Mimi J den Uijl1,2, Yorn J H L van der Wijst1,2, Iris Groeneveld2,3, Peter J Schoenmakers1,2, Bob W J Pirok1,2, Maarten R van Bommel1,2,4. 1. van 't Hoff Institute for Molecular Sciences, Analytical-Chemistry Group, University of Amsterdam, Science Park 904, 1098 XH Amsterdam, The Netherlands. 2. Centre for Analytical Sciences Amsterdam (CASA), Amsterdam, The Netherlands. 3. Amsterdam Institute for Molecular and Life Sciences, Division of Bioanalytical Chemistry, Vrije Universiteit Amsterdam, De Boelelaan 1108, 1081HZ Amsterdam, The Netherlands. 4. Amsterdam School for Heritage, Memory and Material Culture, Conservation and Restoration of Cultural Heritage, University of Amsterdam, P.O. Box 94552, 1090 GN, Amsterdam, The Netherlands.
Abstract
Photodegradation greatly affects everyday life. It poses challenges when food deteriorates or when objects of cultural heritage fade, but it can also create opportunities applied in advanced oxidation processes in water purification. Studying photodegradation, however, can be difficult because of the time needed for degradation, the inaccessibility of pure compounds, and the need to handle samples manually. A novel light-exposure cell, based on liquid-core-waveguide (LCW) technology, was embedded in a multiple-heart-cut two-dimensional liquid chromatography system by coupling the LCW cell to the multiple-heart-cut valve. The sample was flushed from the heart-cut loops into the cell by an isocratic pump. Samples were then irradiated using different time intervals and subsequently transferred by the same isocratic pump to a second-dimension sample loop. The mixture containing the transformation products was then subjected to the second-dimension separation. In the current setup, about 30-40% of the selected fraction was transferred. Multiple degradation products could be monitored. Degradation was found to be faster when a smaller sample amount was introduced (0.3 μg as compared to 1.5 μg). The system was tested with three applications, that is, fuchsin, a 19th-century synthetic organic colorant, annatto, a lipophilic food dye, and vitamin B complex.
Photodegradation greatly affects everyday life. It poses challenges when food deteriorates or when objects of cultural heritage fade, but it can also create opportunities applied in advanced oxidation processes in water purification. Studying photodegradation, however, can be difficult because of the time needed for degradation, the inaccessibility of pure compounds, and the need to handle samples manually. A novel light-exposure cell, based on liquid-core-waveguide (LCW) technology, was embedded in a multiple-heart-cut two-dimensional liquid chromatography system by coupling the LCW cell to the multiple-heart-cut valve. The sample was flushed from the heart-cut loops into the cell by an isocratic pump. Samples were then irradiated using different time intervals and subsequently transferred by the same isocratic pump to a second-dimension sample loop. The mixture containing the transformation products was then subjected to the second-dimension separation. In the current setup, about 30-40% of the selected fraction was transferred. Multiple degradation products could be monitored. Degradation was found to be faster when a smaller sample amount was introduced (0.3 μg as compared to 1.5 μg). The system was tested with three applications, that is, fuchsin, a 19th-century synthetic organic colorant, annatto, a lipophilic food dye, and vitamin B complex.
Photodegradation is the
process in which molecules undergo a change
due to UV/vis irradiation. It poses challenges in many different fields,
but also some opportunities. Food products, especially those stored
over time, can be susceptible to light, which affects their shelf-life
time. For other products pulsed-light processing is used as a disinfection
technique.[1−4] Light may cause healthy food ingredients, for example, vitamins,
to degrade and thus lose their nutritional value.[5] The problem becomes even larger when food ingredients are
transformed into products that can be harmful.[6] Understanding photodegradation and its pathways can help prevent
food spoilage and aid in the development of packaging materials that
reduce light-induced transformation. Another field in which light
degradation poses serious problems is cultural heritage. Organic colorants
applied as dyes degrade over time, resulting in art objects, such
as textiles, furniture, drawings, and paintings, changing color and
losing their historical value.[7−11] Knowledge about the degradation pathways and the influence of other
parameters on light degradation can help us preserve valuable objects
for future generations and understand the original appearance of objects
of art.Conversely, there are fields that exploit photodegradation
advantageously.
