Maarten L S van der Geest1, Lucie McGovern2, Stefan van Vliet1, Hanya Y Zwaan1, Gianluca Grimaldi2,3, Jeroen de Boer2, Roland Bliem1,4, Bruno Ehrler2, Peter M Kraus1,5. 1. Advanced Research Center for Nanolithography, Science Park 106, 1098 XG Amsterdam, The Netherlands. 2. Center for Nanophotonics, AMOLF, Science Park 102, 1098 XG Amsterdam, The Netherlands. 3. Cavendish Laboratory, University of Cambridge,CB2 1TN Cambridge, United Kingdom. 4. Institute of Physics, University of Amsterdam, Science Park 904, 1098 XH Amsterdam, The Netherlands. 5. Department of Physics and Astronomy, and LaserLaB, Vrije Universiteit, De Boelelaan 1105, 1081 HV Amsterdam, The Netherlands.
Abstract
Inorganic-Organic lead halide materials have been recognized as potential high-energy X-ray detectors because of their high quantum efficiencies and radiation hardness. Surprisingly little is known about whether the same is true for extreme-ultraviolet (XUV) radiation, despite applications in nuclear fusion research and astrophysics. We used a table-top high-harmonic generation setup in the XUV range between 20 and 45 eV to photoexcite methylammonium lead bromide (MAPbBr3) and measure its scintillation properties. The strong absorbance combined with multiple carriers being excited per photon yield a very high carrier density at the surface, triggering photobleaching reactions that rapidly reduce the emission intensity. Concurrent to and in spite of this photobleaching, a recovery of the emission intensity as a function of dose was observed. X-ray photoelectron spectroscopy and X-ray diffraction measurements of XUV-exposed and unexposed areas show that this recovery is caused by XUV-induced oxidation of MAPbBr3, which removes trap states that normally quench emission, thus counteracting the rapid photobleaching caused by the extremely high carrier densities. Furthermore, it was found that preoxidizing the sample with ozone was able to prolong and improve this intensity recovery, highlighting the impact of surface passivation on the scintillation properties of perovskite materials in the XUV range.
Inorganic-Organic lead halide materials have been recognized as potential high-energy X-ray detectors because of their high quantum efficiencies and radiation hardness. Surprisingly little is known about whether the same is true for extreme-ultraviolet (XUV) radiation, despite applications in nuclear fusion research and astrophysics. We used a table-top high-harmonic generation setup in the XUV range between 20 and 45 eV to photoexcite methylammonium lead bromide (MAPbBr3) and measure its scintillation properties. The strong absorbance combined with multiple carriers being excited per photon yield a very high carrier density at the surface, triggering photobleaching reactions that rapidly reduce the emission intensity. Concurrent to and in spite of this photobleaching, a recovery of the emission intensity as a function of dose was observed. X-ray photoelectron spectroscopy and X-ray diffraction measurements of XUV-exposed and unexposed areas show that this recovery is caused by XUV-induced oxidation of MAPbBr3, which removes trap states that normally quench emission, thus counteracting the rapid photobleaching caused by the extremely high carrier densities. Furthermore, it was found that preoxidizing the sample with ozone was able to prolong and improve this intensity recovery, highlighting the impact of surface passivation on the scintillation properties of perovskite materials in the XUV range.
Inorganic–Organic lead halide perovskites
(APbX3, with A = methylammonium (MA, CH3NH3+), formamidinium (FA, CH(NH2)2+),
X = Cl, Br, I) have come to the scientific forefront through their
potential in many diverse applications, such as solar cell materials,[1] memristors,[2,3] and high-energy radiation
detectors.[4] This application potential
arises from their favorable and unique properties, including high
photoluminescence (PL) quantum efficiency,[5] long charge carrier and exciton diffusion lengths,[6−8] ion migration and associated memory effects,[9,10] strong
absorbance in the near IR/visible to ultraviolet range[11] and the hard X-ray range[12] as well as facile and cheap solution-based fabrication
into both single-crystal and polycrystalline structures.[13] It is therefore not surprising that lead halide-based
perovskites, both fully inorganic and organic–inorganic, were
reported as promising materials for X-rays (>1 keV) and high-energy
particle detection.[14,15] Of particular interest for high-energy
photon and particle detectors are the excellent downconversion quantum
efficiencies, radiation hardness, and intrinsic tunable visible wavelength
emission through mixing halides and adding interstitial atoms in appropriate
ratios.[4,15] Equally important, however, compared to
traditional radiation detection materials such as CsI:Tl, lead halide
perovskites have a very low production cost.[4] Combined, these properties could make for excellent indirect scintillators.
In an indirect scintillator, high-energy radiation or particles excite
photoelectrons, which in turn excite charge carriers that eventually
decay to emit detectable photons. A good indirect XUV scintillator
has a high conversion efficiency of excitation photons to carriers
and of carriers to emitted photons and can maintain this conversion
efficiency as the dose of ionizing XUV radiation accumulates. The
number of emitted photons versus accumulated dose is known as the
photostability.[16] The photostability of
an indirect XUV scintillator is important because photobleaching causes
the emission to decrease over time, making the scintillator less effective.
