Dominic J Wales1, Meysam Keshavarz1, Carmel Howe2, Eric Yeatman3. 1. Hamlyn Centre, Institute of Global Health Innovation, Imperial College London, London SW7 2AZ, U.K. 2. Department of Bioengineering, Imperial College London, London SW7 2AZ, United Kingdom. 3. Department of Electrical and Electronic Engineering, Imperial College London, London SW7 2AZ, U.K.
Abstract
Herein, we present the first example of 3D printing with poly(octamethylene maleate (anhydride) citrate) (POMaC), a bio-adhesive material which has shown particular promise for implantable biomedical devices. The current methods to fabricate such devices made from POMaC are hindered by the imposed constraints of designing complex molds. We demonstrate the feasibility of exploiting additive manufacturing to 3D print structural functional materials consisting of POMaC. We present 3D printing of biomaterial copolymers consisting of mixtures of poly(ethylene glycol) diacrylate (PEGDA) and POMaC at different ratios. The required parameters were optimized, and characterization of the printing fidelity and physical properties was performed. We have also demonstrated that a range of mechanical properties can be achieved by tuning the POMaC/PEGDA ratio. The biocompatibility of the copolymers was ascertained via a cell viability assay. Such tunable 3D printed biomaterials consisting of POMaC and PEGDA will have significant potential application in the development of functional biomaterial tissue scaffolds and biomedical devices for the future of personalized medicine.
Herein, we present the first example of 3D printing with poly(octamethylene maleate (anhydride) citrate) (POMaC), a bio-adhesive material which has shown particular promise for implantable biomedical devices. The current methods to fabricate such devices made from POMaC are hindered by the imposed constraints of designing complex molds. We demonstrate the feasibility of exploiting additive manufacturing to 3D print structural functional materials consisting of POMaC. We present 3D printing of biomaterial copolymers consisting of mixtures of poly(ethylene glycol) diacrylate (PEGDA) and POMaC at different ratios. The required parameters were optimized, and characterization of the printing fidelity and physical properties was performed. We have also demonstrated that a range of mechanical properties can be achieved by tuning the POMaC/PEGDA ratio. The biocompatibility of the copolymers was ascertained via a cell viability assay. Such tunable 3D printed biomaterials consisting of POMaC and PEGDA will have significant potential application in the development of functional biomaterial tissue scaffolds and biomedical devices for the future of personalized medicine.
Citric acid-based elastomers (CABEs) are
a class of biodegradable
elastomers, the structures of which are based around the multifunctional
citrate monomer.[1] The citrate moiety can
participate in ester cross-link formation between the elastomer chains,
can form hydrogen bonding interactions, and can also facilitate bioconjugation
for tuning of the material properties.[1] Such versatility, combined with the non-toxic nature of citric acid,
has led to a significant growth in research attention focused on the
use of CABEs for biomedical applications. Since the early work on
CABEs by Yang et al. in 2004,[2,3] there have been many
examples of biomedical materials and devices fabricated from citrate-based
biomaterials.[3−18] In particular, a type of CABEs that exhibit exceptional tunable
multifunctionality are poly(alkylene maleate citrates). These polymers
consist of alkylene, maleate, (introducing a polymerizable alkene
group into the backbone of the polymer) and citrate moieties and can
be cross-linked to form thermoset elastomers, either by one or both
of the following two mechanisms: thermal cross-linking via ester formation
and/or photocross-linking via the alkene moieties. One poly(alkylene
maleate citrate) that has received much research attention is poly(octamethylene
maleate (anhydride) citrate) citrate (POMaC) (Figure ). Formed from a co-condensation reaction
of citric acid, 1,8-octanediol, and maleic anhydride, it has been
demonstrated that a high degree of control over the desired mechanical,
adhesive, and biodegradability properties of the resulting cross-linked
thermoset elastomer is possible, by tuning the amount of both thermal
and/or photocross-linking.[8] Undeniably,
this high degree of tunability means that POMaC is well suited for
use as a scaffold for minimally invasive delivery of functional tissues
and for the development of biodegradable sticky “AngioChip”
patches for direct surgical anastomosis, both of which have been demonstrated
in animal models.[6,17] Currently, thermoset POMaC elastomer-based
devices are fabricated using molding processes, however, this methodology
is more suited for scaled-up fabrication due to the high cost of the
design and subsequent fabrication of the mold and there are constraints
in achieving complex geometries.[3] On the
other hand, 3D printing offers rapid and facile fabrication of intricate
structures in smaller volumes, which is highly desirable when making
bespoke biomedical devices for personalized medical applications.[19] Therefore, 3D printing has been widely utilized
to print advanced functional materials[20−26] and biomedical implants or devices.[27−30] 3D bioprinting also allows for
the fabrication of structures containing living cells within biomaterial
matrices.[31−34] To this end, there is a need for 3D printing of POMaC. However,
to date there are no examples of 3D printed POMaC, and only a few
examples in the literature of 3D printing of the wider range of CABEs,
some of which use sacrificial co-printed mold materials.[35−41] Direct ink writing of CABEs can be achieved through the use of filler
materials, such as chitin nanocrystals, as this increases the viscosity
of the CABE ink to achieve the desired printing resolution; however,
filler material loadings of up to 40 weight percentage (wt %) are
required which can affect the preferred mechanical and physiochemical
properties.[40] Specifically, for POMaC,
the cross-linking time to achieve gelation is on the order of several
minutes, which makes the use of POMaC in 3D printing impractical,
without changing the chemical structure of the elastomer as reported
by Savoji et al.[39] If this cross-linking
time was decreased sufficiently, POMaC could be a desirable material
for 3D printing of biofunctional materials. As a solution, we have
investigated the use of poly(ethylene glycol) diacrylate (Mn = 700) (PEGDA700, Figure ) as a copolymer cross-linker for direct
ink writing of POMaC-based materials. The reasoning for using PEGDA
as a copolymer cross-linker is fourfold: i) it increases the concentration
of available double bonds to undergo radical polymerization, thus
decreasing the time needed to achieve sufficient cross-linking for
solidification in the timescale for extrusion 3D printing; (ii) the
polymerization time of PEGDA when photocross-linked in the presence
of ≥2 wt % of photoinitiator is on the order of seconds;[42] (iii) PEGDA has been demonstrated as a suitable
copolymer for fabrication of materials with tunable physical properties;[43−45] and (iv) the resulting PEG-based polymer formed from polymerization
of PEGDA is bioinert.[46,47]
Figure 1
Structures of the polymers POMaC and PEGDA700.
Structures of the polymers POMaC and PEGDA700.Herein, 3D printing of a copolymer biomaterial
with different ratios
of POMaC and PEGDA700 (POMaC/PEGDA) is explored. PEGDA700 was chosen
as it is a liquid at room temperature and pressure, unlike commercially
available PEGDA polymers with molecular weights ≥1000 Da which
are solids. Furthermore, compared to other PEGDA polymer that are
liquid at room temperature and pressure, PEGDA700 has been shown to
demonstrate the fastest polymerization rates.[48] The optimal parameters for (extrusion-based) 3D printing of the
copolymer biomaterial were determined, as well as analysis of the
printing fidelity and elucidation of the achievable range of mechanical
and physical properties through tuning the POMaC/PEGDA ratio. The
3D printed copolymer biomaterials were determined to be biocompatible
by a cell viability assay. To the best of our knowledge, this is the
first example of 3D printing of POMaC with tunable physical/mechanical
attributions, and this work is a systematic study into 3D printing
of POMaC-based copolymer biomaterials which allows for POMaC to be
used in future biomedical applications.