For example, in water-purification systems, advanced-oxidation processes
are applied to reduce the concentration of organic matter.[12,13] In such processes, hydrogen peroxide can be used in combination
with UV light to eradicate toxic compounds in drinking water. The
photodegradation reactions occurring in such systems (i.e., radical
reactions) require further study, since the byproducts formed could
possibly be harmful.[14]To study photodegradation,
several methods exist.[7,15−17] The simplest
method is to illuminate a solution of
a single dissolved compounds with a lamp. Unfortunately, degradation
studies often report exclusively on the degradation of the starting
compound, rather than studying the pathways and concentrations of
all different degradation products.[5,18] Introducing
more compounds and parameters inevitably leads to a tangle of different
degradation pathways that occur simultaneously. To establish unambiguous
links between a compound and its degradation products, it is preferable
to isolate it prior to degradation, so as to circumvent tedious (or
even impossible) measurements and data analysis. Unfortunately, few
compounds are available in pure form or are very difficult and expensive
to obtain. To determine the reaction rates, different time points
should be recorded, so sufficient material must be available.Photodegradation techniques generally require time and manual handling
(e.g., degradation during a week).[7] Often
a subsequent manual extraction is needed, introducing more sources
of error. If more compounds and more parameters (e.g., time, catalysts,
oxygen supply) are studied, the experimental procedure becomes very
cumbersome and lengthy.An alternative method to perform photodegradation
research more
efficiently was recently published.[7,19,20] In this study a liquid-core-waveguide (LCW) photodegradation
cell was developed. This LCW cell comprises of an amorphous-Teflon
(AF2400) tubing, which has a lower refractive index (RI = 1.29) than
typical solvents. When the LCW cell is filled, the light is guided
through the cell by total internal reflection, irradiating the sample
inside.[21] However, just as with conventional
techniques, this approach focused on solutions of pure compounds without
prior liquid chromatography (LC) separation and the accompanying dilution.[7,19] Investigating mixtures is feasible with this system, however, it
is complicated, since the link between parent molecules and their
degradation products will be difficult. One way to reduce the complexity
of the sample undergoing degradation is to separate the compound of
interest from the mixture with LC separation. If the reaction products
are to be characterized with the aid of LC, the resulting setup would
be akin to a two-dimensional liquid chromatography (2D-LC) system
with a light cell implemented as the modulator.[22] Photodegradation is then a specific form of reaction modulation
or transformative modulation.[23−25] To our knowledge, such a 2D-LC
system with light modulation has never been realized before. Normally,
in 2D-LC, a high orthogonality should be achieved between the two
different (first-dimension, 1D, and second-dimension, 2D) separation systems.[26] However,
in reaction modulation, the situation might be opposite. The difference
between the two separations is due to the sample being transformed,
and the separation systems do not need to be different.In this
work an online photodegradation system was developed that
allows a fast and efficient study of solutions of a number of individual
compounds from a complex mixture. This approach is based on a (LCW)
reactor embedded in a multiple-heart-cut 2D-LC with reversed-phase
liquid chromatography (RPLC) in the 1D and 2D separations. The aim was to investigate the effects of a number
of 2D-LC method parameters and the concentration of compounds on the
design and application of the LCW reactor and to evaluate its performance
by comparing the products obtained at varying residence times with
control samples. Finally, it was aspired to demonstrate the versatility
of the new approach by applying it to examples from cultural heritage
and food ingredients. These include fuchsin,[27−29] which is a
mixture of organic colorants that was extensively used in the late
1900s as dye for textile, food, and wine, but is now used as a biological
staining agent, annatto-seeds extract,[30,31] used widely
in the food industry to color hydrophobic products, such as cheese
and butter, and a vitamin-B-complex formulation,[32] a mixture of essential vitamins.
Materials and Methods
Chemicals
HPLC-quality water was
obtained from a purification system (R = 18.2 MΩ
cm; Arium 611UV, Sartorius, Göttingen, Germany). Methanol (MeOH,
LC-MS grade) was obtained from Biosolve (Valkenswaard, The Netherlands).
Crystal violet (CV; ≥90%), eosin Y (EY; 99%), riboflavin (RF;
≥98%), ammonium formate (AF; ≥99.0%), and formic acid
(FA; ≥95%) were purchased from Sigma-Aldrich (Zwijndrecht,
The Netherlands).Fuchsin was a gift from the Cultural Heritage
Agency of The Netherlands (RCE, Amsterdam, The Netherlands). Annatto
seeds were purchased from De Peperbol (Amsterdam, The Netherlands),
and the Davitamon vitamin-B pills were purchased from Etos (RF concentration
2.8 mg per pill, Amsterdam, The Netherlands).
