The oxygen sensitivity of a multipoint antimony electrode for tissue pH measurements. A study of the sensitivity for in vivo PO2 variations below 6 kPa.

F Sjöberg, G Edwall, N Lund.   

Abstract

Monocrystalline micro antimony electrodes in a multipoint arrangement as described by Lund et al. were placed on the skeletal muscle surface of the rabbit. Tissue oxygen levels were measured simultaneously with the MDO (Mehrdraht Dortmund Oberfläche) oxygen electrode. The sensitivity for variations in tissue PO2 (PO2(t)) was evaluated for the antimony metal-metal oxide sensor. The sensitivity (delta E/delta log10 PO2)+/- SE was found to be 21.8 +/- 1.2 mV in the interval between 0.1 kPa and 1 kPa and 53 +/- 5 mV in the interval between 1 kPa and 6 kPa. These results are not consistent with the oxygen sensitivity of monocrystalline antimony described in vitro, but are in agreement with the findings of Nilsson & Edwall. A plausible explanation for the S-shaped oxygen sensitivity curve of antimony at oxygen levels below 10 kPa could be an interaction, at the electrode surface, between the dissolved oxygen and the oxygen bound to haemoglobin. If this is the case, the use of an antimony electrode would make possible the determination of the dissociation of oxyhaemoglobin in tissues.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3576105     DOI: 10.1080/00365518709168863

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  3 in total

1.  New approach for elucidating the oxygen sensitivity and calibration of the antimony electrode.

Authors:  F Sjöberg; G Nilsson; U Gustafsson
Journal:  Med Biol Eng Comput       Date:  1997-05       Impact factor: 2.602

2.  Variations in esophageal oxygen tension measured with intraluminal antimony electrodes.

Authors:  L Tibbling; F Sjöberg
Journal:  Dysphagia       Date:  1995       Impact factor: 3.438

3.  A Multi-Parametric Device with Innovative Solid Electrodes for Long-Term Monitoring of pH, Redox-Potential and Conductivity in a Nuclear Waste Repository.

Authors:  Jordan Daoudi; Stephanie Betelu; Theodore Tzedakis; Johan Bertrand; Ioannis Ignatiadis
Journal:  Sensors (Basel)       Date:  2017-06-13       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.