Abstract
Objective: To explore the validity and accuracy of electrochemiluminescence immunoassay (ECLIA) and enzyme-linked immunosorbent assay (ELISA) in the serological detection of the hepatitis B virus.Entities:
Year: 2022 PMID: 35966724 PMCID: PMC9371851 DOI: 10.1155/2022/9371497
Source DB: PubMed Journal: Evid Based Complement Alternat Med ISSN: 1741-427X Impact factor: 2.650
The reference value range of five serological indexes of hepatitis B.
| Serological indicators | Reference value (more than this value is positive) |
|---|---|
| HBeAg | ≥ 0.05 NCU/Ml |
| HBeAb | ≥ 2.00 NCU/mL |
| HBsAg | ≥ 10.00 mU/mL |
| HBsAb | ≥ 0.20 ng/mL |
| HBcAb | ≥ 1.50 NCU/mL |
Comparison of key baseline data between the two groups ( ± s).
| Groups |
| Gender | Age (years-old) | BMI | |
|---|---|---|---|---|---|
| Male | Female | ||||
| EA | 48 | 26 | 22 | 47.87 ± 6.26 | 24.74 ± 1.66 |
| EB | 48 | 20 | 28 | 44.22 ± 8.68 | 24.93 ± 1.57 |
| Χ2/ | 0.032 | 0.824 | 0.525 | ||
|
| 0.846 | 0.126 | 0.557 | ||
Comparison of the five serum indicators tested by ECLIA and ELISA ( ± s).
| Methods | Serological indicators | ||||
|---|---|---|---|---|---|
| HBeAg (NCU/mL) | HBeAb (NCU/mL) | HBsAg (mU/mL) | HBsAb (ng/mL) | HBcAb (NCU/mL) | |
| GA (ECLIA) | 0.12 ± 0.04 | 2.84 ± 0.52 | 10.78 ± 1.67 | 0.36 ± 0.08 | 1.71 ± 0.13 |
| GB (ELISA) | 0.09 ± 0.02 | 2.19 ± 0.49 | 9.86 ± 1.58 | 0.30 ± 0.05 | 1.69 ± 0.11 |
|
| 6.254 | 14.967 | 6.331 | 11.151 | 1.471 |
|
| < 0.05 | < 0.05 | < 0.05 | < 0.05 | 0.084 |
Comparison of the positive detection rates of the five serum indicators tested by ECLIA and ELISA [n (%)].
| Methods |
| Serological indicators | ||||
|---|---|---|---|---|---|---|
| HBeAg | HBeAb | HBsAg | HBsAb | HBcAb | ||
| GA (ECLIA) | 48 | 48 (100.00) | 44 (91.67) | 46 (96.33) | 43 (89.58) | 45 (93.75) |
| GB (ELISA) | 48 | 40 (83.33) | 37 (77.08) | 39 (81.25) | 36 (75.00) | 42 (87.50) |
| Χ2 | 5.912 | 5.887 | 6.024 | 5.545 | 1.283 | |
|
| 0.011 | 0.014 | 0.009 | 0.021 | 0.256 | |
Comparison of diagnostic efficacy between ECLIA and ELISA [n (%)].
| Methods |
| Accuracy ( | Sensitivity ( | Specificity ( |
|---|---|---|---|---|
| GA (ECLIA) | 50 | 48 (96.00) | 28 (100.00) | 21 (95.45) |
| GB (ELISA) | 50 | 45 (90.00) | 25 (89.29) | 18 (81.82) |
| Χ2 | 0.457 | 1.582 | 1.338 | |
|
| 0.365 | 0.047 | 0.31 |