| Literature DB >> 35774208 |
Jiabo Shen1, Longfu Jiang1, Lifang Pan2.
Abstract
Entities:
Keywords: Conduction system pacing; Intracardiac electrogram; Isoelectric interval; Left bundle branch pacing; Retrograde atrial potential
Year: 2022 PMID: 35774208 PMCID: PMC9237273 DOI: 10.1016/j.hrcr.2022.03.009
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1A,B: Retrograde right atrial potential (PoRA) was gradually decreased as the pacing electrode was screwed from the right ventricular septum to the left ventricular septum. C: When the electrode was screwed into the left bundle branch (LBB) area, PoRA abruptly decreased and the stimulus-to–left ventricle activation time was suddenly shortened. D: An isoelectric interval was observed during decreased in pacing output. E: LBB potential and PoRA were recorded. LVSP = left ventricular septal pacing; ns-LBBP = nonselective left bundle branch pacing; RA = right atrium; RVSP = right ventricular septal pacing; s-LBBP = selective left bundle branch pacing.
Figure 2Retrograde right atrial potential (PoRA) was gradually decreased as the pacing electrode was screwed from the right ventricular septum to the left ventricular septum. When the electrode was screwed into the left bundle branch (LBB) area, PoRA abruptly decreased and the stimulus-to–left ventricle activation time was suddenly shortened. An isoelectric interval was observed during decrease in pacing output. Abbreviations as in Figure 1.