Chemical and physical changes can cause either a decrease or an increase
in photostability, as both can affect the number of emitting centers
available.[4]Lead halide perovskites
were found to be effective intrinsic indirect
scintillators for β-particles[17] and
γ-rays produced in positron emission tomography[18] as well as for soft and hard X-rays.[12,15] Little has been done to investigate whether this effectiveness extends
to lower energy ionizing radiation. While performing well under hard
X-rays,[14] the performance of methylammonium
lead bromide (MAPbBr3) under extreme ultraviolet (XUV)
radiation (10–120 eV) exposure is therefore not extensively
studied and is thus poorly understood. Understanding this behavior
is critical for evaluating the performance of these perovskites as
XUV scintillator, but it can also be critical, for instance, in space-based
solar cells,[19] where XUV radiation and
low-energy particles originating from the sun form a potential damage
source for photovoltaic materials, both conventional and those based
on APbX3 perovskites.Here, we report the investigation
of the visible light emission
of polycrystalline MAPbBr3 thin films excited by broadband
XUV pulses. This allows us to assess the potential of MAPbBr3 thin films as indirect XUV scintillators. Radiation in the XUV spectral
range is strongly absorbed by MAPbBr3 due to the ionizing
energy of the radiation,[20] which can induce
rapid chemical and physical changes of the material. A major chemical
change observed in the material surface was an oxidation reaction,
resulting in passivation of the surface. The result of this oxidation
is a spectral intensity recovery that counteracts the photobleaching,
which otherwise leads to nonradiative recombination. The spectral
intensity recovery is explored in detail, and we propose methods of
using the oxidation to further improve the scintillation properties
of MAPbBr3.
Methods
Thin-Film Preparation
The samples consist of MAPbBr3 spin-coated on quartz to a thickness of 180 ± 20 nm
and a grain size of 200–500 nm, as estimated from the morphology
in scanning electron microscopy micrographs in an earlier publication
using the same recipe.[10]Quartz substrates
(12 × 12 mm2, with 1 mm thickness) were cleaned by
ultrasonication for 15 min, followed by a cleaning and ultrasonication
with detergent, deionized water, acetone, and isopropyl alchohol,
and finally the cleaned substrates were further cleaned with an O2 plasma cleaner for 15 min. The MAPbBr3 perovskite
precursor solution was prepared by dissolving 1.1 M of methylammonium
bromide (98.0%, Sigma-Aldrich) and lead(II) bromide (PbBr2 > 98.0Z > 98.0%, TCI) with a 1:1 molar ratio into a 4:1 dimethylformamide
(anhydrous, 99.8%, Sigma-Aldrich)/dimethyl sulfoxide (anhydrous, ≥99.9%,
Sigma-Aldrich) solvent mix. The dissolution took place overnight on
a hot plate at 60 °C. After having cooled, 100 μL of the
MAPbBr3 precursor solution was spun onto the quartz substrates
at 6000 rpm for 30 s in a nitrogen-filled glovebox. Fifteen seconds
after the beginning of the rotation, 100 μL of chlorobenzene
antisolvent (anhydrous, Sigma-Aldrich) was pipetted onto the substrate.
After the spin-coating procedure, the substrates were annealed at
100 °C for 1 h. During all steps mentioned, the atmosphere inside
the glovebox was maintained at an O2 level below 1 ppm.
For the MAPbBr3 samples that were treated with ozone as
discussed in the corresponding section, samples were exposed using the UV ozone cleaner (Ossila)
at a temperature of 20 °C.
Thin-Film Characterization
The X-ray photoelectron
spectroscopy (XPS) analyses were performed with an HiPP-3 spectrometer
(Scienta Omicron) using a monochromatic aluminum Kα source (1486.6
eV, 20.0 mA, 14 kV). The HiPP-3 analyzer is used with a 0.8 mm cone
and a slit setting of 1.0 mm. An XPS data analysis was done through
the fitting of Voigt profiles and performed using KolXPD. Surface
charging corrections were applied based on the apparent binding energy
of the Br 3d5/2 peak.[21] Relative
peak areas were obtained by dividing the peak areas of peaks obtained
with the same pass energy by their tabulated 1486.6 eV photoemission
cross sections and setting the area of the Pb 4f peak to one.Ultraviolet–Visible (UV–vis) spectroscopy measurements
were conducted with a UV-2600 UV–vis spectrophotometer (Shimadzu)
with a 1 nm step size and measured against a blank quartz substrate.X-ray diffraction (XRD) measurements were conducted using a D2
Phaser (Bruker) with a copper Kα2 (8027.8 eV) source,
with a low-pass angle pinhole to ensure that only the area of interest
is measured and not the entire spot size of the XRD beam. Because
of this, the spectra have to be integrated over many hours to achieve
the required signal-to-noise ratio.
