Vitali Koch1, Gerald Loos1, Leon D Gruenewald1, Katrin Eichler1, Christian Booz1, Tommaso D'Angelo2, Ibrahim Yel1, Scherwin Mahmoudi1, Simon S Martin1, Marc Harth1, Moritz H Albrecht1, Stephan Zangos1, Simon Bernatz1, Axel Thalhammer1, Jan-Erik Scholtz1, Thomas J Vogl1, Tatjana Gruber-Rouh1.
Abstract
Purpose: To assess the diagnostic precision of three different workstations for measuring thoracic aortic aneurysms (TAAs) in vivo and ex vivo using either pre-interventional computed tomography angiography scans (CTA) or a specifically designed phantom model.Entities:
Keywords: CTA, Computed tomography angiography; Computed tomography; Diagnostic accuracy; Endovascular aortic repair; PACS, Picture archiving and communication system; PVC, Polyvinyl chloride; TAA, Thoracic aortic aneurysm; TEVAR, Thoracic endovascular aortic repair; Thoracic aortic aneurysm; Workstation comparison
Year: 2022 PMID: 35757233 PMCID: PMC9213697 DOI: 10.1016/j.ejro.2022.100427
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
Fig. 1Three-dimensional illustration of the phantom using A) TeraRecon, B) Siemens after inserting the central axis (shown as purple line), and C) General Electric with the depiction of the central axis (shown as green line) prior to manual adjustments. The greyish zone in the middle of the phantom (C) is notable which corresponds to numerous air bubbles collected at the aneurysmatic phantom lesion.
Fig. 2Schematic illustration of the phantom showing its true diameters at different predefined measurement points.
Fig. 3Illustration of measurements acquired at the Siemens workstation on coronal and axial images. The purple line represents the central axis (A) and the inner diameter of the aorta (B) in CTA scans. Measurements of phantom tube diameters are shown in coronal (C) and axial views (D). The blue lines represent the inner tube diameter.
Baseline characteristics of the study cohort.
| Number of overall patients (women; men) | 23 (7; 16) |
| Overall mean age (y) ± SD, range | 58 ± 10, 42–83 |
| Overall mean BMI (kg/m²) ± SD, range | 27 ± 3, 23–32 |
| Mean age of women (y) ± SD, range | 57 ± 9, 46–70 |
| Mean age of men (y) ± SD, range | 58 ± 11, 42–83 |
Abbreviations: BMI, body mass index.
Diagnostic accuracy of the phantom at different predefined and fixed measurement points.
| Tube diameter at the subclavian artery | 35 | 35.50 | 34.89–36.11 | 0.0957 |
| Tube diameter at the beginning of the aneursym | 35 | 36.40 | 36.03–36.77 | < 0.0001 |
| Maximum tube diameter of the aneurysm | 75 | 74.90 | 74.67–75.13 | 0.3434 |
| Tube diameter at the end of the aneurysm | 35 | 36.30 | 35.95–36.65 | < 0.0001 |
| Tube diameter at the coeliac trunk | 35 | 36.00 | 36.00–36.00 | < 0.0001 |
| Distance of the aneurysm to the subclavian artery | 50 | 51.42 | 49.55–53.29 | 0.1194 |
| Length of the aneurysm | 300 | 290.00 | 281.15–298.85 | 0.0309 |
| Distance of the aneurysm to the coeliac trunk | 50 | 52.10 | 50.95–53.52 | 0.0085 |
| Total distance between the origin of the subclavian artery and the coeliac trunk | 400 | 393.52 | 384.38–402.66 | 0.1431 |
| Tube diameter at the subclavian artery | 35 | 36.23 | 35.39–37.06 | 0.0088 |
| Tube diameter at the beginning of the aneursym | 35 | 37.07 | 36.45–37.68 | < 0.0001 |
| Maximum tube diameter of the aneurysm | 75 | 76.44 | 74.91–77.97 | 0.0953 |
| Tube diameter at the end of the aneurysm | 35 | 37.28 | 36.82–37.75 | 0.0726 |
| Tube diameter at the coeliac trunk | 35 | 36.17 | 35.78–36.64 | 0.0004 |
