that high-containment facilities face in achieving effective large volume decontamination are well understood. The Biosecurity Research Institute at Kansas State University sought to evaluate a novel system within their biosafety level 3 (BSL-3) and animal biosafety level 3 agriculture (ABSL-3Ag) facility.Entities:
Keywords: decontamination; fogging; high level disinfectant; hydrogen peroxide vapor; sterilant
Year: 2022 PMID: 36032318 PMCID: PMC9402250 DOI: 10.1089/apb.2021.0012
Source DB: PubMed Journal: Appl Biosaf ISSN: 1535-6760
Figure 1.Diagram of Laboratory 1 (2,281 ft[3] [65 m3]). BIs of Geobacillus stearothermophilus (1.7 × 106) and H2O2 CIs were placed in pairs throughout each laboratory. Indicator pairs were placed in sets of three: low near the ground, high near the ceiling, and within the interstitial space of the drop ceiling as numbered. Numbers with an asterisk (*) indicate placement of a single indicator pair. Also shown are approximate locations of the removed ceiling tiles, fans, and HHP device. BI, biological indicator; CI, chemical indicator; H2O2, hydrogen peroxide; HHP, Hybrid Hydrogen Peroxide.
Cycle parameters
| Laboratory | Run | Volume (ft[ | Primary injection (min) | Pulse (min) | BI total | |
|---|---|---|---|---|---|---|
| Study 1 | 1 | 1 | 2,281 ft[ | 16.88 | 40 | 60 |
| 2 | 16.88 | 40 | ||||
| 2 | 1 | 4,668 ft[ | 23.5 | 40 | 192 | |
| 2 | 23.5 | 40 | ||||
| 3 | 23.5 | 40 | ||||
| Study 2 | Necropsy | 1 | 44,212 ft[ | 60.9 | 40 | 206 |
| 2 | 60.9 | 40 |
Parameters of the HHP system for each laboratory decontamination cycle.
BI, biological indicator; HHP, Hybrid Hydrogen Peroxide.
Study 1 biological indicator results
| Location no. | Heights | Location no. | Results | ||
|---|---|---|---|---|---|
| Low | High | Interstitial space | |||
| Laboratory 1 (2,281 [65 m3]) | |||||
| 1 | Pass | Pass | Pass | 9* | Pass |
| 2 | Pass | Pass | Pass | 10* | Pass |
| 3 | Pass | Pass | Pass | 11* | Pass |
| 4 | Pass | Pass | Pass | 12* | Pass |
| 5 | Pass | Pass | Pass | 13* | Pass |
| 6 | Pass | Pass | Pass | 14* | Pass |
| 7 | Pass | Pass | Pass | ||
| 8 | Pass | Pass | Pass | ||
| Laboratory 2 (4,668 ft[ | |||||
| 1 | Pass | Pass | Pass | 19* | Pass |
| 2 | Pass | Pass | Pass | 20* | Pass |
| 3 | Pass | Pass | Pass | 21* | Pass |
| 4 | Pass | Pass | Pass | 22* | Pass |
| 5 | Pass | Pass | Pass | 23* | Pass |
| 6 | Pass | Pass | Pass | 24* | Pass |
| 7 | Pass | Pass | Pass | 25* | Pass |
| 8 | Pass | Pass | Pass | 26* | Pass |
| 9 | Pass | Pass | Pass | 27* | Pass |
| 10 | Pass | Pass | Pass | 28* | Pass |
| 11 | Pass | Pass | Pass | ||
| 12 | Pass | Pass | Pass | ||
| 13 | Pass | Pass | Pass | ||
| 14 | Pass | Pass | Pass | ||
| 15 | Pass | Pass | Pass | ||
| 16 | Pass | Pass | Pass | ||
| 17 | Pass | Pass | Pass | ||
| 18 | Pass | Pass | Pass | ||
BIs of Geobacillus stearothermophilus (1.7 × 106) in locations 1–8 (laboratory 1) and 1–18 (laboratory 2) were placed in sets of three: low near the ground, high near the ceiling, and within the interstitial space of the drop ceiling. In locations 9*–14* (laboratory 1) and 19*–28* (laboratory 2), single indicators were placed on and in equipment throughout the laboratory.
