Performance evaluation of an LED flatbed scanner for triple channel film dosimetry with EBT3 and EBT-XD film.

Marchant Van der Walt1, Linda Marsh2, John Baines2, Stephen Gibson2, Ariadne Shoobridge2, Glenn de Vine2.   

Abstract

We investigate the properties of a light emitting diode (LED) flatbed scanner for use with EBT3 and EBT-XD film types in a clinical radiochromic film (RCF) dosimetry program with modern treatment techniques. The flatbed scanner was characterised in terms of lateral and longitudinal response, X-Y scaling integrity, scanning reproducibility, scanner warm up dependence and film orientation dependence. The preferred lateral response artefact (LRA) corrections are investigated for the LED light source. Supporting evidence is provided regarding the dose independent nature of the corrections while also providing results suggesting a potential film type independence. Results from 2D gamma analysis of four patient treatments were compared between the new 12000XL and existing 10000XL model. Lastly, a dose uncertainty analysis was performed for the film-scanner system combination. It may be concluded that the lateral response variation requires correction while the longitudinal response variation is insignificant. The linear scaling in the lateral and longitudinal directions are within 0.5% and the scanner reproducibility is stable. Scanner warm up dependence no longer exists, and effort should be made to maintain all film orientation in a study set within 15°. The LRA corrections are as reported substantially dose independent and there is evidence to support film type independence. Comparative gamma analysis of patient specific dose maps between the EPSON 10000XL (xenon fluorescent lamp) and 12000XL (LED) scanners showed that results are indistinguishable for both film types across the two scanner models when the necessary corrections are applied. Dose uncertainty is in agreement with the literature and can be kept below 3% with necessary corrections applied.
© 2022. Crown.

Entities:  

Keywords:  LED; Radiochromic film dosimetry; Radiotherapy; Triple channel corrections

Mesh:

Year:  2022        PMID: 35997925     DOI: 10.1007/s13246-022-01161-z

Source DB:  PubMed          Journal:  Phys Eng Sci Med        ISSN: 2662-4729


  21 in total

1.  Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry.

Authors:  David Lewis; Maria F Chan
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

2.  Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification.

Authors:  Antony L Palmer; Alexis Dimitriadis; Andrew Nisbet; Catharine H Clark
Journal:  Phys Med Biol       Date:  2015-10-29       Impact factor: 3.609

3.  Response characterization of EBT-XD radiochromic films in megavoltage photon and electron beams.

Authors:  E Yazmin León-Marroquín; Daniel Mulrow; Arash Darafsheh; Rao Khan
Journal:  Med Phys       Date:  2019-08-01       Impact factor: 4.071

4.  Dose-response linearization in radiochromic film dosimetry based on multichannel normalized pixel value with an integrated spectral correction for scanner response variations.

Authors:  Saad Aldelaijan; Slobodan Devic; Pavlos Papaconstadopoulos; Hamed Bekerat; Robert A Cormack; Jan Seuntjens; Ivan M Buzurovic
Journal:  Med Phys       Date:  2019-09-26       Impact factor: 4.071

5.  Technical Note: On GAFChromic EBT-XD film and the lateral response artifact.

Authors:  David F Lewis; Maria F Chan
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

6.  Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG-55.

Authors:  Azam Niroomand-Rad; Sou-Tung Chiu-Tsao; Michael P Grams; David F Lewis; Christopher G Soares; Leo J Van Battum; Indra J Das; Samuel Trichter; Michael W Kissick; Guerda Massillon-Jl; Paola E Alvarez; Maria F Chan
Journal:  Med Phys       Date:  2020-10-30       Impact factor: 4.506

7.  Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films.

Authors:  Ning Wen; Siming Lu; Jinkoo Kim; Yujiao Qin; Yimei Huang; Bo Zhao; Chang Liu; Indrin J Chetty
Journal:  Radiat Oncol       Date:  2016-10-04       Impact factor: 3.481

8.  Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy.

Authors:  Hideharu Miura; Shuichi Ozawa; Fumika Hosono; Naoki Sumida; Toshiya Okazue; Kiyoshi Yamada; Yasushi Nagata
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

9.  Methodology for radiochromic film analysis using FilmQA Pro and ImageJ.

Authors:  Michelle E Howard; Michael G Herman; Michael P Grams
Journal:  PLoS One       Date:  2020-05-21       Impact factor: 3.240

10.  Multichannel Film Dosimetry for Quality Assurance of Intensity Modulated Radiotherapy Treatment Plans Under 0.35 T Magnetic Field.

Authors:  Gorkem Gungor; Latif Korkmaz; Namik Kayalilar; Gokhan Aydin; Bulent Yapici; Teuta Zoto Mustafayev; Banu Atalar; Enis Ozyar
Journal:  Cureus       Date:  2020-03-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.