Experimental Section
General Experimental Procedure
Unless otherwise stated,
all reagents were purchased from Merck, TCI, Acros Organics, and Fisher
Scientific and used without additional purification. HiPerSolv CHROMANORM
HPLC grade water from VWR was used in all purification experimental
steps. Nitrogen gas refers to oxygen free nitrogen, 99.9% purity from
BOC (Guildford, UK). FT-IR spectra were collected on a Thermo Scientific
Nicolet iS50 FT-IR spectrometer fitted with a diamond ATR module,
4000–400 cm–1, 64 scans, and 0.5 cm–1 resolution. 1H NMR spectra were recorded on a JEOL 400YH
(400 MHz) spectrometer at 25 °C, using DMSO-d6 NMR solvent, 96 scans, and a concentration of POMaC
of approx. 5–10 mg.mL–1. MALDI mass spectrometry
was performed in positive ionization mode using a Waters Micromass
MALDI-ToF mass spectrometer and α-cyano-4-hydroxy-cinnamic acid
as the matrix. The 3D printer used in this work was an Allevi 3 (Allevi,
Philadelphia, USA) equipped with a UV LED (λ = 365 nm, 20 mW.cm–2 at 100% intensity) for photopolymerization. The photoinitiator
used was Irgacure 2959 (2-hydroxy-4′-(2-hydroxyethoxy)-2-methylpropiophenone),
the chemical structure of which is shown in Figure D. Irgacure 2959 is a commonly used photoinitiator
for 3D printing of hydrogels and biopolymers.[49] The optimized printing conditions are described in the Results and Discussion section. Thermogravimetric
analysis (TGA) was performed using a Mettler Toledo TGA/DSC 1 instrument
and the samples were heated in air up to 800 °C with a heating
rate of 5 °C·min–1. Tensile stress–strain
and adhesive testing experiments were performed using a Sauter FH
5 Force Gauge (Max. = 5 N, resolution = 0.001 N).
Figure 2
(A) POMAC pre-polymer
was synthesized through a co-condensation
reaction of the three different monomers (citric acid, maleic anhydride,
and 1,8-octanediol) in a 1:4:5 M ratio at 140 °C for 2 h under
a nitrogen atmosphere. (B) 1H-NMR spectrum of the POMaC
pre-polymer confirming the structure with the structural assignments
as defined in (A). (C) FT-IR spectrum of the POMaC pre-polymer highlighting
the key spectral features of the key functional group moieties. (D)
POMaC-containing inks, consisting of POMaC pre-polymer (100–Y) wt % combined with Y wt % PEGDA700 (where Y = 0, 5, 10, 20, 30, 40, or 50) and 5 wt % Irgacure 2959
photoinitiator, were 3D printed or molded (not depicted) using the
UV LED (λ = 365 nm) curing system on the 3D printer.
(A) POMAC pre-polymer
was synthesized through a co-condensation
reaction of the three different monomers (citric acid, maleic anhydride,
and 1,8-octanediol) in a 1:4:5 M ratio at 140 °C for 2 h under
a nitrogen atmosphere. (B) 1H-NMR spectrum of the POMaC
pre-polymer confirming the structure with the structural assignments
as defined in (A). (C) FT-IR spectrum of the POMaC pre-polymer highlighting
the key spectral features of the key functional group moieties. (D)
POMaC-containing inks, consisting of POMaC pre-polymer (100–Y) wt % combined with Y wt % PEGDA700 (where Y = 0, 5, 10, 20, 30, 40, or 50) and 5 wt % Irgacure 2959
photoinitiator, were 3D printed or molded (not depicted) using the
UV LED (λ = 365 nm) curing system on the 3D printer.
Synthesis of POMaC Pre-polymer
POMaC pre-polymer was
synthesized using a modified version of a synthesis procedure reported
by Tran et al.[8] In the work reported herein,
the reaction time for POMaC was reduced from 3 to 2 h, as it was determined
that the POMaC pre-polymer synthesized for 3 h led to a liquid which
was qualitatively too viscous for direct ink writing. The synthesis
process was repeated several times to obtain a large quantity of POMaC
pre-polymer for 3D printing optimization experiments. In a typical
synthesis, a dry round-bottom flask, containing a magnetic stirrer
bar, was heated to 160 °C under a nitrogen purge. Once at temperature,
citric acid (3.84 g, 2.00 × 10–2 mol), maleic
anhydride (7.84 g, 8.00 × 10–2 mol), and 1,8-octanediol
(14.62 g, 1.00 × 10–1 mol) (reagent molar ratio
= 1:4:5) were added to the flask in that order. The reagents were
then left to stir vigorously at 160 °C until all the reagents
were fully melted, forming a homogeneous clear and colorless liquid
reaction mixture. At this point, the reaction temperature was lowered
to 140 °C and the reagents were left to react for 2 h under a
nitrogen gas purge. After 2 h, the reaction mixture was cooled to
room temperature under a nitrogen purge. When cooled, the nitrogen
purge was stopped and 1,4-dioxane (∼150 mL) was added to the
flask, and the product was dissolved with vigorous stirring resulting
in a clear and colorless solution. This solution was then added dropwise
to vigorously stirred deionized water (5 L) forming a cloudy white
precipitate. This mixture was left to stir for at least 15 h, before
allowing the cloudy white gel-like hydrophobic precipitate to settle
to the bottom of the beaker and then decanting off the partially cloudy
supernatant liquid. This purification process was repeated two times
before a cloudy white viscous gel-like liquid was obtained. The product
was then lyophilized in a FreeZone 4.5 L Console Freeze Dry System
(Labconco, USA) for 3 days to obtain POMaC pre-polymer as a clear
and viscous liquid. 1H NMR (400 MHz, DMSO-d6): δ/ppm = 6.76–6.29 (m, 5 (maleic
moiety alkenyl protons)); 4.20–3.88 (m, 1 (octamethylene
−CH2 backbone protons)); 2.92–2.61 (m, 4 (citric acid moiety −CH2 protons); 1.69–1.47
(m, 2 (octamethylene −CH2 backbone
protons)); 1.27 (broad s., 3 (octamethylene −CH2 backbone protons)). The bold numbers refer to the structural
assignments defined in Figure A. NMR data are consistent with spectra reported in the literature.[6]
POMaC/PEGDA Printing Ink Blends
The preparation of
the ink blends was performed in the dark, and brown glass vials were
used. First, the photoinitiator Irgacure 2959 was weighed out (5,
or 20 wt % for pure POMaC) into the vial, then the appropriate wt
% of PEGDA700 was added, followed by the weighed addition of the corresponding
wt % of POMaC so that the wt % of PEGDA700 and POMaC added up to 100
wt %. The vials of inks were then sonicated in an ultrasonic bath
(35 kHz) for varying lengths of time (specific times given in Table ) until the photoinitiator
was fully dissolved and a homogeneous ink was obtained. However, care
was taken to maintain the temperature of the water bath below 45 °C
during sonication to avoid thermal polymerization. The ink compositions
are detailed below in Table .
