Diego Mannhart1,2, Elisa Hennings1,2, Mirko Lischer1,2, Claudius Vernier1,2, Jeanne Du Fay de Lavallaz1,2, Sven Knecht1,2, Beat Schaer1,2, Stefan Osswald1,2, Michael Kühne1,2, Christian Sticherling1,2, Patrick Badertscher1,2.
Abstract
Introduction: The Withings Scanwatch (Withings SA, Issy les Moulineaux, France) offers automated analysis of the QTc. We aimed to compare automated QTc-measurements using a single lead ECG of a novel smartwatch (Withings Scanwatch, SW-ECG) with manual-measured QTc from a nearly simultaneously recorded 12-lead ECG.Entities:
Keywords: QTc; artificial intelligence; digital health; intelligent ECG; remote patient monitoring (RPM); single-lead ECG; smartwatch
Year: 2022 PMID: 35811720 PMCID: PMC9259864 DOI: 10.3389/fcvm.2022.906079
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
FIGURE 1Flowchart showing pathway of pre-analysis data management and acquisition. AI, artificial intelligence.
Baseline characteristics of patients available for analysis.
| Overall population | QT-interval | QT-interval not detectable in SW-ECG | ||
| Female sex-no. (%) | 151 (48) | 93 (56) | 58 (41) | |
| Mean age -yr | 63.3 (± 17.2) | 60.9 (± 18.5) | 66.3 (± 14.9) | |
| ≥65 yr-no. (%) | 171 (54) | 85 (48) | 86 (61) | |
| BMI, kg/m2 | 26.7 (± 5.2) | 26.4 (± 5) | 27.0 (± 5.5) | |
| Hypertension-no. (%) | 159 (50) | 77 (44) | 82 (59) | |
| DM-no. (%) | 71 (22) | 38 (22) | 33 (24) | |
| Stroke/TIA-no. (%) | 16 (5) | 10 (6) | 6 (4) | |
| History HF-no. (%) | 46 (15) | 16 (9) | 30 (21) | |
| CAD-no. (%) | 87 (27) | 42 (24) | 45 (32) | |
| Sinus rhythm-no. (%) | 242 (90) | 165 (93) | 119 (85) | |
| AF-no. (%) | 33 (10) | 12 (7) | 21 (15) | |
|
| ||||
| Ischemia-no. (%) | 4 (1) | 1 (1) | 3 (2) | |
| Rhythm-no. (%) | 24 (8) | 13 (7) | 11 (8) | |
| QT-no. (%) | 16 (5) | 7 (4) | 9 (6) | |
| Routine check-up-no. (%) | 230 (73) | 125 (71) | 105 (75) | |
| Other -no. (%) | 43 (14) | 31 (18) | 12 (9) | |
|
| ||||
| Right BBB-no. (%) | 20 (6) | 10 (6) | 10 (7) | |
| Left BBB-no. (%) | 28 (9) | 9 (5) | 19 (14) | |
| Left AFB-no. (%) | 9 (3) | 7 (4) | 2 (1) |
AF, atrial fibrillation; AFB, anterior fascicular block; BBB, bundle branch block; BMI, body mass index; CAD, coronary artery disease; DM, diabetes mellitus; ECG, electrocardiogram; HF, heart.
FIGURE 2Comparison of QT-interval measurements, manually measured QTc on a 12-lead ECG and automated SW-ECG. Measurement Agreement was analyzed using the Bland-Altmann Method. Bias is represented by the solid black line, limit of agreement is represented via d dashed red line. Sections of SW-ECG from outliners are shown in red circles. SW-ECG, Single-Lead Withings Electrocardiogram; QTc, corrected QT-interval; AI, Artificial Intelligence.