Instrumentation
LC-LCW-LC-DAD
An Agilent 1290 Infinity
2D-LC system (Agilent, Waldbronn, Germany) was used for all experiments
in this study. The system was comprised of two binary pumps (G4220A),
two diode-array detectors (DAD, G4212A) equipped with Agilent Max-Light
Cartridge Cells (G4212–6008, V0 = 1.0 μL), and a G4226A autosampler as injector. A 2-position
10-port valve configured as an 8-port valve (G4243A) was used with
two multiple-heart-cut (mHC) valves (G64242–64000) with 40
μL loops installed. Two RPLC Zorbax Eclipse Plus C18 columns
(Agilent) were used, with dimensions of 2.1 × 150 mm (3.5 μm
particle size) and 4.6 × 50 mm (1.8 μm) in the first- and
second-dimension, respectively. An 1100 Agilent isocratic pump (G1310A)
was used to transfer the sample from the mHC valve to the LCW cell
and eventually to the loop.The LCW cell (i.d. 800 μm,
o.d. 1000 μm, length 120 mm, volume 60 μL, pressure limit
0.5 MPa) was placed in a light box created by DaVinci Laboratory Solutions
(Rotterdam, The Netherlands), which is described and validated in
previous research[19,20] and shown in Supporting Information, Section S-1. The light source for
degradation was a cold-white (400–700 nm) LED lamp (MCWHF2,
Thorlabs, Newton, NJ, United States), which was coupled to a light
fiber cable (M113L01, Thorlabs) with a 400 μm core diameter,
a UV/vis collimator from Avantes (COL-UV/vis, Apeldoorn, The Netherlands)
and a plano-convex lens (f = 35 mm, Thorlabs, LA4052-ML)
to couple the light from the source into the LCW cell. The LED was
controlled by a LED driver (LEDD1B, Thorlabs). The lightbox was also
equipped with a 6-port switching valve to transfer the sample from
the light box to the 2D separation. The 6-port switching
valve was equipped with a loop of 20 μL.
Methods
Analytical Methods
In this work
a multiple-heart-cut two-dimensional liquid-chromatography setup is
developed and expanded to allow photodegradation in a LCW cell between
the two separation dimensions. A binary pump, autosampler, 1D column, the 1D DAD, and 10-port valve were connected
in series. This latter valve was equipped with two multiple-heart-cut
(mHC) valves. An isocratic pump was used to transfer the 1D fractions stored in the loops of the mHC valves to the LCW cell.
This latter cell was coupled to a 6-port valve, equipped with a 20
μL loop, which connected the 2D binary pump to the 2D column. A schematic overview of the setup is shown in Figure .
Figure 1
Schematic illustration
of liquid chromatography–liquid-core-waveguide
cell–liquid chromatography setup. The three different flow
paths are indicated with a different shade of gray (white, gray, and
black). On the top left, the 1D chromatogram of the test
mix is shown with the peak cuts for the blank, riboflavin, crystal
violet, and eosin Y (semitransparent rectangles).
Schematic illustration
of liquid chromatography–liquid-core-waveguide
cell–liquid chromatography setup. The three different flow
paths are indicated with a different shade of gray (white, gray, and
black). On the top left, the 1D chromatogram of the test
mix is shown with the peak cuts for the blank, riboflavin, crystal
violet, and eosin Y (semitransparent rectangles).For the LC analysis of all samples, mobile-phase
components A and
B consisted of mixtures of aqueous buffer and MeOH in ratios of 95/5
[v/v] for mobile phase A and 5/95 [v/v] for mobile phase B. The aqueous
buffer contained 10 mM ammonium formate at pH = 3, prepared by adding
0.390 g of formic acid and 0.095 g of ammonium formate to 1 L of water.