Figure 2.Diagram of the Necropsy Suite (44,212 ft[3] [1,252 m3]). BIs of Geobacillus stearothermophilus (1.7 × 106) and H2O2 CIs were placed in pairs throughout the necropsy suite. Indicator pairs were placed at three different heights: low pairs at floor level, center pairs ∼10 ft (3 m) high, and ceiling pairs at 21 ft (6.4 m). Single indicator pairs (*) were placed on the digester and on the countertop. Also shown are approximate locations of the fans and HHP device.
Figure 3.Synchronized HHP devices at primary injection in the necropsy suite.
Necropsy suite biological indicator results
| Anteroom | Necropsy laboratory | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Location no. | Low | Center | Ceiling | Location no. | Low | 10 ft (3 m) above | 21 ft (6.4 m) ceiling | ||||||
| (BI) | (CI) | (BI) | (CI) | (BI) | (CI) | (BI) | (CI) | (BI) | (CI) | (BI) | (CI) | ||
| 1 | Pass | Pass | Pass | Pass | Pass | Pass | 10 | Pass | Pass | Pass | Pass | Pass | Pass |
| 2 | Pass | Pass | Pass | Pass | Pass | Pass | 11 | Pass | Pass | Pass | Pass | Pass | Pass |
| 3 | Pass | Pass | Pass | Pass | Pass | Pass | 12 | Pass | Pass | Pass | Pass | Pass | Pass |
| 4 | Pass | Pass | Pass | Pass | Pass | Pass | 13 | Pass | Pass | Pass | Pass | Pass | Pass |
| 5 | Pass | Pass | Pass | Pass | Pass | Pass | 14 | Pass | Pass | Pass | Pass | Pass | Pass |
| 6 | Pass | Pass | Pass | Pass | Pass | Pass | 15 | Pass | Pass | Pass | Pass | Pass | Pass |
| 7 | Pass | Pass | Pass | Pass | Pass | Pass | 16 | Pass | Pass | Pass | Pass | Pass | Pass |
| 8 | Pass | Pass | Pass | Pass | Pass | Pass | 17 | Pass | Pass | Pass | Pass | Pass | Pass |
| 9 | Pass | Pass | Pass | Pass | Pass | Pass | 18 | Pass | Pass | Pass | Pass | Pass | Pass |
| 30 | Pass | Pass | Pass | Pass | Pass | Pass | 19 | Pass | Pass | Pass | Pass | Pass | Pass |
| 31 | Pass | Pass | Pass | Pass | Pass | Pass | 20 | Pass | Pass | Pass | Pass | Pass | Pass |
|
| 21 | Pass | Pass | Pass | Pass | Pass | Pass | ||||||
|
|
|
|
| 22 | Pass | Pass | Pass | Pass | Pass | Pass | |||
|
|
|
|
|
|
| 23 | Pass | Pass | Pass | Pass | Pass | Pass | |
| 24 | Pass | Pass | Pass | Pass | Pass | Pass | 28 | Pass | Pass | Pass | Pass | Pass | Pass |
| 25 | Pass | Pass | Pass | Pass | Pass | Pass | 29 | Pass | Pass | Pass | Pass | Pass | Pass |
| 26 | Pass | Pass | Pass | Pass | Pass | Pass | 32 | Pass | Pass | Pass | Pass | Pass | Pass |
| 27 | Pass | Pass | Pass | Pass | Pass | Pass | 33 | Pass | Pass | Pass | Pass | Pass | Pass |
|
| 34 | Pass | Pass | Pass | Pass | Pass | Pass | ||||||
|
|
|
| |||||||||||
| 35 | Placed on digester | Pass | Pass | ||||||||||
| 36 | Placed on countertop | Pass | Pass | ||||||||||
BIs of Geobacillus stearothermophilus (1.7 × 106) and H2O2 CIs were placed in pairs throughout the necropsy suite. Indicator pairs were placed in sets of three: low near the floor, center at 10 ft (3 m), and ceiling at 21 ft (6.4 m). Pairs are shown beside one another according to the locations within the facility where they were placed.
CI, chemical indicator; H2O2, hydrogen peroxide.