Abstract
PURPOSE: This study aimed to evaluate the image quality and diagnostic value of compressed sensing-sensitivity encoding (CS-SENSE) accelerated 3-dimensional (3D) T2-weighted turbo spin-echo (T2W TSE) sequence in patients with rectal cancer compared with conventional 3D and 2-dimensional (2D) sequences.Entities:
Keywords: Compressed sensing-sensitivity encoding; Magnetic resonance imaging; Rectal neoplasms; Three-dimensional
Mesh:
Year: 2022 PMID: 35997800 PMCID: PMC9560929 DOI: 10.1007/s00261-022-03636-9
Source DB: PubMed Journal: Abdom Radiol (NY)
Fig. 1Flow diagram of participant selection in the study. RC rectal cancer, MRI magnetic resonance imaging
Imaging protocols for 3 types of T2WI sequences in rectal cancer
| Parameters | 2D TSE | 3D TSE | CS-SENSE 3D TSE | ||
|---|---|---|---|---|---|
| Orientation | Sagittal | Oblique axial | Oblique coronal | Oblique axial | Oblique axial |
| Repetition time (ms) | 4445 | 4167 | 4794 | 1250 | 1250 |
| Echo time (ms) | 110 | 100 | 100 | 148 | 148 |
| TSE factor | 25 | 25 | 25 | 120 | 120 |
| Flip angle (degree) | 90 | 90 | 90 | 90 | 90 |
| Field of view (mm) | 240 × 240 × 79 | 260 × 260 × 96 | 260 × 260 | 300 × 422 × 180 | 300 × 422 × 180 |
| Matrix | 300 × 252 × 24 | 324 × 299 × 48 | 324 × 307 × 32 | 376 × 519 × 450 | 376 × 519 × 450 |
| Slice thickness/gap (mm) | 3/0.3 | 3/0.8 | 3/0 | 0.8/–0.4 | 0.8/–0.4 |
| Voxel dimensions (mm) | 0.8 × 0.8 × 3.0 | 0.8 × 0.8 × 3.0 | 0.8 × 0.8 × 3.0 | 0.8 × 0.81 × 0.8 | 0.8 × 0.81 × 0.8 |
| Voxel volume (mm3) | 1.92 | 1.92 | 1.92 | 0.5184 | 0.5184 |
| Number of slices | 24 | 48 | 32 | 450 | 450 |
| Phase sampling (%) | 84 | 92.4 | 94.7 | 98.1 | 98.2 |
| SENSE factor | 2 | 2 | 2 | 2 | NA |
| CS-SENSE factor | NA | NA | NA | NA | 10 |
| In-plane frequency encoding direction | Right-to-left | Right-to-left | Right-to-left | Right-to-left | Right-to-left |
| Acquisition time | 2 min 4 s | 4 min 43 s | 1 min 46 s | 10 min 43 s | 6 min 33 s |
CS-SENSE compressed sensing-sensitivity encoding, NA not applicable, 2D two-dimensional, 3D three-dimensional, TSE turbo spin echo, T2WI T2-weighted imaging
Qualitative scores and interreader agreement of three types of TSE T2WI sequences in rectal cancer
| Parameters | 2D images | 3D images | CS-SENSE 3D images | Friedman test | Multiple comparison analysis | Interreader agreement (kappa) | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| 2D images | 3D images | CS-SENSE 3D images | ||||||||
| Contrast resolution | 4.60 ± 0.58 | 4.74 ± 0.49 | 4.64 ± 0.65 | 0.193 | NA | NA | NA | 0.852 | 0.872 | 0.803 |
| Edge sharpness | 4.38 ± 0.62 | 4.83 ± 0.37 | 4.67 ± 0.64 | < 0.001 | 0.023* | 0.243 | 1.000 | 0.928 | 0.914 | 0.873 |
| Noise | 4.95 ± 0.21 | 4.21 ± 0.46 | 4.02 ± 0.51 | < 0.001 | < 0.001* | < 0.001* | 0.756 | 0.653 | 0.758 | 0.656 |
| Peristaltic artifact | 4.83 ± 0.43 | 4.73 ± 0.49 | 4.62 ± 0.58 | 0.005 | 1.000 | 0.468 | 1.000 | 1.000 | 1.000 | 0.890 |
| Rectal wall layers | 4.48 ± 0.70 | 4.48 ± 0.73 | 4.48 ± 0.73 | 1.000 | NA | NA | NA | 0.876 | 0.833 | 0.848 |
