Meghan Reading Turchioe1, Rezwan Ahmed2, Ruth Masterson Creber3, Kelly Axsom1, Evelyn Horn3, Gabriel Sayer1, Nir Uriel1, Kenneth Stein2, David Slotwiner3.
Abstract
Background: Cardiac implantable electronic devices (CIEDs) may enable early identification of COVID-19 to facilitate timelier intervention. Objective: To characterize early physiologic changes associated with the onset of acute COVID-19 infection, as well as during and after acute infection, among patients with CIEDs.Entities:
Keywords: COVID-19; Cardiac pacemaker; Implantable defibrillators; Physiologic monitoring
Year: 2022 PMID: 35942055 PMCID: PMC9349024 DOI: 10.1016/j.cvdhj.2022.07.070
Source DB: PubMed Journal: Cardiovasc Digit Health J ISSN: 2666-6936
Figure 1Timeline of COVID-19 test dates during the study period among patients included in the study (n = 286).
Figure 2Sampling strategy of patient cohorts: COVID-positive (n = 20), COVID-negative (n = 166), and untested controls (n = 100).
Demographic characteristics of COVID-positive, COVID-negative, and control patient cohorts
| COVID-positive (n = 20) | COVID-negative (n = 166) | Control (n = 100) | ||
|---|---|---|---|---|
| Age | 68.2 (13.3) | 71.3 (15.1) | 67.9 (14.8) | .18 |
| Sex | .31 | |||
| Male | 10 (50) | 98 (59) | 61 (61) | |
| Female | 10 (50) | 68 (41) | 37 (37) | |
| Other/unknown | 0 (0) | 0 (0) | 2 (2) | |
| Race | .04∗ | |||
| White | 5 (25) | 86 (52) | 42 (42) | |
| Black / African American | 4 (20) | 30 (18) | 19 (19) | |
| Asian | 1 (5) | 14 (8) | 10 (10) | |
| Other | 7 (35) | 23 (14) | 11 (11) | |
| Unknown | 3 (15) | 13 (8) | 18 (18) | |
| Ethnicity | .004∗ | |||
| Hispanic/Latino | 9 (45) | 32 (19) | 11 (11) | |
| Non-Hispanic/Latino | 9 (45) | 107 (65) | 64 (64) | |
| Unknown | 2 (10) | 27 (16) | 25 (25) | |
| BMI | 30.3 (7.0) | 27.8 (5.4) | 27.0 (5.1) | .09 |
| Unknown | 3 (15) | 48 (29) | 43 (43) | |
| Heart failure | 7 (35) | 74 (45) | 36 (36) | .48 |
| Unknown | 0 | 10 | 7 | |
| Device type | <.001∗ | |||
| CRT | 3 (15) | 35 (21) | 28 (28) | |
| ICD | 13 (65) | 48 (29) | 47 (47) | |
| Pacemaker | 4 (20) | 83 (50) | 25 (25) | |
| Medicine admission | 16 (80) | 53 (32) | --- | <.001∗ |
| Unknown | 0 (0) | 12 (7) | --- | |
| ICU admission | 0 (0) | 5 (3) | --- | .86 |
| Unknown | 3 (18) | 22 (13) | --- | |
| Death | 0 (0) | 0 (0) | --- | .37 |
| Unknown | 0 (0) | 12 (7) | --- |
Values are mean (SD) or n (%) test for categorical measures.
Statistically significant P values are indicated by an asterisk.
BMI = body mass index; CRT = cardiac resynchronization therapy; ICD = implantable cardioverter-defibrillator; ICU = intensive care unit.
P values were calculated using ANOVA for continuous measures and χ2
Refers to events at the time of the COVID-19 test. Medicine admissions exclude surgical admissions for which routine preoperative COVID-19 testing was conducted. Not evaluated for the control group, who did not have a COVID-19 test.
Figure 3Average sensor changes among COVID-positive (n = 20), COVID-negative (n = 166), and control (n = 100) patient cohorts.
