Jintao Han1, Chunyang Lu1, Mengzhu Shen2, Xiaoyi Sun1, Xiaodong Mo2, Gen Yang1,3. 1. State Key Laboratory of Nuclear Physics and Technology, School of Physics, Peking University, Beijing 100871, China. 2. Beijing Key Laboratory of Hematopoietic Stem Cell Transplantation, Peking University Institute of Hematology, Peking University People's Hospital, Beijing 100044, China. 3. Wenzhou Institute, University of Chinese Academy of Sciences, Wenzhou 352001, China.
Abstract
Isolation of circulating tumor cells (CTCs) is of great significance for the diagnosis, prognosis, and treatment of metastatic cancer. Among CTC capture methods independent of antibodies, membrane filtration-based methods have the advantages of simplicity, rapidity, and high throughput but usually have problems such as clogging, high pressure drop, and impaired cell viability. In this study, we designed and tested a reusable device that used horizontal rotor and fluid-assisted separation to capture CTCs by centrifugal membrane filtration, achieving simple, fast, highly efficient, and viable cell capture on traditional centrifuge. The average capture efficiency was 95.8% for different types of cancer cells with >90% survival, and the removal of white blood cells can reach 99.72% under four times cleaning of the membrane after filtration. A further clinic demo was performed using the device to detect residual leukemic cells in patients; the results showed a 10-fold enrichment of the leukemic cells in peripheral blood samples. Taken together, the simple, robust, and efficient CTC capture device may have the potential for clinic routine detection and analysis of circulating tumor cells.
Isolation of circulating tumor cells (CTCs) is of great significance for the diagnosis, prognosis, and treatment of metastatic cancer. Among CTC capture methods independent of antibodies, membrane filtration-based methods have the advantages of simplicity, rapidity, and high throughput but usually have problems such as clogging, high pressure drop, and impaired cell viability. In this study, we designed and tested a reusable device that used horizontal rotor and fluid-assisted separation to capture CTCs by centrifugal membrane filtration, achieving simple, fast, highly efficient, and viable cell capture on traditional centrifuge. The average capture efficiency was 95.8% for different types of cancer cells with >90% survival, and the removal of white blood cells can reach 99.72% under four times cleaning of the membrane after filtration. A further clinic demo was performed using the device to detect residual leukemic cells in patients; the results showed a 10-fold enrichment of the leukemic cells in peripheral blood samples. Taken together, the simple, robust, and efficient CTC capture device may have the potential for clinic routine detection and analysis of circulating tumor cells.
Cancer metastases cause about 90% of cancer-related deaths.[1] Circulating tumor cells (CTCs) are cancer cells
that shed from primary tumor sites to blood or lymph vessels and have
the potential to induce metastasis in distant tissues or organs through
the blood circulatory system.[2,3] In many cancers, the
number of CTCs correlates with the progression-free survival (PFS)
and overall survival (OS) of patients;[4,5] hence, the
diagnostic and prognostic prediction by CTCs is prospective in cancer
control of patients,[6,7] CTCs are recognized as a biomarker
for tumor progress and therapeutic response,[8] the surveillance of CTCs is of great significance for clinical applications,
such as early metastatic detection and individual treatment plan choice,[9,10] and CTCs from patients can also be used for drug resistance tests
and therapeutic targets’ searches.[11]The first step for CTC analysis is the capture of CTCs, and
it
is crucial to obtain intact and viable CTCs for various applications.