Table 1
Compositions of the POMaC/PEGDA Biomaterial
Inks
biomaterial ink name
POMaC quantity/wt %
PEGDA700 quantity (Y)/wt %
PI quantity/wt %a
approx. sonication time until full dissolution/mins
50/50
50
50
5
30
60/40
60
40
5
30
70/30
70
30
5
60
80/20
80
20
5
60
90/10
90
10
5
120
95/5
95
5
5 (or 20)b
180
100/0
100
0
5 (or 20)c
180
PI = photoinitiator. The photoinitiator
used was Irgacure 2959.
To achieve 3D printing of the 95/5
dog bone pieces, 20 wt % PI was necessary for satisfactory printability.
For the 3D printing of the 95/5 rings, 5 wt % PI was used.
To achieve 3D printing of the 100/0
dog bone pieces, 20 wt % PI was necessary for satisfactory printability.
For the 3D printing of the 100/0 rings, 5 wt % PI was used.
PI = photoinitiator. The photoinitiator
used was Irgacure 2959.To achieve 3D printing of the 95/5
dog bone pieces, 20 wt % PI was necessary for satisfactory printability.
For the 3D printing of the 95/5 rings, 5 wt % PI was used.To achieve 3D printing of the 100/0
dog bone pieces, 20 wt % PI was necessary for satisfactory printability.
For the 3D printing of the 100/0 rings, 5 wt % PI was used.
3D Printing Method
For 3D printing, the inks were loaded
into syringes (PSYR5, Allevi, Philadelphia, USA) fitted with 25-gauge
Luer Lock dispensing tips (12.7 mm length, 0.25 mm inner diameter,
0.51 mm outer diameter, AD725050PK, Adhesive Dispensing Ltd, Milton
Keynes, UK). The geometries of the 3D printed pieces were defined
using standard 3D design software (SolidWorks, Dassault Systèmes,
France) and converted to .stl files for printing.
The desired geometries of the .stl models (a “dog
bone” and a ring) are shown in Figure S1. The pieces were then printed using an Allevi 3 bioprinter (Allevi,
Philadelphia, USA) equipped with a UV LED, λ = 365 nm, 20 mW·cm–2 at 100% intensity (5 cm distance between UV LED and
print specimen). The printing temperature was in the range of 35–40
°C and the print bed temperature was ambient room temperature.
The printing conditions are given in Table and discussed in detail in the Results and
Discussion section. The 3D printed ring pieces were used for tensile
stress–strain experiments.
Table 2
3D Printing Parameters for the Different
POMaC/PEGDA Biomaterial Ink Compositions for Ring and “Dog
Bone” 3D Printed Pieces
UV intensity
and duration
biomaterial ink composition
extrusion pressure/PSI
print speed/mm·s–1
during print
after print
50/50
10
8
15%
N/A
60/40
10
5
15%
N/A
70/30
30
6
0%
100% for 300 s
80/20
45
6
50%
100% for 300 s
90/10
60
8
100%
N/A
95/5a
80
4
100% for 1200 s per slice
100% for 1800 s
100/0b
90
4
100% for 1200 s per
slice
100% for 1800 s
Printing of 95/5 “dog bone”
pieces used these printing parameters, but 20 wt % photoinitiator
was required.
Printing of
100/0 “dog bone”
pieces used these printing parameters, but 20 wt % photoinitiator
was required.
Printing of 95/5 “dog bone”
pieces used these printing parameters, but 20 wt % photoinitiator
was required.Printing of
100/0 “dog bone”
pieces used these printing parameters, but 20 wt % photoinitiator
was required.
Tensile Stress–Strain Experiments
The tensile
stress–strain data of the 3D printed ring specimens were measured
using a vertical tensile stress–strain experimental setup as
shown in Figure S2. 3D printed ring pieces
were placed on the hooks so that the ring pieces were not under tension,
but were on the limit of tension upon an increase of displacement
in the tension force direction (Figure S2A). Displacement in the tension force direction was applied at a constant
rate. The experiments were video recorded and at each displacement
value (in mm) the adhesive peak load values (in N) were extracted
manually from the videos by transcribing the values from the force
gauge readout screen until the ring piece was observed to break. The
data were processed as per the formulae defined in ISO-37 to determine
average values of elongation % at break (Eb/%) and tensile strength at break (TSb/MPa) for each ink
composition.[50] To determine the true stress–elongation
% curves as per ISO-37, the cross-sectional areas of the ring pieces
were measured using optical microscopy and Fiji image analysis software.[51] Average Young’s Modulus (E) values were calculated by averaging the gradients of linear sections
of each true stress–elongation % curve for the 3D printed ring
pieces. Representative true stress–elongation % curve examples
are shown in Figure S2B. The values of E, Eb, and TSb are
presented as the means ± standard deviation.
Adhesive Test Experiments
The adhesive strength data
of the specimens, which had been formed by molding, were measured
by the pull-off tensile adhesion test method using the experimental
setup shown in Figure S3A. Cuboid specimens
[7.5 mm × 7.5 mm × (3.9 ± 0.8) mm, W × L × H] were fabricated by molding of
the different composition inks in a 12-well silicone chamber (Thistle
Scientific, Glasgow, UK) under the UV LED of the Allevi 3 printer
for at least 1 h for 50/50 and 60/40, ∼4 h for 70/30 and 80/20,
∼12 h for 90/10 and up to 24 h for 95/5 and 100/0, with the
“power” of the UV set to 100% on the software interface,
which corresponds to an intensity of 20 mW.cm–2.[52] The adhesive test experiments were recorded
using a camera and the adhesive peak force was extracted manually
from the videos by transcribing the highest magnitude value registered
on the force gauge readout screen before pull-off (due to adhesive
failure and not cohesive failure) of the specimen from the stainless-steel
dolly was achieved. The experimental method is described schematically
in Figure S3B-D. Any specimens that underwent
cohesive failure were discarded and the values not recorded.
