A cheap, versatile, readily modified, and reusable glass probe system enabling delivery of solid air-/moisture-sensitive samples for mass spectrometric (MS) analysis using an Atmospheric pressure Solids Analysis Probe (ASAP) is described. The simplicity of the design allows quick and easy ASAP MS analyses of sensitive solid and liquid samples without the need for any modifications to commercially available vertically loaded ASAP mass spectrometers. A comparison of ASAP mass spectra obtained for metal complexes under air and an inert atmosphere is given.
A cheap, versatile, readily modified, and reusable glass probe system enabling delivery of solid air-/moisture-sensitive samples for mass spectrometric (MS) analysis using an Atmospheric pressure Solids Analysis Probe (ASAP) is described. The simplicity of the design allows quick and easy ASAP MS analyses of sensitive solid and liquid samples without the need for any modifications to commercially available vertically loaded ASAP mass spectrometers. A comparison of ASAP mass spectra obtained for metal complexes under air and an inert atmosphere is given.
Atmospheric pressure Solids
Analysis Probe Mass Spectrometry (ASAP
MS) is a direct ionization MS technique that exploits the many recent
advances in ion generation at ambient pressure. This enables native
sample analysis, removing the need for sample preparation through
solvation or functionalization.[1] In ASAP
MS, the sample is subject to elevated temperatures in an external,
heated, nitrogen-filled vaporization/ionization chamber at ambient
pressure. The resulting vaporized molecules are then singly ionized
(via charge transfer) before being introduced into a mass analyzer.[1] This sequence circumvents problems often found
with traditional MS techniques where ionization is achieved within
the spectrometer, namely, ionization sensitivity and chemical specificity.[2,3] Additionally, not only can ASAP MS be applied to both solid and
liquid samples, it enables analysis of materials of low volatility.
For example, low molecular weight oligomeric materials are increasingly
being analyzed by ASAP MS in preference to other ionization techniques,
such as MALDI or ESI, since using ASAP MS avoids complications that
can result from the need to add a metal cation to promote ionization.[4,5] ASAP MS has also been shown to be particularly useful in the analysis
of homogeneous samples that suffer from solvent suppression (dampening
of the sample signal due to interactions with the solvent) in electrospray
ionization (ESI) or atmospheric pressure chemical ionization (APCI).[5,6] Together with the features described above, the fact that the ASAP
MS technique requires no prior sample preparation through dissolution
in solvent makes this a particularly attractive method for the analysis
of air-/moisture-/solvent-sensitive samples, with the need for rigorously
dried and degassed solvents being eliminated.While the sample
analysis ionization system of a mass spectrometer
is under a completely inert atmosphere (high vacuum or fed with a
dry, inert gas supply), often the method by which the analyte sample
is introduced necessitates brief exposure of the sample to air. This
is a significant limitation for the analysis of air-/moisture-sensitive
samples, and hence, a number of methods have been developed for a
range of mass spectrometry (MS) ionization techniques including liquid
injection field desorption ionization (LIFDI),[7,8] electrospray
ionization (ESI),[9−11] electron impact (EI),[12,13] and matrix-assisted
laser desorption/ionization (MALDI).[14,15] This has been
extended to ASAP MS, leading to the development of so-called iASAP MS, which has included the commercialization of a
specialized dedicated attachment by Advion, Inc. working in collaboration
with Krossing and co-workers.[16] More recently,
a related approach, “Paraffin-Inert ASAP” was been reported
by Afonso, Giusti, and co-workers, which involves sealing a capillary
loaded with sample with paraffin in a glovebox to protect the sample
during transport to the machine.[17] Building
on our prior work around developing methods for the analysis of air-/moisture-sensitive
samples via ASAP MS,[18] we describe herein
the design and use of a cheap, simple, and reusable glass probe delivery
sample system whose configuration can be very easily modified to fit
any vertically loading ASAP mass spectrometer providing an alternative
methodology for analysis of environmentally sensitive solid and liquid
materials. This approach is straightforward, readily applicable, and
requires no alterations to the Waters mass spectrometry instruments
used in this study and can easily be modified for use with other instruments.
