| Test method | Test description | Types of contaminations | Approximate contaminant size |
|---|---|---|---|
| ISO 16603 | Determination of the resistance of protective clothing materials to penetration by blood and body fluids | Synthetic blood | - |
| ISO 16604 | Determination of the resistance of protective clothing materials to penetration by blood-borne pathogens | Phi-X 174 bacteriophage | 0.03 mm[ |
| ISO 22610 | Determination of the resistance to wet bacterial penetration | Bacillus atrophaeus contaminated liquid | 1.0–1.6 mm (length); 0.6–0.9 mm (diameter)[ |
| ISO/DIS 22611 | Determination of the resistance to penetration by biologically contaminated aerosols | Staphylococcus aureus contaminated aerosols | 1 mm[ |
| ISO 22612 | Determination of resistance to dry microbial penetration | Bacillus subtilis contaminated talcum powder | 0.9–1.5 mm (length); 0.4–0.7 mm (diameter)[ |
Test methods for ANSI/AAMI PB 70:2012
| Test method | Test description | Challenge liquid |
|---|---|---|
| AATCC 42 | Water resistance: impact penetration | Water |
| AATCC 127 | Water resistance: hydrostatic pressure | Water |
| ASTM F1670 | Determination of the resistance of protective clothing materials to penetration by synthetic blood | Synthetic blood |
| ASTM F1671 | Determination of the resistance of protective clothing materials to penetration by blood-borne pathogens | Phi-X 174 bacteriophage |
ANSI/AAMI PB70 and EN14126:2003 classification of the level of barrier performance
| Level | Test methods | Criteria |
|---|---|---|
| ANSI/AAMI PB70 standard | ||
| 1 | AATCC 42 | ≤ 4.5 g |
| 2 | AATCC 42 | ≤ 1.0 g |
| AATCC 127 | ≥ 20 cm | |
| 3 | AATCC 42 | ≤ 1.0 g |
| AATCC 127 | ≥ 50 cm | |
| 4 | ASTM F1670 | No penetration at 13.8 kPa |
| ASTM F1671 | No penetration at 13.8 kPa | |
| EN14126:2003 standard | ||
| 1 | ISO 16603 & ISO 16604 | 0.0 kPa |
| 2 | ISO 16603 & ISO 16604 | 1.75 kPa |
| 3 | ISO 16603 & ISO 16604 | 3.5 kPa |
| 4 | ISO 16603 & ISO 16604 | 7.0 kPa |
| 5 | ISO 16603 & ISO 16604 | 14.0 kPa |
| 6 | ISO 16603 & ISO 16604 | 20.0 kPa |