Degradation Experiments
The degradation experiments
were performed in NaOH solution (1 M). This accelerated method was
chosen as the degradation of POMaC and PEGDA in phosphate-buffered
saline (PBS) or physiologically similar serum-containing media has
been extensively established in the literature.[8,53,54] Thus, in this work, it only needed to be
established if the POMaC/PEGDA 3D printed materials hydrolytically
degraded. Accelerated degradation using NaOH(aq.) was performed
to enable screening of the polymer degradation rates for the different
3D printed parts in a short period of time, as has been done previously
in the literature.[8] Cuboid specimens [7.5
mm × 7.5 mm × (3.9 ± 0.8) mm, W × L × H] were fabricated by molding
of the different composition inks in a 12-well silicone chamber (Thistle
Scientific, Glasgow, UK) under the UV LED of the Allevi 3 printer
as per the conditions stated above in the description of the adhesive
testing method. The specimens, which had been formed by molding, were
then weighed and placed in pre-weighed 40 μm mesh cell strainers
(for ease of handling), with the mass of the specimen plus strainer recorded as the starting mass (Starlab, Milton Keynes,
UK) before being placed into an NaOH solution (1 M) and incubated
at 37 °C for 72 h (corresponding to the point at which all samples
had degraded). During incubation, the strainers containing the specimens
were periodically removed, thoroughly dried, and the strainers containing
the degrading specimens were weighed, before being placed back into
the 1 M NaOH solution until 72 h. The mass difference of a cell strainer
over 72 h in 1 M NaOH at 37 °C was determined to be ±1%.
The mass difference of the specimens at the different time points
was calculated by comparing the initial mass of the specimen (calculated
by subtracting the mass of the strainer at t = 0
from the mass of specimen + strainer at t = 0) with
the mass of the specimens at the different time points (calculated
by subtracting the mass of the strainer + specimen at t = X from the mass of strainer at t = 0). Therefore, positive values of % mass change corresponds to
uptake of 1 M NaOH solution within the specimen (resulting in swelling)
and negative values of % mass change corresponds to the degradation
of a part or the whole of a specimen—the hydrolysis degradation
products remain solubilized within the 1 M NaOH solution. The results
are presented as the means ± standard deviation.
Biocompatibility Experiments
Cell permeable resazurin-based
solution (PrestoBlue), which is used to quantitatively measure the
cell proliferation, was used to determine the cell viability of fibroblast
cells (3T3-L1, American Type Culture Collection (ATCC) no. CL-173)
in presence of the varying ratio copolymers of POMaC/PEGDA. In brief,
prior to the cell viability assessment, cells were seeded into a 96-well
plate at a density of 2.5 × 104 per well and supplied
with 90 μL/well cell culture media [Dulbecco’s Modified
Eagle Medium (DMEM) containing 10% heat-activated fetal bovine serum
with 1% penicillin–streptomycin antibiotics (Pen-strep)] and
L-glutamine and incubated overnight at 37 °C and with 5% CO2 incubator to allow cell adherence. On the consecutive day,
the POMaC/PEGDA copolymer specimens of different wt % ratio compositions
were washed three times with PBS to remove any excess uncross-linked
inks or photoinitiator and then added into the wells and cultured
for 48 h. To evaluate the cell viability of the 3T3-L1, 10 μL
PrestoBlue was added to each well and incubated for 15 min at 37 °C.
The absorbance of the reagent at 570 nm was measured using a Thermo
Scientific Varioskan plate reader. Cells in absence of copolymer specimens
and cell culture medium were used as controls. This experiment was
repeated at least four times (n ≥ 4) and the
specimens were 3D printed.
Results and Discussion
A POMaC pre-polymer was synthesized
through a co-condensation reaction
of three different monomers (citric acid, maleic anhydride, and 1,8-octanediol)
in a 1:4:5 M ratio at 140 °C for 2 h under a N2 atmosphere
(Figure A). In contrast
to the reported reaction conditions in the literature, in this work
the reaction time for POMaC was reduced from 3 to 2 h, as it was qualitatively
determined that the POMaC pre-polymer synthesized after 3 h was too
viscous for molding and 3D printing.[8] Furthermore,
based on previous reports on POMaC, the 1:4:5 M ratio of reactants
was chosen (referred to as “POMaC8” in the literature)
to result in the highest number of double bonds within the pre-polymer
chain to increase the photocross-linking rate during molding and 3D
printing.[8] After lyophilization, the resulting
clear and colorless viscous product was characterized by 1H NMR and FT-IR spectroscopy (Figure B,C). The NMR data and the FT-IR data were consistent
with the spectra reported in the literature[6,8] and
thus it was deduced that the desired POMaC pre-polymer product had
been successfully synthesized. The POMaC pre-polymer was also characterized
with MALDI-TOF mass spectrometry and the most intense peak (100% relative
abundance) at m/z 546 g·mol–1 was attributed to the [M]+• species,
which corresponds to a pre-polymer oligomer of the structure shown
in Figure A where
R1 and R2 are −OH moieties (Figure S4). MALDI-TOF mass spectrometry was used
for characterization of the POMaC pre-polymer because it was utilized
in the first reported synthesis of POMaC by Tran et al.,[8] and it has been utilized for characterization
of other CABEs.[1,10,37,55,56]After
confirmation of the desired pre-polymer product, the biomaterial
inks consisting of POMaC pre-polymer (100–Y) wt % combined
with Y wt % PEGDA700 (where Y = 0, 5, 10, 20, 30,
40, or 50) and 5 wt % Irgacure 2959 photoinitiator (Table and Figure D) were mixed. Due to the desirable properties
highlighted in the introduction (vide supra), POMaC was the major
wt % component in all the ink compositions, apart from the 50/50 ink,
which was included to evaluate the effect of PEGDA on the properties
of the 3D printed materials across the full wt % range in which POMaC
is the major component. These biomaterial inks were then used for
either molding or 3D printing.For molding, silicone molds were
used to allow for facile handling
and all the biomaterial inks were successfully molded into cuboids
of 7.5 mm × 7.5 mm x (3.9 ± 0.8) mm, using the UV LED source
of the 3D printer. It was qualitatively determined that with increasing
wt % of POMaC, longer UV curing times were required. Following the
definition of effective energy as stated by Stowe[57] and knowing that the “100% power” setting
on the software interface corresponds to 20 mW·cm–2,[52] the effective energy required for
curing the 50/50 copolymer was calculated as 72 J·cm–2 for 1 h, and the effective energy required for curing the 100% POMaC
was calculated as 1728 J·cm–2 over 24 h. Knowing
that the area of the cuboid molds in this work was 0.5625 cm2, the total energies for curing the 50/50 polymer mold and the 100/0
polymer mold were 40.5 and 972 J, respectively. However, it must be
considered that the average thickness of the molds in this work were
(3.9 ± 0.8) mm, thus if the thickness difference is considered,
the normalized energy for the 50/50 ink is ∼1 J and for the
100/0 ink is ∼24 J. The lower energy value for the 50/50 ink
compared to pure POMaC in the work of Montgomery et al.[58] is to be expected as the addition of PEGDA700
increases the number of photocross-linkable groups per unit volume
and also increases the rate of cross-linking by decreasing the viscosity
of the ink. While the energy value for the 100/0 ink is larger than
the value reported in the literature,[58] it is assumed that there is a uniform intensity of UV light through
the molded piece when normalizing for the thickness difference, but
this is not physically representative of the absorption of UV light,
as described by the Beer–Lambert Law. Furthermore, it should
be noted that Montgomery et al. also used a further thermal cross-linking
treatment after the photocuring step,[58] whereas in this work only photocuring was used. Nonetheless, these
results demonstrated that the curing of POMaC-based inks is possible
with the UV LED curing system of a desktop off-the-shelf 3D printer.The 3D printing of the different composition inks was then investigated.
