Under-deployed TAVI with occluded left-main and severe aortic regurgitation; bailout chimney-in-chimney stenting and aggressive postdilatation saves the day.

Pandula Athauda-Arachchi1,2.   

Abstract

Narrow calcific aortic roots with low coronary heights present challenges for a successful trans-catheter aortic valve replacement. We describe a case where pre-emptive coronary protection and chimney stent-in-stent procedure was successfully performed during a cardiac arrest secondary to coronary occlusion and restricted TAVI valve deployment in such an anatomy.
© 2022 The Author. Clinical Case Reports published by John Wiley & Sons Ltd.

Entities:  

Keywords:  TAVI under‐deployment; complication management; low‐lying left‐main coronary artery occlusion; paravalvular AR

Year:  2022        PMID: 35957777      PMCID: PMC9361718          DOI: 10.1002/ccr3.6224

Source DB:  PubMed          Journal:  Clin Case Rep        ISSN: 2050-0904


CLINICAL PRESENTATION AND INVESTIGATIONS

A 78‐year‐old woman had presented with a 4‐month history of worsening shortness of breath and presyncope with an ejection systolic murmur and was found to have critical aortic stenosis (Peak Gradient >120 mmHg, AVA <0.5 cm2), with a small annulus and LVOT(<18 mm) and mild–moderate aortic regurgitation with preserved LV systolic function (EF > 50%). ECG showed pre‐existing right bundle branch block but no evidence of second or third degree heart block on monitoring despite history of presyncope. Coronary angiogram demonstrated a left dominant coronary system with no significant obstructive disease. Patient was referred for cardiothoracic surgical opinion for surgical aortic valve replacement, but intermittent cognitive/ behavioral issues, diabetes, reduced mobility, and refusal by patient and family resulted in consideration for TAVI. Her calculated Euroscore II (mortality) for root enlargement and a SAVR was 10.99%. The CT of aorta confirmed the difficult anatomy for TAVI, with a small annulus of 16 × 19.6 mm and a low coronary height to dominant left main of 7 mm and small left coronary sinus(21.8 mm), with a potential for left‐main occlusion. Area (251.4 sqmm) derived annular diameter was 17.9 mm (see Figure 1). Common femoral arteries were small (6.8 mm). TAVR in hospital mortality risk was calculated at 2.17%. A further multidisciplinary discussion was offered and following this patient and family decision for TAVR was finally honored.
FIGURE 1

3‐Mensio reconstructions of preprocedure CT aorta, illustrating shallow left coronary sinus and low left coronary height of 7 mm (left dominant system)

INTERVENTIONS

Transfemoral TAVI with a Balloon expandable MyVal 20 mm device (smallest available device) with transradial coronary protection was performed under local anesthesia. Valve crossing was difficult as well as retaining of a Amplatz Extra stiff wire due to small LVOT and “whipping effect” with AR. A Lundequist wire gave better support and a 16 mm balloon was used for predilatation, which surprisingly did not demonstrate any tendency for coronary occlusion by native valve leaflets. Despite this, a coronary guidewire and a 4 × 12 mm Xience DES were parked as a precaution in the LAD. The 20 mm valve was deployed with 2 cc less in the navigator system, under rapid pacing. However, the TAVI valve failed to expand fully (see figure 2) with severe drop in pressures, AR, and slow flow in left main culminating in cardiac arrest. CPR was promptly commenced and the waiting stent in LAD was deployed in ostium of left main as a “Chimney stent” (see figure 3). This resulted in stabilization of situation. A root angiogram showed torrential AR and a decision was made to recross and postdilate the device with the navigator system with 2 cc extra volume. The concern was that the chimney may collapse on postdilatation against the shallow sinus. The initial coronary guidewire, which had, unfortunately, come off to the aorta during the resuscitation attempt now had to be recrossed quickly through the ostial LM stent, avoiding any TAVI or coronary stent struts. An additional 4 × 8 mm DES was parked beyond the previous stent and the optimal postdilatation of TAVI device was achieved with only a trace of residual AR. Due to the concern of compromise of the previous coronary stent in this limited space, a “chimney‐in‐chimney” stenting was performed with another 4 × 8 mm DES (see figure 4), postdilated to 4.5 mm, to fully expand the struts and maintain left‐main access and patency with better radial strengthening of the proximal portion of the chimney. The final result was excellent with no AR and good flow in left main coronary (See figure 5). Fluoroscopy and stent boost appearance was useful to confirm stent and valve expansion (see Figure 6). The femoral access was sealed with Proglide percutaneous sutures. Patient made an uneventful recovery Figures 1, 2, 3, 4, 5, 6.
FIGURE 2

Restricted expansion of TAVI device with left‐main occlusion (trace of flow restored only on engaging coronary guide whilst pulling stent forwards) and severe AR, leading to cardiac arrest

FIGURE 3

First Chimney stenting of Left main with 4 × 12 mm DES

FIGURE 4

Postdilatation of TAVI and chimney‐in‐chimney stenting(4 × 8 mm) for additional protection of the left main

FIGURE 5

Final appearance of left mainstem and TAVI device. Bulky leaflet buttressed under the chimney‐in‐chimney stents (black arrow)

FIGURE 6

TAVI device and the overlapping distal segments of stents visible (black arrow) under fluoroscopy

