Jhoan Danilo Arcos Rosero1, Daniel Camilo Bolaños Rosero1, Luis Fernando Alape Realpe1, Andrés Felipe Solis Pino1,2, Elizabeth Roldán González3.
Abstract
Injuries in the elbow area, such as lateral and medial epicondylitis, are the leading causes of consultation with health specialists. Therefore, this research proposes the mechatronic design of an orthosis with a graphic interface that supports professionals in the rehabilitation of the elbow joint through the execution of flexion-extension and pronation-supination movements. For the development of the rehabilitation prototype, mechatronic design, co-design, and IDEF0 methodologies are used, performing activities such as actuator characterization, simulations, and modeling, among others. Through the execution of a case study in a real environment, the device was validated, where the results suggest a functional and workable prototype that supports the treatment of pathologies in the elbow area through the execution of the mentioned movements, supposing that this is a low-cost alternative with elements to improve, such as the industrial design and new functionalities. The developed proposal shows potential as an economical product that health professionals can use. However, some limitations related to the design and functionalities in the application domain were identified.Entities:
Keywords: elbow injury; mechatronic design; orthosis; rehabilitation prototype
Year: 2022 PMID: 35877338 PMCID: PMC9311914 DOI: 10.3390/bioengineering9070287
Source DB: PubMed Journal: Bioengineering (Basel) ISSN: 2306-5354
Figure 1General activities followed for the design and implementation of the prototype rehabilitator for upper torso areas.
Normal elbow mobility values provided by the Association for the Study of Osteosynthesis and the American Academy of Orthopaedic Surgeons.
| Motion | AO | AAOS |
|---|---|---|
| Flexion | 0–150 | 0–150 |
| Extension | 0–10 | 0 |
| Supination | 0–90 | 0–80 |
| Pronation | 0–90 | 0–80 |
Hand and forearm measurements and weights considered.
| Range | Minimun [cm] | Maximun [cm] |
|---|---|---|
| Hand length | 15.24 | 21.34 |
| Hand width | 6.86 | 9.91 |
| Hand thickness | 2.36 | 3.81 |
| Handle diameter | 3.18 | 3.81 |
| Handgrip width | 7.62 | 10.41 |
| Fist length | 9.65 | 14.22 |
| Forearm length | 21.08 | 27.43 |
| Forearm diameter | 6.1 | 8.13 |
Figure 2Methodology for the design and construction of the prototype.
Figure 3General system diagram IDEF A-0.
Figure 4Detailed IDEFA0 system diagram (detail view).
Figure 5CAD design of an upper torso rehabilitative orthosis prototype.
Stress analysis of the mechanisms of the prototype orthosis to support upper torso pathologies.
| Mechanism | Tension of Von Mises [N/m | Displacement [mm] |
|---|---|---|
| Prone-supination of the wrist | ||
| Prone-supination of the wrist |
Figure 6Graphical interface of the shoulder rehabilitation orthosis for the health care professional.
Figure 7Graphical interface of the database of patients with upper torso pathologies.
Absolute error for the angles of movement measured with the goniometer and the proposed prototype.
| Movement | Measured Value | Actual Value |
|---|---|---|
| Flexo-extension | 70 | 69 |
| Flexo-extension | 135 | 134 |
| Flexo-extension | 20 | 19 |
| pronation-supination | 90 | 90 |
| pronation-supination | 120 | 120 |
Figure 8Photographic evidence of the orthosis prototype for upper torso rehabilitation.
Figure 9Prototype orthosis to support upper torso rehabilitation in case study.
List of activities conducted in the case study.
| Activities | Planned Duration | Support Tools |
|---|---|---|
| Socialization and contextualization of the research project | 30 min | Presentation of the introduction to the case study and the conceptual elements of the proposal. |
| Presentation and introduction of the orthosis prototype | 30 min | Document with the description of the prototype and presentation of the tool. |
| Implementation of the orthosis on the study subjects | 45 min | Requirements document, orthosis guidance document, scenarios, and results templates for execution. |
| Evaluation of the implementation | 20 min | Perception survey |
Figure 10Results of the satisfaction survey for users of the upper torso rehabilitation support orthosis prototype.