The inks were loaded into syringes fitted with 25-gauge Luer Lock
dispensing tips and then placed into the Allevi CORE printhead. The
desired .stl models, that had been designed using
standard 3D design software, were uploaded to the Allevi web-based
interface and an iterative printing process was performed. Both a
“dog bone” .stl model and a ring .stl model were used (Figure S1). An iterative process of trial-and-error 3D printing, informed
by the parameters needed to print the previous biomaterial ink composition,
was used for each different biomaterial ink. The 3D printing parameters
that were altered were extrusion pressure, print speed, UV intensity,
and UV exposure times during and after printing. Finally, a set of
parameters for each biomaterial ink composition, which resulted in
3D printing of the .stl models with good printability,
assessed by qualitative visual assessment, was achieved (Table ). Examples of the
resulting 3D printed “dog bone” and ring pieces are
shown in Figure A,B.
However, it should be noted that the 95/5 and 100/0 “dog bone”
pieces were only 3D printed to a qualitative visually acceptable printability
using 20 wt % photoinitiator and when using the same printing parameters
optimized for the 3D printing of 95/5 and 100/0 ring pieces.
Figure 3
(A) Photographs
of a selection of the 3D printed “dog bone”
pieces, with the biomaterial ink composition indicated below each
piece in yellow text, highlighting the high fidelity of the 3D printing
of the .stl “dog bone” model. “Dog
bone” pieces 95/5 and 100/0 in this figure were printed using
20 wt % photoinitiator. Scale bar = 1 cm. (B) Photographs of some
of the 3D printed ring pieces (80/20, left and 70/30, right) further
highlighting the 3D printing of complex shapes (white grid is equal
to 1 cm).
(A) Photographs
of a selection of the 3D printed “dog bone”
pieces, with the biomaterial ink composition indicated below each
piece in yellow text, highlighting the high fidelity of the 3D printing
of the .stl “dog bone” model. “Dog
bone” pieces 95/5 and 100/0 in this figure were printed using
20 wt % photoinitiator. Scale bar = 1 cm. (B) Photographs of some
of the 3D printed ring pieces (80/20, left and 70/30, right) further
highlighting the 3D printing of complex shapes (white grid is equal
to 1 cm).During this process, several trends between printing
parameters
and the composition of the biomaterial inks were observed. For instance,
there was a linear relationship between the required extrusion pressure
and wt % of POMaC pre-polymer in the biomaterial ink and the resulting
linear regression equation [extrusion pressure = 1.671(Weight % POMaC)—83.643]
enables prediction of the required extrusion pressure required for
inks with any wt % of POMaC between 50–100 wt % POMaC (Figure S5A). This can be explained by the viscosity
of the biomaterial inks, as with increasing wt % of POMaC the viscosities
were qualitatively observed to increase. The viscosity of PEGDA700
at room temperature is 98 mPa·s,[59] whereas that of structurally similar poly(alkyl citrates) is known
to be >210 mPa·s,[60] thus the greater
the wt % of POMaC, the more viscous the biomaterial ink and thus a
greater extrusion pressure was required. Additionally, the required
UV intensity during 3D printing generally increased with increased
wt % of POMaC (Figure S5C). By decreasing
the wt % of PEGDA700, the concentration of available double bonds
available to undergo radical polymerization decreased, thus this increased
both the UV curing time and the UV intensity required for sufficient
cross-linking to polymerize in a suitable timescale for extrusion
printing. However, it was determined that there was no clear relationship
between the print nozzle movement speed and wt % POMaC (Figure S5B).Then, the printability of
the different biomaterial inks with different
compositions was quantitatively assessed. The 3D printed ring pieces
(Figures S1C,D and 4A) were halved with a scalpel so that microscopy images of the cross
sections of each 3D printed ring (Figure B,C) could be captured, to enable the realized
height, width, and area of each cross section (Figure D) to be measured using Fiji image analysis software,[51] and compared
to the design height, width, and area of the cross section of the
ideal ring structure (.stl model) (Figure E).
Figure 4
(A) Photograph of an
80/20 3D printed ring piece (B) Example microscopy
image of the left-hand cross-sectional area of a 3D printed ring piece.
Scale bar = 1 mm. (C) Example microscopy image of the right-hand cross-sectional
area of a 3D printed ring piece. Scale bar = 1 mm. (D) Example microscopy
image of a cross-sectional area with an overlaid schematic defining
the measured cross-sectional area (red semi-transparent hemi-ellipse),
the cross-sectional height (vertical red arrow), and the cross-sectional
width (horizontal arrow). (E) Cross-sectional area of the ideal ring .stl model with an overlaid schematic defining the measured
cross-sectional area (red semi-transparent rectangle), the height
(vertical red arrow), and the width (horizontal arrow).
(A) Photograph of an
80/20 3D printed ring piece (B) Example microscopy
image of the left-hand cross-sectional area of a 3D printed ring piece.
Scale bar = 1 mm. (C) Example microscopy image of the right-hand cross-sectional
area of a 3D printed ring piece. Scale bar = 1 mm. (D) Example microscopy
image of a cross-sectional area with an overlaid schematic defining
the measured cross-sectional area (red semi-transparent hemi-ellipse),
the cross-sectional height (vertical red arrow), and the cross-sectional
width (horizontal arrow). (E) Cross-sectional area of the ideal ring .stl model with an overlaid schematic defining the measured
cross-sectional area (red semi-transparent rectangle), the height
(vertical red arrow), and the width (horizontal arrow).The desired ideal height of the cross section of
the ring was 3.75
mm, the desired ideal width of the cross section of the ring was 1.5
mm and thus the ideal cross-sectional area was 5.625 mm2. To quantify the printability, the ratio of the measured values
for the 3D printed ring pieces to the ideal values are calculated,
with a ratio of 1 corresponding to the measured value being equivalent
to the ideal dimension value. The ratios between the ideal and measured
dimensions (cross-sectional height, cross-sectional width, and cross-sectional
area) are shown in Figure S6A-C. It can
be noted that the ratios of the area, width, and height are different
from the desired design value = 1. However, from all the plots, 70/30
had the best printability, that is, the values of the ratios of area,
width, and height were closest to the design value. Therefore, with
further optimization of the 70/30 composition, better printability
is expected. This experiment implies that to achieve an ideal printability,
a compensation factor should be considered.The 3D printed pieces
were then characterized with FT-IR (Figures S7, 5A–D,
and S8) and Raman spectroscopy (Figure S9), and TGA (Figure S10 and Table ). There were clear differences in the FT-IR spectra of the copolymer
3D printed pieces with decreasing PEGDA700 wt % content (Figure S7); for example, in the C–H stretch
region (approx. 3000–2800 cm–1), which corresponds
to methylene C–H asymmetric (νas) and symmetric
(νs) stretches (Figure A) and in the fingerprint region between
approx. 1200–1000 cm–1, which corresponds
to C–O stretches of both ester and aliphatic ether functional
groups (Figure B).