Experimental Section
Materials
[Pd(acac)2] (1) was
obtained from Alfa Aesar and used without further purification. 1,2-bis(di-tert-butylphosphinomethyl)benzene
was obtained from Lucite International and was recrystallized before
use. Complexes 2(24) and 3(25,26) were prepared in accordance to literature
procedures.
General Considerations
Air-sensitive manipulations
were carried out under an atmosphere of dry nitrogen using standard
Schlenk-line and glovebox (Saffron Scientific; Innovative Technologies)
techniques. Hexanes were dried using Innovative Technologies Solvent
Purification System facilities. Tetrahydrofuran (THF) was dried by
heating at reflux over sodium wire with a benzophenone indicator,
followed by distillation. All solvents were degassed prior to use
using freeze–pump–thaw techniques. ASAP (Atmospheric
pressure Solids Analysis Probe) mass spectra were acquired with an
LCT Premier XE mass spectrometer or Waters Xevo QToF mass spectrometer.
Mass spectrometer conditions for ASAP analysis using the glass probe
delivery system were optimized using a series of air-stable organometallic
complexes. Details of optimization experiments can be found in the Supporting Information. Infrared spectra were
recorded in the solid state using a PerkinElmer Frontier ATR-FTIR
instrument.
Synthesis of CoBr2(dtbpx) (4)
To a solution of cobalt(II) bromide (0.27 g, 1.24 mmol) in THF (10
mL), 1,2-bis(di-tert-butylphosphinomethyl)benzene
(0.50 g, 1.28 mmol) in THF (15 mL) was added dropwise. The reaction
was stirred at room temperature for 16 h after which time precipitation
was mediated through addition of hexanes (20 mL). The solids were
collected via filtration and washed with hexanes (3 × 10 mL)
before drying to yield 4 as a turquoise powder (0.69
g, 1.12 mmol, 90%) (found: C, 46.9; H, 7.15. C24H44Br2CoP2 requires: C, 47.0; H, 7.20%). MS (ASAP+, anaerobic) m/z 532.15
([M-Br]+, 3%), 395.31 (100). IR: ν cm–1 3064w (C–H), 2972–2868s (C–H), 1601w (C=C),
1472–1370s (C–H). Raman (532 nm): ν cm–1 282m (Co–Br), 580s (C–C), 690m (C–P), 812m
(C–C), 942m (C=C), 1027w (C–C), 1057m (C=C),
1235s (CH2), 1476m (CH3) and 1603m (C=C).
Silent to 31P NMR spectroscopy. UV–vis (THF): λmax 658, 699, and 744 nm.
Mass Spectrometric Measurement
An open-ended glass
capillary (50 mm × 1 mm) is dipped into a powdered sample to
give a plug of height 1 mm and placed within the glass delivery probe
(either within a glovebox or open to air depending on sample), and
both PTFE taps on the glass delivery probe closed such that the sample
is isolated. The probe is connected to an N2 gas supply
set to a flow rate of 5, 10, 15, or 20 mL min–1 (controlled
by Dywer Variable Area Flowmeter) and purged via the side arm for
20 s. The glass delivery probe is connected to an LCT Premiere XE
or Waters QToF Xevo mass spectrometer through the Delrin holder to
the standard ASAP adaptor and time allowed for the probe tip to reach
temperature (∼1 min). Sample measurement conditions are set
using MassLynx software and data collection started for a 2 or 3 min
run. Here, 10–30 s into data collection, the PTFE stopcocks
on the glass delivery probe are opened simultaneously to allow the
capillary
tube containing the sample to drop into the ionization chamber of
the mass spectrometer; the sample exits the probe through the sample
release hole allowing necessary ionization prior to data collection.