Extreme-Ultraviolet Excited Luminescence with a High-Harmonic
Source
The perovskite thin film is exposed to XUV radiation
using a table-top setup. A more detailed overview of this setup and
its characteristics is given in a previous publication.[22] Briefly, 1.1 mJ is split from the output of
a 2 kHz titanium:sapphire (Ti:Sa) laser (Solstice ACE, Spectra Physics,
800 nm, 35 fs, 7 W) and focused into a noble gas target in vacuum.
This generates broadband XUV radiation through the high-order harmonic
generation (HHG) process.[23−26] A representative spectrum generated at peak intensities
of ∼1.5 × 1014 W/cm2 is shown in Figure A. The XUV radiation
is refocused with a toroidal mirror onto the sample mounted on a sample
stage. The pulse energy is estimated to be ∼0.5 nJ/pulse on
target.[22,27] The spot size of the XUV on the sample is
∼45 μm full-width at half-maximum (fwhm), measured with
a knife edge. Together this yields a single pulse fluence of 26 μJ/cm2. An accumulated broadband XUV dose of 15 J/cm2 is thus achieved within 5 min using our table-top source. Different
excitation wavelengths between 220 and 2400 nm are generated in a
wavelength-tunable optical parametric amplifier (OPA) pumped by 1.3
mJ from the same Ti:Sa laser. This was used to compare the XUV-excited
emission with emission excited by longer-wavelength pulses, in this
series of experiments between 260 and 400 nm center wavelength. The
spot size of the output of the wavelength-tunable OPA is ∼150
μm fwhm at the sample position, measured with a knife edge.
Figure 1
(A) Calculated
absorption spectrum (blue) of 180 nm MAPbBr3 based on atomic
scattering factors and the reported density,
and XUV spectrum (red) from HHG, with a calibrated intensity axis
in number of photons per second per 0.1 eV. (B) UV–Vis absorption
(black) and emission spectrum when excited by broadband XUV radiation
(red) and 400 nm pulsed radiation (blue). The feature at 560 nm originates
from random lasing in the polycrystalline MAPbBr3 thin
film. (C) Normalized spectral emission intensity as a function of
accumulated dose on target in the case of MAPbBr3 excited
with 400 nm (green), XUV (red), and sodium salicylate (blue), also
excited by XUV pulses.
(A) Calculated
absorption spectrum (blue) of 180 nm MAPbBr3 based on atomic
scattering factors and the reported density,
and XUV spectrum (red) from HHG, with a calibrated intensity axis
in number of photons per second per 0.1 eV. (B) UV–Vis absorption
(black) and emission spectrum when excited by broadband XUV radiation
(red) and 400 nm pulsed radiation (blue). The feature at 560 nm originates
from random lasing in the polycrystalline MAPbBr3 thin
film. (C) Normalized spectral emission intensity as a function of
accumulated dose on target in the case of MAPbBr3 excited
with 400 nm (green), XUV (red), and sodium salicylate (blue), also
excited by XUV pulses.The emission from the sample is collimated by a
short focal length
parabolic mirror and guided out of the vacuum chamber into a streak
camera (C10910, Hamamatsu) for time-resolving the decay kinetics or
into a spectrometer (Maya 2000 PRO, Ocean Optics) to follow the emission
and bleaching on longer time scales (up to tens of minutes). The sample
is raster-scanned to expose larger areas where required.
Results and Discussion
XUV Photostability
MAPbBr3 shows intense
luminescence when excited with XUV radiation, as expected due to earlier
reported hard X-ray excited luminescence in MAPbBr3.[14] The XUV-excited visible luminescence spectrum
is shown as the red curve in Figure B and peaks at ∼540 nm. Because of XUV-induced
photobleaching the intensity of this luminescence decreases rapidly
over time (Figure C). Using atomic scattering factors,[20] we find that the absorption of XUV radiation in the range between
20 and 45 eV is particularly strong for MAPbBr3 due to
the proximity of the photon energies to the ionization threshold and
due to its constituent heavily absorbing elements and having a high
density of ∼3.5 g/cm3.[28] The XUV absorbance of a 180 nm thin film of MAPbBr3 is
depicted in blue in Figure A. When matched with a typical XUV spectrum from our HHG-based
XUV source (Figure A, in red), one can see that the absorbance is highest where the
XUV photon flux is likewise highest. The XUV absorbance is larger
than anywhere within the UV–vis absorption range, which is
shown in the black curve in Figure B, and also larger than anywhere else in the X-ray
range.[20] A calculation with Beer–Lambert’s
law, using the known XUV spectrum, suggests that 99% of all photons
are absorbed within the first 75 nm of the MAPbBr3 thin
film, as elaborated on in the Supporting Information. In addition, a single XUV photon has more than 10 eV excess energy,
when compared to visible or (D)UV photons, with which it can indirectly
excite multiple carriers through secondary electron cascades, further
contributing to a high carrier density.[29] This yields an estimated carrier density of between 6.2 × 1019 (at one carrier per XUV photon) and 6.7 × 1020 cm–3 (maximum discrete amount of carriers per
XUV photon). In the surface region even higher carrier densities are
estimated (between 2.3 × 1020 and 2.4 × 1021 cm–3). See the Supporting Information for details on the calculation. The large estimation
range originates from the unknown XUV photon to carrier conversion
factor. Figure C depicts