| Tumor significance | 4.71 ± 0.66 | 4.71 ± 0.48 | 4.74 ± 0.54 | 0.976 | NA | NA | NA | 0.919 | 1.000 | 1.000 |
| Overall image quality | 4.76 ± 0.53 | 4.64 ± 0.48 | 4.52 ± 0.55 | 0.022 | 0.900 | 0.243 | 1.000 | 0.933 | 0.941 | 0.889 |
Values are mean ± standard deviation. Significance values were adjusted by the Bonferroni correction for multiple tests. P < 0.05 indicated a statistically significant difference. The Pa value is for the comparison between 2D images and 3D images, Pb value is for the comparison between 2D images and CS-SENSE 3D images, and Pc value is for the comparison between 3D images and CS-SENSE 3D images
CS-SENSE compressed sensing-sensitivity encoding, NA not applicable, 2D two-dimensional, 3D three-dimensional, TSE turbo spin echo, T2WI T2-weighted imaging
*Statistically significant difference
Fig. 2A 68-year-old male patient with rectal cancer. The scores of overall image quality on a 2D image, b 3D image and c CS-SENSE 3D image were 5, 4 and 4, respectively. The edge sharpness on CS-SENSE 3D image was similar to that on 2D image and 3D image. The noise on CS-SENSE 3D image and 3D image was similar, but they were more evident than the noise on 2D image. CS-SENSE compressed sensing-sensitivity encoding, 2D two-dimensional, 3D three-dimensional
Quantitative evaluation and interreader agreement of three types of T2WI sequences in rectal cancer
| Parameters | 2D images | 3D images | CS-SENSE 3D images | Friedman test | Multiple comparison analysis | ICC (95% CI) | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| 2D images | 3D images | CS-SENSE 3D images | ||||||||
| SNRtumor | 16.62 ± 3.19 | 7.07 ± 1.22 | 5.68 ± 1.00 | < 0.001 | < 0.001* | < 0.001* | < 0.001* | 0.948 (0.905–0.972) | 0.971 (0.946–0.977) | 0.957 (0.921–0.977) |
| SNRrectal wall | 18.49 ± 3.32 | 8.56 ± 1.63 | 7.05 ± 1.61 | < 0.001 | < 0.001* | < 0.001* | 0.004* | 0.964 (0.934–0.981) | 0.876 (0.778–0.932) | 0.967 (0.938–0.982) |
| CNR | 2.53 ± 2.01 | 1.49 ± 1.11 | 1.39 ± 1.03 | 0.016 | 0.033* | 0.045* | 1.000 | 0.802 (0.659–0.890) | 0.808 (0.663–0.893) | 0.917 (0.826–0.958) |
Values are mean ± standard deviation. Significance values were adjusted by the Bonferroni correction for multiple tests. P < 0.05 indicated a statistically significant difference. The Pa value is for the comparison between 2D images and 3D images, Pb value is for the comparison between 2D images and CS-SENSE 3D images, and Pc value is for the comparison between 3D images and CS-SENSE 3D images
95% CI 95% confidence interval, CNR contrast-to-noise ratio, CS-SENSE compressed sensing-sensitivity encoding, ICC intraclass correlation coefficient, SNR signal-to-noise ratio, 2D two-dimensional, 3D three-dimensional, T2WI T2-weighted imaging
*Statistically significant difference
Diagnostic capability and interreader assessment for three TSE T2WI sequences in rectal cancer
| Parameters | 2D images | 3D images | CS-SENSE 3D images | Friedman test | Multiple comparison analysis | Interreader agreement (kappa) | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 1 | Reader 2 | Reader 1 | Reader 2 | 2D images | 3D images | CS-SENSE 3D images | |||||