Mobile-phase components A and B were used for both the 1D and 2D separation. The 1D flow rate was set
to 0.4 mL/min. The 1D gradient program started isocratically
at 100% A from 0 to 1 min, followed by a linear gradient to 100% B
in 7 min, maintained for 2 min at 100% B, and finally returned to
100% A in 2 min. The composition was kept at 100% A for the remainder
of time before starting a new run. The 2D flow rate was
set to 0.3 mL/min. The 2D gradient program started isocratically
at 100% A from 0 to 1 min, followed by a linear gradient to 100% B
in 7 min, 100% B for 1 min, and back to 100% A in 0.01 min. The mobile
phase was kept at 100% A for at least 12 min, depending on the degradation
time. The isocratic pump delivered a 50/50 [v/v] mixture of A and
B at a flow rate of 0.05 mL/min when in operation.In the multiple-heart-cut
method, a 2D gradient stop
time of 11.95 min and a cycle time of 12.00 min were used for both
the 0 min (10 min residence time, no illumination) and 10 min (10
min residence time with illumination) degradations. In general, the
total duration of the 2D method was 2 min longer than the
residence time. For example, for a 30 min degradation the 2D cycle time was 32 min and the gradient stop time 31.95 min. The
time-based heart-cut method was used, and the cut times for all compounds
studied in this work can be found in te Supporting Information, Section S-2. In all cases, an extra blank cut
was taken before the elution of the first peak.The isocratic
pump (isoP, Figure ) was used to transfer the sample from the
mHC loop (indicated in Figure ) to the LCW cell and to transfer the degraded sample from
the cell to the 2D injection loop (see Figure ). When the 2D method
was started, the isoP was operated for 1 min (50 μL)
to bring the sample from the mHC loop to the LCW cell. The flow rate
of the isoP was then set to zero to degrade the sample
until 1.6 min (80 μL) before the start of the next modulation,
where the sample was flushed to the 2D injection loop.
For example, in a method where the sample was degraded for 10 min
and the heart-cut of the last peak ended at 8.81 min, the isocratic
pump was operated at a flow rate of 0.05 mL/min from 8.81 until 9.81
min, 0 mL/min from 9.82 min until 19.20 min, and 0.05 mL/min from
19.21 until 20.81 min. This 12 min cycle was then repeated for the
number of cuts taken.
Sample Preparation
A test mixture
was used of riboflavin (RF, 25 ppm), eosin Y (EY, 20 ppm), and crystal
violet (CV, 5 ppm) in a H2O/MeOH (70/30%) solution. Fuchsin
was dissolved in in H2O/MeOH (50/50%) with a concentration
of 50 ppm.Five annatto seeds (0.1620 g) were extracted in 5
mL of MeOH/H2O (75/25%) and sonicated for 20 min in an
ultrasonic bath. After sonicating, the solution was passed through
a PTFE filter (0.45 μm).[33]Two pills of vitamin B (0.8820 g) were grinded, dissolved in 20
mL of MeOH, and sonicated in an ultrasonic bath for 20 min. After
this, the liquid was passed through a PTFE filter (0.45 μm)
and then diluted in a vial to reach a H2O/MeOH ratio of
50/50%.[34]
Data Processing
The chromatograms
were processed with Agilent OpenLAB CDS software (Agilent, Santa Clara,
CA, U.S.A.). Calculations and figures were performed with MATLAB R2018a
(Mathworks, Woodshole, MA, U.S.A.) and Microsoft Excel.
Results and Discussion
In this research,
an online system was developed that allowed fast
and efficient study of the in-solution photodegradation of a number
of individual compounds obtained from the separation of a mixture.
The system was based on a liquid-core-waveguide (LCW) reactor embedded
in a multiple-heart-cut 2D-LC (mHC-2D-LC) setup. In this section,
the setup and optimizable parameters will first be discussed (Section ). The degradation
efficiency will be evaluated (Section ). To demonstrate the potential of the
system three applications will be described (Sections –3.5).
Design of the System
To facilitate
quick and efficient light degradation, an LCW cell was incorporated
in a mHC-2D-LC setup. This LCW cell was previously coupled to an LC-DAD
for studying degradation products.[7,19] In those studies,
the sample was directly introduced into the light cell either manually[7,19] or by automated sampling with a multipurpose sampler.[20] In the setup used in the present research, however,
the sample was automatically transferred to the LCW from the 1D separation (Figure ). In the so-called time-based-sampling method, the time frame
of the eluting compounds must be inserted in the method. Therefore,
the retention times of the targeted analytes in the 1D
system had to be determined first. The cut times for all compounds
studied in this work can be found in Supporting Information, Section S-2. After the cut times were established,
a sample was injected by an autosampler into the 1D column.