Figure 5
(A) Upon
reduction of the PEGDA700 wt % content, the magnitude
of the asymmetric methylene C–H stretch (νas, ∼2930 cm–1) increased, whereas the symmetric
methylene C–H stretch (νs, ∼ 2855 cm–1) decreased, indicating a change in polymer chain
packing density. (B) Upon reduction of the PEGDA700 wt % content,
the magnitude of the C–O stretch of the ester functional groups
(νC-O, ester, ∼1160 cm–1) increased due to more ester groups, whereas the magnitude of the
C–O stretch of the aliphatic ether groups (νC-O, aliphatic ether, ∼1093 cm–1) decreased due to fewer aliphatic
ether functional groups. (C) Ratio of the intensity of the asymmetric
methylene C–H stretch (νas, ∼2930 cm–1) to the intensity of the symmetric methylene C–H
stretch (νs, ∼2855 cm–1)
increased with decreasing PEGDA700 wt % content until PEGDA700 ≤
40 wt % whereupon the ratio remained a constant value. The red line
is a guide to the reader indicating an intensity ratio value of 1.0.
(D) Ratio of the intensity of the C–O stretch of the ester
functional groups (νC–O, ester, ∼1160
cm–1) to the intensity of the C–O stretch
of the aliphatic ether groups (νC–O, aliphatic ether, ∼1093 cm–1) increased linearly with decreasing
PEGDA700 wt % content due to fewer ethylene glycol repeating units
present and a subsequent increase in the number of ester linkages
(see polymer structures in Figure ).
Table 3
TGA of the Copolymer Pieces 3D Printed
with the Different Biomaterial Inksa
composition
To/°C
Tp/°C
total mass loss/%
est. water content/%
0–100 (100% PEGDA)
171
357
98.5
0.2
50/50
190
397
97.8
0.5
60/40
176
407
98.6
0.7
70/30
177
389
99.6
0.9
80/20
177
392
99.3
0.9
90/10
167
381
99.2
0.8
95/5
161
395
98.6
0.6
100/0 (100% POMaC)
163
395
98.7
0.6
There were few differences between
the different compositions, apart from onset temperature, T, which decreased with increasing
POMaC content. The water mass content for all compositions was less
than <1%. T is defined
as per ASTM-E2550-17.[62]
(A) Upon
reduction of the PEGDA700 wt % content, the magnitude
of the asymmetric methylene C–H stretch (νas, ∼2930 cm–1) increased, whereas the symmetric
methylene C–H stretch (νs, ∼ 2855 cm–1) decreased, indicating a change in polymer chain
packing density. (B) Upon reduction of the PEGDA700 wt % content,
the magnitude of the C–O stretch of the ester functional groups
(νC-O, ester, ∼1160 cm–1) increased due to more ester groups, whereas the magnitude of the
C–O stretch of the aliphatic ether groups (νC-O, aliphatic ether, ∼1093 cm–1) decreased due to fewer aliphatic
ether functional groups. (C) Ratio of the intensity of the asymmetric
methylene C–H stretch (νas, ∼2930 cm–1) to the intensity of the symmetric methylene C–H
stretch (νs, ∼2855 cm–1)
increased with decreasing PEGDA700 wt % content until PEGDA700 ≤
40 wt % whereupon the ratio remained a constant value. The red line
is a guide to the reader indicating an intensity ratio value of 1.0.
(D) Ratio of the intensity of the C–O stretch of the ester
functional groups (νC–O, ester, ∼1160
cm–1) to the intensity of the C–O stretch
of the aliphatic ether groups (νC–O, aliphatic ether, ∼1093 cm–1) increased linearly with decreasing
PEGDA700 wt % content due to fewer ethylene glycol repeating units
present and a subsequent increase in the number of ester linkages
(see polymer structures in Figure ).There were few differences between
the different compositions, apart from onset temperature, T, which decreased with increasing
POMaC content. The water mass content for all compositions was less
than <1%. T is defined
as per ASTM-E2550-17.[62]The FT-IR spectra of the 3D printed pieces confirm
that upon reduction
of the PEGDA700 wt % content the intensity of the asymmetric methylene
C–H stretch (νas) peak increased, whereas
the intensity of the symmetric methylene C–H stretch (νs) peak decreased (Figure A). This was further confirmed by comparison of the
intensity of the νas peak to the intensity of the
νs peak for the different biomaterial ink compositions
as shown in Figure C. It was also concluded that the νas/νs ratio increases when the PEGDA700 wt % content decreases—up
to PEGDA700 ≤ 40 wt % whereas it remained at a constant value
of ∼1.4. An increase in the magnitude of the νas/νs ratio can indicate that there are larger changes
in the dipole moments of the vibrations due to an increase in free
volume allowing more space for polymer segmental motions.[61] An increase in free volume can be attributed
to the increased number of side chains due to the citric acid unit
in the polymer chain of POMaC (see Figure ) upon decreasing PEGDA700 wt %, which results
in poorer packing of the polymers chains and thus affords an increase
in free volume until PEGDA700 ≤ 40 wt %. In addition, the intensity
of the C–O stretch of the ester functional groups (νC–O, ester) increases with decreasing PEGDA700
wt % content, whereas the intensity of the C–O stretch of aliphatic
ether groups (νC-O, aliphatic ether) decreased (Figure B), as confirmed by the linear increase in the ratio of the intensities
of the C–O stretch peaks (νC–O, ester/νC–O, aliphatic ether) in Figure D. This is attributed
to fewer ethylene glycol repeating units (repeating unit of PEGDA,
see Figure ) in the
3D printed copolymers with decreasing PEGDA700 wt % content and increase
in the number of ester functional groups, thus confirming the difference
in wt % content of both polymers in each 3D printed copolymer piece.
Finally, it was determined that for each copolymer piece consisting
of the different 3D printed biomaterial ink compositions that there
were still unpolymerized C=C double bonds present due to the
presence of the double bond (C=C) stretch peak (νC=C, ∼1643 cm–1) in the FT-IR
spectra of the pieces (Figure S7), but
it is noted that the presence of remaining unpolymerized C=C
double bonds in POMaC-based polymers has been reported in the literature.[8] Upon further investigation of the ratio of the
intensity of the νC=C peak to the intensity
of the carbonyl (C=O) stretch peak (νC=O, ∼1724 cm–1) there was no clear difference
between the different biomaterial ink compositions (Figure S8), but the presence of remaining C=C bonds
was confirmed. Finally, analysis of the Raman spectra revealed that
there was an increase in the carbonyl C=O Raman shift (ΔṽC=O) with decreasing PEGDA700 wt % with a resulting
increase in POMaC wt % content, which has a greater number of carbonyl
functional groups (Figure S9). Thus, the
Raman spectral results further confirm the difference in wt % content
of both polymers in each 3D printed copolymer piece.The 3D
printed pieces were also characterized with TGA (Figure S10 and Table ). For all biomaterial ink compositions,
there was near complete mass loss after heating to 800 °C and
the estimated water mass content, but there was no apparent trend
between the different compositions and the peak temperature (Tp). However, there was a trend of decreasing
onset temperature (To), as defined as
per ASTM E2550-17,[62] with increasing POMaC
wt %. This can be attributed to poorer packing between the polymer
chains with increasing wt % of POMaC due to a greater number of side
chains from the citric acid moieties. Finally, the water content for
all biomaterial compositions was <1 wt %, which is corroborated
by the hydrophobic nature of POMaC pre-polymer, as noted during the
synthesis (vide supra) and in the literature.[58] In the case of the low water content of 0/100 (100% PEGDA), neat
PEGDA was used instead of forming a PEGDA hydrogel, as is typically
described in the literature.[63]For
potential utilization of these 3D printed materials for wound
sealing or wearable biomedical technologies, the mechanical and adhesive
properties need to be measured and understood. Therefore, the mechanical
properties of 3D printed ring pieces of each biomaterial ink composition
were evaluated through tensile stress–strain experiments as
per the experimental setup in Figure S2. Displacement in the tension force direction was applied at a constant
rate, and at each displacement value (in mm) the adhesive peak load
values (in N) were recorded until the ring piece was observed to break.
