Cherry Kim1, Thomas Kwack1, Wooil Kim2, Jaehyung Cha3, Zepa Yang4, Hwan Seok Yong5.
Abstract
No published studies have evaluated the accuracy of volumetric measurement of solid nodules and ground-glass nodules on low-dose or ultra-low-dose chest computed tomography, reconstructed using deep learning-based algorithms. This is an important issue in lung cancer screening. Our study aimed to investigate the accuracy of semiautomatic volume measurement of solid nodules and ground-glass nodules, using two deep learning-based image reconstruction algorithms (Truefidelity and ClariCT.AI), compared with iterative reconstruction (ASiR-V) in low-dose and ultra-low-dose settings. We performed computed tomography scans of solid nodules and ground-glass nodules of different diameters placed in a phantom at four radiation doses (120 kVp/220 mA, 120 kVp/90 mA, 120 kVp/40 mA, and 80 kVp/40 mA). Each scan was reconstructed using Truefidelity, ClariCT.AI, and ASiR-V. The solid nodule and ground-glass nodule volumes were measured semiautomatically. The gold-standard volumes could be calculated using the diameter since all nodule phantoms are perfectly spherical. Subsequently, absolute percentage measurement errors of the measured volumes were calculated. Image noise was also calculated. Across all nodules at all dose settings, the absolute percentage measurement errors of Truefidelity and ClariCT.AI were less than 11%; they were significantly lower with Truefidelity or ClariCT.AI than with ASiR-V (all P<0.05). The absolute percentage measurement errors for the smallest solid nodule (3 mm) reconstructed by Truefidelity or ClariCT.AI at all dose settings were significantly lower than those of this nodule reconstructed by ASiR-V (all P<0.05). Furthermore, the lowest absolute percentage measurement errors for ground-glass nodules were observed with Truefidelity or ClariCT.AI at all dose settings. The absolute percentage measurement errors for ground-glass nodules reconstructed with Truefidelity at ultra-low-dose settings were significantly lower than those of all sizes of ground-glass nodules reconstructed with ASiR-V (all P<0.05). Image noise was lowest with Truefidelity (all P<0.05). In conclusion, the deep learning-based algorithms were more accurate for volume measurements of both solid nodules and ground-glass nodules than ASiR-V at both low-dose and ultra-low-dose settings.Entities:
Mesh:
Year: 2022 PMID: 35737734 PMCID: PMC9223620 DOI: 10.1371/journal.pone.0270122
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.752
Characteristics of ten spherical synthetic nodules.
| Nodule types | Diameter (mm) | Reference volume (mm3) | Density (HU) |
|---|---|---|---|
| SN | 3 | 14.1 | 100 |
| 5 | 65.5 | ||
| 8 | 268.1 | ||
| 10 | 523.6 | ||
| GGN | 5 | 65.5 | −630 |
| 8 | 268.1 | ||
| 10 | 523.6 | ||
| 5 | 65.5 | −800 | |
| 8 | 268.1 | ||
| 10 | 523.6 |
Note—SN, solid nodule; GGN, ground-glass nodule.
Summary of the study protocol and radiation dose data.
| Radiation dose | CT dose index volume (mGy) | Dose length product (mGy×cm) | Effective dose (mSv) |
|---|---|---|---|
| 120 kVp/220 mA | 3.39 | 147.56 | 2.07 |
| 120 kVp/90 mA | 1.39 | 60.37 | 0.85 |
| 120 kVp/40 mA | 0.62 | 26.83 | 0.38 |
| 80 kVp/40 mA | 0.2 | 8.49 | 0.12 |
*Tissue weighting factor for calculating the effective dose: 0.014 mSv/mGy×cm
Absolute percentage measurement error according to nodule type.