Average number of days prior to COVID-19 testing with significant changes from baseline, comparing COVID-positive to COVID-negative patients
| Sensors | COVID-positive (n = 20) | COVID-negative (n = 166) | |||
|---|---|---|---|---|---|
| N | Number of days (mean ± SD) | N | Number of days (mean ± SD) | ||
| Respiratory rate | 14 | -8.5 ± 8.5 | 113 | -3.9 ± 7.5 | .01∗ |
| Activity | 18 | -8.2 ± 8.3 | 154 | -3.5 ± 6.1 | .008∗ |
| Night heart rate | 8 | -2.1 ± 2.8 | 35 | -2.3 ± 4.2 | .80 |
| 24-Hour heart rate | 15 | -3.5 ± 4.3 | 98 | -3.3 ± 5.1 | .62 |
| Temperature | 17 | -2.3 ± 2.7 | 139 | -1.0 ± 1.5 | .08 |
| RSBI | 8 | -3.1 ± 4.0 | 35 | -4.3 ± 8.9 | .75 |
| Thoracic impedance | 8 | -7.1 ± 11.1 | 35 | -11.4 ± 11.3 | .29 |
| Heart sound S1 | 8 | -2.6 ± 4.6 | 34 | -3.0 ± 5.5 | .80 |
| Heart sound S3 | 8 | -2.6 ± 3.5 | 33 | -1.3 ± 2.6 | .29 |
| HeartLogic index | 8 | -16.4 ± 12.5 | 33 | -9.2 ± 12.6 | .08 |
Mann-Whitney U test (Wilcoxon rank sum test) was used to compare COVID-positive to COVID-negative.
Statistically significant P values are indicated by an asterisk.
RSBI = rapid shallow breathing index.
N represents the number of patients with devices containing specified sensors and with valid sensor value during baseline and test period.
Temperature represents the pulse generator temperature at the location of the implant.
Odds ratios and 95% confidence intervals for mean cardiac implantable electronic device sensor changes during the 7 days prior to COVID-19 testing associated with COVID-19 test status (positive vs negative patients), unadjusted and adjusted for age, sex, race, and device type
| Sensor | Unadjusted OR (95% CI) | Adjusted OR (95% CI) |
|---|---|---|
| Respiratory rate | 1.75 (1.29–2.44)∗ | 2.31 (1.33–5.13)∗ |
| Activity | 0.36 (0.15–0.85)∗ | 2.88 (0.32–27.4) |
| Night heart rate | 1.06 (0.92–1.21) | --- |
| 24-Hour heart rate | 1.01 (0.90–1.14) | --- |
| Temperature | 2.15 (1.40–3.49)∗ | 2.96 (0.97–10.99) |
| RSBI | 1.36 (0.99–1.96) | --- |
| Thoracic impedance | 0.99 (0.85–1.19) | --- |
| Heart sound S1 | 0.53 (0.05–2.98) | --- |
| Heart sound S3 | 212.90 (0.12–833264.27) | --- |
| HeartLogic index | 1.09 (1.00–1.22) | --- |
Odds ratios indicated by asterisks are statistically significant. Only variables with significant associations with the outcome variable in unadjusted models were included in adjusted models.
CI = confidence interval; OR = odds ratio; RSBI = rapid shallow breathing index.
Temperature represents the pulse generator temperature at the location of the implant.
Average number of days after COVID-19 testing that sensors returned to baseline, comparing COVID-positive to COVID-negative patients
| Sensors | COVID-positive (n = 20) | COVID-negative (n = 166) | |||
|---|---|---|---|---|---|
| N | Number of days (mean ± SD) | N | Number of days (mean ± SD) | ||
| Respiratory rate | 14 | 13.0 ± 16.5 | 112 | 4.0 ± 9.6 | .006∗ |
| Activity | 18 | 18.2 ± 20.5 | 154 | 6.1 ± 12.4 | .007∗ |
| Night heart rate | 8 | 4.5 ± 8.3 | 35 | 4.5 ± 12.0 | .67 |
| 24-Hour heart rate | 15 | 6.7 ± 10.4 | 98 | 5.1 ± 11.5 | .23 |
| Temperature | 17 | 1.3 ± 1.3 | 139 | 1.5 ± 2.6 | .35 |
| RSBI | 8 | 2.4 ± 3.1 | 35 | 3.7 ± 10.9 | .67 |
| Thoracic impedance | 8 | 12.8 ± 20.5 | 35 | 11.9 ± 16.5 | .75 |
| Heart sound S1 | 8 | 13.1 ± 21.1 | 34 | 4.6 ± 13.4 | .79 |
| Heart sound S3 | 8 | 3.1 ± 3.2 | 33 | 2.4 ± 6.5 | .23 |
| HeartLogic index | 8 | 22.4 ± 21.5 | 33 | 13.5 ± 22.0 | .08 |
Mann-Whitney U test (Wilcoxon rank sum test) was used to compare COVID-positive to COVID-negative.
RSBI = rapid shallow breathing index.
N represents the number of patients with devices containing specified sensors and with valid sensor value during baseline and test period.
Temperature represents the pulse generator temperature at the location of the implant.