The data were then processed as per the formulae defined in ISO-37
to calculate average values of tensile strength at break (TSb/MPa) and elongation % at break (Eb/%)
for each biomaterial ink composition.[50] The cross-sectional areas measured with optical microscopy during
the printability evaluation experiments (vide supra) were used to
plot the true stress–elongation % curves from which average
Young’s modulus values (E) were determined.
The average values of TSb, Eb, and E for each biomaterial ink composition are
given in Table and
the relationship between the average Young’s modulus (E) values and PEGDA700 wt % is shown in Figure .
Table 4
Average Values of Young’s Modulus
(E), Tensile Strength at Break (TSb),
and Elongation at Break (Eb) Obtained
through Mechanical Property Analysis of 3D Printed Ring Pieces Consisting
of the Different Biomaterial Ink Compositionsa
composition
avg. Young’s modulus, E/MPa
avg. tensile strength at break,TSb/MPa
avg.
elongation at break, Eb/%
50/50
1.22 ± 0.01
0.163 ± 0.010
15.44 ± 0.05
60/40
0.75 ± 0.05
0.093 ± 0.046
21.06 ± 9.60
70/30
0.70 ± 0.06
0.080 ± 0.101
18.34 ± 2.08
80/20
0.89 ± 0.03
0.103 ± 0.008
32.98 ± 8.49
90/10
0.05 ± 0.22
0.007 ± 0.010
30.57 ± 6.46
95/5
0.05 ± 0.15
0.016 ± 0.016
55.90 ± 5.44
100/0 (100% POMaC)
0.12 ± 0.19
0.030 ± 0.034
44.83 ± 1.62
Average values of TSb and Eb are defined as per ISO-37.[50] Values are mean ± standard deviation.
Figure 6
With decreasing PEGDA700
wt %, there is a decrease in the Young’s
moduli (E), which is attributed to a decrease in
cross-link density with decreasing PEGDA700 wt %. The dashed blue
line is a guide for the eye only.
With decreasing PEGDA700
wt %, there is a decrease in the Young’s
moduli (E), which is attributed to a decrease in
cross-link density with decreasing PEGDA700 wt %. The dashed blue
line is a guide for the eye only.Average values of TSb and Eb are defined as per ISO-37.[50] Values are mean ± standard deviation.There was no clear trend observed between the ink
compositions
and Eb or TSb values, which
was attributed to intra-variability in terms of the difference of
ratios between the ideal and measured dimensions (the ideal design
values = 1) for different specimens consisting of the same ink composition.
Nonetheless, despite no clear linear regression relationship between
average E values and PEGDA700 wt %, there is an overall
decrease of E with decreasing PEGDA700 wt %. This is attributed to
fewer cross-links with decreased PEGDA700 wt %. In addition, it is
highlighted that the values of E for all the POMaC/PEGDA inks were
in the region of E values for human tissues[64] and in the region of typical E values for hydrogels
in the literature,[65] which further suggests
the suitability of 3D printed POMaC/PEGDA copolymer for biomedical
applications.Then, the adhesive strength of specimens of each
biomaterial ink
composition was evaluated using the pull-off tensile adhesion test
method. The highest magnitude value registered on the force gauge
readout screen (Figure S3) before pull-off,
due to adhesive failure and not cohesive failure, of a specimen from
the bottom stainless-steel dolly was recorded and an average adhesive
strength for each composition was calculated (n ≥
3) (Figure ).
Figure 7
Average adhesive
strengths measured, using the pull-off tensile
adhesion test method, for each different biomaterial ink composition
cured material, showing that the 90/10 composition was the “stickiest”,
that is, has the highest average adhesive strength. The results are
presented as the mean ± standard deviation, n ≥ 3. Statistical significance was evaluated using one-way
ANOVA followed by post-hoc t-tests with the Holm–Bonferroni
correction applied (* = P ≤ 0.05; ** = P ≤ 0.01; *** = P ≤ 0.001;
otherwise not significant). The magnitude of the values and error
bars for both the 50/50 and 60/40 compositions are too small to be
visible on this plot and thus the values are indicated on the plot.
Average adhesive
strengths measured, using the pull-off tensile
adhesion test method, for each different biomaterial ink composition
cured material, showing that the 90/10 composition was the “stickiest”,
that is, has the highest average adhesive strength. The results are
presented as the mean ± standard deviation, n ≥ 3. Statistical significance was evaluated using one-way
ANOVA followed by post-hoc t-tests with the Holm–Bonferroni
correction applied (* = P ≤ 0.05; ** = P ≤ 0.01; *** = P ≤ 0.001;
otherwise not significant). The magnitude of the values and error
bars for both the 50/50 and 60/40 compositions are too small to be
visible on this plot and thus the values are indicated on the plot.As shown in Figure , the largest magnitude average adhesive strength was
measured for
the 90/10 biomaterial ink composition specimen—four times larger
in magnitude than the average adhesive strength for the 100/0 “100%
POMaC” biomaterial ink composition specimen. Indeed, there
was a high statistically significant difference between the average
adhesive strength of the 90/10 and 50/50 biomaterial ink compositions,
very statistically significant differences between 90/10 and 60/40
and between 90/10 and 70/30, and a statistically significant difference
between 90/10 and 80/20. More crucially, there was a statistically
significant difference between 90/10 and 100/0, thus the 90/10 copolymers
were on average stickier than pure POMaC 100/0. It is proposed that
the enhanced adhesive strength of 90/10 compared to the other copolymers
is due to inclusion of an optimal wt % of hydrophilic PEGDA to the
relatively hydrophobic POMaC. Indeed, it is known in the literature
that PEGDA is hydrophilic,[66] and it has
been successfully used as a co-monomer for enhancement of adhesive
strength in other polymer systems.[67,68] The lack of
a statistically significant difference between 95/5 and 100/0 is attributed
to the large error caused by poor cohesive strength of the 95/5 copolymer
samples; adhesive failure without cohesive failure was difficult to
achieve. Nonetheless, these results suggest that another benefit of
copolymers of POMaC and PEGDA is tunable, and in some cases, increased
adhesive strengths are achieved compared to pure POMaC (100/0).For potential biomedical applications, ideally the materials need
to degrade so that the 3D printed material or a part is not permanently
stuck on or within the body, which can cause further complications.