The components of interest were transferred to loops in the mHC valve.
In this work the number of peaks collected did not exceed the capacity
of a single “deck” (multiloop valve) and the first 2D method only started after the last fraction was stored.When the mHC is used to store the separated components prior to exposure
in the LCW, an additional loop is required to transfer the last exposed
component to the 2D separation. Normally in mHC-2D-LC,
the 2D pump is directly connected to the heart-cut valve.
In the present setup an additional pump (isoP in Figure ) was inserted to
transfer the component of interest from the mHC valve to the LCW cell.
The outlet of the LCW was connected to a six-port valve that served
as modulation valve and injection valve for the 2D system.
The storage loops were eluted one by one to the LCW cell, where they
were either stored or degraded for a selected time interval. The exact
timing had to be adapted, depending on the degradation time. When
the sample was injected to the 2D separation, the next
component in a storage loop was transferred to the LCW and the cycle
repeats. To ensure the start of the last 2D separation,
an extra 1D cut was required, since the presence of the
additional flow path and the light cell could not be detected by the
software. In the present setup, a blank 2D run was performed
before the components of interest were transferred to the second dimension,
since all stored analyte peaks were first transferred to the LCW cell
before being transported to the 2D system. This extra cut
was always taken in the method between 5 and 5.1 min. The timings
established for the different degradation times are listed in Supporting Information, Section S-3 and Tables S-2–S-4.Because the LCW
cell has a pressure limit of 0.5 MPa, the flow
rates used in the 1D and 2D LC cannot be used
to fill and empty the light cell. This is circumvented by using wider
tubing (between the LCW and the six-port valve and for the loop in
this valve) and a low operating flow rate for the isocratic pump (see Figure ). The volume of
the mHC storage loops (40 μL) is smaller than that of the LCW
cell (60 μL), which in turn is larger than the injection loop
of the 2D system (20 μL). This will result in an
inevitable loss of sample when a peak is transferred from the 1D column to the 2D system. The highest yields were
observed with a flush volume of either 80 or 85 μL (pump operating
at 1.6 or 1.7 min at 50 μL/min). The ratio of 2D
peak area to 1D peak area for an 80 μL flush volume
was found to be 36% for RF, 42% for CV, and 30% for EY. For an 85
μL flush volume, the ratios were 34% for RF, 37% for CV, and
26% for EY. These ratios are seen to be significantly different for
the different compounds. This is due to the limited size of the storage
loops in the mHC valve (40 μL, corresponding to an elution window
of 0.1 min at a 1D flow rate of 400 μL/min). A smaller
fraction of the sample is transferred when the 1D peak
is broader. This can be corrected for, resulting in a reduction of
the variation between the different analytes. The corrected ratios
for an 80 μL flush volume were found to be 39% for RF, 45% for
CV, and 43% for EY. For the 85 μL flush volume, the ratios were
found to be 37% for RF, 39% for CV, and 36% for EY. Slightly higher
ratios were observed for a transfer volume of 80 μL. For riboflavin
the effect of the transfer time is seen to be small, whereas it is
larger for the later-eluting compounds.It may be argued that
peak-based cutting (based on a detection
threshold) is less prone to variations in chromatographic conditions
than time-based cutting. However, in the present setup peak-based
cutting was only possible with one specific wavelength. Moreover,
a blank (first) cut should be additionally programmed (time-based)
and the entire program of the isocratic pump must be anchored on the
peak-based cutting times. This was not possible with the present setup.
Feasibility of the System
To assess
the system, three compounds in a mixture were studied, that is, riboflavin,
crystal violet, and eosin Y. The first step in this process was to
explore how the analytes behave in the cell, with and without irradiation.
The results are shown in Figure .
Figure 2
Overlay of chromatograms obtained after 10 min residence
time of
a fraction cut from the 1D effluent in the LCW cell with
the light off (blue) and the light on (pink). A, riboflavin (254 nm);
B, crystal violet (590 nm); C, eosin Y (520 nm). Note that the x-axis is following the 1D separation order.
