Several investigators have approximated the scatter component of digital radiographic images by applying a constant weighting factor to a convolved version of the detected image. The scatter approximation is then subtracted from the detected image. When a constant weighting factor is used, the scatter in areas of low transmission is underestimated. We have extended this technique by allowing for a spatially variable weighting factor based on the local image intensity. This technique improves the scatter estimate and should provide better results for general videodensitometric applications.

Mesh:

Year:  1987        PMID: 3600520     DOI: 10.1118/1.596088

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  A model-based scatter artifacts correction for cone beam CT.

Authors:  Wei Zhao; Don Vernekohl; Jun Zhu; Luyao Wang; Lei Xing
Journal:  Med Phys       Date:  2016-04       Impact factor: 4.071

2.  A Deep Unsupervised Learning Model for Artifact Correction of Pelvis Cone-Beam CT.

Authors:  Guoya Dong; Chenglong Zhang; Xiaokun Liang; Lei Deng; Yulin Zhu; Xuanyu Zhu; Xuanru Zhou; Liming Song; Xiang Zhao; Yaoqin Xie
Journal:  Front Oncol       Date:  2021-07-16       Impact factor: 6.244

  2 in total