| Literature DB >> 36015970 |
Maria Pinelli1, Emanuele Lettieri1, Andrea Boaretto2, Carlo Casile3, Giuseppe Citro4, Bernardino Zazzaro5, Adriana Ravazzoni5.
Abstract
Background: Self-monitoring of blood glucose (SMBG) is of paramount relevance for type 2 diabetes mellitus (T2DM) patients. However, past evidence shows that there are physical and cognitive issues that might limit the usage of glucometers by T2DM patients aged 65 years and over. Objective: Our aim was to investigate the physical and cognitive issues related to the usage of glucometers by T2DM patients aged 65 years and over. Materials andEntities:
Keywords: diabetes; glucometer; glucose monitoring; self-monitoring of blood glucose (SMBG); usability
Mesh:
Substances:
Year: 2022 PMID: 36015970 PMCID: PMC9416294 DOI: 10.3390/s22166202
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.847
Sample Characteristics.
| Sample Characteristics | |
|---|---|
| Gender | 40% women, 60% men |
| Age | 65–80 years old |
| Origin | Sicily, Basilicata and Piedmont. |
Figure 1Logical connections about quality of life and issues in the usage of glucometers for SMBG.
SMBG features solving T2DM 65+ patients’ issues.
| Issue | SMBG Features to Address It |
|---|---|
| Visual impairment |
Larger screen Appropriate font ratio Auditory and tactile feedback |
| Haptic deterioration |
Large, silvery/metallic, raised, squared and not too sensitive buttons Big enough glucometers Non-slippery and easy-to-clean material |
| Low speed of execution of activities | Longer screen-dimming |
| Memory loss | Guided operations |
| Difficulty in coordinating activities | Auditory and tactile feedback |
| Difficulty in interpreting results | Dynamic control indicators |
Patients’ answers regarding the features of Accu-Chek® Instant glucometer.
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|
| 1 | I believe the backlit display makes it easy to read the results on the screen | 38% | 48% | 17% | 0% | 0% | 86% | 4.1 |
| 2 | The wide dosing area will make it easy to apply the blood at the end of the strip | 34% | 45% | 17% | 3% | 0% | 79% | 4.1 |
| 3 | The strip ejector makes it easy to remove the strip hygienically | 48% | 28% | 21% | 3% | 0% | 76% | 4.2 |
| 4 | The colour scale (target range indicator) makes it easy to understand the results quickly | 17% | 38% | 10% | 3% | 31% | 55% | 3.1 |
| 5 | I believe it is easy to learn and use without HCP training | 45% | 34% | 17% | 3% | 0% | 79% | 4.2 |