Image quality improvement and motion degradation reduction in shoulder MR imaging: comparison of BLADE and rectilinear techniques at 3-Tesla scanning.

Ajay Kohli1,2, David T Pilkinton1, Yin Xi1, Gina Cho1, Daniel Moore1, Daniel Mohammadi1, Avneesh Chhabra3,4.   

Abstract

OBJECTIVE: MR imaging of joints, particularly shoulder, requires a high degree of spatial resolution to ascertain anatomy and pathology. Unfortunately, motion artifacts can reduce the clinical quality of the examinations. BLADE sequence reduces motion degradation improving overall diagnostic imaging quality. The objective was to compare standard, rectilinear k-space coverage turbo spin echo (TSE), and BLADE sequences.
MATERIAL AND METHODS: Over a 4-month period, fifty-seven consecutive patients (22 males, 35 females; mean age: 48.5 years, range: 23-64 years) were scanned using traditional intermediate-weighted spin echo and BLADE sequences. Qualitative evaluation was performed by three musculoskeletal fellowship trained radiologists, each with more than 5 years of experience. Image sequences were evaluated using a Likert scale for each of the following five categories: motion degradation, ghosting/phase misregistration artifacts, star/radial encoding artifacts, fat suppression quality, and overall diagnostic quality. Additionally, image sequences were evaluated for signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) using manually drawn regions of interest (ROI) analysis.
RESULTS: Ghosting and phase artifacts were lower within BLADE sequence while streak artifacts were higher (p < 0.001). Image fat suppression, tendon and labral appearances, and the overall SNR and CNR were comparable on both sequences (p > 0.05).
CONCLUSION: Addition of BLADE reduces motion degradation and improves overall diagnostic imaging quality. Application of BLADE in patient scans suspected of motion artifacts can reduce the frequency of repeat imaging in patients with claustrophobia or those where motion is a concern. By reducing overall imaging time and call backs, it could reduce the cost burden to patients and healthcare system.
© 2022. The Author(s), under exclusive licence to International Skeletal Society (ISS).

Entities:  

Keywords:  Artifacts; Blade technique; Motion reduction; Propeller technique; Shoulder MRI

Mesh:

Year:  2022        PMID: 35751690     DOI: 10.1007/s00256-022-04085-7

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.128


  1 in total

1.  Elimination of motion and pulsation artifacts using BLADE sequences in shoulder MR imaging.

Authors:  E Lavdas; M Vlychou; E Zaloni; K Vassiou; A Tsagkalis; Z Dailiana; I Fezoulidis
Journal:  Skeletal Radiol       Date:  2015-08-26       Impact factor: 2.199

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.