the integrated and normalized spectral emission intensity of a single-pulse
excitation fluence of 8 mJ/cm2 excited by 400 nm pulses
(green curve), which shows a very small decrease of the emission intensity
for large exposure doses (2% decrease for 78 J/cm2 accumulated
dose), despite an estimated carrier density of (6.7 ± 0.2) ×
1020 cm–3. This stands in sharp contrast
to the more than 75% decrease in emission intensity for XUV excitation
with an accumulated dose of 15 J/cm2 and 26 μJ/cm2 single-pulse fluence. This compares favorably with several
plastic scintillators as can be seen in the Supporting Information, Figure S2.As
the estimated surface carrier densities are very high in both cases,
but nevertheless comparable, the rapid photobleaching has to be explained
in a different manner. The major difference between the two is that,
in the case of XUV excitation, the photons have a lot of excess energy
with which to trigger chemical reactions at the surface and in this
way bleach the emitting state. This high carrier density accelerates
reactions responsible for photobleaching and modifies the surface,
where most of the XUV-excited emission originates, thus quenching
the luminescence. The resulting carrier density distribution does
not affect the shape of the emission spectrum, as can be seen in Figure B, although the sharp
amplified spontaneous emission (ASE) feature that results in random
lasing at 560 nm is not clearly visible when MAPbBr3 is
excited with XUV radiation, despite our carrier density estimates
suggesting that it should be visible.[30] The reason for the absence of the ASE is unknown. It is likely a
combination of inefficient pumping of gain channels due to the large
separation between excitation and emission wavelength, making it less
likely to form the required conditions for ASE, in combination with
the previously mentioned rapid photobleaching making the conditions
even more unfavorable.The photobleaching of the emission of
MAPbBr3 as a function
of accumulated dose shows anomalous behavior, not seen in the commonly
used XUV scintillator sodium salicylate (Figure C, blue curve).[31] In the case of sodium salicylate, the emission intensity drops rapidly
after receiving an accumulated dose of just 2 J/cm2, due
to high excited-state densities, triggering photobleaching of the
emitting state.[32] This is followed by a
slow further photobleaching as the remaining emitting states within
the excited sample volume are also bleached. Similar to sodium salicylate,
the XUV-excited PL intensity initially rapidly decreases in MAPbBr3, as seen in the red curve in Figure C. This is however followed by a remarkable
recovery of the PL intensity. After receiving three times the dose
it received at its lowest PL intensity (reaching 15% of its initial
value), the emission recovers to 40% of the initial intensity. This
remarkable recovery of PL intensity does not correspond to a change
in line shape of the emission spectrum. The initial emission spectrum
only shows an increase or decrease as a function of accumulated dose.
Intermittent radiation exposure, where the XUV source was turned on
and off by blocking the generating fundamental beam with a shutter,
did not affect the observed emission intensity recovery in terms of
minimum or maximum (Figure S1 in the Supporting Information), as the maximum was achieved
at approximately the same accumulated dose. We hypothesize that the
reason for this increase of emission intensity over time, after the
initial rapid photobleaching, is a comparatively slow oxidation reaction
that takes place at the surface, stimulated by the continuous XUV
exposure, resulting in surface passivation. This hypothesis is verified
in another section, vide infra, where we characterize an area of a
pristine sample that has been exposed to ∼10 J/cm2 of accumulated XUV radiation. Passivation through ozone oxidation
was reported to increase emission intensity in the case of 50 keV
hard X-ray excitation of a single crystal,[14] where it was attributed to ozone treatment leading to a reduced
trap state density at the surface. Trap states are a major contributor
to the reduced external quantum efficiency due to their ability to
prevent radiative recombination of the excited carriers.[29,33] In the following sections we confirm this oxidation-induced photostability
by time-resolved luminescence and postexposure characterization.
Time-Resolved Luminescence under XUV Exposure
The decay
kinetics of the emission are strikingly affected by the excitation
wavelength, as shown in Figure . The decay kinetics of an XUV-excited MAPbBr3 thin
film with a single-pulse fluence of only 26 μJ/cm2 (blue) is comparable to the decay kinetics of a sample excited at
a wavelength of 330 nm with a fluence of 17 mJ/cm2 (red),
despite having a factor of 650 times lower fluence. This is also in
stark contrast to the much slower decrease observed when exciting
the thin film with a more comparable fluence of 50 μJ/cm2 at a deep ultraviolet (DUV) wavelength of 260 nm (cyan).
Fitting a biexponential decay to the time-resolved, 260 nm excited
emission yields τ1 = 0.56 ± 0.2 ns and τ2 = 7.7 ± 0.8 ns with nearly equal proportionality constants
of α1 = 0.49 and α2 = 0.51. The
reason for this contrasting behavior is again the high carrier concentrations,
as elaborated on in the previous section, that are excited at the
surface, which in turn results in rapid quenching of the emission.