Results and Discussion
ASAP Sample Glass Delivery Probe Development
Our previous preliminary report on the analysis of air-/moisture-sensitive
materials using ASAP MS demonstrated successful analysis of two highly
air-sensitive materials through application of a simple inert atmosphere
sample introduction method.[18] This comprised
sealing the analyte sample in a glass capillary tube inside an inert
atmosphere glovebox, transporting the sealed tube to the spectrometer,
and finally introducing the analyte into the spectrometer by using
a rotating baffle (originally intended to divert an internal calibrant
into the spectrometer) to break the sealed capillary located inside
the external, heated vaporization/ionization chamber. Although simple
to achieve, this approach could potentially allow fine particle debris
(e.g., broken glass shards) to block the inlet or contaminate the
first stages of the API interface of the mass spectrometer. This can
only be limited by careful cleaning between samples, something that
necessitates repeated exposure of the chamber to ambient atmosphere
conditions. Additionally, this simple delivery method for the introduction
of air-/moisture-sensitive samples for ASAP MS does, albeit briefly,
result in the exposure of the analyte to air. For many air-/moisture-sensitive
complexes, even this brief exposure is sufficient for oxidation and/or
hydrolysis to occur.The reusable sample delivery probe protocol
described herein for analysis of air-/moisture-sensitive samples by
ASAP MS is based upon the standard method used in vertically loaded
ASAP instruments. To obtain a mass spectrum using ASAP MS for an air
stable sample, a small quantity of the solid or liquid analyte is
loaded onto the end of a capillary tube. This capillary is placed
within a holder, which in turn is then placed within the heated vaporization/ionization
chamber. This chamber is kept at slightly above atmospheric pressure
with nitrogen gas and at an elevated temperature (300–650 °C).
Application of a corona discharge within the chamber ionizes the nitrogen
gas,[19,20] which then transfers charge to the analyte
sample, with the resulting singly charged ions then entering the sample
block, extraction cone, ion optics, and finally into the TOF chamber
via a sample cone (that has an applied voltage gradient).To
circumvent sample contact with air and/or moisture, a simple
low cost “glass delivery probe” was designed. This allows
transportation and delivery of liquid or solid samples from a standard
glovebox directly into the vaporization/ionization chamber of a standard
commercial vertically loading ASAP mass spectrometer without exposure
to the atmosphere (Figure ). The necessary glass probes and Delrin probe holder were
manufactured in house to specifications to fit two commercial vertically
loading mass spectrometers (LCT Premiere XE ToF and Water’s
Xevo QToF mass spectrometers) capable of ASAP analysis. In a glovebox,
a sample-dipped glass capillary is sealed in the “glass delivery
probe” under an inert atmosphere between two closed PTFE stopcocks
for transportation to the mass spectrometer. The end of the “glass
delivery probe” described herein is then inserted into the
instrument such that the tip of the probe resides within the heated
vaporization/ionization chamber. The side arm and the three-way PTFE
tap arrangement allow purging of the lower end of the probe with an
inert gas (Figure a). Subsequently, to allow sample analysis, a flow of inert gas from
the top of the probe (vertical gas flow) is maintained, and both PTFE
taps are opened allowing the sample-loaded capillary to drop down
into the base of the probe (located within the heated zone) where
the analyte is vaporized and then ionized by a corona discharge prior
to entry to the TOF system as normal (Figure b).
Figure 1
Waters Xevo QToF mass spectrometer fitted with
a vertical loading
ASAP inlet port with a glass sample delivery probe in position.
Figure 2
Simplified representation of ASAP glass delivery probe:
(a) during
purge of the probe and (b) for analysis (top). Schematic representation
of the inside of the ionization chamber showing relative positions
of desolvation gas (N2) injection system, corona discharge,
and sample cone (bottom).
Waters Xevo QToF mass spectrometer fitted with
a vertical loading
ASAP inlet port with a glass sample delivery probe in position.Simplified representation of ASAP glass delivery probe:
(a) during
purge of the probe and (b) for analysis (top). Schematic representation
of the inside of the ionization chamber showing relative positions
of desolvation gas (N2) injection system, corona discharge,
and sample cone (bottom).The use of a reusable glass delivery probe to deliver
the air-/moisture-sensitive
samples into the instrument allows quick and easy ASAP mass spectrometric
analysis to be performed with no alterations to existing mass spectrometers.
Using this methodology sample turnaround is only limited by cleaning/drying
of the probe(s) between each use.
