Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems.

Mo'mena Ahmad A Awad-Allah1, Hesham Mohamed Gharieb1, Rania Gamal Eldin Zaki2, Ihab Saad Othman3.   

Abstract

PURPOSE: To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices. PATIENTS AND METHODS: A prospective randomized cohort study, conducted on 47 eyes of 47 healthy orthotropic individuals, with an age range of 18-50 years and a corrected Snellen's distance visual acuity (CDVA) of 0.8 decimal or better. Angle Kappa is assessed directly using Orbscan® III software version 1.8.165.1. (Bausch and Lomb Rochester, New York, United States), while Pentacam® HR 1.21r.65 (Oculus Optikgeräte GmbH, Wetzlar, Germany) and Sirius device (CSO, version 3.2.1.60, Costruzione Strumenti Oftalmici, Florence, Italy) were used to calculate angle kappa indirectly.
RESULTS: Least mean difference of estimated angle Kappa was between Orbscan and Pentacam devices (- 0.18° ± 1.8), and it was statistically insignificant (p value = 0.1294). Differences between both Orbscan and Sirius, and Pentacam and Sirius were statistically significant (p value = 0.0004 and < 0.0001 consecutively). Bland Altman analysis showed a 95% confidence interval between Orbscan III and Pentacam of - 3.76 to 3.4 and between Orbscan III and Sirius of - 3.79 to 2.26.
CONCLUSION: Pentacam parameters can be used as a reliable method to calculate angle kappa indirectly, without usage of any additional measurements from other machine. Sirius device parameters could also be used, but with less accurate results. A simple modification to those devices' software to calculate it, and incorporate it in the printout is possible, and highly recommended.
© 2022. The Author(s).

Entities:  

Keywords:  Kappa angle; Orbscan III; Pentacam; Sirius device

Year:  2022        PMID: 35902424     DOI: 10.1007/s10792-022-02433-z

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.029


  1 in total

1.  Angle Kappa Measurements: Normal Values in Healthy Iranian Population Obtained With the Orbscan II.

Authors:  Hamid Gharaee; Masoud Shafiee; Rafie Hoseini; Mojtaba Abrishami; Yalda Abrishami; Mostafa Abrishami
Journal:  Iran Red Crescent Med J       Date:  2014-12-25       Impact factor: 0.611

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.