At such high carrier concentrations, trimolecular recombination pathways
become relevant, but the radiative emission should still be approximately
biexponential.[34] One can thus fit a biexponential
to the start of the XUV-excited PL decay and get a reasonable fit.
This yields decay constants τ1 = 0.18 ± 0.02
ns and τ2 = 2.6 ± 0.2 ns with proportionality
constants α1 = 0.83 and α2 = 0.17,
in rough agreement with the decay times of cesium lead bromide composites
excited with γ-rays.[18] The high-fluence,
330 nm-excited decay similarly has a rapid decay with decay constants
τ1 = 0.27 ± 0.02 ns and τ2 =
3.8 ± 0.2 ns with proportionality constants α1 = 0.88 and α2 = 0.12, but, as mentioned, an 800
times higher fluence is required to achieve this accelerated decay
for 330 nm excitation. The carrier density in the case of the high-fluence
330 nm excitation is 9.8 × 1020 cm–3, which is in the same order of magnitude as the high estimate for
XUV excitation. See the Supporting Information for details on the carrier density calculation. As is apparent from
the fits in Figure , there is a long-term effect that is not properly captured by a
simple biexponential fit at very high intensities, resulting in underfitting
in the case of both the 330 nm and XUV-excited emissions. A more in-depth
treatment of the PL from MAPbBr3 thin films when excited
in the XUV range or at high fluences is necessary to fully appreciate
the reason for a simple biexponential model being insufficient, which
is beyond the scope of this investigation.
Figure 2
Comparison of the PL
decay of MAPbBr3 when excited with
broadband XUV radiation (blue) with a single-pulse fluence of 26 μJ/cm2, 330 nm excitation with 17 mJ/cm2 (red), and 260
nm excitation with 50 μJ/cm2, including fits.
Comparison of the PL
decay of MAPbBr3 when excited with
broadband XUV radiation (blue) with a single-pulse fluence of 26 μJ/cm2, 330 nm excitation with 17 mJ/cm2 (red), and 260
nm excitation with 50 μJ/cm2, including fits.
Passivation through XUV Exposure
As the photostability
experiments indicate, XUV exposure causes a chemical reaction to occur
at the surface, which causes the beneficial emission recovery (Figure C). Specifically,
we observe oxidation of the polycrystalline MAPbBr3 films
due to the water layer that is present on virtually all surfaces,
even in high-vacuum systems.[35] It is reported
that MAPbBr3 reacts with water in ambient air over time,[36] so a reaction with water, especially when activation
energy is provided through high-energy photons,[21] is not unexpected. We exposed 1 mm2 of an MAPbBr3 thin-film sample to a dose of ∼10 J/cm2. XPS measurements of the unexposed and exposed part verify that
oxidation of the surface has taken place. This is shown in Figure A, where the XUV-exposed
area of a sample (blue curve) shows an O 1s feature, whereas this
peak is absent in the unexposed part of the sample (red curve). The
full XPS spectrum of the unexposed and exposed area of the MAPbBr3 can be seen in Figure S3A in the Supporting Information. Ratios between spectrally
integrated XPS peaks are summarized in Table . As lead does not outgas or otherwise leave
the sample,[21,36] ratios are normalized to the
Pb 4f peak area. These ratios are indicative of the elemental composition
of the surface and deviate from the expected ratio based on the bulk
formula. For the pristine samples, the ratios are comparable to previously
reported elemental compositions.[21,36] A closer look
at the O 1s peak in Figure B shows that at least three species have formed, one being
significantly lower in binding energy than the others. We shall look
more closely at the two large ones first. Both species are oxidized
forms of carbon, as lead oxide would be expected at lower binding
energies. This is further confirmed by there being no noticeable shift
or broadening in the Pb 4f peak (Figure D) and the amount of bromine relative to
lead not changing, as can be seen in Table . Furthermore, the Br 3d peak does not show
any noticeable change either (Figure S3D,E in the Supporting Information), which
is not unexpected in an ultrahigh vacuum.[10,21] The C–N peak in Figure C shows an apparent broadening after XUV exposure,
caused by additional oxidized carbon species forming on the surface.
These species are singly bonded (likely alcohols) and doubly bonded
(ketones, aldehydes) carbon to oxygen, based on their binding energies
and relative peak intensities in both C 1s and O 1s.[37] This is further confirmed by the fact that the N 1s peak
does not show any shift in binding energy, nor does it show any broadening
(Figure S3B,C in the Supporting Information). Although the high binding energy
of 288.4 eV of the smallest C 1s peak suggests a carboxylate (−CO2–) or a carbonate (−CO3–), the total amount of oxygen present in the O
1s peak speaks against this interpretation. The ratio of the total
peak area of carbon, bound to oxygen, to the O 1s peak is closer to
1.4:1, which makes a ketone or aldehyde more likely to correspond
to this high binding energy. The observed carbon bound to oxygen is
higher than expected still, which could possibly result from the higher
kinetic energy and thus larger probing depth of C 1s compared to the
O 1s photoelectrons. The small (6% of total O 1 area) peak at a binding
energy of 529 eV is a metal oxide, and the likely candidate is lead
oxide (PbO), as lead dioxide (PbO2) would appear as a broader
peak due to the presence of a strong satellite at higher binding energies.[38,39] Because of the low amount of PbO, as suggested by the size of the
peak in the O 1s spectrum, it cannot be observed in the Pb 4f spectrum,
as it overlaps with the much stronger Pb–Br peak. The ratio
of the relevant peak area in the O 1s spectrum and the Br 3d doublet
peak area is ∼3:100 (Pb–O/Pb–Br).