| Distance of the aneurysm to the subclavian artery | 50 | 50.06 | 48.67–51.45 | 0.9268 |
| Length of the aneurysm | 300 | 281.81 | 278.75–284.88 | < 0.0001 |
| Distance of the aneurysm to the coeliac trunk | 50 | 51.21 | 48.68–53.74 | 0.3078 |
| Total distance between the origin of the subclavian artery and the coeliac trunk | 400 | 383.08 | 379.17–386.98 | < 0.0001 |
| Tube diameter at the subclavian artery | 35 | 35.73 | 35.43–36.03 | 0.0004 |
| Tube diameter at the beginning of the aneursym | 35 | 37.67 | 37.20–38.14 | < 0.0001 |
| Maximum tube diameter of the aneurysm | 75 | 75.10 | 74.98–75.22 | 0.0957 |
| Tube diameter at the end of the aneurysm | 35 | 36.55 | 36.27–36.83 | < 0.0001 |
| Tube diameter at the coeliac trunk | 35 | 37.27 | 37.11–37.44 | < 0.0001 |
| Distance of the aneurysm to the subclavian artery | 50 | 48.84 | 46.57–51.10 | 0.2749 |
| Length of the aneurysm | 300 | 295.67 | 291.72–299.62 | 0.0349 |
| Distance of the aneurysm to the coeliac trunk | 50 | 42.44 | 40.83–44.75 | < 0.0001 |
| Total distance between the origin of the subclavian artery and the coeliac trunk | 400 | 386.95 | 383.09–390.32 | < 0.0001 |
Correlation analysis between the 1st and 2nd round at every single workstation.
| Aortic diameter at the subclavian artery | 0.845 | < 0.0001 | 0.824 | < 0.0001 | 0.946 | < 0.0001 |
| Aortic diameter at the beginning of the aneursym | 0.841 | < 0.0001 | 0.854 | < 0.0001 | 0.981 | < 0.0001 |
| Diameter of the thrombus within the aortic lesion | 0.598 | 0.1118 | 0.992 | < 0.0001 | 0.956 | < 0.0001 |
| Maximum diameter of the aneurysm | 0.925 | < 0.0001 | 0.895 | < 0.0001 | 0.792 | < 0.0001 |
| Aortic diameter at the end of the aneurysm | 0.729 | < 0.0001 | 0.786 | < 0.0001 | 0.971 | < 0.0001 |
| Aortic diameter at the coeliac trunk | 0.720 | < 0.0001 | 0.793 | < 0.0001 | 0.989 | < 0.0001 |
| Diameter of the right femoral artery | 0.426 | 0.0470 | 0.722 | < 0.0001 | 0.836 | < 0.0001 |
| Diameter of the left femoral artery | 0.395 | 0.0674 | 0.600 | 0.0005 | 0.404 | 0.0568 |
| Distance of the aneurysm to the subclavian artery | 0.544 | < 0.0001 | 0.314 | 0.1216 | 0.987 | < 0.0001 |
| Length of the aneurysm | 0.807 | < 0.0001 | 0.954 | < 0.0001 | 0.990 | < 0.0001 |
| Distance of the aneurysm to the coeliac trunk | 0.805 | < 0.0001 | 0.959 | < 0.0001 | 0.989 | < 0.0001 |
| Total distance between the origin of the subclavian artery and the coeliac trunk | 0.805 | < 0.0001 | 0.892 | < 0.0001 | 0.933 | < 0.0001 |
Abbreviations: r, correlation coefficient.
Correlation assessment per round comprising all three workstations, separated into the 1st and 2nd round.
| Aortic diameter at the subclavian artery | 0.775 | 0.888 |
| Aortic diameter at the beginning of the aneursym | 0.789 | 0.900 |
| Diameter of the thrombus within the aortic lesion | 0.963 | 0.674 |
| Maximum diameter of the aneurysm | 0.854 | 0.521 |
| Aortic diameter at the end of the aneurysm | 0.856 | 0.719 |
| Aortic diameter at the coeliac trunk | 0.839 | 0.749 |
| Diameter of the right femoral artery | 0.509 | 0.410 |
| Diameter of the left femoral artery | 0.522 | 0.445 |
| Distance of the aneurysm to the subclavian artery | 0.519 | 0.564 |
| Length of the aneurysm | 0.891 | 0.912 |
| Distance of the aneurysm to the coeliac trunk | 0.856 | 0.983 |
| Total distance between the origin of the subclavian artery and the coeliac trunk | 0.810 | 0.912 |
Abbreviations: r, correlation coefficient.
