Laboratory and field evaluation of three desktop instruments for assay of cholesterol and triglyceride.

H von Schenck, L Treichl, B Tilling, A G Olsson.   

Abstract

We evaluated three desktop instruments suitable for decentralizing the assays of blood cholesterol and triglycerides to satellite and physician's office testing. The instruments, Ektachem DT 60 (E), Reflotron (R), and Seralyzer (S), were used according to their manufacturers' instructions to assay serum or capillary blood from outpatients at two physicians offices. Accuracy was assessed in the centralized laboratory by using an automated centrifugal analyzer (A). The bias of A was monitored with an international quality-control material. We found all instruments easy to handle. Regression equations for cholesterol determinations were: E = 0.92A + 0.7 (n = 331, r = 0.94), R (capillary blood) = 0.96A + 0.3 (n = 256, r = 0.95), and S = 0.93A + 0.6 (n = 260, r = 0.92). For triglycerides we obtained E = 1.02A (n = 331, r = 0.97), R (cap. blood) = 0.88A (n = 213, r = 0.97), R = 0.94A + 0.1 (n = 90, r = 0.99), and S = 0.96A (n = 266, r = 0.98). Duplicate and within-day precision was less than 8%. Between-day precision (during a month) was less than 10%. We stress the need of both laboratory and field evaluation and emphasize the benefit of quality control.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3594854

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  6 in total

1.  Performance of the Reflotron in Massachusetts' Model System for Blood Cholesterol Screening Program.

Authors:  S Havas; R Bishop; L Koumjian; J Reisman; S Wozenski
Journal:  Am J Public Health       Date:  1992-03       Impact factor: 9.308

2.  Precision and accuracy of a portable blood analyzer system during cholesterol screening.

Authors:  P Greenland; N L Bowley; C A French; B Meiklejohn; S Gagliano; C E Sparks
Journal:  Am J Public Health       Date:  1990-02       Impact factor: 9.308

3.  Comparison of the use of a dry chemistry analyser in primary care in Norway and the United Kingdom.

Authors:  G Thue; S Sandberg; D G Bullock
Journal:  Br J Gen Pract       Date:  1993-01       Impact factor: 5.386

4.  Is there an association between shift work and having a metabolic syndrome? Results from a population based study of 27,485 people.

Authors:  B Karlsson; A Knutsson; B Lindahl
Journal:  Occup Environ Med       Date:  2001-11       Impact factor: 4.402

5.  Increasing glucose concentrations and prevalence of diabetes mellitus in northern Sweden, 1990-2007.

Authors:  Bernt Lindahl; Hans Stenlund; Margareta Norberg
Journal:  Glob Health Action       Date:  2010-10-27       Impact factor: 2.640

6.  A reversal of decreasing trends in population cholesterol levels in Västerbotten County, Sweden.

Authors:  Nawi Ng; Owe Johnson; Bernt Lindahl; Margareta Norberg
Journal:  Glob Health Action       Date:  2012-03-23       Impact factor: 2.640

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.