| Radiation dose (CTDIvol) | Nodule type | ASiR-V | TFI | ClariCT.AI | P-value |
|---|---|---|---|---|---|
| 120 kVp/220 mA (3.39 mGy) | Across all nodules | 7.68 ± 7.12 | 4.12 ± 5.39 | 2.67 ± 3.41 |
|
| SN | 9.49 ± 7.69 | 8.79 ± 7.75 | 4.82 ± 1.09 | 0.081 | |
| GGN | 6.47 ± 1.41 | 1.02 ± 0.22 | 1.23 ± 0.28 |
| |
| 120 kVp/90 mA (1.39 mGy) | Across all nodules | 11.96 ± 9.97 | 5.02 ± 4.51 | 5.88 ± 4.27 |
|
| SN | 9.36 ± 8.28 | 8.18 ± 5.42 | 6.43 ± 4.50 | 0.340 | |
| GGN | 13.70 ± 10.78 | 2.93 ± 1.94 | 5.51 ± 4.16 |
| |
| 120 kVp/40 mA (0.62 mGy) | Across all nodules | 13.58 ± 7.08 | 4.78 ± 5.78 | 8.34 ± 6.78 |
|
| SN | 9.91 ± 6.55 | 8.14 ± 7.47 | 6.20 ± 3.87 | 0.087 | |
| GGN | 16.03 ± 6.44 | 2.54 ± 2.37 | 9.76 ± 7.93 |
| |
| 80 kVp/40 mA (0.2 mGy) | Across all nodules | 18.15 ± 9.72 | 7.39 ± 7.23 | 7.75 ± 6.41 |
|
| SN | 14.21 ± 10.91 | 10.50 ± 8.94 | 7.86 ± 7.67 |
| |
| GGN | 20.78 ± 8.04 | 5.31 ± 5.03 | 7.69 ± 5.59 |
|
Note—CTDI, CT dose index volume, SN, solid nodule; GGN, ground-glass nodule.
“Across all nodules” refers to the mean APEs from all SNs with diameters of 3, 5, 8, and 10 mm and all GGNs with diameters of 5, 8, and 10 mm.
*Comparison with ASiR-V, P<0.05
†Comparison with ClariCT.AI, P<0.05
Absolute percentage measurement error according to solid nodule size.
| Radiation dose (CTDIvol) | Nodule size (mm) | ASiR-V | TFI | ClariCT.AI | P-value |
|---|---|---|---|---|---|
| 120 kVp/220 mA (3.39 mGy) | 3 | 17.24 ± 6.03 | 14.59 ± 5.70 | 4.11 ± 4.13 |
|
| 5 | 5.58 ± 2.52 | 4.05 ± 1.57 | 1.68 ± 1.44 | 0.105 | |
| 8 | 5.41 ± 6.29 | 7.65 ± 4.87 | 4.76 ± 2.77 | 0.368 | |
| 10 | 9.72 ± 7.76 | 8.86 ± 6.40 | 8.72 ± 5.90 | 0.779 | |
| 120 kVp/90 mA (1.39 mGy) | 3 | 18.83 ± 6.20 | 13.88 ± 5.52 | 8.80 ± 5.34 |
|
| 5 | 3.25 ± 3.66 | 5.35 ± 3.24 | 4.28 ± 4.40 | 0.368 | |
| 8 | 11.27 ± 7.45 | 6.81 ± 5.32 | 3.45 ± 2.87 | 0.174 | |
| 10 | 4.09 ± 4.93 | 6.67 ± 4.28 | 9.20 ± 3.07 | 0.472 | |
| 120 kVp/40 mA (0.62 mGy) | 3 | 17.24 ± 5.48 | 6.42 ± 5.58 | 6.90 ± 6.70 |
|
| 5 | 5.88 ± 3.26 | 3.02 ± 2.68 | 5.62 ± 2.49 | 0.174 | |
| 8 | 9.61 ± 4.56 | 5.13 ± 5.11 | 5.80 ± 3.54 | 0.174 | |
| 10 | 6.90 ± 6.89 | 7.01 ± 7.16 | 6.47 ± 3.14 | >0.999 | |
| 80 kVp/40 mA (0.2 mGy) | 3 | 26.89 ± 6.80 | 17.59 ± 6.07 | 17.96 ± 8.58 |
|
| 5 | 5.08 ± 2.61 | 3.63 ± 2.11 | 1.95 ± 2.31 | 0.368 | |
| 8 | 15.40 ± 11.17 | 9.40 ± 10.79 | 5.61 ± 3.69 | 0.420 | |
| 10 | 9.44 ± 7.60 | 5.38 ± 6.68 | 5.90 ± 2.63 | 0.368 |
Note—CTDIvol, CT dose index volume.
*Comparison with ASiR-V, P<0.05
Fig 1CT images of SNs at different radiation dose settings.
Images of SNs (5 mm) at four radiation dose settings reconstructed with ASiR-V, TFI, and ClariCT.AI.
