Nils Correvon1, Lucas Fasquel1,2, Pouyan Yazdani1,3, Jean-Bernard Michotte1, Jonathan Dugernier2, Olivier Contal1.
Abstract
Background: CPAP is the first line treatment of obstructive sleep apnea. Recently, the use of added filters has been debated following the field safety notice of Philips Respironics™ on potential health risks due to foam degradation used in their ventilators. However, the added resistance of filters has never been analyzed.Entities:
Keywords: APAP; CPAP; HEPA filter; antibacterial filter; obstructive sleep apnea syndrome (OSAS)
Year: 2022 PMID: 35968545 PMCID: PMC9364751 DOI: 10.3389/fmedt.2022.891390
Source DB: PubMed Journal: Front Med Technol ISSN: 2673-3129
Figure 1Bench test setup. Configuration was set in the following order: 1. APAP device, 2. Filter, 3. 180 cm SlimLine™ tubing (ResMed™, San Diego, USA), 4. Whisper Swivel™ II exhalation port (Philips-Respironics™, Murrysville, USA), 5. Connector, 6. UIAL valve, 7. Starling Resistance, 8. Connector, 9. ASL 5000™ simulator.
Air leaks flow of the UIAL valve according to device set pressure level.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Air leaks flow | 24.0 | 27.6 | 30.0 | 32.4 | 34.8 | 37.2 | 39.6 | 42.0 | 43.2 | 45.6 | 48.0 | 49.2 | 50.4 | 51.6 | 54.0 | 55.2 | 56.4 |
Figure 2Pressure variation in APAP mode during simulated apnea events, with UIAL.
Pressure variation in APAP mode during simulated respiratory events.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Without UIAL | Obstructive sleep apnea | No filter | 4.0 | 4.0 | 4.9 | 4.9 | 6.0 | 6.0 | 7.0 | 7.0 | 7.9 | 7.9 | 8.9 | 8.9 | |
| Filter 1 | 3.7 | 3.7 | 4.6 | 4.6 | 5.6 | 5.6 | 6.6 | 6.6 | 7.5 | 7.5 | 8.5 | 8.5 | |||
| Filter 2 | 3.6 | 3.6 | 3.6 | 4.5 | 4.5 | 5.4 | 5.4 | 6.4 | 6.4 | 7.3 | 7.3 | 8.3 | |||
| Filter 3 | 3.5 | 3.5 | 3.5 | 4.4 | 4.4 | 5.3 | 5.3 | 6.2 | 6.2 | 7.2 | 7.2 | 8.1 | |||
| Filter 4 | 3.5 | 3.5 | 3.5 | 4.4 | 4.4 | 5.4 | 5.4 | 6.3 | 6.3 | 7.3 | 7.3 | 8.2 | |||
| Hypopnea | No filter | 4.0 | 4.0 | 4.9 | 4.9 | 6.0 | 6.0 | 7.0 | 7.0 | 7.9 | 7.9 | 8.9 | 8.9 | ||
| Filter 1 | 3.6 | 3.6 | 3.6 | 4.6 | 4.6 | 5.6 | 5.6 | 6.5 | 6.5 | 7.5 | 7.5 | 8.5 | |||
| Filter 2 | 3.6 | 3.6 | 4.6 | 4.6 | 5.5 | 5.5 | 6.5 | 6.5 | 7.5 | 7.5 | 8.5 | 8.5 | |||
| Filter 3 | 3.5 | 3.5 | 4.4 | 4.4 | 5.3 | 5.3 | 6.2 | 6.2 | 7.1 | 7.1 | 8.1 | 8.1 | |||
| Filter 4 | 3.5 | 4.4 | 4.4 | 5.3 | 5.3 | 6.3 | 6.3 | 7.2 | 7.2 | 8.1 | 8.1 | 9.1 | |||
| With UIAL | Obstructive sleep apnea | No filter | 3.7 | 3.7 | 4.7 | 4.7 | 5.6 | 5.6 | 6.6 | 6.6 | 7.5 | 7.5 | 8.4 | 8.4 | |
| Filter 1 | 3.0 | 3.0 | 3.0 | 3.1 | 3.9 | 3.9 | 4.8 | 4.8 | 5.7 | 5.7 | 6.6 | 8.1 | |||
| Filter 2 | 2.8 | 2.8 | 2.8 | 2.8 | 2.8 | 2.8 | 3.6 | 3.6 | 4.4 | 4.4 | 5.3 | 5.3 | |||
| Filter 3 | 2.6 | 2.6 | 2.6 | 2.6 | 2.8 | 3.7 | 3.7 | 4.5 | 4.5 | 5.3 | 5.3 | 6.0 | |||
| Filter 4 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | 2.7 | |||
| Hypopnea | No filter | 3.7 | 3.7 | 3.7 | 3.7 | 4.7 | 4.7 | 5.7 | 5.7 | 6.6 | 6.6 | 7.5 | 7.5 | ||
| Filter 1 | 3.1 | 3.1 | 3.9 | 3.9 | 4.8 | 4.8 | 4.8 | 5.7 | 5.7 | 6.5 | 6.5 | 7.4 | |||
| Filter 2 | 2.8 | 2.8 | 3.6 | 3.6 | 4.4 | 4.4 | 5.2 | 5.2 | 6.1 | 6.1 | 6.9 | 6.9 | |||
| Filter 3 | 2.5 | 2.5 | 3.3 | 3.3 | 4.0 | 4.0 | 4.8 | 4.8 | 4.8 | 5.5 | 5.5 | 6.3 | |||
| Filter 4 | 2.7 | 2.7 | 3.5 | 3.5 | 4.3 | 4.3 | 5.1 | 5.1 | 5.9 | 5.9 | 6.7 | 6.7 | |||
UIAL, unintentional air leaks; Filter 1, GVS™ Medguard; Filter 2, Vyaire™ AirLife™; Filter 3, GVS™ ECO filter™; Filter 4, King System™ Virobac II™.