Sample Introduction: Comparison of Capillary
vs Glass Delivery Probe ASAP Methods
In the first instance,
the air-stable coordination and organometallic complexes [Pd(acac)2] (1) and [Rh(COD)(tmh)] (COD = 1,5-cyclooctadiene,
tmh = 2,2,6,6-tetramethyl-3,5-heptanedionato) (2), as
well as moisture-sensitive [CoBr2(ADICy)] (3) (ADICy = N,N-bis(cyclohexyl)-1,2-diimino-1,2-dimethylethane)
and air- and moisture-sensitive [CoBr2(dtpbx) (4) (dtbpx = 1,2-bis(di-tert-butylphosphinomethyl)benzene)
(Figure ) were selected
to allow direct comparison between traditional (where the sample is
exposed to the atmosphere), capillary-loaded, and “glass delivery
probe”-loaded ASAP MS spectra. These samples were also employed
in mass spectrometer optimization experiments. In all the following
ASAP MS experiments where the glass delivery probe is applied, the
probe was allowed to “warm up” for 30 s after introduction
to the ionization chamber before the sample is “dropped”.
Figure 3
Representative
coordination and organometallic complexes (1–4) tested for analysis by ASAP MS using
capillary or glass probe delivery methods.
Representative
coordination and organometallic complexes (1–4) tested for analysis by ASAP MS using
capillary or glass probe delivery methods.The original design of the “glass delivery
probe”
included a single sample release hole positioned so as to face the
sample cone (i.e., ToF system inlet) to maximize the amount of sample
being introduced into the ToF analyzer (Figure a, b; top). However, this arrangement results
in the corona discharge pin facing glass, which is found to limit
the intensity of the resulting signal. We hypothesized that introduction
of a second sample release hole in the probe tip would provide a greater
means of escape for the vaporized sample as a result of direct impingement
of the corona discharge upon the sample within the probe and hence
greater MS signal intensity. However, no significant increase to the
signal intensity in the total ion count (TIC) measured for a sample
of 3 ([M–CoBr2]+ signal)
was observed moving from one to two sample release holes (cf. TIC
intensity ∼20,000 and ∼12,000 for one and two sample
release holes, respectively). It is important to note that in all
the ASAP MS experiments undertaken using the glass delivery probe
method, the TIC signal intensity obtained was significantly lower
than that obtained when using the traditional dipped capillary-loaded
ASAP MS (Figure ),
but the TIC signal intensity remained at a level that did not impact
on analysis performance. Indeed, this reduction to the intensity in
the signal observed may be an advantage due to reducing the likelihood
of overloading the MS ion detector.[21]
Figure 4
Comparison
of total ion count (TIC) signals obtained using ASAP
(top) and glass delivery probe ASAP (bottom) MS techniques with a
sample of [CoBr2(dtbpx)] (4).
Comparison
of total ion count (TIC) signals obtained using ASAP
(top) and glass delivery probe ASAP (bottom) MS techniques with a
sample of [CoBr2(dtbpx)] (4).The signal attenuation found when using the glass
delivery probe
ASAP MS arrangement has been attributed to a cooling effect caused
by the “sweep”/purge gas flow applied to the probe,
which results in less efficient volatilization of the solid sample
in the ASAP chamber. For example, at a high (15 mL/min) purge gas
flow rate, a significantly reduced TIC signal intensity is measured
compared to that observed with a gas purge flow rate of 10 mL/min
due to this cooling effect (Figure ). However, further reduction to the purge gas flow
rate to 5 mL/min shows an attenuation in TIC signal intensity, suggesting
that the purge gas flow also aids release of the sample to the ionization
chamber where it can then be vaporized and analyzed (characterized
by a sharp TIC signal with high intensity immediately upon sample
introduction).
Figure 5
Comparison of total ion count (TIC) signals using the
glass delivery
probe ASAP MS technique with various vertical N2 gas flow
rates at 450 °C from a sample of [CoBr2(ADICy)] (3): top, 5 mL/min; middle, 10 mL/min; bottom, 15
mL/min vaporization/ionization chamber.