However, this has always been challenging since CTCs are rare and
fragile in peripheral blood.[12] Nowadays,
methods using the biological or physical properties of CTCs and methods
combined to detect or capture CTCs are gradually emerging;[13] however, there is still a long way to go to
reach extensive clinical routine use. The immunocapture methods show
good specificity but deeply rely on the expression level of cell surface
biomarkers.[14] The heterogeneity and epithelial–mesenchymal
transition (EMT) of cancer cells are sure to result in loss of detected
CTCs in affinity-based methods,[15] and the
attachment of target cells to the artificial surface will definitely
affect the viability of cells.[16] Besides,
antibody-based methods are usually expensive and time-consuming with
the biochemical marker reagents usage and surface modification process.[17]The CTC enrichment methods based on physical
properties utilize
the size, shape, density, or deformability difference between CTCs
and blood cells, showing advantages for rapid and label-free analysis.[18] Microchip-based streamline or filtration methods
can reach high flow rate but usually are sensitive to it, and these
methods often need special requirements or have poor robustness.[19] On the other hand, the micropore filtration
methods are simple, low cost, and easy to operate and can reach high
throughput and recovery.[20,21] However, there are
usually the problems of clogging, high pressure drop, and cell viability
damage.[22] In order to improve the performance
of membrane filtration, many researches attempted to design a new
filter structure or filtration device, such as a separable bilayer
(SB) microfiltration device or fabric filter device.[23,24] Fluid-assisted separation technology (FAST) is a method inspired
by antifouling membranes with liquid-filled pores in nature, which
has the advantage of low pressure drops in filtration processes and
may have prospects in the application of centrifugal filtration.[25]With regard to the cancer cell identification,
a combination of
antibodies is usually used for different cells; however, the immunobinding
process is rather complex and time-consuming.[26] As widely used in clinical diagnosis, especially in the detection
of hematologic disorders, morphological examination by using Wright–Giemsa
staining and microscope investigation can provide the detailed structure
of cells and has the advantage of simpleness, quickness, and good
accuracy.[27,28]Due to the shortness and limited application
of different CTC capture
methods, there is still urgent need for simple, high-throughput, highly
efficient and viable CTC capture methods for large volume clinical
samples. Here we design a device that realizes self-adapting to a
uniform cell distribution on a membrane by horizontal rotor and fluid-assisted
separation. Using the device, we can achieve label-free CTC capture
with high throughput, high capture efficiency, and high viability,
and combined with morphological examination, the separation and identification
of cancer cells can be realized in the meantime during filtration.
For clinical applications such as point-of-care testing of patient
blood samples, the device may be used for routine cancer detection
and analysis since it is simple, rapid, robust, and reusable.
Results
Design Principles of the
Filtration Device
As shown in Figures a, d, and S1, the filtration device
was designed for the
realization of fluid-assisted CTC separation in a traditional centrifuge
with the device assembled in a 50 mL centrifuge tube, and the horizontal
rotor was used to avoid uneven cell stress during centrifugation;
thus, the cells distributed uniformly on the membrane and simple,
rapid, highly efficient CTC capture was achieved (Figure b).
Figure 1
Operating principles
of the device. (a) Schematic illustration
of the centrifugal process. (b) Schematic illustration of the performance
of blood cells on the membrane during centrifugation. (c) Distribution
of red fluorescent protein (RFP)-stained HeLa cells on the membrane
during centrifugation. The lower image is a zoomed picture of the
upper image in the dotted box. (Scale bar: 100 μm.)
Figure 2
Design of the filtration device. (a) Design parameter of the device.
(b) Schematic illustration of the device. (c) Picture of the device.
Photograph courtesy of Jintao Han. Copyright 2020. (d) Schematic illustration
of the liquid flow in the device during the filtration process.
Operating principles
of the device. (a) Schematic illustration
of the centrifugal process. (b) Schematic illustration of the performance
of blood cells on the membrane during centrifugation. (c) Distribution
of red fluorescent protein (RFP)-stained HeLa cells on the membrane
during centrifugation. The lower image is a zoomed picture of the
upper image in the dotted box. (Scale bar: 100 μm.)Design of the filtration device. (a) Design parameter of the device.
(b) Schematic illustration of the device. (c) Picture of the device.
Photograph courtesy of Jintao Han. Copyright 2020. (d) Schematic illustration
of the liquid flow in the device during the filtration process.In order to compare the cells’ performance
on the membrane
when centrifuged in a different way, HeLa cells stained red were filtrated
in a fixed or horizontal rotor and imaged with a low-magnification
microscope. The results are shown in Figure c: the distribution of cells using the fixed
rotor was uneven, and cells were found at the edge of the membrane,
while cells using the horizontal rotor were uniformly distributed
on the membrane, avoiding cell capture efficiency reduction and cell
viability damage caused by clogging.
Design
Parameters of the Filtration Device
The design parameters
of the device are shown in Figure a, and it consisted of four
parts (Figure b and
c). The top cylindrical chamber was used for sample injection, the
bottom tank was a liquid reservoir, and the middle component comprised
of upper and lower parts was used to place the membrane. There were
grooves or notches in the two parts, which were used for the flow
of the liquid to the 50 mL tube (Figure d). The sample volume injected to the device
can be as large as 15 mL, and the centrifugation process was usually
finished in 1 min. Therefore, rapid, high-throughput liquid sample
analysis can be achieved.