Figure 3
XPS spectra of exposed (blue) and pristine (red) surfaces of an
MAPbBr3 thin film. (A) Survey spectrum with labeled peaks,
showing the appearance of an O 1s peak and a strong shape change in
the C 1s peak. (B) The O 1s peak of a pristine and exposed part of
the sample. (C) The C 1s peak of exposed and pristine MAPbBr3 shows the formation of oxidized carbon species, in addition to a
strong increase in the amount of C–C bonded carbon at low binding
energies. (D) The Pb 4f peak, showing no change.
Table 1
Measured Ratios of Spectrally Integrated
XPS Peaks Divided by the Photoemission Cross Sections of the O 1s,
C 1s, N 1s, Pb 4f, and Br 3d Core Levels, Normalized to the Spectrally
Integrated Pb 4f Peak, for a Pristine and an XUV-Exposed Area of a
MAPbBr3 Perovskite Sample
atom. species orbital
pristine
XUV exposed
Pb 4f
1
1
O 1s
0
1.1
C
1s
1.6
4.7
N
1s
0.8
0.8
Br
3d
2.4
2.3
XPS spectra of exposed (blue) and pristine (red) surfaces of an
MAPbBr3 thin film. (A) Survey spectrum with labeled peaks,
showing the appearance of an O 1s peak and a strong shape change in
the C 1s peak. (B) The O 1s peak of a pristine and exposed part of
the sample. (C) The C 1s peak of exposed and pristine MAPbBr3 shows the formation of oxidized carbon species, in addition to a
strong increase in the amount of C–C bonded carbon at low binding
energies. (D) The Pb 4f peak, showing no change.XRD spectra show that the peak ratio between the
(001) and (002)
planes, which are normally approximately fixed over a chemically homogeneous
sample, is significantly different when comparing an exposed and an
unexposed part on the same sample, as can be seen in Figure A. Where the (002) peak is
∼50% of the height of the peak at (001) in the unexposed part
of the sample, the exposed part shows a ratio closer to 80%, while
for both the exposed and unexposed cases the 2θ angles agree
with reported literature values.[40] This
again supports chemical change of the surface, rather than in the
bulk, as the peaks do not broaden or shift from their 2θ angle.
Indeed, formation or introduction of other atomic species at low percentages
at the surface has been reported to change the ratio of XRD peaks
without broadening or affecting the 2θ angle.[41,42] A small change in UV–vis absorbance is also observable (Figure B,C), starting at
∼395 nm, as well as a small blue-shift of the exciton feature.
This does not point to a particular species being formed but could
result from the small amount of PbO that is formed, as the orthorhombic
polymorph of PbO absorbs strongly in the blue to ultraviolet region
and could thus also explain the small apparent blue-shift of the exciton
feature.[43]
Figure 4
(A) XRD spectra of an unexposed and exposed
area on the same sample
of MAPbBr3 thin film, with peak positions in agreement
with previously reported values[40] but with
vastly different ratios between the (100) and the (200) directions.
(B) UV–Vis absorption spectra of an unexposed and exposed area.
(C) Difference in absorbance between the exposed and unexposed areas,
showing an increased absorbance around 395 nm.
(A) XRD spectra of an unexposed and exposed
area on the same sample
of MAPbBr3 thin film, with peak positions in agreement
with previously reported values[40] but with
vastly different ratios between the (100) and the (200) directions.
(B) UV–Vis absorption spectra of an unexposed and exposed area.
(C) Difference in absorbance between the exposed and unexposed areas,
showing an increased absorbance around 395 nm.It is thus clear that the anomalous photostability
in Figure C is caused
by XUV
induced passivation of the surface region, which removes trap states
that would otherwise inhibit radiative decay of the excited state
in untreated MAPbBr3. XUV radiation has a high enough photon
energy to trigger the necessary chemical reactions. Nevertheless,
at higher accumulated doses, the emission intensity still trends toward
zero. For this reason, preoxidizing the surface can potentially also
have a positive effect on the emission intensity over time, as trap
states are removed before XUV excitation, instead of through XUV excitation.
We explore this concept in the next section.
Improved XUV Photostability through UV Ozone Treatment
Samples of MAPbBr3 thin films were exposed to progressively
longer durations of UV ozone treatment to passivate the surface before
XUV excitation and to determine if there is an optimal ozone-treatment
time before ozone starts negatively impacting XUV photostability through
excessive oxidation. The results are summarized in Figure and Table . The dose required to achieve maximum recovery
increases when treating samples with ozone, at the cost of a slightly
lower absolute PL intensity. This is apparent in Figure F: a near-doubling of the accumulated
dose is required to achieve the maximum of the emission-intensity
recovery, while it peaks to slightly lower values than before the
sample was treated (from 35% to 32%). The minimum of the emission
has decreased significantly (from 25% to 15%) after ozone treatment.