Overlay of chromatograms obtained after 10 min residence
time of
a fraction cut from the 1D effluent in the LCW cell with
the light off (blue) and the light on (pink). A, riboflavin (254 nm);
B, crystal violet (590 nm); C, eosin Y (520 nm). Note that the x-axis is following the 1D separation order.These data were obtained with a modulation time
of 12 min, leading
to a residence time in the LCW cell of about 10 min. The sample was
irradiated with an LED “white” spectrum, including wavelengths
ranging from 400 to 800 nm. This illumination source was chosen to
mimic indoor conditions without UV light as encountered, for example,
in museums or supermarkets. Clear degradation can be seen for riboflavin
(Figure A). For crystal
violet and eosin Y, no degradation products are observed (Figure B,C). Some variations
are observed in the peak area of the main compounds, but this is likely
due to slight variations in the 1D retention times between
runs, leading to variations in sample transfer. For riboflavin, a
number of different degradation products are formed after 10 min.
The most intense peak (eluting at 51 min) corresponds to lumichrome,
as confirmed by its UV-absorbance spectrum. The complete identification
of the degradation products of riboflavin is beyond the scope of this
research.To study the effect of the degradation time, a time
series was
recorded with irradiation ranging from 0 to 30 min (Figure A). Longer residence times
in the LCW result in a further degradation. Besides the increase in
peak area for the lumichrome (LCH in Figure ) and other degradation products, a peak
arose before the RF peak for longer irradiation times, indicating
the formation of a product with a nearly identical 2D retention
time. For quantitative interpretation (vide infra) the two compounds were sufficiently separated and treated independently.
Figure 3
Timeseries
of riboflavin (RF) degradation in the LCW cell for 1D injection
volumes of 15 μL (A) and 3 μL (B):
0 min (10 min with lamp off, blue) 10 min (purple), 20 min (pink, Figure A only), 30 min (yellow).
Timeseries
of riboflavin (RF) degradation in the LCW cell for 1D injection
volumes of 15 μL (A) and 3 μL (B):
0 min (10 min with lamp off, blue) 10 min (purple), 20 min (pink, Figure A only), 30 min (yellow).The results presented in Figure A were obtained by using a 1D
injection
volume of 15 μL, which leads to large sample amounts and, therefore,
possible “saturation” of the LCW cell. The intention
is to irradiate the collected fraction along the entire length of
the cell. However, when saturated, a significant fraction of the light
may be absorbed by the sample at the front of the LCW cell, leading
to an axial illumination gradient. To test this hypothesis, the 1D injection volume of the same sample was reduced by 80% to
3 μL. The results are shown in Figure B. While the two sets of overlaid chromatograms
in Figure are seemingly
similar, the areas of the degradation products are much larger relative
to the RF peak in Figure B. Clearly, the lower amount of sample introduced into the
LCW leads to faster degradation. However, smaller amounts of degradation
product are formed, and identification of minor products may be more
difficult in the current setup. In Figure the area of the degradation products relative
to the total peak area of RF is shown for both the 15 and 3 μL 1D injection volumes. Degradation product a is formed in the first 10 min, but its relative concentration stays
similar thereafter. Hence, its absolute concentration passes through
a maximum. The relative peak areas of the other three degradation
products all increase over time. Products b and c were only detected after 30 min degradation at the lower
injected concentration. However, for LCH, there is a clear increase
visible. The results confirm that a lower analyte concentration results
in faster degradation. The injection volume or the absolute amount
of sample introduced, should be optimized, so as to avoid saturation
of the LCW cell, while exceeding the limits of detection for the degradation
products. A more-sensitive detector shifts the optimum toward lower
injected amounts.
Figure 4
Ratios of peak areas relative to riboflavin for the various
degradation
products (a, b, c, and lumichrome, LCH) for 1D injection
volumes of 15 μL (left) and 3 μL (right) for 10- and 30
min irradiation.
Ratios of peak areas relative to riboflavin for the various
degradation
products (a, b, c, and lumichrome, LCH) for 1D injection
volumes of 15 μL (left) and 3 μL (right) for 10- and 30
min irradiation.