Characterization of the
Filtration Membrane
Used in the Device
The characterization of the filtration
membrane by SEM is demonstrated in Figure S2a, and the size distribution of the membrane pore is shown in Figure S2b, with an average pore size of 6.881
± 0.592 μm (n = 100) and porosity of 3.3%.
Besides, membranes with different pore sizes or porosities can also
be assembled in the device for CTC capture. The images of the membrane
before and after filtration of the blood sample are shown in Figure S2c, demonstrating the device can be used
for simple and rapid liquid sample analysis.
Validation
of the Device Using Cancer Cells
in Culture Medium or Spiked in Blood
The capture efficiency
of cancer cells in culture medium or spiked in blood was measured
using different cell densities and cell lines. RFP-labeled cancer
cells were added in the device for centrifugal filtration, and the
cells on the membrane and flowing out from the device were counted.
The results of capture efficiency are shown in Figure S3a and b for culture medium sample and Figure a and b for the blood sample.
In order to simulate the clinical CTC concentration in patient blood,
the densities of the cancer cells were set at the range 1–600
cells/mL. The average capture efficiency showed good linearity and
was 96.2% and 95.1% in the culture medium and blood, respectively.
The capture efficiencies of four cell lines, including cervical cancer
HeLa cells, colon cancer SW620 cells, breast cancer MDA-MB-231 cells,
and lung cancer NCl-H226 cells, were detected, and the average efficiencies
in blood were 94.6%, 97.8%, 95.7%, and 94.9%, respectively. For the
cancer cells in the culture medium or spiked in blood, the average
capture efficiency for the four cell lines was 95.4 ± 0.8% and
95.8 ± 1.4%, indicating the device can be used to capture different
types of cancer cells with different sizes and deformabilities.
Figure 3
Device validation
using cells spiked in blood. (a) Capture efficiency
for different cell densities using HeLa cells spiked in blood. Regression
analysis is shown. (b) Capture efficiency for cervical cancer HeLa
cells, colon cancer SW620 cells, breast cancer MDA-MB-231 cells, and
lung cancer H226 cells spiked in blood (n = 3). (c)
Merged view of cancer cells (red) and white blood cells (WBCs) (blue)
on the membrane after filtration of blood spiking sample. (Scale bar:
100 μm.) (d) Capture efficiency and WBC depletion ratio of HeLa
cells spiked in blood under different PBS wash times of the membrane
after centrifugation (n = 3).
Device validation
using cells spiked in blood. (a) Capture efficiency
for different cell densities using HeLa cells spiked in blood. Regression
analysis is shown. (b) Capture efficiency for cervical cancer HeLa
cells, colon cancer SW620 cells, breast cancer MDA-MB-231 cells, and
lung cancer H226 cells spiked in blood (n = 3). (c)
Merged view of cancer cells (red) and white blood cells (WBCs) (blue)
on the membrane after filtration of blood spiking sample. (Scale bar:
100 μm.) (d) Capture efficiency and WBC depletion ratio of HeLa
cells spiked in blood under different PBS wash times of the membrane
after centrifugation (n = 3).Purity was important for CTC detection and analysis, so rapid and
efficient WBCs’ removal in blood samples was significant. The
fluorescent images of cancer cells (red) and WBCs (blue) on the membrane
and flow out of the device after filtration are shown in Figures c and S3c, respectively. It was evident that most of
the cancer cells were collected on the membrane while the majority
of WBCs flowed out of the membrane.The wash times of the membrane
using PBS after filtration affected
not only the removal of WBCs but also the detection of CTCs; hence,
the depletion ratio of WBCs and the capture efficiency of cancer cells
were calculated under different washing times. As shown in Figure d, for washing 0–4
times, the log depletion ratio of WBCs was 1.91 ± 0.21, 1.99
± 0.18, 2.18 ± 0.13, 2.34 ± 0.23, and 2.55 ± 0.15.
With WBCs’ removal proportion increased from 98.78% to 99.72%,
the capture efficiency of cancer cells was 94.6%, 94.4%, 96.0%, 93.1%,
and 91.2% at 0–4 washing times, indicating the validity of
the device used for CTC capture.In order to further compare
the performance of CTC detection at
different centrifugal modes, the capture efficiency of cancer cells
and the depletion ratio of WBCs were measured using a horizontal rotor
or fixed rotator and FAST or non-FAST centrifugation. From the results
shown in Figure S3d, we find that when
capturing the cancer cells spiked in blood using the horizontal rotor
and FAST centrifugation, both the capture efficiency and depletion
ratio showed promotion compared with the fixed rotator and non-FAST
centrifugation.