The reason for this is unknown. The optimal time was found to be 120
s of ozone treatment—the associated curve is depicted in Figure F. Longer treatment
times tend to have diminishing returns. See Figure S5B in the Supporting Information for the 10 min of UV ozone treatment case, showing both rapid photobleaching
and no recovery as well as very little initial intensity to start
with. No XUV-excited PL was observed in samples exposed to 20 min
of UV ozone treatment, likely due to the ozone oxidizing the surface
to such an extent and depth that the attenuation length of the XUV
is comparable to the depth of oxidation, resulting in too few emitting
states to be observed. This is also evidenced by the UV–vis
absorption spectrum before and after ozone treatment, with a significant
decrease in absorbance due to the oxidation in the case of 20 min
of ozone treatment, as can be seen in Figure D. Despite the strong decrease in absorbance
in the blue/green region of the spectrum, the increase in absorbance
in the 580–630 nm region of the absorption spectrum actually
causes samples with 20 min of UV ozone treatment to appear somewhat
darker than before to the naked eye. The change is more clearly visible
in the differential absorbance spectrum in Figure E. This absorption feature originates from
the formation of PbO, which absorbs strongly around 600 nm in one
of its polymorphs.[43] Ozone treatment does
not affect the electronic structure of the emitter under XUV excitation,
as only the surface region of the material is affected. This can be
seen in Figure S4A.
Figure 5
(A) XPS spectra of MAPbBr3 thin films at increasing
UV ozone treatment durations. (B, C) O 1s and C 1s peaks of the ozone-treated
samples with assigned species. (D) UV–Vis absorption of the
ozone-exposed and unexposed samples. (E) ΔAbsorbance of ozone-treated
samples. (F) Normalized PL intensities for a 120 s UV/ozone-treated
and pristine thin film sample, in green and red, respectively, showing
a significant shift in accumulated dose required to achieve the maximum
intensity.
Table 2
Measured Ratios of Spectrally Integrated
XPS Peaks at the O 1s, C 1s, N 1s, Pb 4f, and Br 3d Binding Energies,
Normalized to the Pb 4f peak, for Increasing UV Ozone-Treatment Timesa
atom. species
30 s
1 min
2 min
20 min
orbital
Pb 4f
1
1
1
1
O 1s
0
0.05
0.5
1.2
C 1s
1.2
0.9
0.6
1.9
N 1s
0.8
0.7
>0a
0
Br 3d
2.5
2.3
1.2
0.1
The ratio of N to Pb is larger
than zero, as the peak is somewhat visible, but it cannot be properly
fitted for quantification.
(A) XPS spectra of MAPbBr3 thin films at increasing
UV ozone treatment durations. (B, C) O 1s and C 1s peaks of the ozone-treated
samples with assigned species. (D) UV–Vis absorption of the
ozone-exposed and unexposed samples. (E) ΔAbsorbance of ozone-treated
samples. (F) Normalized PL intensities for a 120 s UV/ozone-treated
and pristine thin film sample, in green and red, respectively, showing
a significant shift in accumulated dose required to achieve the maximum
intensity.The ratio of N to Pb is larger
than zero, as the peak is somewhat visible, but it cannot be properly
fitted for quantification.The exact chemical reactions occurring at the surface
due to ozone
treatment are different from those occurring due to XUV exposure,
as evidenced by XPS measurements. This is not unexpected, as oxidation
takes place due to an atmosphere of highly reactive ozone, instead
of reactions with a thin water layer mediated by XUV photons. The
XPS spectra of the ozone-treated samples in the range between binding
energies of 580 and 220 eV are shown in Figure A, normalized to the spectrally integrated
Pb 4f peak in the full spectrum (Figure S4A). Ratios of elements, normalized to Pb 4f, can be seen in Table . Looking at only
the full spectrum and Table , several trends become apparent. First, the Pb 4f peak becomes
broader as Pb–O bonds are formed and Pb–Br bonds are
oxidized. The Pb 4f peak shows a small shift (Figure S4B,C), most apparent after 20 min of ozone treatment,
due to the formation of PbO. The nitrogen bound in methylammonium
is initially stable as evidenced by the presence of a C–N peak
in the C 1s spectrum in Figure C and the single peak in the N 1s spectrum (Figure S4D), but after only 2 min of ozone treatment the remaining
amount of nitrogen can no longer be properly quantified in either
the C 1s or N 1s spectra, and after 20 min it is no longer measured.