Effective end-expiratory pressure at different set CPAP pressure level.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Effective end expiratory pressure (cmH | No filter | 4.0 | 4.9 | 5.9 | 6.9 | 7.9 | 8.8 | 9.8 | 10.8 | 11.8 | 12.8 | 13.8 | 14.8 | 15.8 | 16.8 | 17.7 | 18.8 | 19.7 | −1.5 (0.4)* |
| Filter 1 | 3.6 | 4.6 | 5.5 | 6.4 | 7.4 | 8.3 | 9.3 | 10.3 | 11.3 | 12.2 | 13.2 | 14.1 | 15.1 | 16.0 | 17.0 | 18.0 | 18.9 | −6.2 (2.0)* | |
| Filter 2 | 3.5 | 4.5 | 5.4 | 6.3 | 7.2 | 8.2 | 9.1 | 10.1 | 11.0 | 12.0 | 12.9 | 13.9 | 14.9 | 15.8 | 16.7 | 17.6 | 18.6 | −8.2 (2.4)* | |
| Filter 3 | 3.5 | 4.4 | 5.3 | 6.1 | 7.1 | 8.0 | 9.0 | 9.8 | 10.8 | 11.7 | 12.7 | 13.6 | 14.5 | 15.5 | 16.4 | 17.3 | 18.3 | −10.0 (1.9)* | |
| Filter 4 | 3.5 | 4.4 | 5.3 | 6.2 | 7.1 | 8.1 | 9.0 | 10.0 | 10.9 | 11.8 | 12.8 | 13.7 | 14.7 | 15.6 | 16.6 | 17.6 | 18.4 | −9.2 (2.6)* | |
| Effective end expiratory pressure (cmH | No filter | 3.8 | 4.7 | 5.7 | 6.6 | 7.6 | 8.5 | 9.4 | 10.4 | 11.3 | 12.3 | 13.1 | 14.1 | 15.0 | 16.0 | 16.9 | 17.8 | 18.8 | −5.9 (0.5)* |
| Filter 1 | 3.1 | 4.0 | 4.8 | 5.6 | 6.5 | 7.4 | 8.2 | 9.1 | 10.0 | 10.8 | 11.7 | 12.5 | 13.4 | 14.4 | 15.2 | 16.0 | 16.9 | −16.9 (2.3)* | |
| Filter 2 | 2.9 | 3.7 | 4.5 | 5.3 | 6.1 | 7.0 | 7.8 | 8.7 | 9.4 | 10.3 | 11.1 | 12.0 | 12.8 | 13.7 | 14.5 | 15.3 | 16.1 | −20.9 (3.4)* | |
| Filter 3 | 2.6 | 3.3 | 4.0 | 4.8 | 5.5 | 6.3 | 7.1 | 7.8 | 8.6 | 9.4 | 10.1 | 10.8 | 11.6 | 12.3 | 13.1 | 13.8 | 14.6 | −28.3 (3.3)* | |
| Filter 4 | 2.7 | 3.5 | 4.3 | 5.1 | 6.0 | 6.7 | 7.6 | 8.4 | 9.3 | 10.0 | 10.9 | 11.7 | 12.5 | 13.3 | 14.1 | 15.0 | 15.9 | −23.1 (3.5)* |
CPAP, Continuous positive airway pressure; UIAL, unintentional air leaks; Median diff., the median difference between effective and set pressure at all pressure level; IQR, interquartile range; Filter 1, GVS™ Medguard™; Filter 2, Vyaire™ AirLife™; Filter 3, GVS™ ECO filter™; Filter 4, King System™ Virobac II™, *p-value < 0.001.