| mrT1-2a | 33 | 33 | 35 | 35 | 35 | 35 | 0.905 | NA | NA | NA | 0.921 | 1.000 | 0.957 |
| mrT3a | 2 | 0 | 0 | 0 | 0 | 0 | |||||||
| mrT3b | 4 | 6 | 4 | 4 | 4 | 4 | |||||||
| mrT3c | 2 | 2 | 2 | 2 | 2 | 2 | |||||||
| mrT3d | 0 | 0 | 0 | 0 | 0 | 0 | |||||||
| mrT4a | 1 | 1 | 1 | 1 | 1 | 1 | |||||||
| mrT4b | 0 | 0 | 0 | 0 | 0 | 0 | |||||||
| mrN0 | 34 | 34 | 34 | 34 | 34 | 34 | 1.000 | NA | NA | NA | 1.000 | 1.000 | 0.941 |
| mrN1a | 4 | 4 | 4 | 4 | 4 | 4 | |||||||
| mrN1b | 4 | 4 | 4 | 4 | 4 | 4 | |||||||
| mrN1c | 0 | 0 | 0 | 0 | 0 | 0 | |||||||
| mrN2a | 0 | 0 | 0 | 0 | 0 | 0 | |||||||
| mrN2b | 0 | 0 | 0 | 0 | 0 | 0 | |||||||
Unless otherwise indicated, data are numbers of patients, with percentages in parentheses. Significance values were adjusted by the Bonferroni correction for multiple tests. P < 0.05 indicated a statistically significant difference. The Pa value is for the comparison between 2D images and 3D images, Pb value is for the comparison between 2D images and CS-SENSE 3D images, and Pc value is for the comparison between 3D images and CS-SENSE 3D images
CS-SENSE compressed sensing-sensitivity encoding, NA not applicable, 2D two-dimensional, 3D three-dimensional, TSE turbo spin echo, T2WI T2-weighted imaging
aMRI has limited ability to distinguish T1 and T2 tumors
Diagnostic capability and interreader assessment for three TSE T2WI sequences in rectal cancer
| Parameters | 2D images | 3D images | CS-SENSE 3D images | McNemar test | Interreader agreement (kappa) | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 1 | Reader 2 | Reader 1 | Reader 2 | 2D images | 3D images | CS-SENSE 3D images | ||||
| mrEMVI (−) | 40 | 39 | 39 | 39 | 39 | 39 | 0.333 | 0.333 | 0.333 | 0.788 | 1.000 | 1.000 |
| mrEMVI (+) | 2 | 3 | 3 | 3 | 3 | 3 | ||||||
| mrMRF (−) | 39 | 39 | 39 | 39 | 39 | 39 | 0.333 | 0.333 | 0.333 | 1.000 | 1.000 | 1.000 |
| mrMRF (+) | 3 | 3 | 3 | 3 | 3 | 3 | ||||||
Unless otherwise indicated, data are numbers of patients, with percentages in parentheses. Significance values were adjusted by the Bonferroni correction for multiple tests. P < 0.05 indicated a statistically significant difference. The Pa value is for the comparison between 2D images and 3D images, Pb value is for the comparison between 2D images and CS-SENSE 3D images, and Pc value is for the comparison between 3D images and CS-SENSE 3D images
CS-SENSE compressed sensing-sensitivity encoding, NA not applicable, 2D two-dimensional, 3D three-dimensional, TSE turbo spin echo, T2WI T2-weighted imaging, EMVI extramural vascular infiltration, MRF mesorectal fascia
Fig. 3White arrows indicate the T staging of rectal cancers. a–c A T2 rectal cancer of a 61-year-old man on the 2D, 3D and CS-SENSE 3D images, respectively. d–f A T3 rectal cancer of a 50-year-old man on the 2D, 3D and CS-SENSE 3D images, respectively. g–i A T4 rectal cancer of a 48-year-old woman on the 2D, 3D and CS-SENSE 3D images, respectively. CS-SENSE compressed sensing-sensitivity encoding, 2D two-dimensional, 3D three-dimensional