Literature DB >> 35925340

[First clinical results with the PAUL® Glaucoma Implant at the University Eye Hospital Bonn].

Constance Weber1, Sarah Hundertmark2, Ralf Brinken2, Frank G Holz2, Karl Mercieca2.   

Abstract

BACKGROUND: Glaucoma drainage devices (GDD) are an invasive procedure for the treatment of glaucoma. The PAUL® Glaucoma Implant (PGI) has been developed as a new, innovative therapeutic procedure. The PGI differs from previous GDD with regard to the smaller size of the drainage tube.
OBJECTIVE: This study analyses 6‑months results of the PGI in terms of effectiveness and safety.
METHODS: A database of patients treated with the PGI at the University Eye Hospital Bonn was created and continuously updated based on follow-up controls. Statistical analysis was performed using SPSS Statistics for Windows (IBM Corp., Armonk, NY, USA).
RESULTS: A total of 53 eyes of the first 51 consecutive patients treated with the PGI were included in this study. Mean intraocular pressure was 26.62 mmHg (7-48 mmHg) preoperatively and reduced to 12.20 mmHg (3-22 mmHg) after 6 months. Local pressure-lowering therapy was reduced from 3.37 agents preoperatively to 0.30 agents after 6 months. The complication rate was low; only 3 patients (5.8%) had persistent hypotony. In 16 patients, the intraluminal prolene stent was removed in the postoperative course after an average of 2.9 months. Thereafter, these patients experienced a reduction of intraocular pressure from 22.21 to 11.07 mmHg.
CONCLUSION: The PAUL® Glaucoma Implant is a safe treatment modality that can successfully reduce intraocular pressure to a low level and reduce pressure-lowering local therapy. It has a low complication rate, particularly regarding postoperative hypotony.
© 2022. The Author(s).

Entities:  

Keywords:  Glaucoma drainage device; Glaucoma surgery; Refractory glaucoma; Secondary glaucoma; Therapeutic procedure

Year:  2022        PMID: 35925340     DOI: 10.1007/s00347-022-01669-5

Source DB:  PubMed          Journal:  Ophthalmologie        ISSN: 2731-720X


  2 in total

1.  Surgical technique, perioperative management and early outcome data of the PAUL® glaucoma drainage device.

Authors:  Neeru Amrita Vallabh; Fiona Mason; Jonathan T S Yu; Kenneth Yau; Cecilia H Fenerty; Karl Mercieca; Anne Fiona Spencer; Leon Au
Journal:  Eye (Lond)       Date:  2021-09-20       Impact factor: 4.456

Review 2.  New concepts for glaucoma implants--controlled aqueous humor drainage, encapsulation prevention and local drug delivery.

Authors:  Wolfram Schmidt; Christian Kastner; Katrin Sternberg; Reto Allemann; Marian Löbler; Rudolf Guthoff; Klaus-Peter Schmitz
Journal:  Curr Pharm Biotechnol       Date:  2013       Impact factor: 2.837

  2 in total