It is known that POMaC is degraded under hydrolysis conditions and
is biodegradable and that PEGDA is bioinert. Indeed, after 2 weeks,
approximately 50% mass loss occurs for POMaC in PBS, and 80% mass
loss degradation occurs in 10 weeks.[8] In
addition, the degradation rate of PEGDA-based materials in PBS can
be tuned from hours to weeks,[53] and significant
in vivo degradation of PEGDA-based materials has been shown to occur
within 12 weeks.[54] Therefore, for this
work, all that was required was determination of whether these composite
materials, which contain varying amounts of degradable POMaC and degradable
PEGDA, also degraded (Figure ).
Figure 8
Degradation assay showing the mass changes recorded upon treatment
of cured pieces of the different biomaterial ink composition materials
with 1 M sodium hydroxide solution over a period of 48 h—degradation
was accelerated through pragmatic use of 1 M sodium hydroxide solution.[8] After 48 h, all pieces had fully degraded by
visual observation. The results are presented as the mean ± standard
deviation.
Degradation assay showing the mass changes recorded upon treatment
of cured pieces of the different biomaterial ink composition materials
with 1 M sodium hydroxide solution over a period of 48 h—degradation
was accelerated through pragmatic use of 1 M sodium hydroxide solution.[8] After 48 h, all pieces had fully degraded by
visual observation. The results are presented as the mean ± standard
deviation.The cured pieces of the different biomaterial ink
compositions
were treated under accelerated degradation conditions (1 M sodium
hydroxide solution at 37 °C with gentle agitation)—a pragmatic
choice in terms of experiment time for determination of hydrolytic
degradation.[8] At first, all of the pieces
were visually observed to swell, which was also confirmed for all
compositions by the recorded mass changes, apart from the 100/0 biomaterial
ink composition material, which fully degraded by visual observation
within 2 h under these conditions (and confirmed by the mass change
recorded). After 48 h, all the pieces were observed visually to have
degraded, and this was further confirmed with the mass change results
as all the cured biomaterial ink composition pieces underwent an approximate
−100% mass change. However, before 48 h, the biomaterial ink
composition that was most resistant to swelling and subsequent degradation
was the 70/30 composition. These results allow for design of the composition
of the POMaC/PEGDA biomaterial inks so that degradation time of the
3D printed pieces can be controlled.After confirming that the
different biomaterial copolymer pieces
were degradable, the biocompatibility was investigated. The cytotoxicity
of the copolymer materials was evaluated using the resazurin-based
PrestoBlue reagent. Metabolically active cells reduce the oxidized
non-fluorescent blue resazurin to the red fluorescent dye resorufin
by the mitochondria in live cells, a colorimetric change that is a
quantifiable indicator of viability of cells in culture. The amount
of resorufin is directly proportional to the number of viable cells.
As depicted in Figure , a minimum of 60% viability was measured for all materials relative
to the control, which highlights that these copolymer materials are
suitable for biomedical applications. Furthermore, there are no statistically
significant differences between the different copolymer compositions
and 0/100 (100% PEGDA) and 100/0 (100% POMaC), thus the combinations
of PEGDA with POMaC at different compositional ratios have little
to no toxicity. However, the statistically significant difference
in viability of 50/50, 60/40, 70/30, 80/20, and 95/5 relative to the
control for all materials is attributed to the toxicity of residual
Irgacure 2959 photoinitiator directly. Despite the known low toxicity
of Irgacure 2959, ≥5 wt % photoinitiator was required for successful
3D printing of these biomaterial copolymers. Therefore, it is hypothesized
that residual Irgacure 2959, not consumed during the photocross-linking
step of 3D printing, may generate radical species that are toxic to
the cells, thus resulting in a minimum of 60% cell viability for all
copolymer compositions. However, it is predicted that the biocompatibility
can be improved further as it has been shown previously that POMaC-based
materials shows good biocompatibility up to and beyond 8 weeks in
vivo (Figure ).[69]
Figure 9
Cell viability assay: reduction of PrestoBlue reagent
as a correlate
of cell viability over a period of 48 h is plotted for different biomaterial
ink compositions. Values are normalized to the control and shown as
mean ± standard deviation, n ≥ 4. Statistical
significance was evaluated using one-way ANOVA followed by post-hoc t-tests with the Holm–Bonferroni correction applied
(* = P ≤ 0.05; ** = P ≤
0.01; *** = P ≤ 0.001; otherwise not significant).
Cell viability assay: reduction of PrestoBlue reagent
as a correlate
of cell viability over a period of 48 h is plotted for different biomaterial
ink compositions. Values are normalized to the control and shown as
mean ± standard deviation, n ≥ 4. Statistical
significance was evaluated using one-way ANOVA followed by post-hoc t-tests with the Holm–Bonferroni correction applied
(* = P ≤ 0.05; ** = P ≤
0.01; *** = P ≤ 0.001; otherwise not significant).Overall, this proposed method, through modification
of the ink
formulations, affords a more generalizable approach to achieve 3D
printing of elastomers with a wide range of mechanical and physicochemical
properties, without the need to alter the chemical structure of the
elastomer, nor use of a support material[39] or fillers.
Conclusions
This work is the first example of 3D printing
of POMaC and POMaC-based
copolymer biomaterials. It was determined that extrusion 3D printing
of the POMaC-based inks could be optimized for, and achieved with,
an off-the-shelf lab-based extrusion 3D bioprinter and the use of
PEGDA as a copolymer. Qualitative assessment of the printed parts
demonstrated that complex shapes could be achieved; however, quantitative
assessment revealed that printing fidelity of these POMaC-based copolymer
biomaterial inks needs further optimization. Nonetheless, it was also
determined that the use of PEGDA does not impact the degradability
or biocompatibility, compared to 100/0 (100% POMaC), while allowing
for a significant increase in adhesive strength (90/10). Overall,
it was determined that the use of PEGDA has two benefits: 3D printing
of complex-shaped POMaC-based materials is achievable and the physical
properties of the resulting 3D printed parts can be tuned, optimized,
and altered through altering the ratio. Therefore, this work represents
a significant advance toward the application of 3D printed POMaC-based
inks for the next generation of personalized biomedical devices.
Authors: Dominic J Wales; Qun Cao; Katharina Kastner; Erno Karjalainen; Graham N Newton; Victor Sans Journal: Adv Mater Date: 2018-04-30 Impact factor: 30.849
Authors: Boyang Zhang; Miles Montgomery; M Dean Chamberlain; Shinichiro Ogawa; Anastasia Korolj; Aric Pahnke; Laura A Wells; Stéphane Massé; Jihye Kim; Lewis Reis; Abdul Momen; Sara S Nunes; Aaron R Wheeler; Kumaraswamy Nanthakumar; Gordon Keller; Michael V Sefton; Milica Radisic Journal: Nat Mater Date: 2016-03-07 Impact factor: 43.841