Bromine follows the same trend: it remains stable after up to 1 min
of ozone treatment but disappears rapidly at longer treatment times,
with approximately half remaining after 2 min and almost completely
disappearing after 20 min (Figure S34E–G). Ozone treatment also causes an oxygen peak to appear after 2 min,
as can be seen in detail in Figure B.The C–C peak at low binding energies
in the C 1s spectrum
(Figure C) decreases
faster than this O 1s peak appears, leading us to conclude that an
ozone treatment first of all removes hydrocarbons from the surface
before reacting with the surface itself. The O 1s peak becomes rather
complex after only 2 min of UV ozone treatment, requiring at least
three Voigt profiles to fit in the case of 2 min and four profiles
in the case of 20 min of UV ozone treatment. The O 1s peak at 20 min
of UV ozone treatment matches the O 1s peak reported by Wei et al.
in case of their ozone oxidation treatment.[14] The lowest-energy species at 529.5 eV is again PbO, which is much
stronger than in the XPS spectrum of XUV-exposed MAPbBr3 in Figure B. It
is unlikely to correspond to PbO2, as the ratio of oxygen
present in the O 1s peak is closer to 0.8:1 compared to the total
Pb 4f peak area. The slightly mismatched O–Pb/O ratio is possibly
caused by the Pb 4f peak corresponding to two species, PbO and lead
hydroxide (Pb(OH)2), which sometimes occurs as a shoulder-like
peak in PbO XPS spectra.[38,44] The ratio of the O
1s peak at a binding energy of 530.9 eV to the smaller Pb 4f peak
(Figure S4B, top spectrum) is ∼2:1,
which also supports this assignment, although it is unknown why there
is no significant hydroxide peak in the case of 2 min of ozone treatment.
The remaining two peaks in the O 1s spectrum are again oxidized carbon
species, as there are matching peaks present in the C 1s spectrum.
The lower binding energy peak matches to an equivalent peak in the
carbon spectrum in an almost 1:1 ratio in both cases, suggesting the
formation of a C–O bond, perhaps again in the form of an alcohol.
A rather puzzling difference occurs in the case of the highest energy
peak, however. The ratio of the C 1s peak at 288.4 eV to the O 1s
peak at 533.3 eV is closer to 1:3 in the case of 2 min of UV ozone
treatment, suggesting that a carbonate is formed. This changes in
the case of a 20 min treatment, where the ratio is again closer to
1:1, suggesting a ketone or aldehyde to have formed concurrently with
the lead hydroxide. The reason for this change is unknown but is perhaps
linked to the disappearance of Br and the formation of lead hydroxide.Overall, this leads us to conclude that an ozone treatment first
oxidizes C–C species on the surface and thus cleans and perhaps
already partially passivates the surface, which explains the reduction
of the C 1s signal at 284.6 eV for short ozone-treatment times. After
a tipping point between one and 2 min of ozone treatment, the ozone
starts to react with nitrogen and bromide, causing outgassing, and
the lead, alongside remaining carbon as well as additional carbon
from the environment, is oxidized to form PbO and a series of oxidized
carbon species. This simple demonstration thus shows that a UV ozone
treatment is beneficial for indirect XUV scintillation in MAPbBr3 thin films but only up to a certain duration of UV ozone
treatment that is yet to be properly determined and warrants further
investigation.
Conclusions
We have demonstrated that MAPbBr3 has remarkable XUV-excited
emission properties in terms of emission intensity as a function of
exposure time. Time-resolved XUV-excited luminescence measurements
showed accelerated decays compared to UV-excited luminescence, which
we attributed to an increased carrier concentration at the surface
despite low excitation fluences. This increased carrier density drives
photoinduced oxidation and photobleaching. The oxidation counteracts
the photobleaching, and thus we observed a recovery of emission intensity
as a function of exposure dose. XPS measurements helped attribute
this recovery to the formation of passivating oxidized lead and carbon
compounds. In order to harness this effect for improving the XUV-excited
luminescence of MAPbBr3, we applied a simple ozone treatment
to preoxidize the sample. This controlled passivation improved the
photobleaching tolerance of our samples for XUV excitation, doubling
the dose required to photobleach the sample completely. Our results
demonstrate that MAPbBr3 can be used as an indirect XUV
scintillator, due to its remarkable recovery mechanism, especially
when pretreated with ozone.
Authors: Yanqi Luo; Parisa Khoram; Sarah Brittman; Zhuoying Zhu; Barry Lai; Shyue Ping Ong; Erik C Garnett; David P Fenning Journal: Adv Mater Date: 2017-09-29 Impact factor: 30.849
Authors: Samuel D Stranks; Giles E Eperon; Giulia Grancini; Christopher Menelaou; Marcelo J P Alcocer; Tomas Leijtens; Laura M Herz; Annamaria Petrozza; Henry J Snaith Journal: Science Date: 2013-10-18 Impact factor: 47.728
Authors: Johannes M Richter; Mojtaba Abdi-Jalebi; Aditya Sadhanala; Maxim Tabachnyk; Jasmine P H Rivett; Luis M Pazos-Outón; Karl C Gödel; Michael Price; Felix Deschler; Richard H Friend Journal: Nat Commun Date: 2016-12-23 Impact factor: 14.919
Authors: Sylvianne D C Roscam Abbing; Filippo Campi; Alexandra Zeltsi; Peter Smorenburg; Peter M Kraus Journal: Sci Rep Date: 2021-12-20 Impact factor: 4.379