Comparison of total ion count (TIC) signals using the
glass delivery
probe ASAP MS technique with various vertical N2 gas flow
rates at 450 °C from a sample of [CoBr2(ADICy)] (3): top, 5 mL/min; middle, 10 mL/min; bottom, 15
mL/min vaporization/ionization chamber.Physical sample ejection from the probe/capillary
is further inferred
from the observation of sprayed/deposited samples upon the walls,
door, and sample cone of the vaporization/ionization chamber post
analysis when very high (>15 mL/min) purge gas flow rates are used,
which resulted in the bulk of the sample being outside of the main
heated zone and thus not undergoing efficient vaporization. Thus,
a moderate purge gas flow rate is therefore optimal, such that cooling
effects are minimized, and gas-entrained physical sample ejection
from the probe is enhanced. Notably, however, the signal intensity
of MS TIC obtained when using the glass probe delivery system can
be increased to an intensity comparable to that of a sample run under
standard ASAP conditions (loaded capillary) simply by increasing the
temperature of the vaporization gas within the purge gas flow. This
temperature increase combats the cooling effects of the purge gas
flow. Varying the sample cone voltage used for both traditional capillary-loaded
and “glass delivery probe” ASAP MS experiments was found
to have an identical and significant impact upon the fragmentation
observed within the resulting mass spectra obtained as reported previously.[22] Thus, standard cone voltages (e.g., 30 V) were
used for all experiments.The ASAP mass spectra obtained following
introduction of the sample
into the spectrometer using the glass delivery probe under anaerobic
conditions are comparable to those obtained with standard dipped capillary
loading. For example, Figure compares the spectra obtained from complex 3 with dipped capillary and glass delivery probe sample loading methods.
Of note, both techniques show the [M–Br]+ ion as
the highest intensity (100%) ion present. In addition, the spectra
resulting from both loading techniques show some recombination of
ions within the mass spectrometer generating the ion corresponding
to [2M–Br]+ at m/z 851.0 Da (Figure ), as a result of a high concentration of sample. When the glass
delivery probe sample introduction method is used to load the sample,
the relative intensity of this dimer (recombination) product (compared
to the [M–Br]+ ion) produced within the mass spectrometer
is reduced relative to when the sample is introduced using the dipped
capillary approach (cf. 16% with dipped capillary and 3% with glass
delivery probe). It is proposed that this difference is the result
of a smaller amount of sample being released into the ion source region
(as indicated by a reduced TIC) when the glass probe delivery method
is used, vide supra. It is likely that lower sample concentration
will result in a corresponding reduction in the production of in-source
recombination products such as [2M–Br]+ observed
for complex 3, an effect that has been observed upon
reducing sample concentration during ESI-MS analysis.[23]
Figure 6
Comparison of the ASAP mass spectra obtained for a sample of [CoBr2(ADICy)] (3) using a traditional dipped
capillary (450 °C; top) and glass sample delivery probe (500
°C; N2 flow rate = 10 mL/min; bottom) sample introduction
methods. For both ionization methods, m/z 851.0 corresponds to [2M–Br]+. Insets show experimental
isotope patterns.
Comparison of the ASAP mass spectra obtained for a sample of [CoBr2(ADICy)] (3) using a traditional dipped
capillary (450 °C; top) and glass sample delivery probe (500
°C; N2 flow rate = 10 mL/min; bottom) sample introduction
methods. For both ionization methods, m/z 851.0 corresponds to [2M–Br]+. Insets show experimental
isotope patterns.
ASAP MS of an Air-Sensitive Sample
Having demonstrated that mass spectra of comparable quality could
be obtained using both standard dipped capillary and glass delivery
probe loading during ASAP MS experiments, the glass delivery probe
was tested for the ASAP MS analysis of air-sensitive samples. As an
example of the benefits of the glass probe-loaded sample over traditional
capillary loading, the mass spectra obtained of an air-sensitive,
reaction product mixture (solid state) in which the predominant species
is [CoBr2(dtbpx)] (4) (as determined by Raman,
UV–vis, and NMR spectroscopies) using both sample introduction
methods are shown in Figure .
