Lars Müller1, Judith Kannenberg1, Ronald Biemann2, Mario Hönemann1, Grit Ackermann3, Christian Jassoy4.
Abstract
BACKGROUND: The concentration of antibodies against the SARS-CoV-2 spike protein is frequently being measured for clinical and epidemiological purposes. The aim of this study was to examine whether the results of different quantitative SARS-CoV-2 spike antibody assays are comparable.Entities:
Keywords: Coronavirus disease 2019 (COVID-19); Quantitative SARS-CoV-2 antibody assay; SARS-CoV-2 spike protein; Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2)
Mesh:
Substances:
Year: 2022 PMID: 36029637 PMCID: PMC9388276 DOI: 10.1016/j.jcv.2022.105269
Source DB: PubMed Journal: J Clin Virol ISSN: 1386-6532 Impact factor: 14.481
Minimal and maximal SARS-CoV-2 spike protein-specific antibody concentrations measured with four assays
| 50 | 7.1 | 1 | 21.8 | 12.0 | 12.0 | 11 | 35.2 | |
| 2877.2 | 408.6 | 20.3 | 441.5 | 1964.0 | 1964.0 | 145.8 | 466.5 | |
AU: Arbitrary unit.
BAU: Binding antibody unit.
U: Unit.
RE: Relative Unit.
Correlation of the antibody concentrations (BAU/ml) obtained with SARS-CoV-2 spike antibody assays.
| 76 | 0.953 (0.927–0.970) | |||||
| 106 | 0.829 (0.758–0.881) | 76 | 0.729 (0.603–0.820) | |||
| 63 | 0.826 (0.727–0.891) | 63 | 0.789 (0.673–0.867) | 63 | 0.788 (0.671–0.867) | |
Pearson coefficient of correlation; CI: 95% confidence interval
Fig. 1Comparison of the antibody values with assay pairs. Antibody values in BAU/ml from subjects 6-9 months after infection with SARS-CoV-2 were compared with the Wilcoxon signed rank test. The bars represent the geometric mean values. N: number of sera. Abbott: Abbott SARS-CoV-2 IgG II Quant assay; Siemens: Siemens Atellica IM SARS-CoV-2 IgG (sCOVG) assay; Roche: Roche ElecsysT Anti-SARS-CoV-2-S assay; Euroimmun: Euroimmun Anti-SARS-CoV-2-QuantiVac-ELISA (IgG) assay.
Fig. 2Antibody concentrations in pairs of SARS-CoV-2 antibody assays. Passing-Bablok regression lines and linear equations for pairs of antibody assays. N: number of sera. Dashed lines represent the regression lines; thin lines represent the 95% confidence intervals; dotted lines represent the 45° centerline. Abbott: Abbott SARS-CoV-2 IgG II Quant assay; Siemens: Siemens Atellica IM SARS-CoV-2 IgG (sCOVG) assay; Roche: Roche ElecsysT Anti-SARS-CoV-2-S assay; Euroimmun: Euroimmun Anti-SARS-CoV-2-QuantiVac-ELISA (IgG) assay.
Fig. 3Antibody concentrations of the WHO Anti-SARS-CoV-2 antibody standard 20/136. The concentration of the WHO antibody in BAU/ml obtained with the Abbott, Siemens and Roche assay; Abbott: Abbott SARS-CoV-2 IgG II Quant assay; Siemens: Siemens Atellica IM SARS-CoV-2 IgG (sCOVG) assay; Roche: Roche ElecsysT Anti-SARS-CoV-2-S assay; The Euroimmun assay was not tested.