Pairwise correlation analysis of two workstations, respectively.
| Aortic diameter at the subclavian artery | 0.710 | < 0.0001 | 0.827 | < 0.0001 | 0.795 | < 0.0001 | 0.859 | < 0.0001 | 0.868 | < 0.0001 | 0.937 | < 0.0001 |
| Aortic diameter at the beginning of the aneursym | 0.745 | < 0.0001 | 0.784 | < 0.0001 | 0.838 | < 0.0001 | 0.912 | < 0.0001 | 0.868 | < 0.0001 | 0.923 | < 0.0001 |
| Diameter of the thrombus within the aortic lesion | 0.966 | < 0.0001 | 0.949 | < 0.0001 | 0.973 | < 0.0001 | 0.972 | < 0.0001 | 0.615 | 0.0460 | 0.562 | 0.0573 |
| Maximum diameter of the aneurysm | 0.854 | < 0.0001 | 0.798 | < 0.0001 | 0.917 | < 0.0001 | 0.919 | < 0.0001 | 0.372 | 0.0186 | 0.415 | 0.0189 |
| Aortic diameter at the end of the aneurysm | 0.807 | < 0.0001 | 0.579 | 0.0011 | 0.659 | 0.0003 | 0.646 | 0.0002 | 0.964 | < 0.0001 | 0.521 | 0.0042 |
| Aortic diameter at the coeliac trunk | 0.732 | < 0.0001 | 0.812 | < 0.0001 | 0.954 | < 0.0001 | 0.649 | 0.0002 | 0.951 | < 0.0001 | 0.619 | 0.0002 |
| Diameter of the right femoral artery | 0.516 | 0.0461 | 0.507 | 0.0437 | 0.504 | 0.0445 | 0.305 | 0.0478 | 0.315 | 0.0069 | 0.614 | 0.0002 |
| Diameter of the left femoral artery | 0.718 | < 0.0001 | 0.518 | 0.0412 | 0.338 | 0.1749 | 0.324 | 0.0212 | 0.346 | 0.0033 | 0.660 | 0.0003 |
| Distance of the aneurysm to the subclavian artery | 0.274 | 0.1028 | 0.887 | < 0.0001 | 0.196 | 0.1505 | 0.819 | < 0.0001 | 0.465 | 0.0017 | 0.538 | 0.0009 |
| Length of the aneurysm | 0.870 | < 0.0001 | 0.876 | < 0.0001 | 0.931 | < 0.0001 | 0.947 | < 0.0001 | 0.886 | < 0.0001 | 0.905 | < 0.0001 |
| Distance of the aneurysm to the coeliac trunk | 0.813 | < 0.0001 | 0.794 | < 0.0001 | 0.962 | < 0.0001 | 0.982 | < 0.0001 | 0.988 | < 0.0001 | 0.977 | < 0.0001 |
| Total distance between the origin of the subclavian artery and the coeliac trunk | 0.692 | 0.0002 | 0.863 | < 0.0001 | 0.881 | < 0.0001 | 0.907 | < 0.0001 | 0.909 | < 0.0001 | 0.918 | < 0.0001 |
Abbreviations: r, correlation coefficient.
Error report comprising the need for manual correction of measurements as well as system errors.
| Correction of the center axis | 1 | 2 | 1 |
| Correction of the captured vessel lumen | 5 | 1 | 1 |
| Other | 8 | 0 | 0 |
| While loading data sets | 1 | 0 | 5 |
| During image processing | 1 | 4 | 1 |
| Other | 0 | 0 | 1 |
| Tube diameter at the artificial subclavian artery | 35 |
| Tube diameter at the beginning of the aneurysm | 35 |
| Maximum tube diameter of the aneurysm | 75 |
| Tube diameter at the end of the aneurysm | 35 |
| Tube diameter at the coeliac trunk | 35 |
| Distance between the aneurysm and the artificial subclavian artery | 50 |
| Length of the aneurysm | 300 |
| Distance between the aneurysm and the artificial coeliac trunk | 50 |
| Total distance between the origin of the artificial subclavian artery and the coeliac trunk | 400 |