Figure 7
Comparison of the ASAP mass spectra obtained from a sample of an
as-obtained reaction mixture (solid state) containing [CoBr2(dtbpx)] (4) introduced via a standard dipped capillary
(top), m/z 411.3 Da corresponds
to [M–CoBr2+OH]+, and using a glass sample
delivery probe (bottom), m/z 395.3
Da corresponds to [M–CoBr2+H]+. Insets
show experimental isotope patterns.
Comparison of the ASAP mass spectra obtained from a sample of an
as-obtained reaction mixture (solid state) containing [CoBr2(dtbpx)] (4) introduced via a standard dipped capillary
(top), m/z 411.3 Da corresponds
to [M–CoBr2+OH]+, and using a glass sample
delivery probe (bottom), m/z 395.3
Da corresponds to [M–CoBr2+H]+. Insets
show experimental isotope patterns.From the ASAP MS spectrum obtained using the glass
sample delivery
probe, there is a significantly lower degree of oxidation of the diphosphine
(dtbpx) (cf. 15% vs 70% oxidation of [M-CoBr2]+) observed as shown by the highest intensity ion at m/z 395.3 Da, which corresponds to [M–CoBr2+H]+. Here, the strong signal associated with this
ligand results from the ionization process, as no free phosphine is
observed in the corresponding 31P NMR spectrum (THF) obtained
of this crude mixture. In contrast, the ASAP MS spectrum obtained
from the reaction sample introduced using a dipped capillary shows
the highest intensity ion as being the partial oxidation product [M–CoBr2+OH]+ at m/z 411.3
Da. Further, use of the glass delivery probe allows observation of
the [M–Br]+ ion of [CoBr2(dtbpx)] (4) at m/z 532.1 Da within
the resulting ASAP-MS spectrum, whereas this ion signal is very weak
when the same sample is analyzed using the dipped capillary technique
(Figure ). It is also
noted that use of the dipped capillary gives rise to an additional,
but weak, signal corresponding to [M–Br+O]+ at m/z 548.1 Da (calc. m/z 548.1 Da). Irrespective of the sample introduction method,
the expected molecular ion (m/z 611.1
Da) arising from ionization of 4 is not observed.
Conclusions
This work describes a very
cheap, straightforward, easily manufactured
and modified, reusable delivery system to enable the analysis of air-/moisture-sensitive
samples by ASAP MS. This method can be viewed as an alternative to iASAP methods reported previously.[16−18] The use of
a simple “glass probe” to transport and deliver samples
from a glovebox to an ASAP-capable mass spectrometer avoids the need
to either house an MS instrument or MS sample introduction inlet in
a glovebox while maintaining inert conditions and is beneficial in
terms of cost (cf. glovebox-coupled MS[9,10]), does not
require deconvolution of the resulting spectra (cf. paraffin inert
ASAP–piASAP[17]), and removes the
risks associated with broken glass fragments collecting in the mass
spectrometer source (cf. our previous preliminary report on air-sensitive
ASAP MS[18]).It was found that a higher
temperature is required to achieve satisfactory
vaporization of the coordination and organometallic complexes tested
for ASAP MS analysis when using the glass probe delivery system (450–550
°C) compared to traditional dipped capillary ASAP MS analysis
(350–450 °C). This is necessitated by the need to compensate
for the cooling effect of the flow of nitrogen purge gas through the
glass probe into the ionization chamber. The TIC in mass spectra generated
while using the glass delivery probe is lower than that obtained under
standard ASAP conditions. However, the signal obtained when the glass
delivery probe is employed remains sufficient for confident mass spectral
assignments to be made. The ASAP mass spectra obtained when using
the glass delivery probe are comparable to those obtained using a
capillary-loaded ASAP MS technique.Further studies are underway
and are investigating preheating of
the inert gas used within the “sweep”/purge gas flow
to reduce the observed cooling effect and the subsequent decrease
in TIC. In addition, the anaerobic sample delivery technique described
herein does not currently offer the option of high-resolution mass
determination (Da, 4 d.p.) due to the aqueous-based lock-mass solutions
applied within Waters mass spectrometers. Work is currently ongoing
to investigate suitable compounds for use in anaerobic MS to allow
high resolution mass determination of air-/moisture-sensitive samples.