Performance of high-resolution monitors for digital chest imaging.

S A Suddarth, G A Johnson, R H Sherrier, C E Ravin.   

Abstract

High-resolution cathode-ray tubes (CRT's) are currently the most viable soft-copy display for digital radiography. We present here methods for measuring large-area contrast ratio and detail contrast ratio. A two-dimensional charge coupled device (ccd) array signal-averaged with a video frame buffer permits linear microradiometric measure of individual beam lines. Results from three different 1000-line monitors demonstrate the shift variance of resolution. The detail contrast ratio (or modulation depth) was found to vary from 100% to less than 10% across the face of one CRT. Dynamic focus in both the horizontal and vertical deflection circuitry proved effective in reducing this shift variance. Comparisons of three phosphors demonstrate the utility of long persistence phosphors (P164) for static display in producing brighter images with less flicker. Recommendations for CRT design and selection for high-resolution digital radiography are included.

Mesh:

Year:  1987        PMID: 3587149     DOI: 10.1118/1.596079

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

Review 1.  Literature review: picture archiving and communication system.

Authors:  U P Schmiedl; A H Rowberg
Journal:  J Digit Imaging       Date:  1990-11       Impact factor: 4.056

2.  The digital imaging workstation. 1990.

Authors:  Ronald L Arenson; Dev P Chakraborty; Sridhar B Seshadri; Harold L Kundel
Journal:  J Digit Imaging       Date:  2003-03       Impact factor: 4.056

3.  Detection of interstitial lung abnormalities on picture archive and communication system video monitors.

Authors:  T L Washowich; S C Williams; L A Richardson; G E Simmons; N V Dao; T W Allen; G C Hammet; M J Morris
Journal:  J Digit Imaging       Date:  1997-02       Impact factor: 4.056

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.