Kafui C Kouassi1,2, Améyo M Dorkenoo3,4, Komivi Gbada3,5, Yaovi-Gameli Afanyibo3,6, Minogblon Têko3,7, Adjane Koura8.
Abstract
Background: A national proficiency test (PT) programme is not currently implemented in most low-income countries. However, participation in such PT programmes assists improves test performance and result accuracy. Objective: This study assessed how well 11 government hospital laboratories performed 18 basic clinical chemistry tests and identified areas needing improvement.Entities:
Keywords: Togo; biochemistry; laboratory; performance; quality control
Year: 2022 PMID: 35811750 PMCID: PMC9257935 DOI: 10.4102/ajlm.v11i1.1565
Source DB: PubMed Journal: Afr J Lab Med ISSN: 2225-2002
Clinical chemistry tests with two different methods used in Togo from July 2016 to December 2016.
| Tests | Group A methods | Group B methods | ||
|---|---|---|---|---|
| Laboratory method | Number of PLs | Laboratory method | Number of PLs | |
| Total Bilirubin | Sulfanilic acid/Dimethylsulfoxyl | 10 | Sulfanilic acid/Caffeine-sodium benzoate | 1 |
| Calcium | O-cresolphthalein | 3 | Arsenazo III | 8 |
| Magnesium | Calmagite | 7 | Xylidyl Blue | 4 |
PLs, participating laboratories.
Spectrophotometers and brands of analysers and reagents used by the participating laboratories in Togo from July 2016 to December 2016.
| Category of spectrophotometer | Manufacturer | Country of origin | Instrument model | Reagents |
|---|---|---|---|---|
| Fully automated | Vital Scientific® | the Netherlands | SELECTRA® PRO M | Elitech® |
| Spinreact® | Spain | SPIN 120® | Chemelex® | |
| QUIMICA CLINICA® | Spain | QUIMICA CLINICA® BIO 100 | Quimica clinica® | |
| HUMAN Diagnotics Wordwide® | Germany | HUMASTAR 200 | Human® | |
| Mindray® | China | MINDRAY® BS-120 | Chemelex® | |
|
| ROBERT RIELE® | Germany | RIELE® 5010 | Chemelex® |
| Hospitex Diagnostics® | Italy | Master T- HOSPITEX® | Cypress® | |
| URIT® | China | URIT®-810 | Chemelex® | |
| Sinnowa Medical Science & Technology® | China | SINNOWA® BS-3000 | Chemelex® |
FIGURE 1Analytical performance score of participating laboratories for each test performed in Togo from July 2016 to December 2016.
Comparison of participating laboratories’ performance between cycles in Togo from July 2016 to December 2016.
| PT cycles | Acceptable results |
| |||
|---|---|---|---|---|---|
|
| % |
| % | ||
| Cycle 1 versus Cycle 2 | 156 | 71 | 139 | 78 | 0.092 |
| Cycle 2 versus Cycle 3 | 139 | 78 | 156 | 76 | 0.721 |
| Cycle 3 versus Cycle 4 | 156 | 76 | 141 | 81 | 0.287 |
Note: Cycle 1 (71%) versus Cycle 2 (78%) versus Cycle 3 (76%) versus Cycle 4 (81%); P = 0.069.
PT, proficiency test; n, number of acceptable results.
Comparison of participating laboratories’ performance based on the use of internal quality control in Togo from July 2016 to December 2016.
| PT cycle | Acceptable results |
| |||
|---|---|---|---|---|---|
| PLs using IQC serum ( | PLs using no IQC serum ( | ||||
|
| % |
| % | ||
| Cycle 1 ( | 61 | 76 | 95 | 67 | 0.110 |
| Cycle 2 ( | 55 | 86 | 81 | 71 | 0.020 |
| Cycle 3 ( | 66 | 89 | 94 | 72 | 0.010 |
| Cycle 4 ( | 62 | 98 | 87 | 79 | 0.002 |
PT, proficiency test; PL, participating laboratory; N, Number of laboratories; n, number of results; IQC, internal quality control.