Application to Fuchsin
To demonstrate
the versatility of the LC-LCW-LC system several different applications
were studied. The first of these concerned fuchsin, which is one of
the earliest synthetic organic colorants from the 19th century. It
was used to dye textile and as a food colorant in the beginning of
the 20th century. Nowadays it is used as a biological staining agent
for bacteria and sometimes as a disinfectant.[30,31] The structure of fuchsin is shown in Supporting Information, Section S-4, Figure S-3. It is sold as a mixture of four compounds, which are the triply,
doubly, and singly methylated derivatives (known as Magenta III, M3,
Magenta II, M2, and Magenta I, M1, respectively) and pararosaniline
(M0), which is not methylated (see Figures S-2–S-5). The chromatogram of the mixture (detection wavelength 555 nm)
is shown in Figure A. The four main compounds can clearly be seen, but impurities are
also present. Out of the four main compounds, M0 and M3 are commercially
available as pure components, allowing their degradation pathways
to be studied individually.[35,36] In contrast, it has
not been possible so far to study the degradation of M1 and M2, which
are not available as (more or less) pure compounds. The present LC-LCW-LC
setup offers a unique possibility for degradation studies of such
individual compounds. Separating the fuchsin mixture prior to compound
degradation is essential for this purpose. Figure B illustrates the successful degradation
of M1.
Figure 5
(A) RPLC chromatogram of a fuchsin sample. 1D detection
wavelength 555 nm. The shaded fraction (M1) is transferred to the
LCW and degraded for 4 h. (B) Second-dimension chromatograms of the
M1 fraction after 4 h degradation shown in yellow and pink (overlay
of two repeat experiments). M1 in 2D indicated by the yellow
shaded bar. 2D detection wavelength 254 nm.
(A) RPLC chromatogram of a fuchsin sample. 1D detection
wavelength 555 nm. The shaded fraction (M1) is transferred to the
LCW and degraded for 4 h. (B) Second-dimension chromatograms of the
M1 fraction after 4 h degradation shown in yellow and pink (overlay
of two repeat experiments). M1 in 2D indicated by the yellow
shaded bar. 2D detection wavelength 254 nm.After 1 h, no degradation products were observed.
The present setup
readily allows more extensive degradation studies. For that reason,
it was chosen to degrade fuchsin (M1) for a longer period. After 4
h, at a wavelength of 254 nm, a number of degradation products were
detected (Figure B).
The present system creates many opportunities to study these products
in detail, for example by attaching the setup to a mass spectrometer.
The entire experiment (1D separation, 4 h degradation, 2D separation) was repeated, and the 2D chromatograms
are overlaid in Figure B. Excellent repeatability of the entire process is demonstrated,
despite the long degradation times. Slight variations in peak areas
(e.g., for M1) can be explained from minor variations in the 1D retention times.
Application to Annatto
Annatto is
a natural food-coloring agent. It is obtained from the seeds of the
Achiote tree and used for coloring many different lipophilic products,
such as cheese, ice cream, and margarine. The main coloring compound
in the annatto extract is bixin, an apocarotenoid, but there are many
other coloring compounds in the extract, as can be seen in the two
chromatograms of Figure A (recorded at 254 and 450 nm). Studying the photodegradation pathways
of bixin can help the food industry prevent food spoilage and improve
food packaging. Bixin is one of many compounds present and it is difficult
and expensive to obtain it in a purified form. Therefore, degradation
studies of bixin have been only been performed on bixin from the annatto
extract instead on the pure form.[37,38]
Figure 6
(A) RPLC chromatogram
of an annatto seed extract. Detection wavelength
254 nm (black) and 450 nm (gray). The shaded gray fraction is transferred
to the LCW and degraded using different time intervals. (B1 and B2) 2D chromatograms representing a timeseries of bixin during
0 (blue), 30 (purple), 60 (pink), and 120 min (yellow), recorded at
254 nm. (C) is the same as (B), except recorded at 450 nm.
(A) RPLC chromatogram
of an annatto seed extract. Detection wavelength
254 nm (black) and 450 nm (gray). The shaded gray fraction is transferred
to the LCW and degraded using different time intervals. (B1 and B2) 2D chromatograms representing a timeseries of bixin during
0 (blue), 30 (purple), 60 (pink), and 120 min (yellow), recorded at
254 nm. (C) is the same as (B), except recorded at 450 nm.Figure B,C show
a series of overlaid 2D chromatograms of degraded bixin
(0, 30, 60, and 120 min), recorded at 254 and 450 nm, respectively.
At 450 nm, a small shoulder peak starts to appear before the bixin
peak after 30 min exposure. After 60 min, a number of small peaks
are visible at 254 nm. This shows the complexity of the degradation
of a single compound, and it underlines the value of the current LC-LCW-LC
setup. Without the prior separation it would be impossible to deduce
which degradation products arise from bixin. This shows how important
a preseparation to the photodegradation reaction can be. Again, adding
a mass spectrometer to the system will allow detailed interpretation
of the structures and pathways.