Viability and Growth Verification
of Cancer
Cells after Separation
Since survival of captured cells is
also vital for further cell analysis and testing, the viability and
growth of cancer cells after centrifugal filtration were detected.
As shown in Figure a-b, green fluorescence demonstrated good viability of captured cancer
cells and the survival rate was 90.2% and 94.4% at the first and tenth
day, respectively. The difference of cell viability between experiment
and control group was not significant, indicating that the filtration
process had little damage on cancer cells.
Figure 4
Cancer cells collected
by the device. (a) Images of the cell viability
of the control and the collected cells at days 1 and 10. (Scale bar:
100 μm.) (b) Statistics of the cell survival rate of the control
and the collected cells (n = 3). (c) Images of cell
growth of the control and the collected cells over time. (Scale bar:
100 μm.) (d) Growth curves of the control and the collected
cells (n = 3).
Cancer cells collected
by the device. (a) Images of the cell viability
of the control and the collected cells at days 1 and 10. (Scale bar:
100 μm.) (b) Statistics of the cell survival rate of the control
and the collected cells (n = 3). (c) Images of cell
growth of the control and the collected cells over time. (Scale bar:
100 μm.) (d) Growth curves of the control and the collected
cells (n = 3).Figure c-d showed
the images of captured cells at 1–5 day after filtration and
corresponding growth curve. Cells grew well in both experiment and
control group, the cell counts had no significant difference and the
curve both fitted growth curve model well.
Isolation
of Residual Leukemic Cells from
Patients Received Treatment
Detected immature cells in peripheral
blood indicated emergence or relapse of the blood disease. We analyzed
the peripheral blood of acute leukemia patients who had received treatment,
and the residual leukemic cells were enriched by the device and the
staining process was finished during filtration. As demonstrated in Figure a and b, the cells
stained by Wright–Giemsa clearly showed the size and structure
of the cell and its nuclei, and the leukemic cell percentage of patients
by counting 100–800 cells in three parallel groups is shown
in Figure c. For patient-1
and patient-3, whose cancer cells were undetectable in peripheral
blood, the leukemic cell percentage after enrichment was 2.3% and
13.9% respectively, and for patient-2 the leukemic cell percentage
increased from 1% to 10.7%, indicating 10-fold improvement for cancer
cell detection.
Figure 5
Leukemic cell detection of patients by the device. (a)
Image of
the membrane after Wright–Giemsa staining. (Scale bar: 100
μm.) (b) Representative images of blood cells in patients. (Scale
bar: 20 μm.) (c) Percentage of the leukemic cells in the peripheral
blood of patients before and after enrichment.
Leukemic cell detection of patients by the device. (a)
Image of
the membrane after Wright–Giemsa staining. (Scale bar: 100
μm.) (b) Representative images of blood cells in patients. (Scale
bar: 20 μm.) (c) Percentage of the leukemic cells in the peripheral
blood of patients before and after enrichment.
Discussions and Conclusions
Although many
researches revealed the potential value of CTCs in
cancer diagnosis, prognosis, and treatment, the application of CTC
detection in the clinical environment still faces major challenges.
Large sample volumes and multiple blood draws may be requirements
for CTC analysis in the future and may provide more information about
patients. Membrane filtration and other size-based methods used for
CTC capture have shown great prospects in application with the advantages
of simpleness, rapidness, and high throughput. However, in the face
of complex liquid samples and high standard requirements, more preclinical
tests and large-scale, multicenter clinical trials are needed.In this study, we designed and tested a membrane filtration device
using a horizontal rotor and fluid-assisted separation to capture
cancer cells in blood. The device can be assembled into a 50 mL centrifuge
tube and was able to realize simple, fast (1 min), cheap (reusable),
efficient (∼95%), and viable (>90%) cell capture on a conventional
centrifuge, indicating the potential of the device used for clinic
routine detection and analysis of circulating tumor cells. Besides,
the ability of the filtration device to analyze and process large
volumes of liquid with high efficiency and high throughput was sure
to have important clinical significance. Furthermore, different types
of filtration membranes can be placed in the device and membranes
with higher porosity may be used to achieve more effective WBC removal
and more high-throughput liquid sample analysis.The enrichment
of cancer cells by centrifugal filtration was combined
with morphological examination of cells on the membrane by Wright–Giemsa
staining in our study, realizing both cell separation and identification
at the same time. For the detection of residual leukemic cells in
the peripheral blood of patients post treatment, the device realized
10-fold enrichment of the cancer cells, making close surveillance
of patients possible in routine. With regard to the clinical blood
samples of different patients, the critical size for separating leukemic
cells and normal WBCs needs to be adjusted according to the unique
size distribution of different patients’ blood cells for better
device performance. Besides, more blood sample analysis of different
types of leukemia is necessary to explore the application potential
of the method, and for further application, cells collected after
enrichment can be used to investigate drug resistance or search for
potential therapeutic targets.