3‐Mensio reconstructions of preprocedure CT aorta, illustrating shallow left coronary sinus and low left coronary height of 7 mm (left dominant system) Restricted expansion of TAVI device with left‐main occlusion (trace of flow restored only on engaging coronary guide whilst pulling stent forwards) and severe AR, leading to cardiac arrest First Chimney stenting of Left main with 4 × 12 mm DES Postdilatation of TAVI and chimney‐in‐chimney stenting(4 × 8 mm) for additional protection of the left main Final appearance of left mainstem and TAVI device. Bulky leaflet buttressed under the chimney‐in‐chimney stents (black arrow) TAVI device and the overlapping distal segments of stents visible (black arrow) under fluoroscopy

DISCUSSION

Potential for occlusion of low‐lying left main during TAVI in native aortic stenosis could be anticipated (to an extent) based on the findings on a CT aorta. Performance of a BASILICA procedure or selection of new generation supra‐annular aortic valves that grasp the native valve leaflets may reduce the incidence of acute coronary occlusion, but the availability of these options is varied in different countries. Whether the predilatation may or may not show the tendency for coronary occlusion, provision of coronary protection with wire/stent could be life‐saving practice, but it may be difficult to implement in narrow anatomies with small annuli and small sinuses. When the TAVI devices may not fully expand and complications such as left‐main occlusion and a cardiac arrest occurs, rapid restoration of left‐main flow with chimney stenting is paramount and subsequent optimization of the TAVI prosthesis is mandatory. In narrow coronary sinuses, there is always a risk of the chimney stent becoming deformed or occluded, with poor outcomes perioperatively. We described our case where “chimney‐in‐chimney” stenting had to be performed following postdilatation of the TAVI device to optimize the left‐main blood supply. To our knowledge, this situation of chimney‐in‐chimney stenting has only very rarely been described in literature but could be vital to augment radial strength of the chimney in narrow anatomies in emergency situations as per our case. It is important to ensure good immediate outcomes by anticipation of potential complications, in setting up new structural heart intervention programmes to alleviate the growing heart failure epidemic in many countries with aging populations.

FOLLOW‐UP

Patient had substantial reduction of her transaortic gradients down to 28 mmHg (Peak) and 14 mmHg(mean) with only a trace of residual paravalvular AR. The pre‐existing conduction disturbance with intermittent brief episodes of heart block persisted, hence a backup permanent pacemaker was implanted. There was no chest pain/ischemia or LV dysfunction noted at 1 week. Patient remains on long‐term dual antiplatelet therapy.

CONCLUSIONS

Suboptimal deployment of a TAVI device in a narrow aortic annulus with proximity to left‐main origin could present a challenge in balancing patient safety from complications and optimization of device parameters. Pre dilatation test of aortic valve did not appear to predict potential of complications experienced in this TAVI procedure. Chimney stenting or chimney‐in‐chimney stenting of left main could be life‐saving in this instance but careful follow‐up of patients will be required in due course.

AUTHOR CONTRIBUTION

Sole author, responsible for procedure, all data, preparation, and submission of the manuscript.

CONFLICT OF INTEREST

None.

ETHICAL APPROVAL

Not applicable.

CONSENT

Written informed consent was obtained from the patient to publish this report in accordance with the journal's patient consent policy. Video S1 Click here for additional data file.
  4 in total

Review 1.  Imaging for Predicting, Detecting, and Managing Complications After Transcatheter Aortic Valve Replacement.

Authors:  Frank E Corrigan; Patrick T Gleason; Jose F Condado; John C Lisko; John H Chen; Norihiko Kamioka; Patricia Keegan; Sharon Howell; Stephen D Clements; Vasilis C Babaliaros; Stamatios Lerakis
Journal:  JACC Cardiovasc Imaging       Date:  2019-05

2.  Real-world experience using the ACURATE neo prosthesis: 30-day outcomes of 1,000 patients enrolled in the SAVI TF registry.

Authors:  Helge Möllmann; Christian Hengstenberg; Michael Hilker; Sebastian Kerber; Ulrich Schäfer; Tanja Rudolph; Axel Linke; Norbert Franz; Thomas Kuntze; Holger Nef; Utz Kappert; Thomas Walther; Michael O Zembala; Stefan Toggweiler; Won-Keun Kim
Journal:  EuroIntervention       Date:  2018-02-02       Impact factor: 6.534

3.  The BASILICA Trial: Prospective Multicenter Investigation of Intentional Leaflet Laceration to Prevent TAVR Coronary Obstruction.

Authors:  Jaffar M Khan; Adam B Greenbaum; Vasilis C Babaliaros; Toby Rogers; Marvin H Eng; Gaetano Paone; Bradley G Leshnower; Mark Reisman; Lowell Satler; Ron Waksman; Marcus Y Chen; Annette M Stine; Xin Tian; Danny Dvir; Robert J Lederman
Journal:  JACC Cardiovasc Interv       Date:  2019-06-12       Impact factor: 11.195

4.  Chimney Stenting for Coronary Occlusion During TAVR: Insights From the Chimney Registry.

Authors:  Federico Mercanti; Liesbeth Rosseel; Antoinette Neylon; Rodrigo Bagur; Jan-Malte Sinning; Georg Nickenig; Eberhard Grube; David Hildick-Smith; Davide Tavano; Alexander Wolf; Giuseppe Colonna; Azeem Latib; Satoru Mitomo; Anna Sonia Petronio; Marco Angelillis; Didier Tchétché; Chiara De Biase; Marianna Adamo; Mohammed Nejjari; Franck Digne; Ulrich Schäfer; Nicolas Amabile; Guy Achkouty; Raj R Makkar; Sung-Han Yoon; Ariel Finkelstein; Danny Dvir; Tara Jones; Bernard Chevalier; Thierry Lefevre; Nicolo Piazza; Darren Mylotte
Journal:  JACC Cardiovasc Interv       Date:  2020-03-23       Impact factor: 11.195

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.