Literature DB >> 36001170

Comparing the accuracy of three-dimensional mini-optical portable navigation and accelerometer-based portable navigation system for acetabular cup placement during total hip arthroplasty.

Shinya Hayashi1, Yuichi Kuroda2, Naoki Nakano2, Tomoyuki Matsumoto2, Tomoyuki Kamenaga2, Toshihisa Maeda2, Ryosuke Kuroda2.   

Abstract

INTRODUCTION: This study compared the accuracy of three dimensional (3D) mini-optical navigation and accelerometer-based portable navigation systems for cup positioning during a total hip arthroplasty (THA) in the supine position.
MATERIALS AND METHODS: This retrospective cohort study assessed data for 77 hips using 3D mini-optical navigation (n = 37) and accelerometer-based portable navigation (n = 40). The patients underwent THA through the mini-anterolateral approach in the supine position using a portable navigation system. We assessed the preoperative target angles, recorded intraoperative cup angles, postoperative CT imaging angles, cup angle measurement errors, and other clinical parameters.
RESULTS: The mean absolute differences in radiographic inclination were similar between 3D mini-optical navigation and accelerometer-based portable navigation systems during THA in the supine position (2.8° ± 1.7° vs 2.8° ± 1.9°, p = 0.637). The mean absolute differences in radiographic anteversion were also similar (2.6° ± 2.3° vs 2.5° ± 1.9°, p = 0.737). Cup malalignment (absolute difference of inclination or anteversion between postoperative CT and preoperative target angle of > 5°) was significantly associated with body mass index (BMI) in accelerometer-based portable navigation but not in 3D mini-optical navigation.
CONCLUSIONS: This is the first study to compare the accuracy of cup positioning between 3D mini-optical and accelerometer-based navigations in THA in the supine position. Both portable navigation systems accurately identified the orientation of cup placement. The accuracy of 3D mini-optical navigation is not affected by high BMI and may be preferred over other options in such patients.
© 2022. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Entities:  

Keywords:  3D mini-optical navigation; Accelerometer-based navigation; Computer-assisted surgery; Total hip arthroplasty

Year:  2022        PMID: 36001170     DOI: 10.1007/s00402-022-04583-2

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   2.928


  1 in total

1.  Accuracy of Acetabular Component Positioning Using Computer-assisted Navigation in Direct Anterior Total Hip Arthroplasty.

Authors:  Michael P Bradley; Jessica R Benson; Jeffrey M Muir
Journal:  Cureus       Date:  2019-04-16