Materials and Methods
Device Fabrication
The device is
fabricated according to the design from metal aluminum (Al) and polyamide
(PA) using Turning technology. The device was assembled with a membrane
and fit into 50 mL centrifuge tubes (Coring) for cell separation experiments.
Cell Culture and Blood Collection
HeLa,
SW620, and MDA-MB-231 cells were cultured in DMEM high-glucose
medium (HyClone), and NCI-H226 cells were cultured in RPMI 1640 medium
(HyClone). The cell culture media were all supplemented with 10% fetal
bovine serum (FBS, Bai Ling Biotechnology) and 1% penicillin–streptomycin
(P/S) (HyClone). All of the cells were cultured in the incubator at
the conditions of 37 °C and 5% CO2. The rabbit blood
was used for preclinical cancer cell spiking experiments since the
components in rabbit blood are similar to human blood and it is convenient
to obtain animal blood.[29] The rabbit blood
was collected in lithium heparin tubes (BD Vacutainer), and 10 mg/mL
heparin sodium solution (Yuanye) was added to avoid coagulation.
Cell Isolation Process
For the formation
of fluid-assisted separation and prevention of cell adhesion, a device
assembled with a membrane was added 3 mL of 1% BSA (bovine serum albumin,
Amresco), centrifuged at 300 rpm for 1 min, and then left at room
temperature for 30 min. As shown in Figure S1b, a 1 mL cell suspension of spiked blood was added to the device
and centrifuged at 300 rpm for 1 min with or without PBS wash. The
device was then disassembled, and the membrane and liquid at the bottom
of the 50 mL centrifuge tube were collected for cell imaging.
Viability and Growth Curve of Cells before
and after Filtration
The Calcein-AM/PI staining kit (BestBio)
was used to assay the viability of isolated cancer cells. Live cells
and dead cells showed green and red fluorescence, respectively. For
the verification of cell proliferation ability before and after filtration,
the isolated cancer cells were cultured at 37 °C in a humidified
incubator (5% CO2). The number of cells was counted at
days 1, 2, 3, 4, and 5, and the growth curve was plotted and fit by
a growth formula.
Leukemic Cells’
Isolation from Patients’
Peripheral Blood
The peripheral blood of leukemia patients
was obtained from Peking University People’s Hospital. All
participants gave written informed consent in accordance with the
Declaration of Helsinki, and approval was given by the Peking University
People’s Hospital Institutional Review Board (2021PHB125-001).
ACK lysing buffer (Gibco) was added to blood at 4 °C to remove
red blood cells, and the cell solution was transferred to a freezing
tube for storage. When performing experiments, the diluted patient
sample was stained by Wright–Giemsa stain and filtrated by
the device with an Isopore Polycarb membrane. The cells on the membrane
were imaged by a microscope under a bright field.
Authors: Elisabeth Trapp; Wolfgang Janni; Christian Schindlbeck; Julia Jückstock; Ulrich Andergassen; Amelie de Gregorio; Marianna Alunni-Fabbroni; Marie Tzschaschel; Arkadius Polasik; Julian G Koch; Thomas W P Friedl; Peter A Fasching; Lothar Haeberle; Tanja Fehm; Andreas Schneeweiss; Matthias W Beckmann; Klaus Pantel; Volkmar Mueller; Brigitte Rack; Christoph Scholz Journal: J Natl Cancer Inst Date: 2019-04-01 Impact factor: 13.506
Authors: Sunitha Nagrath; Lecia V Sequist; Shyamala Maheswaran; Daphne W Bell; Daniel Irimia; Lindsey Ulkus; Matthew R Smith; Eunice L Kwak; Subba Digumarthy; Alona Muzikansky; Paula Ryan; Ulysses J Balis; Ronald G Tompkins; Daniel A Haber; Mehmet Toner Journal: Nature Date: 2007-12-20 Impact factor: 49.962