Application to Vitamin B
Vitamin
B is a group of eight essential vitamins, including thiamine (B1), riboflavin (B2), niacin (B3), pantothenic
acid (B5), pyridoxine (B6), biotin (B7), folic acid (B11), and cobalamins (B12),
with many different health benefits. All kinds of vitamin B products
are available on the market. As already described in section , riboflavin is susceptible
to light. The riboflavin for the study described in section was obtained as a chemical
standard which can be studied without prior separation. The LC-LCW-LC
setup allows study of riboflavin, and all other compounds present
in a vitamin B complex. In Figure A, the 1D chromatogram is shown for a vitamin-B
formulation. Many compounds with vastly different peak areas are observed.
The chromatographic behavior of the molecules with many acid groups
is poor under these chromatographic conditions.
Figure 7
(A) RPLC chromatogram
of a vitamin-B-complex formulations, recorded
at 254 nm (black) and 300 nm (gray). (B) 2D chromatograms
corresponding to a timeseries of degradations of RF recorded at 254
nm (0 min, blue, 10 min, purple, 30 min, yellow).
(A) RPLC chromatogram
of a vitamin-B-complex formulations, recorded
at 254 nm (black) and 300 nm (gray). (B) 2D chromatograms
corresponding to a timeseries of degradations of RF recorded at 254
nm (0 min, blue, 10 min, purple, 30 min, yellow).RF was isolated from the mixture using the 1D separation,
introduced into the LCW and irradiated using different time intervals.
The 2D chromatograms after degradation of RF are shown
in Figure B. This
illustrates that RF (or, indeed, any compound) can be studied from
complex mixtures, yielding results that are comparable to those obtained
with pure standards (see section and Figure ). It should be noted, however, that the conditions in the
light cell are different from those experienced in real-world situations.
This matrix is acidified and has MeOH present as organic modifier.
It is possible that these reaction conditions could lead to different
reaction kinetics than present in real-world degradations. It is of
importance to compare these degradation studies to the literature
present on these subjects. As such, this system should not be considered
as a replacement of other aging techniques, but rather as a new tool
providing information about degradation mechanisms in a much faster
manner. However, understanding the differences in degradation between
these matrices proved to be important.[7,8,39]
Concluding Remarks
In this work, a
2D-LC-based photodegradation workflow was developed
to study isolated compounds from mixtures using a liquid-core-waveguide
(LCW) cell as reactor. This allowed rapid and efficient studies of
components of interest isolated from mixtures. Fractions of the first-dimension
effluent were collected by a multiple-heart-cut valve, specifically
targeting compounds of interest. A method was developed to transfer
the analyte fractions from the first-dimension separation to the light
exposure cell and subsequently to the second-dimension separation.
About 30% to 40% of the isolated and degraded fraction was transferred.
The fastest degradations could be performed with low concentrations
of sample (total amount of analyte is approximately 0.3 μg),
and care should be taken to avoid saturation of the LCW, which can
result in an inhomogeneous irradiation. The system allows us to generate
degradation profiles by irradiation of the sample using different
time intervals. To our knowledge, this is the first time light-induced
reaction modulation has been employed. The method proved to be stable
over longer degradation times (up to 4 h). To demonstrate the versatility
of the system several samples that contained compounds of interests,
as well as many other components, were studied by selecting one of
the components present in the mixture. This illustrates the relevance
of a preseparation having the advantage that the first separation,
the exposure inside the LCW, and the second separation can be carried
out online and fully automated. As an example, according to our knowledge,
this was the first time a degradation profile of magenta I isolated
from fuchsin was demonstrated.
Authors: Bert Wouters; Bob W J Pirok; Dionysios Soulis; Rocío C Garmendia Perticarini; Sander Fokker; Rick S van den Hurk; Maciej Skolimowski; Ron A H Peters; Peter J Schoenmakers Journal: Anal Chim Acta Date: 2018-12-12 Impact factor: 6.558
Authors: Bob W J Pirok; Noor Abdulhussain; Tom Aalbers; Bert Wouters; Ron A H Peters; Peter J Schoenmakers Journal: Anal Chem Date: 2017-08-08 Impact factor: 6.986