Absolute percentage measurement error according to nodule size and attenuation of ground-glass nodules.
| Radiation dose (CTDIvol) | Attenuation of GGN | Nodule size (mm) | ASiR-V | TFI | ClariCT.AI | P-value |
|---|---|---|---|---|---|---|
| 120 kVp/220 mA (3.39 mGy) | −630 HU | 5 | 9.97 ± 5.59 | 2.25 ± 1.72 | 1.49 ± 1.50 |
|
| 8 | 2.36 ± 0.56 | 0.37 ± 0.04 | 0.30 ± 0.18 |
| ||
| 10 | 2.36 ± 0.29 | 0.49 ± 0.45 | 0.93 ± 1.14 | 0.057 | ||
| −800 HU | 5 | 19.02 ± 3.30 | 1.26 ± 0.40 | 1.26 ± 0.89 |
| |
| 8 | 0.95 ± 0.78 | 0.37 ± 0.40 | 2.05 ± 2.04 | 0.319 | ||
| 10 | 4.17 ± 1.61 | 1.36 ± 1.15 | 1.36 ± 1.74 | 0.058 | ||
| 120 kVp/90 mA (1.39 mGy) | −630 HU | 5 | 16.31 ± 14.08 | 2.64 ± 1.50 | 8.98 ± 7.91 | 0.472 |
| 8 | 7.21 ± 5.90 | 3.64 ± 2.38 | 5.02 ± 2.68 | 0.057 | ||
| 10 | 6.37 ± 3.22 | 0.87 ± 0.25 | 1.50 ± 2.01 | 0.174 | ||
| −800 HU | 5 | 28.23 ± 8.22 | 2.64 ± 2.25 | 6.80 ± 3.30 |
| |
| 8 | 8.70 ± 3.94 | 3.47 ± 0.68 | 5.35 ± 1.84 | 0.368 | ||
| 10 | 15.38 ± 10.43 | 4.34 ± 2.42 | 5.41 ± 2.13 | 0.472 | ||
| 120 kVp/40 mA (0.62 mGy) | −630 HU | 5 | 22.35 ± 5.90 | 3.02 ± 1.99 | 15.62 ± 0.08 |
|
| 8 | 7.86 ± 0.30 | 1.50 ± 1.34 | 4.13 ± 1.52 |
| ||
| 10 | 10.48 ± 0.18 | 2.16 ± 1.26 | 5.77 ± 2.47 |
| ||
| −800 HU | 5 | 24.60 ± 0.89 | 4.51 ± 1.53 | 19.67 ± 13.10 |
| |
| 8 | 15.42 ± 0.36 | 3.57 ± 4.37 | 6.28 ± 3.16 |
| ||
| 10 | 15.49 ± 0.11 | 0.48 ± 0.29 | 7.11 ± 5.15 |
| ||
| 80 kVp/40 mA (0.2 mGy) | -630 HU | 5 | 15.66 ± 7.86 | 5.12 ± 3.85 | 5.19 ± 6.03 |
|
| 8 | 12.34 ± 4.82 | 6.61 ± 5.79 | 9.06 ± 3.53 |
| ||
| 10 | 18.93 ± 1.44 | 0.77 ± 0.72 | 1.68 ± 1.08 |
| ||
| -800 HU | 5 | 26.59 ± 7.03 | 4.47 ± 5.69 | 13.90 ± 7.13 |
| |
| 8 | 19.99 ± 5.53 | 6.99 ± 4.20 | 9.43 ± 2.61 |
| ||
| 10 | 31.15 ± 2.17 | 7.92 ± 7.18 | 6.84 ± 4.07 |
|
Note—CTDIvol, CT dose index volume, GGN, ground-glass nodule.
*Comparison with ASiR-V, P<0.05
†Comparison with ClariCT.AI, P<0.05
Fig 2CT images of GGNs at different radiation dose settings.
Images of GGNs (8 mm) with an attenuation of −630 HU at four radiation dose settings reconstructed with ASiR-V, TFI, and ClariCT.AI.
Fig 3CT images of GGNs at different radiation dose settings.
Images of GGNs (8 mm) with an attenuation of −800 HU at four radiation dose settings reconstructed with ASiR-V, TFI, and ClariCT.AI.
Comparison of image noise values of different algorithms.
| Radiation dose (CTDIvol) | ASiR-V | TFI | ClariCT.AI | P-value |
|---|---|---|---|---|
| 120 kVp/220 mA (3.39 mGy) | 7.95±1.32 | 3.34±0.47 | 6.56±0.53 | 0.002 |
| 120 kVp/90 mA (1.39 mGy) | 11.48±1.44 | 4.65±0.92 | 7.71±0.96 | 0.001 |
| 120 kVp/40 mA (0.62 mGy) | 16.69±0.63 | 6.71±0.53 | 12.16±1.37 | <0.001 |
| 80 kVp/40 mA (0.2 mGy) | 18.73±2.83 | 10.73±1.83 | 13.15±2.21 | 0.014 |
Note—CTDIvol, CT dose index volume
*Comparison with ASiR-V, P<0.05
†Comparison with ClariCT.AI, P<0.05