Abstract
PURPOSE: To evaluate microscopically whether the print quality and accuracy of sizing of Landolt ring near vision charts are adequate for the calibration of reading charts.Entities:
Keywords: Landolt rings; Near vision charts; Near visual acuity; Print quality; Reading acuity; Sizes of optotypes
Year: 2022 PMID: 35965343 PMCID: PMC9377069 DOI: 10.1186/s40662-022-00302-5
Source DB: PubMed Journal: Eye Vis (Lond) ISSN: 2326-0254
Fig. 1C-Test Landolt ring test (left column) and Landolt rings of the near vision chart from the RADNER Charts (right column); original magnification, ×40. Photographs of Landolt rings taken of print sizes ranging from 0.1 to − 0.2 logMAR. Nominal sizes are given between the columns. The C-Test shows deficiencies in print quality and aberrations in the optotype heights, thicknesses of the lines, and widths of the openings. The shape of the Landolt rings and endings that configure the opening of the Landolt rings are distorted. The lines are blurred and frayed. The openings are too narrow; the heights and thicknesses of the lines are > 5% compared with the nominal values. The Landolt rings of the near vision charts from the RADNER charts are more accurate. The edges of the lines are sharper, and the endings that form the openings are in parallel. The heights are within the 5% tolerance, but the widths of the openings and lines lie outside that tolerance
Fig. 2Landolt ring near vision chart from Precision Vision (left column) and Landolt rings graphically optimized for print quality (right column; RADNER optimized); original magnification, ×40. Photographs of Landolt rings taken of print sizes ranging from 0.1 to − 0.3 logMAR. Nominal sizes are given between the columns. The Landolt ring near vision charts of Precision Vision show deficiencies in print quality and aberrations in the optotype heights, thicknesses of the lines, and widths of the openings. The endings that configure the opening of the Landolt rings are parallel from 0.0 to 0.1 logMAR, but the lines are blurred. The heights of the Landolt rings are within the 5% tolerance at all sizes. The openings are narrower, and the widths of the lines are wider than the nominal values (outside 5%). The Landolt rings optimized for printed quality show the best print quality. The edges of the lines are sharp, and the endings at the openings are clearly parallel. The Landolt rings closely match the nominal values in height and line width. Only two openings are outside the 5% tolerance, i.e., − 0.1 and 0.1 logMAR
Differences in percent from the calculated nominal values
| Parameters | C-Test | Landolt-C RADNER charts | Landolt-C ETDRS format | Landolt-C custom made |
|---|---|---|---|---|
| 0.1 logMAR | ||||
| Height | 1.90 | 1.52 | − 0.14 | 1.24 |
| Opening | − 18.49 | − 16.44 | − 10.27 | − 8.22 |
| Line thickness | 14.38 | 9.59 | 6.16 | 2.74 |
| 0.0 logMAR | ||||
| Height | − 0.52 | 0.34 | 0.00 | 0.00 |
| Opening | − 21.55 | − 15.52 | − 19.83 | − 3.45 |
| Line thickness | 18.97 | 11.21 | 11.21 | 2.59 |
| − 0.1 logMAR | ||||
| Height | 2.16 | 1.08 | 0.00 | 0.00 |
| Opening | − 14.13 | − 11.97 | − 17.29 | − 11.96 |
| Line thickness | 23.91 | 14.13 | 19.57 | 3.26 |
| − 0.2 logMAR | ||||
| Height | 15.23a | 3.27 | 1.36 | − 1.63 |
| Opening | − 8.22a | − 17.81 | − 17.81 | − 1.47 |
| Line thickness | 36.99a | 24.66 | 24.66 | 5.48 |
| − 0.3 logMAR | ||||
| Height | – | – | − 0.44 | − 1.71 |
| Opening | – | – | − 22.41 | − 10.44 |
| Line thickness | – | – | 24.14 | 3.45 |
aThe original visual acuity grade of the C-Test is − 0.15 logMAR, which causes a bigger difference in the calculated nominal values for − 0.2 logMAR